The Bilimplant® dental implants and abutments, which are models within the Proimtech Implant and Abutment system, are intended for oral implantation to provide a support structure for connected prosthetic devices. Bone Level Implant: Thanks to its design features, it is placed completely at the bone level. It is produced to be applied in different bone types and different regions (anterior and posterior) in the lower and upper jaw. Since it is at the bone level, it can be used for more aesthetic results, especially in the front areas of the jaws (in the areas on the smile line). Tissue Level Implant: The Tissue Level Implant has a 2.3 mm machined collar. It can be used in posterior applications in the upper and lower jaw where there are no aesthetic concerns, in different bone types and especially in cases with high gingival amount in order to facilitate prosthetic stages. Straight Abutments: It is the superstructure part that supports fixed partial dentures manufacture on a straight implant. It is used in single member or bridge cemented restorations. Cement retained abutments have a different gingival height of 1-5 mm, diameters of 3.5, 4.5, 6 mm depending on the platform diameters. Healing Caps: Following the second surgery of the gingiva in two-stage surgical procedures, and after the placement of the implant in single-stage surgeries, it is screwed into the implant body and protects the internal structure of the implant. It is not used to support a prosthetic superstructure. It is used for transgingival healing and shaping of soft tissue during the healing process of soft tissue. There are two different designs for healing caps in dental implant systems. These are manufactured to be compatible with tissue level and bone level implants. Abutment Screws: Connects and fixes the abutment and implant body.
Device Story
Proimtech Dental System consists of endosseous dental implants, abutments, healing caps, and abutment screws. Implants are placed into the maxillary or mandibular bone to provide a foundation for prosthetic restorations (crowns/bridges). The system is used by dental clinicians in clinical settings. Implants are made of Grade 4 CP titanium (ISO 5832-2) with an SLA (sand-blasted, large-grit, acid-etched) surface. Abutments and screws are made of Ti-6Al-4V ELI (ASTM F136). The device provides a stable anchor for prosthetic components, restoring masticatory function and aesthetics for patients with missing teeth. The clinician selects the appropriate implant/abutment size based on bone type and anatomical region. Healing caps are used during the soft tissue healing phase to protect the internal implant structure and shape the gingiva. The system is intended for single-use only.
Clinical Evidence
No clinical data were included in this submission. Evidence is based on bench testing, including biocompatibility (ISO 10993-5/12), sterility validation (ISO 11137-1/2, ISO 17665-1/2), bacterial endotoxin testing (USP <85>), and surface characterization via SEM and EDS.
Technological Characteristics
Implants: Grade 4 CP titanium (ISO 5832-2), SLA surface. Abutments/Screws: Ti-6Al-4V ELI (ASTM F136). Connection: Internal/External conical/quadrilateral. Sterilization: Gamma irradiation (implants) or moist heat/steam (abutments/caps). Dimensions: Various diameters (3.5-6.5mm) and lengths (6-18mm). Single-use only.
Indications for Use
Indicated for functional and esthetic oral rehabilitation of upper or lower jaw in edentulous or partially edentulous patients. Implants can be used for immediate, early, or late implantation following tooth extraction or loss. Implants can be placed with immediate function for single-tooth or multiple-tooth applications given good primary stability and appropriate occlusal loading. Abutments and prosthetic parts are indicated for use with implants to support crowns or bridges.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
CAMLOG Dental Implant Abutments, Healing Caps, and Accessories (K051636)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2024
Proimtech Saglik Urunleri Anonim Sirketi Hakan Cevik General Manager Imes Sitesi, No:3 Dudullu Osb Mahallesi Imes 305.Sokak Istanbul. Istanbul 34773 TURKEY
# Re: K231100
Trade/Device Name: Proimtech Dental Body Implant, Proimtech Abutment, Proimtech Healing Cap, Proimtech Abutment Screw Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 25, 2023 Received: July 31, 2024
Dear Hakan Cevik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231100
#### Device Name
Proimtech Dental Body Implant, Proimtech Abutment, Proimtech Healing Cap, Proimtech Abutment Screw
#### Indications for Use (Describe)
Indications for use Bilimplant® dental implants are indicated for the functional and esthetic oral rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients. They can be used for immediate, early or late implantation following the extraction or loss of natural teeth. The implants can be placed with immediate function for single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function.
Bilimplant Abutments and Prosthetic parts are intended for use with Implants in the maxillary and / or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary 510k K231100 Dental Body Implant Proimtech Saglik Urunleri Anonim Sirketi August 23, 2024
## ADMINISTRATIVE INFORMATION
| Manufacturer Name: | PROIMTECH SAGLIK URUNLERI ANONIM SIRKETI |
|--------------------|-------------------------------------------------------------------------------------|
| Adress: | Dudullu OSB Mahallesi İmes Sanayi Sitesi 305. Sokak C Blok NO:3, Ümraniye, İstanbul |
| Telephone: | (0216) 999 50 41 |
| Fax: | |
| Official Contact: | Hakan Cevik, General Manager of Proimtech Saglik Urunleri Anonim Sirketi |
| Phone Number: | +90 535 355 51 00 |
| E-Mail: | hakancevik@bilimplant.com |
#### DEVICE NAME AND CLASSIFICATION
Device Trade Name: Proimtech Dental Body Implant, Proimtech Abutment, Proimtech Healing Cap, Proimtech Abutment Screw
Common Name: Dental Body Implant, Abutment, Healing Cap, Abutment Screw,
Classification Name: Dental Body Implant, Endosseous dental implant (21 CFR 872.3640),
| Device Class: | II |
|--------------------|----------|
| Product Code: | DZE |
| Regulation Number: | 872.3640 |
| Review Panel: | Dental |
1
{5}------------------------------------------------
## Legally Marketed Device to Which Claim Substantial Equivalence:
for; Primary Predicate Device Implant Body: K140878, STRAUMANN BONE LEVEL TAPERED IMPLANT
for; Reference Device Implant Body: K173961, Straumann® BLX Implant System
for; Reference Device Implant Body: K212533, BLX WB Ø5.0 (L18), Ø5.5 and Ø6.5 mm (L14 and L16) Implants
for; Reference Device Implant Body: K130222, Straumann Dental Implant System SLActive and Roxolid Product Families
for; Reference Device Abutment: K160850, Nucleoss Abutment System SANLILAR TIBBI CIHAZLAR MED. KIM.SAN.TIC.LTD.STI
for; Reference Device Abutment Screw: K182448, MegaGen Implant Co., Ltd.
for; Reference Device Healing Cap: K052957, Dentium Implantium® & SuperLine® Prosthetics
for; Reference Healing Cap: K051636, CAMLOG Dental Implant Abutments, Healing Caps, and Accessories
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
# Dental Body Implant:
Indications for use Bilimplant® dental implants are indicated for the functional and esthetic oral rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients. They can be used for immediate, early or late implantation following the extraction or loss of natural teeth. The implants can be placed with immediate function for single-tooth restorations and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function.
For Abutment Healing Cap Abutment Screw:
## Indications:
Bilimplant Abutments and Prosthetic parts are intended for use with Bilimplants in the maxillary and / or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.
{7}------------------------------------------------
# 2. Intended Use
The Bilimplant® dental implants and abutments, which are models within the Proimtech Implant and Abutment system, are intended for oral implantation to provide a support structure for connected prosthetic devices.
| PRODUCT NAME | INTENDED USE |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bone Level Implant | Thanks to its design features, it is placed completely at the bone level. It is produced to be applied in different bone types and different regions (anterior and posterior) in the lower and upper jaw. Since it is at the bone level, it can be used for more aesthetic results, especially in the front areas of the jaws (in the areas on the smile line). |
| Tissue Level Implant | The Tissue Level Implant has a 2.3 mm machined collar. It can be used in posterior applications in the upper and lower jaw where there are no aesthetic concerns, in different bone types and especially in cases with high gingival amount in order to facilitate prosthetic stages. |
| PRODUCT GROUP | INTENDED USE |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Straight Abutments | It is the superstructure part that supports fixed partial<br>dentures manufacture on a straight implant. It is used in single<br>member or bridge cemented restorations. Cement retained<br>abutments have a different gingival height of 1-5 mm,<br>diameters of 3.5, 4.5, 6 mm depending on the platform<br>diameters. |
| Healing Caps | Following the second surgery of the gingiva in two-stage<br>surgical procedures, and after the placement of the implant<br>in single-stage surgeries, it is screwed into the implant body<br>and protects the internal structure of the implant. It is not<br>used to support a prosthetic superstructure. It is used for<br>transgingival healing and shaping of soft tissue during the<br>healing process of soft tissue. There are two different designs<br>for healing caps in dental implant systems. These are<br>manufactured to be compatible with tissue level and bone<br>level implants. |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
4
{8}------------------------------------------------
| Abutment Screws | Connects and fixes the abutment and implant body. |
|-----------------|---------------------------------------------------|
|-----------------|---------------------------------------------------|
{9}------------------------------------------------
# Dental Implant
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | REFERENC E DEVICE | REFERENCE DEVICE | REFERENCE DEVICE |
|------------------------------------------|--------------------------------------------|----------------------------------------------------|-----------------------------------------------------|--------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Technologi<br>cal<br>Characteris<br>tics | Proimtech Dental Body Implant<br>(K231100) | Straumann Bone Level Tapered<br>Implants (K140878) | Straumann® Bone Level<br>Tapered Implants (K140878) | Straumann® BLX Implant<br>System (K173961) | BLX WB<br>Ø5.0 (L18),<br>Ø5.5 and Ø6.5<br>mm (L14 and L16)<br>Implants (K212533) | Straumann Dental Implant<br>System SLActive and Roxolid<br>Product Families (K13022<br>2) |
| Product<br>Code | DZE | DZE | DZE | DZE | DZE | DZE |
{10}------------------------------------------------
Indications for use Bilimplant® dental implants are indicated for the functional and esthetic oral rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients. They can be used for immediate, early or late implantation following the extraction or loss of natural teeth. The implants can be placed with immediate function for singletooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function.
Indications
for Use
Indicated for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients. Straumann dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases
Straumann® dental implants are indicated for oral endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially dentate patients. Straumann dental implants can also be used for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants by the corresponding elements (abutments). In cases of fully edentulous patients, 4 or more implants must be used in immediately loaded cases.
Straumann BLX Implants are suitable for endosteal implantation in the upper and lower jaw and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. BLX Implants can be placed with immediate function on single-tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components.
Straumann® dental implants are indicated for the functional and esthetic oral rehabilitation of the upper or lower jaw of edentulous or partially edentulous patients. They can be used for immediate, early or late implantation following the extraction or loss of natural teeth. The implants can be placed with immediate function for single-tooth and/or multiple-tooth restorations when good primary stability is achieved and with appropriate occlusal loading to restore chewing function.
Straumann® Dental implants are indicated for oral endosteal implantation in the upper and lower jaw arches for the functional and aesthetic oral rehabilitation of edentulous and partially dentate patients. Straumann® Dental implants are also indicated for immediate or early implantation following extraction or loss of natural teeth. Implants can be placed with immediate function on single-tooth and/or multiple tooth applications when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations used are single crowns, bridges and partial or full dentures, which are connected to the implants through the corresponding components (abutments).
7
{11}------------------------------------------------
| Material | Grade 4 commercially pure titanium<br>conforming with ISO 5832-<br>2 | Grade 4 commercially pure<br>titanium conforming with ISO<br>5832-2 | Commercially pure grade 4<br>titanium & Titaniun-<br>13Zirconium alloy (Roxolid®) | Titanium<br>-13<br>Zirconium alloy (Roxolid<br>®) | Titanium-13<br>Zirconium alloy<br>(Roxolid®) | Titanium-13 Zirconium alloy<br>(Roxolid®) |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Surface<br>Treatment | Sand blasted Large grit Acid etched<br>(SLA) | Sand blasted Large grit Acid<br>etched (SLA) | SLA, SLActive | Hydrophilic SLActive® | Hydrophilic<br>SLActive® and SLA® | SLA,<br>SLActive |
| Design | Image: Implant | Image: Implant | Image: Implant | Image: Implant | Image: Implant | Image: Implant |
| Single<br>Use<br>Only | Yes | Yes | Yes | Yes | Yes | Yes |
| Transfer<br>piece | Fixture-mount transfer part, intended<br>to support the implant while in the<br>primary package, to aid in the<br>removal of the implant from the<br>primary package, and to aid in<br>placement of the implant into the<br>osteotomy site. | Snap fit mount Loxim™<br>transfer piece, intended to<br>support the implant while in<br>the primary package, to aid in<br>the removal of the implant<br>from the primary package, and<br>to aid in placement of the<br>implant into the osteotomy<br>site. | Fixture-mount transfer part,<br>intended to support the<br>implant while in the primary<br>package, to aid in the<br>removal of the implant from<br>the primary package, and to<br>aid in placement of the<br>implant into the osteotomy<br>site. | Fixture-mount transfer part,<br>intended to support the<br>implant while in the primary<br>package, to aid in the<br>removal of the implant from<br>the primary package, and to<br>aid in placement of the<br>implant into the osteotomy<br>site. | Fixture-mount<br>transfer part,<br>intended to support<br>the implant while in<br>the primary package,<br>to aid in the removal<br>of the implant from<br>the primary package,<br>and to aid in<br>placement of the<br>implant into the<br>osteotomy site. | Fixture-mount transfer part,<br>intended to support the<br>implant while in the primary<br>package, to aid in the<br>removal of the implant from<br>the primary package, and to<br>aid in placement of the<br>implant into the osteotomy<br>site. |
| | | | | | | |
{12}------------------------------------------------
| Diameter<br>(mm) | 3.7, 4.1, 4.8, 5.5, | 3.3, 4.1, 4.8 | 4.1, 4.8 | Ø 4.5, Ø5.5, Ø6.5 mm | Ø5.0, Ø5.5, Ø6.5<br>mm | Ø4.1, Ø4.8<br>mm |
|------------------|---------------------|---------------|----------|----------------------|------------------------|------------------|
{13}------------------------------------------------
| Length<br>(mm) | 8, 10,<br>12, 14 | 8, 10, 12, 14 | 8, 10, 12, 14 | 6 to 18 mm | Ø5.0 mm: 18 mm<br>Ø5.5 and 6.5 mm: 14<br>and 16mm | 6, 8, 10, 12 mm |
|-------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|---------------------------------------------------|-----------------------------|
| Sterilizatio<br>n | Sterile (Gamma irradiation) | Sterile (Gamma irradiation) | Sterile (Gamma irradiation) | Sterile (Gamma irradiation) | Sterile (Gamma irradiation) | Sterile (Gamma irradiation) |
Substantial Equivalence Discussion
#### Similarities
The device has the same characteristics compared to the following aspects. Indication for Use, Design, Material, Surface Transfer Part and Sterlization.
#### Difference
The diameter and length directions of the different from the reference device. However, as a esult of the tests conduced, it was cecided hat there was no significat difference in product performance and that this difference did not affect substantial equivalence.
Reference to fully edentulous patients is not necessary because the subject device is intended for fully edertulous patients, it is not intended for full-arch restorations.
#### Discussion
On the basis of the above discussion, it is concluded that the device in question is substantially equivalent to the reference device.
{14}------------------------------------------------
## Abutment
| Technological Characteristics | SUBJECT DEVICE | REFERENCE DEVICE |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Proimtech Abutment | Nucleoss Abutment System (K160850) |
| Product Code | NHA | NHA |
| Indications for Use | Bilimplant Abutments and Prosthetic parts are intended for use with Bilimplant<br>Dental Implants in the maxillary and / or mandibular arches to provide support<br>for crowns or bridges for edentulous or partially edentulous patients.<br>Straight: Internal quadrilateral connection for single or multiple cement-<br>retained restorations. | Nucleoss Abutments and Prosthetic parts are intended for use with Nucleoss Tpure<br>Dental Implants in the maxillary and/or mandibular arches to provide support for<br>crowns, bridges or overdenture for edentulous or partially edentulous patients.<br>Straight: For single or multiple cement retained restorations Internal hex<br>connection. |
| Abutment Design | | |
| Material | | |
| Titanium alloy | Titanium alloy | |
| Abutment Platform Diameter (mm) | Straight: 3.5, 4.5, 5, 6 | Straight: 4.2, 5.0 6.0 |
| Abutment Platform Heights (mm) | Straight: 5.5, 7 | Straight: 1.0, 2.0, 3.0, 5.0 |
| Gingival Height | 1, 2, 3, 4 and 5 mm | 1, 1.5, 2, 3, 4, 4.5 and 5 mm |
| Single Use Only | Yes | Yes |
| Sterilization | Provided non- sterile | Provided non-sterile |
| Recommended Sterilization Method | Autoclave Moist Heat or Steam | Autoclave Moist Heat or Steam |
{15}------------------------------------------------
{16}------------------------------------------------
| Design | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Substantial Equivalence Discussion | | |
| <b>Similarities</b><br>The device in question has the same characteristics compared to the reference device in the following aspects. Abutment Design, Material, Single use and Sterilization.<br><b>Difference</b><br>The connection type of the device in question is Four Corner Connection Type. Abutment Platform Diameter and Abutment Platform Heights are slightly different from the reference device.<br>However, as a result of the tests, it was decided that there was no significant difference in product performance and that this difference did not affect the substantial equivalence. | | |
Discussion
On the basis of the above discussion, it is concluded that the device in question is substantially equivalent to the reference device.
{17}------------------------------------------------
# Healing Cap
| | SUBJECT DEVICE | REFERENCE DEVİCE | REFERENCE DEVICE |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | NHA | NHA | NHA |
| Device Name | Proimtech Healing Cap | Implantium Abutments<br>(Healing Abutment) | CAMLOG Dental<br>Implant<br>Abutments, Healing<br>Caps, and<br>Accessories |
| Manufacturer | PROIMTECH SAGLIK URUNLERI ANONIM SIRKETI | Dentium Co., Ltd. | Altatec GmbH |
| 510(k) Number | K231100 | K052957 | K051636 |
| Indication for use | Bilimplant Abutments and Prosthetic parts are intended for use with Bilimplant Dental Implants in the maxillary and / or<br>mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients. | Implantium Prosthetics is<br>intended for use as an aid in<br>prosthetic rehabilitation. | CAMLOG<br>(Cylindrical, Wide<br>Body. Bottleneck)<br>displace the gingiva<br>from the space<br>above the<br>CAMLOG implant<br>or bar abutment<br>during the CAMLOG<br>implant healing<br>time and serve for<br>proper gingiva<br>shaping. |
| Materials | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Titanium Alloy Ti-<br>6Al-4V ELI |
| Connection type | Internal | Internal | Internal |
| Sterilization | Non-sterile | Non-sterile | Sterile |
| Recommended<br>Sterilization Method | Autoclave Moist Heat or Steam | Autoclave Moist Heat or Steam | Gamma Irradiation |
| Single Use Only | Yes | Yes | Yes |
| Diameter | 3.5, 4.5, 4.8, 5, 5.4, 6.5 | 4.04 / 4.10 / 4.14 / 4.20 / 4.50<br>/ 4.54 / 4.64 / 4.70 / 5.50 /<br>5.54 / 5.64 / 5.75<br>6.50 / 6.54 / 6.64 / 6.75 / 7.64<br>/ 8.64 / 9.64 | 3.3. 3.8, 4.3. 5 and<br>6mm<br>Cylindrical |
| Length | 4.70, 5.10, 5.40, 6.20, 6.50, 6.90, 7.20, 7.50, 8, 8.20, 8.40, 8.50, 9, 9.40, 9.70, 10, 10.20, 10.40, 10.50, 11, 11.20,<br>11.40, 11.50, 11.70, 12, 12.50 | 8.70 / 10.91 / 10.93 / 11.04 /<br>11.15 / 12.41 / 12.44 / 12.55 /<br>12.65 / 12.66<br>14.42 / 14.44 / 14.55 / 14.66 | 2, 4 and 6mm |
| Gingival Height | 1, 2, 2.5, 3, 4, 5, and 6 mm | 3, 4, 5 and 7 mm | 2, 4 and 6 mm |
| Design | | Image: silver colored implant abutment | Image: gray colored implant abutment |
{18}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{19}------------------------------------------------
## Similarities
The device in question has the same characteristics compared to the following aspects. Design, Material, Connection Type, Single use and Sterlization. Difference
The diameter and length different in question are sightly different from the reference in significant difference in product performance and this ofference does not affect the substantial equivalence.
## Discussion
On the basis of the above discussion, it is conduded that the device in question is substantially equivalent to the reference device.
{20}------------------------------------------------
# Abutment Screw
| | SUBJECT DEVICE | REFERENCE DEVICE |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Product Code | NHA | NHA |
| Device Name | Proimtech Abutment Screw | Abutment Screw |
| Manufacturer | PROIMTECH SAGLIK URUNLERI ANONIM SIRKETI | MegaGen Implant Co., Ltd. |
| 510(k) Number | K231100 | K182448 |
| Indication for use | Bilimplant Abutments and Prosthetic parts are intended for use with Bilimplant Dental<br>Implants in the maxillary and / or mandibular arches to provide support for crowns or bridges<br>for edentulous or partially edentulous patients. | Abutment Screw is used for securing the abutment to the<br>endosseous implant. |
| Materials | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Connection Interface | External Conical Connection | Internal Conical Connection |
| Sterilization | Non-sterile | Non-sterile |
| Recommended Sterilization Method | Autoclave Moist Heat or Steam | Autoclave Moist Heat or Steam |
| Single Use Only | Yes | Yes |
| Surface Treatment | Machined, Anodizing | Machined |
| Diameter | 2, 2.2, 2.3 | 2.2 |
| Length | 4.50, 5.50, 7.90, 8.10 | 7.9 |
| Design | | |
| Substantial Equivalence Discussion | | |
#### Similarities
The device in question has he same charace device in the following appects. Indication for Use, Jesign, Material, Surface Treation. Difference
The dianeter, length dinerisons and connection are slight) different from the reference device. Howeve, there in o significant difference in product performance and this difference does not affect the substantial equivalence.
#### Discussion
On the basis of the above discussion, it is concluded that the device in question is substantially equivalent to the reference device.
{21}------------------------------------------------
## PERFORMANCE DATA
The subject device was evaluated and tested as recomments Root Form Endosseous Dental Implants and Endosseous Dertal Implant Alauthents lissued May 12,2004), Use of International Standation of medical devices- Part 1: Evaluation and testing within a risk management process" (issued September 4,2020), Reprocessing Meath Care settings: Validation Methods and Labelling (issued March 17, 2015), Submission and Review of Sterlity Information in Peemarket Notification (510(k) Submissions of Devices Labeled as Sterile (issued January 21,2016), and Progen and Answer (issued June 2012).
Non-dinical data presented to demonstration waldet Gamma sterlization validation for device implants and valve screws according to O 1137-1 and ISO 11137-2; Bacterial endotoxin (BET) testing, including time (LAL) testing according to USP-43- NF38; 2020 <85>, in accordance with ASTM F1980 and shel life validation through packaging vith ISO 1160; moist heat sterilization in accordance with ISO 17665-1 and ISO TS 17665-2 (to be performed by the end used; biocompatibility testing in accordance with ISO 10993-5 and ISO 10993-12.
The endosseus threaded surface of the subject is grit-blasted with non-resorbable aluminum oxide (AI20) particles; this surface was valiated by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) characterization. Bacterial endotoxins in the subject devices provided sterile to the end user will be monitored and BET level of ≤ 20 EU/device.
No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICES
The subject devices are broadly equivalent in inciples to the primary predecessor and reference devices listed above. At the beginning of this unmary is a table compaing the Indications for Use and table tharacteristics of the subject device, the primary precursor devices.
The Indications for Use Statement (FUS) of the substantially equivalent to that of the primary precursor devices KL7396, K212553, K18022, K160850, K182448, K152957, K05166510. Minor differ its intentified intended use as an endosseus dertal inqlant and dental implant abutments to support a prosthesis to restore masticatory function.
There are minor differences between the FUS of the primary predicate device. These minor differences do not raise new safety or efficacy questions as the IFUS states equivalent intended use.
## CONCLUSION
The subject device, the primary predicate devices have the same intended use, have similar technological characteristics, and are made of similar materials. The subject device, the primary predicate devices encompass the same rage of physical dimensions, are paraged in similar materials, and are sterilized using similar method. This submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.