The Aurelia Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Aurelia Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, provided nonsterile.
Device Story
Aurelia Surgical Mask is a three-layer, single-use, non-sterile disposable face mask. Construction consists of inner and outer spun-bonded polypropylene layers and a middle melt-blown polypropylene filtration layer. Features include polyurethane/nylon ear loops and a 3mm aluminum flat wire nose clip for fit. Available in ASTM Level 1 and Level 2 specifications. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against blood, body fluids, and particulates. Device functions as personal protective equipment (PPE) to reduce cross-contamination risk. Performance is verified through fluid resistance, bacterial/particulate filtration efficiency, differential pressure, and flammability testing.
Clinical Evidence
Bench testing only. Performance verified against ASTM F2100-19 standards, including fluid resistance (ASTM F1862), bacterial filtration efficiency (ASTM F2101), particulate filtration efficiency (ASTM F2299), differential pressure, and flammability (16 CFR 1610). Biocompatibility testing conducted per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization and irritation). Accelerated aging study (50°C for 229 days) supports a 5-year shelf life.
Technological Characteristics
Three-layer nonwoven polypropylene mask. Inner/outer layers: spun-bonded polypropylene; middle layer: 25gsm melt-blown polypropylene. Nose clip: 3mm aluminum flat wire. Ear loops: polyurethane and nylon. ASTM Level 1 and 2 specifications. Non-sterile. Biocompatibility: ISO 10993-5, ISO 10993-10. Standards: ASTM F2100-19, ASTM F1862, ASTM F2101, ASTM F2299, 16 CFR 1610.
Indications for Use
Indicated for use by healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 5, 2023
Supermax Healthcare Canada-Supermax Medical % Alicja Wojewnik CEO dicentra, Inc. 7 St Thomas Street, Unit 603 Toronto. Ontario M5S 2B7 Canada
Re: K231005
Trade/Device Name: Aurelia Surgical Mask ASTM Level 1: Aurelia Surgical Mask ASTM Level 2 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: November 1, 2023 Received: November 1, 2023
# Dear Alicja Wojewnik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231005
Device Name
Aurelia Surgical Mask ASTM Level 1: Aurelia Surgical Mask ASTM Level 2
Indications for Use (Describe)
The Aurelia Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Aurelia Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, provided nonsterile.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Registration Use (Part 21 CFR 201.2 Subpart D) | ☑ |
|------------------------------------------------|---|
| OTC Therapeutic Use (21 CFR 201.2 Subpart G) | ☐ |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Dicentra logo. The logo features the word "dicentra" in a stylized font, with the "d" in purple and the rest of the letters in black. To the left of the "d" is a purple sphere made up of interconnected dots. Below the word "dicentra" are the words "GUIDANCE | INNOVATION | TRUST" in a smaller font.
# 510(k) Summary
Aurelia Surgical Mask
| Applicant | Supermax Healthcare Canada – Supermax Medical<br>1001 rue Jean-Talon bureau 100<br>Saint-Bruno-de-Montarville, Quebec, J3V 0N3<br>Canada |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Nicolas Bergeron<br>Executive Vice President<br>Tel. +1 (866) 428-9188<br>Cell: +1 (514) 512-4123<br>nbergeron@supermaxcanada.com |
| Correspondent: | dicentra, Inc.<br>603 - 7 St. Thomas Street<br>Toronto, Ontario M5S2B7 Canada<br>Alicja Wojewnik<br>Tel. (416) 361-3400 Ext 228<br>radicentra@dicentra.com |
| Summary Date | October 20, 2023 |
| Proprietary Name | Aurelia Surgical Mask ASTM Level 1; Aurelia Surgical Mask ASTM Level 2 |
| Common Name | Surgical Mask |
| Classification Regulation | 21 CFR 878.4040. |
| Product Codes | FXX |
| Device Classification | Class II |
| Predicate Device | a. Crosstex® Isofluid® Earloop Face Mask (K082258) Primary Predicate<br>b. Crosstex® Procedural Earloop Face Mask (K082258) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for "dicentra". The logo features the word "dicentra" in a stylized font, with the "d" in purple and the rest of the letters in black. To the left of the "d" is a purple graphic that resembles a network of connected dots. Below the word "dicentra" are the words "GUIDANCE | INNOVATION | TRUST" in a smaller font.
# Indications for Use
The Aurelia Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Aurelia Surgical Mask is intended for use in infection control practices to reduce the posure to blood and body fluids. This is a single use, disposable device, provided non-sterile.
## Device Description
The proposed device, Aurelia Surgical Masks are three-layer masks constructed of nonwoven polypropylene. The inner and outer layers are made of spun-bonded polypropylene. The middle layer consists of 25gsm melt-blown polypyrene. The mask contains an ear loop to secure the mask over the users' mouth and face and includes a nosepiece to provide a firm fit over the nose. Ear loops are made of polyurethane and nylon. The nose clip which is made of 3mm aluminum flat wire. The disposable surgical face mask is available in two different specifications: ASTM Level 1 and ASTM Level 2. The masks meet the following ASTM 1 and ASTM 2 Technical standards.
| Test | ASTM 1 | ASTM 2 |
|-----------------------------------------|-----------------|----------------|
| Fluid<br>Resistance | 80mmHg | 120mmHg |
| Bacterial<br>Filtration<br>Efficiency | ≥95% | ≥98% |
| Particulate<br>Filtration<br>Efficiency | ≥95% | ≥98% |
| Differential<br>Pressure | <5.0 mm H2O/cm² | <6.0 mm 2O/cm² |
| Flammability | Class 1 | Class 1 |
#### Table 1: ASTM Standards
The device is single use and provided non-sterile. Both masks are blue in color and have the following dimensions: Length: 17.5 cm, Height: 9.5 cm, Ear loop Length: 15.2 cm, Nosepiece: 11.5 cm
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for "dicentra". The logo features the word "dicentra" in a stylized font, with the "d" in purple and the rest of the letters in black. To the left of the "d" is a purple graphic that resembles a network of interconnected nodes. Below the word "dicentra" are the words "GUIDANCE | INNOVATION | TRUST" in a smaller font.
# Table 2: Material Information
| Raw Material | Product Details | Supplier | Chemical<br>Composition | Colorant |
|--------------|----------------------------------------------|----------------------------------------------|------------------------------------------------------------|-----------------------------------------|
| Inner Layer | White 25 gsm Spunbonded<br>non-woven fabric | Anhui Medysen<br>Import & Export<br>Co., LTD | Polypropylene<br>CAS:9003-07-0 | N/A |
| Middle Layer | White 25 gsm Melt-<br>Blown non-woven fabric | Anhui Medysen<br>Import & Export<br>Co., LTD | Polypropylene<br>CAS:9003-07-0<br>Irganox<br>CAS:6683-19-8 | N/A |
| Outer Layer | Blue 25 gsm Spun-bonded<br>nonwoven fabric | Anhui Medysen<br>Import & Export<br>Co., LTD | Polypropylene<br>CAS:9003-07-0 | Phthalocyanine<br>blue CAS:147-<br>14-8 |
| Nose Piece | 3mm Aluminum Flat Wire<br>Nose Strip | Anhui Medysen<br>Import & Export<br>Co., LTD | Aluminum<br>CAS:7429-90-5 | N/A |
| Ear Loops | White 4MM Elastic Ear Loop | Anhui Medysen<br>Import & Export<br>Co., LTD | Nylon CAS:32131-<br>17-2<br>Polyurethane<br>CAS:51852-81-4 | N/A |
# Table 3: Comparison to Predicate Device (LV1)
| Comparison to<br>Predicate Device | Predicate Device:<br>Crosstex® Isofluid® Earloop Face Mask<br>(K082258) | Candidate Device:<br>Aurelia Surgical Mask Lv. 1 | Conclusion |
|--------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------|------------|
| Product Code | FXX | FXX | Same |
| Regulation<br>Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Fluid Resistance | 80mmHg | 80mmHg | Same |
| Particulate Filtration<br>Efficiency | 95% | 99.5% | Different |
| Bacterial Filtration<br>Efficiency | 95% | 99.7% | Different |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Dicentra logo. The logo features the word "dicentra" in a stylized font, with the "d" in purple and the rest of the letters in black. To the left of the "d" is a purple graphic that resembles a network of connected dots. Below the word "dicentra" are the words "GUIDANCE | INNOVATION | TRUST" in a smaller font.
| Differential Pressure | <4.0 mm H2O/cm² | 4.5 mm H2O/cm² | Different |
|-----------------------|------------------|------------------|-----------|
| Flammability | Class 1 | Class 1 | Same |
| Biocompatibility | | | |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Sensitization | No Sensitization | No Sensitization | Same |
| Irritation | No Irritation | No Irritation | Same |
#### Table 4: Comparison to Predicate Device (LV2)
| Comparison to<br>Predicate Device | Predicate Device:<br>Crosstex® Procedural Earloop Face<br>Mask (K082258) | Candidate Device:<br>Aurelia Surgical Mask Lv. 2 | Conclusion |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------------------------------|
| Product Code | FXX | FXX | Same |
| Regulation<br>Number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Fluid Resistance | 120mmHg | 120mmHg | Same |
| Particulate Filtration<br>Efficiency | 98% | 99.6% | Different |
| Bacterial Filtration<br>Efficiency | 98% | 99.8% | Different |
| Differential Pressure | <5.0 mm H2O/cm² | 4.1 mm H2O/cm² | Different |
| Flammability | Class 1 | Class 1 | Same |
| Biocompatibility | | | |
| Cytotoxicity | No Cytotoxicity | No Cytotoxicity | Same |
| Sensitization | No Sensitization | No Sensitization | Same |
| Irritation | No Irritation | No Irritation | Same |
| Test | Purpose | Acceptance<br>Criteria<br>per ASTM F2100-<br>19 | Results |
| Fluid Resistance<br>(ASTM F1862) | FXX This test is performed<br>to<br>Evaluate Personal<br>Protective Equipment<br>against fluid penetration.<br>An arterial spray is<br>simulated to test the<br>PPE. | Level1:<br>80mmHg | Aurelia Mask Level 1:<br>80mmHg |
| | | Level 2:<br>120mmHg | Aurelia Mask Level 2:<br>120 mmHg |
| Differential Pressure<br>(Delta-P)<br>(ASTM F2100-19<br>EN 14683:2019) | The purpose of this test<br>is to ensure the drop<br>in pressure is not too<br>great so the user may<br>breathe through the<br>subject device. | Level1:<br>< 5.0 mm H2O/cm | Aurelia Mask Level 1:<br>< 4.5 mm H2O/cm² |
| | | Level 2:<br>< 6.0 mm H2O/cm² | Aurelia Mask Level 2:<br>< 4.1 mm H2O/cm² |
| Bacterial Filtration<br>Efficiency (BFE)<br>(ASTM F2101) | The BFE test is<br>performed to determine<br>the bacterial filtration<br>efficiency of test articles. | Level1:<br>≥ 95% | Aurelia Mask Level 1:<br>≥ 99.7% |
| | | Level 2:<br>≥ 98% | Aurelia Mask Level 2:<br>≥ 99.7% |
| Particulate<br>Filtration Efficiency<br>(PFE)<br>(ASTM F2299) | The PFE test is<br>performed to evaluate<br>the non-viable particle<br>filtration efficiency of the<br>test article. | Level1:<br>≥ 95% | Aurelia Mask Level 1:<br>≥ 99.5% |
| | | Level 2:<br>≥ 98% | Aurelia Mask Level 2:<br>≥ 99.6% |
| Flammability<br>(16 CFR 1610) | The purpose of this test<br>is to ensure the subject<br>device does not ignite<br>when exposed to flame. | Level1:<br>Class 1 | Aurelia Mask Level 1:<br>Class 1 |
| | | Level 2:<br>Class 1 | Aurelia Mask Level 2:<br>Class 1 |
| Biocompatibility | Test Method/ Standard | Acceptance Criteria | Results |
| Cytotoxicity | ISO 10993-5 | Non- Cytotoxicity | Pass |
| Sensitization | ISO 10993-10 | Non- Sensitization | Pass |
| Irritation | ISO 10993-10 | Non- Irritating | Pass |
# Performance Testing Summary
Non-clinical performance tests and Biocompatibility were conducted to verify that the Aurelia Surgical Masks met or exceeded all the design specifications as the predicate device K082258.The results were indicated in the Predicate Comparison Section above and restated below.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Dicentra logo. The logo features the word "dicentra" in a stylized font, with the "d" in purple and the rest of the letters in black. To the left of the word is a purple sphere made up of interconnected dots. Below the word "dicentra" are the words "GUIDANCE | INNOVATION | TRUST" in smaller, black font.
## Table 5: Summary of Performance Testing
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for "dicentra". The logo features the word "dicentra" in a stylized font, with the "d" in purple and the rest of the letters in black. To the left of the "d" is a purple graphic that resembles a network of interconnected nodes. Below the word "dicentra" are the words "GUIDANCE | INNOVATION | TRUST" in a smaller font.
# Shelf-Life
The proposed shelf life for the Aurelia Surgical masks is 5 years. An accelerated aging test was performed to support this claim and a real time aging study has been underway since December 2021.
The Accelerated Aging Test was initiated on 10th January 2022 and completed on 16th September 2022. The test was conducted in compliance with USFDA regulations (21 CFR Parts 58, 210, 211, and 820). 60 boxes/samples were subjected to an incubation temperature of 50°C ± 20°C for 229 days (5 years simulated shelf life)
The test report does not specify any failures of the packaging or loss of sterile integrity after the accelerated aging. Therefore, the results indicate the packaging, maintained integrity and sterile barrier properties up to 5 years of simulated shelf life under the test conditions.
## Conclusion
The Aurelia Surgical masks are substantially equivalent to the predicate device [K082258]. The design and technological characteristics of the devices are the same and the proposed device meets or exceeds the efficacy and safety tests results of the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.