K230947 · Vivos Therapeutics, Inc. · LRK · Nov 28, 2023 · Dental
Device Facts
Record ID
K230947
Device Name
C.A.R.E. Appliance (DNA, mRNA, mmRNA)
Applicant
Vivos Therapeutics, Inc.
Product Code
LRK · Dental
Decision Date
Nov 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K230947 · Nov 28, 2023
C.A.R.E. Appliance (DNA, mRNA, mmRNA)
Vivos Therapeutics, Inc.
Research database (2018-2023)
Retrospective analysis of real-world clinical data from a research database was used to demonstrate the efficacy of C.A.R.E. appliances in treating moderate and severe obstructive sleep apnea, including patients treated with the appliance alone or in combination with PAP/myofunctional therapy.
Obstructive Sleep Apnea; Retrospective analysis; AHI reduction; Real-world clinical data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Real World Clinical Data Analysis; Retrospective cohort analysis; Follow-up/Duration: At least six months between pre- and post-sleep studies; Study Period: 2018-2023
Adults 18+ with moderate (15≤AHI<30) or severe (AHI ≥ 30) OSA treated with C.A.R.E. appliances; Sample Size: 108 patients (73 severe, 35 moderate)
Not applicable for this study
Change in AHI, Transpalatal Width, and OSA classification improvement
Indications for Use
The C.A.R.E. Appliances are intended to reduce nighttime snoring and to treat mild and moderate obstructive sleep apnea in adults, 18 years of age and older. The C.A.R.E. Appliances are also intended to treat moderate and severe obstructive sleep apnea (OSA) in adults, 18 years of age and older along with positive airway pressure (PAP) devices and/or myofunctional therapy, as needed.
Device Story
C.A.R.E. appliances (DNA, mRNA, mmRNA) are custom-fabricated oral devices consisting of upper and lower trays; used to reduce snoring and treat OSA. Devices are fitted by dentists for home use. Principle of operation involves jaw expansion and mid-facial redevelopment to increase nasal cavity volume and oropharyngeal airway patency; mRNA and mmRNA models additionally utilize mandibular advancement. DNA model lacks tray connection; mRNA/mmRNA use flange/hinge connections. Devices feature adjustment mechanisms for antero-posterior, transverse, and vertical dimension of occlusion (VDO) customization. Optional back-of-device extender prevents airway collapse. Clinical benefit includes reduced apnea-hypopnea index (AHI) and improved airway patency. Healthcare providers use clinical assessment and sleep study results to guide treatment; device output is the physical maintenance of airway patency during sleep.
Clinical Evidence
Evidence includes a randomized, blinded study (n=15) and real-world data (RWD) (2018-2023). Randomized study showed 64% combined AHI reduction. RWD (n=108) for moderate/severe OSA patients showed 80% of patients improved by at least 45% or one classification. Mean AHI reduction was 50.8% for severe and 44.4% for moderate cohorts. 10% of patients experienced inadvertent tooth movement/bite changes managed with orthodontics.
Technological Characteristics
Custom-fabricated oral appliances using standard orthodontic acrylics and wires. Features include upper/lower trays, adjustment mechanisms for 6 degrees of freedom (AP, TV, VDO), and optional airway extenders. DNA model is non-connected; mRNA/mmRNA use flange/hinge connections. Non-sterile, custom-manufactured per patient dental models. No software or electronic components.
Indications for Use
Indicated for adults 18+ with snoring, mild/moderate obstructive sleep apnea (OSA), or moderate/severe OSA when used adjunctively with PAP devices or myofunctional therapy.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
November 28, 2023
Vivos Therapeutics, Inc % Colette Cozean Regulatory Consultant Eyedeas Company 21581 Midcrest Dr Lake Forest, California 92630
Re: K230947
Trade/Device Name: C.A.R.E. Appliance (DNA, mRNA, mmRNA) Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK, LQZ Dated: November 26, 2023 Received: November 27, 2023
Dear Colette Cozean:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230947
Device Name
C.A.R.E. Appliances (DNA, mRNA, mmRNA)
Indications for Use (Describe)
The C.A.R.E. Appliances are intended to reduce nighttime snoring and to treat mild and moderate obstructive sleep apnea in adults, 18 years of age and older. The C.A.R.E. Appliances are also intended to treat moderate and severe obstructive sleep apnea (OSA) in adults, 18 years of age and older along with positive airway pressure (PAP) devices and/or myofunctional therapy, as needed.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 5. 510(k) Summary - K230947
| Applicant: | Vivos Therapeutics<br>300 S. 5th Street<br>Murray, KY, 42017, USA<br>(270)-226-9237 |
|---------------------|-------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Colette Cozean, PhD<br>21581 Midcrest Drive<br>Lake Forest, CA<br>92630 (949) 855-2885<br>colettecozean@gmail.com |
| Date Prepared | November 18, 2023 |
| Proprietary Name | C.A.R.E. Appliances (DNA, mRNA, mmRNA) |
| Common Name | Dental Device - Anti Snoring /Obstructive Sleep Apnea Device |
| Classification Name | Anti Snoring/ Obstructive Sleep Apnea Device |
| Product Code | LRK, LQZ |
| Primary Predicate | DNA (K222872) |
| Reference Device | mRNA (K130067) |
| Reference Devices | mmRNA (K210203) |
# Device Description
The C.A.R.E. appliances are intended to reduce nighttime snoring and to treat mild and moderate obstructive sleep apnea in adults, 18 years of age and older. The C.A.R.E. appliances are also intended to treat moderate and severe obstructive sleep apnea (OSA) in adults, 18 years of age and older along with positive airway pressure (PAP) devices and/or myofunctional therapy, as needed. They consist of an upper tray and a lower tray and is designed to open the airway during sleep. The device is customized to each patient, and features an adjustment mechanism to allow it to be further customized to each patient.
The DNA appliance does not connect the upper and lower trays, while the mRNA and mmRNA appliances connect the trays with a flange and hinge respectively.
The devices are identical to the already-marketed predicates of the same name.
{5}------------------------------------------------
### Scientific Principles
During sleep, the muscles in the tongue and back of the throat relax, which can cause them to sag and narrow the airway. Airflow through a narrow airway is the cause of snoring. When this narrowing of the airway is severe, it results in Obstructive Sleep Apnea (OSA), where the airway closes. This can happen up to hundreds of times during the night, lasting for a minute or longer.
With these closures, the brain detects the lack of oxygen and disturbs sleep to draw breath. In many cases, the individual isn't completely aware of the stoppages, which don't fully awaken the sleeper. Sleep apnea has been linked to major medical conditions, including hypertension, headaches, heart disease, diabetes, depression, and more.
#### Device Function
The C.A.R.E. appliances are customized oral devices featuring a lower tray and an upper trav. These trays put gentle pressure on the back of the throat to prevent the airway from collapsing during sleep.
Studies have shown that customized oral devices that function by increasing the patency of the airway show comparable efficacy to continuous positive airway pressure (CPAP) devices, considered the gold standard of treatment for OSA (Oral appliance therapy in Olstructive Sleep Aprea-Hypopnea syndrome - A clinical study on therapeutic outcomes Hoekema A PhD thesis, University Medical Centre Groningen Department of Oral and Maxillofacial Surgery. pp 110, 2007). On the basis of these studies, use of oral devices has been recommended by the American Academy of Sleep Medicine for patients with mild or moderate OSA, or for those with severe OSA who are unable to tolerate the CPAP device.
The C.A.R.E. Appliances aim to expand the nasal airway through jaw expansion and mid-facial redevelopment. In doing so, an oral device may be able to permanently improve the oropharyngeal airway. Studies have shown that the C.A.R.E. appliances can increase nasal cavity volume and reduce the incidence of apnea-hypopnea episodes. The mRNA and mmRNA also increase air cavity volume by mandibular advancement.
The C.A.R.E. appliances are customized on models of the patient's teeth, using standard orthodontic acrylics and standard orthodontic wires for clasps and retention. The C.A.R.E. appliances allow for six degrees of freedom in customization, including antero-posterior (AP) adjustment, transverse (TV) adjustment, as well as permitting adjustments of the vertical dimension of occlusion (VDO).
The addition of an optional extender on the back of the device further prevents the patient's airway from collapsing during sleep.
# Intended Use
The C.A.R.E. appliances are intended to reduce nighttime snoring and to treat mild and moderate obstructive sleep apnea in adults, 18 years of age and older. The C.A.R.E. appliances are also intended to treat moderate and severe obstructive sleep apnea (OSA) in adults, 18 years of age and older along with positive airway pressure (PAP) devices and/or myofunctional therapy, as needed.
Target Population: Patients 18 years of age and older with obstructive sleep apnea.
{6}------------------------------------------------
Environment of Use: Fitting of the C.A.R.E. appliances in the dental office for patient use at home.
Comparison to Predicate Devices: Each of the C.A.R.E. appliances (DNA, mRNA and mmRNA) are identical to the previously cleared predicate devices of the same name. This is a request for an expanded Indication of Use only.
The cleaning instructions, instructions for use, and labeling are those currently used for the already-marketed predicate device. The have been reworded, but the indications for use, contraindications, warnings and cautions are the same. The precautions, warnings, risk analysis and other critical statements have been changed only slightly to allow for simultaneous treatment with other modalities as was done with the DNA appliance in K222872) according to the Vivos Method.
Shelf Life: The devices are provided non-sterile. Shelf life will be identical to the predicate devices. No shelf life is required as the device is custom-manufactured and immediately fitted to the patient by the dentist.
Non-clinical Testing: A risk analysis was performed, which considered soreness, obstruction of breathing, tooth movement, and breakage. The product was compared to predicate devices in each area to show the risks were equivalent to the predicate devices. No biocompatibility testing was done as all the components are the same as the predicate device.
Clinical Testing: This submittal relies on peer-reviewed literature and clinical data (RWD) demonstrating that C.A.R.E. Appliances can treat mild and moderate obstructive sleep apnea in adults, 18 years of age and older and treat moderate and severe obstructive sleep apnea (OSA) in adults, 18 years of age and older along with positive airway pressure (PAP) devices and/or myofunctional therapy, as needed. The C.A.R.E. appliances raise no new questions of safety and efficacy as compared to the predicate device.
A peer-reviewed, blinded, controlled, randomized study evaluated the reduction in AHI in 15 consecutive patients randomly assigned to a DNA or a mRNA treatment group. The sleep studies were conducted pre- and post-treatment (the average treatment period was 9.7 months) without an appliance in the mouth. AHI reduction in DNA patients was 70.2%, mRNA patients was 50.6% and C.A.R.E. appliances combined was 64.0%.
Real Word Clinical Data obtained from a research database over a period of 5 years (2018-2023) was filtered to include i) patients 18 years of age and older, ii) treated with a C.A.R.E. appliance, iii) pre-and post-sleep studies at least six months apart and iv) diagnosed with moderate (15≤AHI<30) or severe (AHI ≥ 30) Obstructive Sleep Apnea (OSA). Results were as follows:
| | CARE<br>severe | CARE<br>severe | CARE<br>moderate | CARE<br>moderate |
|----------------------------------------------|----------------|----------------|------------------|------------------|
| | Count | % | Count | % |
| Number of patients | 73 | 100% | 35 | 100% |
| No of Transpalatal Width Same or<br>Improved | 72 | 99% | 35 | 100% |
| Number of AHI Same or Improved | 71 | 97% | 32 | 91% |
{7}------------------------------------------------
| Improved by at least 1 Classification | 57 | 78% | 28 | 80% |
|---------------------------------------|----|-----|----|-----|
| Improved by 45% or 1 Classification | 58 | 80% | 28 | 80% |
| Resolved | 10 | 14% | 7 | 20% |
#### AHI Changes
| | AHI<br>Pre-Treatment<br>(CARE<br>appliances) | AHI<br>Post-Treatment<br>(CARE<br>Appliances) | AHI Change<br>(CARE<br>appliances) | Percent<br>Decrease |
|-------------|----------------------------------------------|-----------------------------------------------|------------------------------------|---------------------|
| Severe CARE | 46.1 ± 15.1 | 21.7 ± 14.8 | -24.3 ± 18.1 | 50.8% |
| Moderate | 21.6 ± 4.5 | 12.0 ± 10.5 | -9.5 ± 10.8 | 44.4% |
Eighty percent (80%) of patients with severe OSA improved by at least 45% or 1 classification (also by 50% or 1 classification). Eighty percent (80%) of patients with moderate OSA improved by at least 45% or 1 classification. Twenty percent (20%) of moderate patients and fourteen percent (14%) of severe patients completely resolved their OSA.
Thirty-seven (37 patients) were treated with the C.A.R.E. appliance alone resulting in an AHI improvement of 52.8%. Thirty-six (36) patients were treated with C.A.R.E. appliances and positive airway pressure and/or myofunctional therapy as prescribed by their dentist for all or a portion of their treatment.
While there were no persistent safety issues, 10% of patients had inadvertent tooth movement and/or changes in bite that were treated with braces or aligners.
Conclusion: The clinical data supports the finding that the C.A.R.E. appliances raise no new questions of safety and efficacy as compared to the predicate devices for the treatment of mild and moderate obstructive sleep apnea in adults, 18 years of age and older and the treatment of moderate and severe obstructuve sleep apnea (OSA) in adults, 18 years of age and older along with positive airway pressure (PAP) devices and/or myofunctional therapy, as needed.
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.