K230950 · Siemens Medical Solutions USA, Inc. · OWB · Dec 14, 2023 · Radiology
Device Facts
Record ID
K230950
Device Name
ARTIS icono (VE30A)
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Dec 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ARTIS is a family of dedicated angiography systems developed for single-plane and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angtography, rotational angiography, multipurpose angiography, and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image-guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions. The ARTIS systems can also support the acquisition of position triggered imaging for spatial data synthesis. The ARTIS systems include also the software option DynaCT with following indications for use: DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures, and treatment follow-up.
Device Story
ARTIS icono (VE30A) is a multi-axis motorized C-arm angiography system for diagnostic and interventional imaging. Inputs: X-ray signals captured via flat panel detector. Processing: High-voltage generator powers X-ray tube; system performs 2D/3D image reconstruction and post-processing. Output: Real-time fluoroscopic and radiographic images displayed on ceiling-mounted monitors. Used in OR/clinic by physicians/technicians; tableside control via footswitch and console. Supports image fusion and navigation. Benefits: Enables precise catheter navigation and visualization of internal structures for surgical planning and treatment follow-up. Modifications in VE30A include updated software, increased memory, new 'Structure Scout' elements (Calcium, Gadolinium, Bismuth), and enhanced usability features (e.g., editing case flow, user profiles, display layouts).
Clinical Evidence
Bench testing only. No clinical data. Verification and validation testing included software functional testing, system validation, image quality assessment (CNR, exposure control), and testing of new 'Structure Scout' elements. All results met acceptance criteria.
Indicated for diagnostic imaging and interventional procedures, including cardiac, neuro, general, rotational, and multipurpose angiography, and whole-body radiographic/fluoroscopic procedures. Suitable for pediatric and obese patients. Includes image-guided surgery and 3D reconstruction (DynaCT) for hard/soft tissue imaging. Not for projection radiography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
December 14, 2023
Siemens Medical Solutions USA, Inc. % Patricia Jones Regulatory Affairs Professional 40 Liberty Boulevard MALVERN, PA 19355
Re: K230950
Trade/Device Name: ARTIS icono (VE30A) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, IZI, JAA, JAK Dated: November 30. 2023 Received: December 1, 2023
Dear Patricia Jones:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230950
Device Name ARTIS icono (VE30A)
#### Indications for Use (Describe)
ARTIS is a family of dedicated angiography systems developed for single-plane and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angtography, rotational angiography, multipurpose angiography, and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography.
Additional procedures that can be performed include angiography in the operating room, image-guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.
The ARTIS systems can also support the acquisition of position triggered imaging for spatial data synthesis.
The ARTIS systems include also the software option DynaCT with following indications for use:
DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures, and treatment follow-up.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
# 510(k) Summary: ARTIS icono (VE30A) 510(K) #: K230950
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65 Malvern, PA 19355
Date Prepared: December 7, 2023
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
- 1. General Information: Importer / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
#### Manufacturing Site:
Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335
#### Contact Person: 2.
Ms. Patricia D. Jones Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Phone: (678) 575-8832 Email: patricia.jones@siemens-Healthineers.com
#### Device Name and Classification: 3.
Trade Name: Classification Name: Classification Panel: Requlation Number: Device Class: Product Codes:
### ARTIS icono (VE30A)
Image-intensified fluoroscopic X-ray System Radiology 21 CFR §892.1650 Class II OWB, IZI, JAA, JAK
- Legally Marketed Primary Predicate Device: 4. Trade Name: ARTIS icono (VE21) 510(k) Clearance K220432 Clearance Date June 29, 2022 Classification Name: Image-intensified fluoroscopic X-ray System Classification Panel: Radiology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
Requlation Number: Device Class: Product Code: Subsequent Product Codes: 21 CFR §892.1650 Class II OWB IZI, JAA, JAK
Total Product Life Cycle:
All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues.
#### 5. Device Description:
The ARTIS icono (VE30A) is a medical device that allows visualization of vessels within the human body. It is of the utmost importance to find the right projections so physicians can navigate catheters and other devices safely. The ARTIS icono (VE30A) consists of a patient table and a multi-axis motorized c-arm that can be positioned around the patient and angulated in a double-oblique fashion iso-centering the region of interest between the x-ray tube and the flat panel detector. The X-ray generator is placed separately. The displays for visualizing the X-ray images are mounted at the ceiling with a movable display suspension system. System operation is executed via control modules table side so that the physician can move and position the table and c-arm adequately for best imaging while manipulating catheters or other devices during the x-ray. X-ray release is tableside via a footswitch.
The ARTIS icono (VE30A), modular angiography systems are designed as sets of components that may be combined into two different configurations (Biplane or Floor) to provide specialized angiography systems. In general, they are equipped with a Carm, stand, flat panel detector, x-ray tube, collimator, high voltage generator, patient table, and image post-processing.
The ARTIS icono (VE30A) covers the complete range of angiographic applications, cardiac angiography, neuro-angiography, general angiography, surgery and surgical multipurpose rotational angiography, angiography, angiography. radiographic/fluoroscopic procedures.
The following components are configured to create a Floor or Biplane configuration: (1) Floor stand with C-arm, X-ray tube assembly, and FD
- (2) Patient table
- (3) Display ceiling suspension with displays
- (4) Footswitch for releasing radiation
- (5) Control console for controlling the stand, patient table, and imaging system
Images and operating elements are displayed on screens. Depending on the ARTIS icono (VE30A) configuration, different display variants are used to visualize image and information content. Displays that visualize single images or large displays that are configurable to visualize multiple images and information contained in various layouts are used.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
Post-processing can be done in the exam room or in the control room that offers monitors as well, with a footswitch location in the exam room or the control room. The ARTIS icono (VE30A) is capable of 2D and 3D imaging. The c-arms can be mounted on the floor or for biplane systems on the floor and the ceiling. Other systems and software syngo Application Software, synqo X Workplace, Sensis, and or third-party systems may also be integrated into the ARTIS icono (VE30A) screen configuration. Different screen configurations and layouts are possible in the examination and control rooms.
The Subject device ARTIS icono with software version VE30A supports the following modifications.
| Software / Hardware changes specific to New System Software VE30A | |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| 1. | Updated system Software from VE21A to VE30A |
| A. | Added an optional Second Workplace |
| B. | Updated memory data storage to 600K |
| C. | Updated of Roadmap Phase 3 |
| D. | Added new elements for Organ Program "Structure Scout": Calcium, Gadolinium, and Bismuth |
| E. | Updated Usability Features: |
| | 1) Updated Case Flow feature to include Editing |
| | 2) Updated Body Region to include Editing |
| | 3) Updated Procedure Name to include Editing |
| | 4) On Screen Display Reconfiguration |
| | 5) Updated Favorites on the Pilot Module to include protocols, frame, and pulse rates |
| | 6) Updated Direct Access Bar on the Pilot Module to include Editing |
| | 7) Updated Favorite on the Pilot Module to include List & Icon View |
| | 8) Updated ARTIS Profiles to include User Profile |
| | 9) Updated Pilot Module to lock |
| | 10) Updated Image Directory to include Custom Filter possible as default |
| | 11) Additional Display Layouts for Cockpit |
| | 12) Added "Inactive" as a function for Dynamic Viewer during PT registration and after system restarts. |
| 2. | New PC for imaging System due to obsolescence |
Table 1: Modification for ARTIS icono (VE30A)
### 6. Indications for Use:
ARTIS is a family of dedicated angiography systems developed for single-plane and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography, and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e., patient extremities. This does not include projection radiography.
Additional procedures that can be performed include angiography in the operating room, image-guided surgery by X-ray, by image fusion, and by navigation systems.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
The examination table as an integrated part of the system can be used for X-ray imaging, surgery, and interventions.
The ARTIS systems can also support the acquisition of position triggered imaging for spatial data synthesis.
The ARTIS systems include also the software option DynaCT with following indications for use:
DynaCT is an X-ray imaging software option, which allows the reconstruction of twodimensional images acquired with a standard angiographic C-arm device into a threedimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures, and treatment follow-up.
#### Substantial Equivalence: 7.
The ARTIS icono (VE30A) is substantially equivalent to the legally marketed predicate listed in the table below:
| Predicate Device Name<br>and Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties |
|----------------------------------------------------|------------------|-------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Primary Predicate<br>ARTIS icono (VE21)<br>Siemens | K220432 | 06/29/2022 | Indications for use Software Version VE21 PC Usability Features Workplace Station Data Memory Storage Roadmap Feature Structure Scout |
### Table 2: Predicate Device Comparable Properties for Subject Device Modifications
#### Summary of Technological Characteristics of the Subiect Device as Compared 8. with the Predicate Device:
The ARTIS icono (VE30A) is designed as a set of components (C-arm, X-ray tube, housing, flat panel detector, digital imaging system, collimator, generator, etc.) that may be combined into two different configurations (Floor & Biplane to provide specialized angiography systems. Components used with ARTIS icono (VE21) are either commercially available with current Siemens systems or include modifications to existing components. Technological differences between the Subject Device and the Predicate Device are provided in Table 3 below for all modifications.
### Table 3: Summary of Comparison of Technological Characteristics
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a graphic of orange dots arranged in a circular pattern.
| Modifications | Subject Device<br>ARTIS icono (VE30A) | Predicate Device<br>ARTIS icono (VE21A)<br>K220432 | Comparison Results |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| New System<br>Software/<br>Hardware<br>Changes | 1. Updated system Software from VE21A to VE30A | System software version VE21 | Comparable:<br>System software VE30A was updated to support modifications 1A-2. All System Software modifications conform to "Guidance for the content of Premarket submission for software in Medical Devices" |
| | A. Added an optional 2nd Workplace | ARTIS icono offered only 1 Workplace | |
| | B. Updated memory data storage to 600K | Memory data storage is 400K | |
| | C. Updated Roadmap Phase 3 | Roadmap with 4 phases | These software and hardware changes do not raise any new risks or any issues regarding the safety or effectiveness of the device. |
| | D. Added new elements for Organ Program "Structure Scout": Calcium, Gadolinium, and Bismuth | Structure Scout cleared elements are platinum, tantalum, iron, iodine, steel and nitinol, and CO2 | |
| | E. Updated Usability Features: | | |
| | 1) Updated Case Flow feature to include Editing | Editing of the Case Flow was not available. | All test results met all acceptance criteria. |
| | 2) Updated Body Region to include Editing | Editing of the Body Region was not available. | |
| | 3) Updated Procedure Name to include Editing | Editing of the Procedure Name was not available. | |
| | 4) On Screen Display Reconfiguration | Reconfiguring of the On-Screen Display was not available. | |
| | 5) Updated Favorites on the Pilot Module to include protocols, frame and pulse rates | Favorites on PM did not include protocols, frame and pulse rates. | |
| | 6) Updated Direct Access Bar on the Pilot Module to include Editing | Editing of the Direct Access Bar was not available. | |
| | 7) Updated Favorite on the Pilot Module to include List & Icon View | Favorites on PM with list & Icon View was not available. | |
| | 8) Updated ARTIS Profiles to include User Profile | ARTIS Profiles to include user Profile was not available. | |
| | 9) Updated Pilot Module to lock | Locking PM was not available. | |
| | 10) Updated Image Directory to | Image Directory to include Custom Filters as default | |
| Modifications | Subject Device<br>ARTIS icono (VE30A) | Predicate Device<br>ARTIS icono (VE21A)<br>K220432 | Comparison<br>Results |
| | include Custom<br>Filter possible<br>as default | was Body Region was not<br>available. | |
| | 11) Additional<br>Display<br>Layouts for<br>Cockpit | Additional Display Layouts<br>for cockpits were not<br>available. | |
| | 12) Added<br>"Inactive" as a<br>function for<br>Dynamic<br>Viewer during<br>PT registration<br>and after<br>system<br>restarts. | Dynamic Viewer was Active<br>during PT registration and<br>after system restarts.<br>Inactive was not available. | |
| New PC | 2. New PC for Imaging<br>System due to<br>obsolescence | PC for Imaging System | |
| | 3. NFJ- 0157: Modifications<br>for the ARTIS icono (floor &<br>biplane) for inclusion of the<br>new ARTIS Transfer<br>Solution in combination<br>with MR Nexaris Dockable<br>Table for system<br>MAGNETOM Vida / Sola | NFJ- 0157:<br>Modifications<br>for the ARTIS icono (floor &<br>biplane) for inclusion of the<br>new ARTIS Transfer<br>Solution in combination<br>with MR Nexaris Dockable<br>Table for system<br>MAGNETOM Vida / Sola | |
| | 4. NFJ #0163-Introduction of<br>optional AVVIGO™<br>Guidance System II to the<br>Artis Family of Medical<br>Devices. The AVVIGO<br>Guidance System II is a<br>3rd Party 510(k) cleared<br>device (K212490) that is<br>integrated (connected)<br>with the ARTIS system | NFJ #0163-Introduction of<br>optional AVVIGO™<br>Guidance System II to the<br>Artis Family of Medical<br>Devices. The AVVIGO<br>Guidance System II is a 3rd<br>Party 510(k) cleared device<br>(K212490) that is integrated<br>(connected) with the ARTIS<br>system family | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
#### 9. Nonclinical Performance Testing:
Non-clinical tests were conducted for the ARTIS icono (VE30A) during product development. The following non-clinical testing was conducted on the presented modifications and relied on: Software functional, verification, and System validation testing with passing results. Configuration, DSA Roadmap, Subtracted Fluoro and transfer images testing was conducted in support of the Updated roadmap Phase 3. CNR image Quality exposure control, fluoro auto exposes values cardiology, DR, CNR, DSA, and neuro testing were conducted for Structure Scout new elements for the Organ program. All testing results passed.
The ARTIS icono (VE30A) was certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance and Electromagnetic Compatibility:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
- AAMI ANSI ES60601-1:2005/(R)2012 ●
- IEC 60601-1-2:2020 ●
- IEC 60601-1-3:2013 ●
- IEC 60825-1:2014 ●
- TR 60878:2015
- IEC 62304:2015 ●
- IEC 80001-1:2010 ●
- IEC 60601-2-28:2017 ●
- IEC 60601-2-43:2019 ●
- IEC 60601-2-54:2018 ●
- ISO 10993-1:2018 ●
- ISO 14971:2019 ●
- ISO 15223-1:2021 ●
- ISO 17664:2018
- ISO 17664-2:2021
- IEC 60601-2-35:2020 ●
- IEC 62366-1:2020 ●
- IEC 62563-1:2021 ●
- NEMA PS 3.1:2022 .
- ANSI UL 2900-1:2017
- ANSI UL 2900-2-1:2017 ●
- AAMI TIR57:2016 ●
- ISO 14155:2020
### Verification and Validation:
Non-clinical tests were conducted on ARTIS icono System software (VE30A) during product development. The modifications described in this Premarket Notification are supported with verification and validation testing.
The Risk analysis was completed, and risk control was implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
ARTIS icono (VE30A) was tested and found to be safe and effective for intended users, uses, and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual. Customer employees are adequately trained in the use of this equipment.
Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse, or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
IEC 80001-1-2010 is the hospital. Provided in the Software Section is the required cybersecurity information.
## Summary:
Performance tests were conducted to test the functionality of the ARTIS icono (VE30A). These tests have been performed to assess the functionality of the subject device. The results of all conducted testing and clinical assessments were found acceptable and did not raise any new issues of safety or effectiveness.
### 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device safely and effectively.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practices, and all equipment is subject to final performance testing. Furthermore, the operators are healthcare professionals who are trained and responsible for evaluating the postprocessing of X-ray images.
## 11. Conclusion as to Substantial Equivalence:
The predicate device was cleared based on non-clinical supportive information. Nonclinical test results demonstrate that the ARTIS icono (VE30A) acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data, and software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.