K230948 · Verathon Medical (Canada) Ulc · EOQ · May 3, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K230948
Device Name
BFlex™ 2 Slim 3.8 Single-Use Bronchoscope
Applicant
Verathon Medical (Canada) Ulc
Product Code
EOQ · Ear, Nose, Throat
Decision Date
May 3, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
BFlex™ 2 Slim 3.8 Single-Use Bronchoscope is intended to work with the video monitor, in conjunction with nonpowered endoscopic accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
Device Story
Single-use flexible bronchoscope; part of BFlex™ 2 system including reusable monitor and cable. Captures real-time airway video via distal tip camera/LED light source; transmits to monitor. Used in hospital environments by pediatric anesthesiologists, adult intensivists, or pulmonologists. Facilitates airway visualization and procedure guidance; supports use of non-powered endoscopic accessories (e.g., biopsy tools) via working channel. Provides clinicians with direct visual feedback for airway management; aids in diagnostic or therapeutic interventions. Benefits include reduced cross-contamination risk due to single-use design and improved procedural efficiency.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and usability validation. Usability study included 30 users (15 pediatric anesthesiologists, 15 adult intensivists/pulmonologists) across 13 simulated scenarios, achieving a 99.97% task success rate.
Technological Characteristics
Flexible endoscope; 3.8mm outer diameter; 1.2mm working channel. Features: LED illumination, 0° view, 120° diagonal FOV, 640x480 resolution. Materials: flexible plastic. Power: rechargeable lithium-ion (via system). Sterilization: Ethylene Oxide (EO). Standards: ES60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 8600-1/3/4, ISO 10993-1.
Indications for Use
Indicated for endoscopy within the airways and tracheobronchial tree in adults and pediatric patients (6 years and older).
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
BFlex™ 2 Slim 3.8 Single-Use Bronchoscope System (K193488)
BFlex™ 2 Ultraslim 2.8 Single-Use Bronchoscope System (K211947)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 3, 2023
Verathon Medical (Canada) ULC Chauneen Wood Director, RA 2227 Douglas Road Burnaby, B.C. V5C 5A9 Canada
Re: K230948
Trade/Device Name: BFlex™ 2 Slim 3.8 Single-Use Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: April 1, 2023 Received: April 4, 2023
Dear Chauneen Wood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230948
Device Name BFlex™ 2 Slim 3.8 Single-Use Bronchoscope
Indications for Use (Describe)
BFlex™ 2 Slim 3.8 Single-Use Bronchoscope is intended to work with the video monitor, in conjunction with nonpowered endoscopic accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6 2.4L5.6 10.4L2.4 7.2L0 9.6L5.6 15.2L16 4L13.6 2.4Z" fill="black"></path> </svg> </span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
This summary of Safety and Effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR 80792.
#### Submitter:
Verathon Medical (Canada) ULC 2227 Douglas Road Burnaby, BC V3W 1P2 Canada
#### Contact Person:
Chauneen Leah Wood M.S., RAC Director, Regulatory Affairs Phone: (425) 419-4114 Email: chauneen.wood@verathon.com
### Date Summary Prepared:
March 31, 2023
#### Establishment Registration Number:
Verathon Medical (Canada) ULC Registration Number: 9615393 Owner/Operator Number: 9095489
### Device Trade or Proprietary Name:
BFlex™ 2 Slim 3.8 Single-Use Bronchoscope
### Device Common or Usual Name:
Flexible Bronchoscope
| Device Trade or Proprietary Name | Device Common or Usual Name |
|-------------------------------------------|-----------------------------|
| BFlex™ 2 Slim 3.8 Single-Use Bronchoscope | Flexible Bronchoscope |
#### Device Classification:
| Device Panel | Ear, Nose, and Throat Devices |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (Medical specialty) | Subpart E - Surgical Devices |
| Classification Regulation | 21 CFR 874.4680 |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Identification | A bronchoscope (flexible or rigid) and accessories is a tubular<br>endoscopic device with any of a group of accessory devices which<br>attach to the bronchoscope and is intended to examine or treat the<br>larynx and tracheobronchial tree. It is typically used with a fiberoptic |
{4}------------------------------------------------
| Device Panel<br>(Medical specialty) | Ear, Nose, and Throat Devices<br>Subpart E – Surgical Devices |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Regulation | 21 CFR 874.4680 |
| | light source and carrier to provide illumination. The device is made of<br>materials such as stainless steel or flexible plastic. This generic type of<br>device includes the rigid ventilating bronchoscope, rigid<br>nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-<br>bronchial telescope, flexible foreign body claw, bronchoscope tubing,<br>flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush,<br>rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope<br>aspirating tube, but excludes the fiberoptic light source and carrier. |
| Product Code | EOQ |
| Classification | Class II |
### Predicate Device:
BFlex™ 2 Slim 3.8, and Ultraslim 2.8 Single-Use bronchoscopes are the legally market devices to which BFlex™ 2 Slim 3.8 Single-Use Bronchoscope claims substantial equivalence. The proposed BFlex™ 2 Slim 3.8 Single-Use Bronchoscope has the same intended use, and technological characteristics of the predicate. The predicates identified along with their 510(k) clearance numbers and respective clearance dates included in the table below:
| Predicate Device | 510(k) Number | Clearance Date |
|-------------------------------------------------------|---------------|-------------------|
| BFlex™ 2 Slim 3.8 Single-Use Bronchoscope System | K193488 | January 16, 2020 |
| BFlex™ 2 Ultraslim 2.8 Single-Use Bronchoscope System | K211947 | November 03, 2021 |
### Device Description:
The BFlex™ 2 Slim 3.8 Single-Use Bronchoscope is a component to the BFlex™ Single-Use Bronchoscope System. The BFlex™ Single-Use Bronchoscope system consists of a single-use flexible bronchoscope, a reusable monitor, and a reusable cable. The BFlex™ 2 Single-Use Bronchoscope System provides real time viewing and recording for a wide range of airway procedures.
### Indication for Use:
The BFlex™ 2 Slim 3.8 Single-Use Bronchoscope is intended to work with a video monitor, in conjunction with non-powered endoscopic accessories and other ancillary equipment, for endoscopy within the airways and tracheobronchial tree.
### Intended Patient Population:
The BFlex™ 2 Single-Use system is for use in a hospital environment. The BFlex™ 2 bronchoscope is a single-use device designed for use in adults, with the BFlex™ 2 Ultraslim 2.8 and BFlex™ Slim 3.8 also designed for pediatric use (Ultraslim 2.8-6 months to 6 years, Slim 3.8-6 years). It has been verified and validated for the following endotracheal tube (ETT) and endoscope accessory (EA) sizes:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "verathon" in a sans-serif font. The word is written in a light blue color. The letters are all lowercase.
| Model | Minimum ETT Internal<br>Diameter | EA Minimum Working Channel<br>Width |
|------------------------|----------------------------------|-------------------------------------|
| BFlex™ 2 Ultraslim 3.8 | 4.0 mm | - |
| BFlex™ 2 Slim 3.8 | 5.0 mm | 1.2 mm |
Note: There is no guarantee that instruments selected solely using these instrument dimensions will be compatible in combination.
#### Technological Characteristics:
The proposed BFlex™ 3.8 Single-Use Bronchoscope when compared to the predicate bronchoscopes has similar technological characteristics. See the comparison table below for similarities and differences between the proposed and predicate devices:
#### Comparison Table
| Technological Characteristic | BFlex™ 2 Ultraslim<br>2.8 Single-Use<br>Bronchoscope<br>(Predicate) | BFlex™ 2<br>Slim<br>3.8 Single-Use<br>Bronchoscope<br>(Predicate) | BFlex™ 2<br>Slim<br>3.8 Single-Use<br>Bronchoscope<br>(Proposed) |
|------------------------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------|
| Flexible Endoscope | Yes | Yes | Yes |
| Size Distinguishing Color<br>(Non-patient contacting) | Teal | Purple | Purple |
| Outside Diameter of Flexible<br>Insertion Tube/Shaft and Distal<br>Tip | 2.8mm | 3.8mm | 3.8mm |
| Minimum Internal Diameter of<br>Working Channel | Not Applicable | 1.2mm | 1.2mm |
| Suction Port | Not Applicable | Yes | Yes |
| Accessory Port | Not Applicable | Yes | Yes |
| Single Use Bronchoscope | Yes | Yes | Yes |
| Sterility | Sterile by Ethylene<br>Oxide (EO) | Sterile by Ethylene<br>Oxide (EO) | Identical |
| Control Button for Tip<br>maneuverability | Yes | Yes | Yes |
| Power Source | Rechargeable<br>Lithium-ion | Rechargeable<br>Lithium-ion | Identical |
| Camera | Yes | Yes | Yes |
| Direction of View, Relative to<br>Center Line of Distal tip | 0° | 0° | Identical |
{6}------------------------------------------------
| Technological Characteristic | BFlex™ 2 Ultraslim<br>2.8 Single-Use<br>Bronchoscope<br>(Predicate) | BFlex™ 2<br>Slim<br>3.8 Single-Use<br>Bronchoscope<br>(Predicate) | BFlex™ 2<br>Slim<br>3.8 Single-Use<br>Bronchoscope<br>(Proposed) |
|------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------|
| Field of view, horizontal/vertical | 85° | Identical | Identical |
| Field of View, diagonal | 120° | Identical | Identical |
| Depth of Field | 5-50mm | Identical | Identical |
| Image Resolution | 640x480 | 640x480 | Identical |
| LED Light Source | Yes | Yes | Yes |
| Image Display | Displays image on a<br>Reusable Video<br>Monitor | Displays image on<br>a Reusable Video<br>Monitor | Identical |
| Extended Viewing | Yes | Yes | Yes |
# Performance Testing:
Performance testing has been completed to demonstrate that the proposed BFlex™ 2 Slim 3.8 Single-Use Bronchoscope meets the safety and performance requirements established in the design specifications. Comprehensive verification and validation testing included the following:
- Full System Requirements Testing
- Electrical Safety according to
- ES60601-1:2005/(R)2012 and A1:2012 Medical electrical equipment Part 1: O General requirements for basic safety and essential performance
- IEC 60601-2-18: Edition 3.0 2009-08 Particular requirements for the basic safety o and essential performance of endoscopic equipment
- Electromagnetic Compatibility according to
- IEC 60601-1-2:2014 Ed.4.0 Medical electrical equipment Part 1-2: General O requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- Optical testing according to ●
- ISO 8600-1:2015 (Ed. 4.0) O Endoscopes - Medical endoscopes and endotherapy devices -- Part 1: General requirements
- ISO 8600-3:2019 (Ed. 2.0) O
{7}------------------------------------------------
# verathon
Optics and Optical instruments - Medical endoscopes and endoscopic accessories - Part 3: Determination of field of view and direction of view of endoscopes with optics
- ISO 8600-4:2014 (Ed. 2.0) o Endoscopes - Medical endoscopes and certain accessories - Part 4: Determination of maximum width of insertion portion
- Biocompatibility according to ●
- ISO 10993-1:2018 о
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Sterile Packaging Integrity Testing ●
- Cleaning Testing ●
- Design Validation
- Usability testing for the proposed BFlex 3.8 Single Use Bronchoscope for use with ● pediatric patients was conducted as part of the validation testing for the BFlex 2 family (all sizes). The usability study performed employed the same methods as was used for validation of the GlideScope BFlex 2.8 Single-Use Bronchoscope (K211947). The usability testing represents aa well-established method according to Section C of the "The Special 510(k) Program", Guidance for Industry ad Food and Drug Administration Staff (September 13, 2019).
This study was conducted with two (2) user groups with thirty (30) representative end users. Primary users included one group of fifteen (15) pediatric anesthesiologists and one group of fifteen (15) adult intensivists or pulmonologists. The assessment was accomplished through thirteen (13) simulated use scenarios with pediatric Anesthesiologist and adult Intensivist/Pulmonologists.
Overall performance using the BFlex 2 system was excellent with 99.97% task success rate for pediatric Anesthesiologists and adult Intensivists or Pulmonologists. The pertinent user and system requirements and results (Table L-5) associated with the pediatric claims for the proposed BFlex 2 Slim 3.8 Single-Use Bronchoscope device can be found in Section L, Table L-5 of the Special 510(k) application. Study data demonstrated based on observed participate performance and subjective assessments, that BFlex 2 bronchoscopes can be used by the intended users, in the intended use environment without patterns of crucial use error or unacceptable negative use-safety assessments.
Results: All testing resulted in acceptance criteria passed.
{8}------------------------------------------------
# verathon
## Summary of Clinical Tests:
The Clinical testing was not performed as non-clinical studies were sufficient to support substantial equivalence.
## Conclusion:
The information in this 510(k) Premarket Notification demonstrates that the proposed BFlex™ 2 Slim 3.8 Single-Use Bronchoscope is substantially equivalent to the previously cleared predicate BFlex™ 2 Slim 3.8 and Ultraslim 2.8 Single-Use Bronchoscopes with respect to safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.