For Mode 0, Ova+ is indicated for temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medication. For Modes 1, 2& 3 Ova+ is indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities. The Ova+ is intended for use in the home. lt should be used only by women aged 18 and above.
Device Story
Ova+ is a battery-powered, single-channel Transcutaneous Electrical Nerve Stimulator (TENS) for home use. It delivers electrical stimulation to superficial nerves via self-adhesive electrodes placed on intact skin. The device features four preset treatment modes controlled by manual push-buttons. It is designed to be worn under clothing to provide continuous, drug-free pain relief. The user selects the mode and intensity; the device provides electrical pulses to modulate pain perception. It benefits patients by offering a non-invasive, portable option for managing menstrual and musculoskeletal pain. The device is intended for use by women aged 18 and older.
Clinical Evidence
No clinical testing performed. Substantial equivalence is supported by non-clinical bench testing, including electrical safety, EMC, and software verification, alongside a usability study conducted per ISO 62366.
Indicated for women aged 18+ for temporary relief of dysmenorrhea (menstrual cramps) when used with OTC pain medication (Mode 0) and temporary relief of sore/aching muscles in shoulder, waist, back, neck, arms, and legs due to exercise or household strain (Modes 1-3).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 15, 2023
TensCare Ltd Saskia Eldridge-Hinmers Regulatory Affairs Associate 9 Blenheim Road Epsom, Surrey KT19 9BE United Kingdom
Re: K230926
Trade/Device Name: Ova+ (K-OVAP-USA) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: November 13, 2023 Received: November 13, 2023
Dear Saskia Eldridge-Hinmers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Doe W. Kumsa -S
for Pamela Scott, MS Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230926
Device Name Ova+ (K-OVAP-USA)
Indications for Use (Describe)
For Mode 0, Ova+ is indicated for temporary relief of pain associated with dysmenorthea (menstrual cramps) when used with over-the-counter pain medication.
For Modes 1, 2& 3 Ova+ is indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">X</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a stylized letter 'e' with a wing extending from its left side. Above the letter, there is a crown. The overall design appears to be a logo or emblem, possibly associated with an organization or individual whose name starts with the letter 'e'.
The Queen's Award
For Enterprise:
International
Trade 2021
Image /page/4/Picture/2 description: The image shows a logo for BSI ISO 13485 Medical Devices Quality Management. The logo is a rounded rectangle with the text "bsi." on the left side and a heart shape below it. On the right side of the logo, the text "ISO 13485 Medical Devices Quality Management" is written in a smaller font. The logo is used to indicate that a company has met the requirements of the ISO 13485 standard for medical device quality management systems.
Image /page/4/Picture/3 description: The image shows the logo for TensCare. The word "TensCare" is written in blue font and is enclosed in a red box with rounded corners. Below the logo, the phrase "Feel better naturally" is written in a smaller, lighter font.
TensCare LTD. PainAway House 9 Blenheim Road Longmead Business Park Epsom, Surrey, KT19 9BE United Kingdom
## 510(k) SUMMARY
## 1. Submitters Identification
| Submitter: | TensCare Ltd |
|------------------------------------|-------------------------------------------------------------|
| Address: | 9 Blenheim Road, Epsom,<br>Surrey, KT19 9BE, United Kingdom |
| Tel : | +44(0)1372 723 434 |
| Fax : | +44(0)1372 745 434 |
| FDA Establishment Registration No: | 3003446042 |
| Contact person: | Saskia Eldridge-Hinmers |
|-----------------|----------------------------------------|
| | Regulatory Affairs Associate |
| Contact Phone: | +44(0)7879424785 |
| Contact Email: | Saskia.Eldridge-Hinmers@tenscare.co.uk |
| Date Prepared: | 15th November 2022 |
### Address of the manufacturing facility:
EasyMed Instruments Co Ltd 3/F, 5F/6F, Block A, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Daliang, 528300 Shunde, Foshan, Guangdong, China,
FDA Establishment Registration No: 3004049909
### Address of American Representative:
| Contact First Name; | SCOTT A |
|---------------------|-----------------------------------------------------------------------|
| Contact Last Name: | BEDNAR |
| Title: | MR |
| Business Name: | QA/RA CONSULTING GROUP, INC. Full |
| Address: | 3335 TUSCARAWAS ROAD<br>BEAVER , PENNSYLVANIA , 15009 , UNITED STATES |
| E-mail: | sbednar@qaraconsultinggroup.com |
| Phone Number: | 412 -4188066 |
Tenscare Ltd., Traditional 510(k) Ova+
18-11-2021
tenscare.co.uk +44 (0)1372 723434 sales@tenscare.co.uk
TENSCARE LTD IS AN EMPLOYEE OWNED BUSINESS MANAGING DIRECTOR - N.C. WRIGHT, SALES DIRECTOR - E.G. SLAVOVA, FINANCE DIRECTOR - C.S.K. SMITH. REGISTERED ADDRESS AS SHOWN ABOVE REGISTERED IN ENGLAND NO. 02839925.
{5}------------------------------------------------
## 2. Subject Device
| Manufacturer | : Tenscare Limited |
|----------------------------|--------------------------------------------------------------|
| Trade/Device Name | : Ova+ (K-OVAP-USA) |
| 510(k) Number | : K230926 |
| Device Classification Name | : Stimulator, Nerve, Transcutaneous, Over-the-counter |
| Common Name: | : Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| Classification Name | : Stimulator, Nerve, Transcutaneous, Over-The-Counter |
| CFR Regulation Number | : 21 CFR 882.5890 |
| Product Code | : NUH |
| FDA Device Classification | : Class II Medical Device |
| Device Type ID | : 3755 |
| Device Definition | : TEMPORARY RELIEF OF PAIN DUE TO SORE/ACHING<br>MUSCLES |
## 3. Predicate Devices
### 3.1. Primary Predicate Device
| Manufacturer | :Life Care Limited. |
|----------------------------|-------------------------------------------------------------|
| Trade/Device Name | :Livia |
| 510(k) Number | :K183110 |
| Device Classification Name | :Stimulator, Nerve, Transcutaneous, Over-the-counter |
| Regulation Description | :Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| CFR Regulation Number | :21 CFR 882.5890 |
| Product Code | :NUH |
| FDA Device Classification | :Class II Medical Device |
| Device Type ID | 3755 |
| Device Definition | : TEMPORARY RELIEF OF PAIN DUE TO SORE/ACHING<br>MUSCLES |
### 3.2. Secondary Predicate Device
| Manufacturer | :TensCare Ltd |
|----------------------------|------------------------------------------------------------------------------------------------------------------------|
| Trade/Device Name | :Perfect EMS_OTC |
| 510(k) Number | :K200694 |
| Device Classification Name | :Stimulator, Nerve, Transcutaneous,<br>Over-the-counter, Powered muscle stimulator |
| Regulation Description | :Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| CFR Regulation Number | :21 CFR 882.5890 |
| Product Code | :NUH (Stimulator, Nerve, Transcutaneous, Over-The-Counter<br>NGX (Stimulator, Muscle, Powered, For Muscle Conditioning |
| Classification Panel: | Neurology; Physical Medicine |
| FDA Device Classification | :Class II Medical Device |
{6}------------------------------------------------
## 4. Device Description
The Ova+ is a non-invasive small and convenient unit that can be worn under clothing to provide safe, continuous, drug-free period pain relief whilst maintaining a normal, active lifestyle.
The device is battery powered, single channel home using Transcutaneous Electrical Nerve Stimulation (TENS).
The device is supplied with self-adhesive electrodes which connect to the control unit by cable and plug and are placed on patients' intact skin by the end user. Electrical stimulation is delivered via these self-adhesive electrodes to superficial nerves.
The unit is intended for home use by the patient and is designed with simplicity and ease of use in mind. It has four preset treatment Modes. The level of electrical stimulation of Ova+ is easily controlled by the end user using manual, push-button controls.
The new device Ova+ uses the same technical principle and substantially equivalent stimulation parameters to the marketed device Livia (K183110) and/or reference predicate device Perfect EMS (K200694)
The Ova+ uses a rechargeable Li-Ion polymer battery and a custom TENS connecting lead with 2mm connecting pins.
The Ova+ is supplied with adhesive electrodes that are similar to those of the predicate devices.
## 5. Indication For Use/Intended Use
For Mode 0, Ova+ is indicated for temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medication.
For Modes 1, 2& 3 Ova+ is indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
### Intended Use
The Ova+ is intended for use in the home. lt should be used only by women aged 18 and above.
{7}------------------------------------------------
## 6. Indication for Use comparison
| Characteristic | New Device<br>Ova+ | Primary Predicate<br>Device<br>Livia | Secondary Predicate<br>Device Perfect EMS<br>OTC | |
|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Manufacturer | Tenscare Limited | Life Care Ltd. | TensCare Ltd | |
| 510(k) No. | K230926 | K183110 | K200694 | |
| Indications for<br>Use | For Mode 0, Ova+ is<br>indicated for temporary<br>relief of pain associated<br>with dysmenorrhea<br>(menstrual cramps) when<br>used with over-the-<br>counter pain medication.<br>For Modes 1, 2& 3 Ova+<br>is indicated for temporary<br>relief of pain associated<br>with sore and aching<br>muscles in the shoulder,<br>waist, back, neck, upper<br>extremities (arm) and<br>lower extremities (leg)<br>due to strain from<br>exercise or normal<br>household work activities. | The Livia is designed for<br>symptomatic relief and<br>management of chronic<br>pain, and for temporary<br>relief of pain associated<br>with sore and aching<br>muscles in the shoulder,<br>waist, back, neck, upper<br>extremities (arm) and<br>lower (extremities) leg<br>due to strain from<br>exercise or normal<br>household work activities.<br>The Livia is also indicated<br>for temporary relief of<br>pain associated with<br>dysmenorrhea (menstrual<br>cramps) when used with<br>over-the-counter pain<br>medication. | TENS: The device is<br>designed to be used<br>for temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist, back,<br>neck, upper<br>extremities (arm),<br>lower extremities (leg),<br>abdomen and bottom<br>due to strain from<br>exercise or normal<br>household work<br>activities | |
| Intended Users | The Ova+ is intended for<br>use in the home.<br>It should be used only by<br>women aged 18 and<br>above. | Livia should be used only<br>by women aged 18 and<br>above. | Keep away from<br>children | |
| Class | ll | ll | ll | |
| Product code | NUH | NUH | NUH | |
| Regulation number | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | |
| OTC Use | Yes | Yes | Yes | |
| Single Use or<br>Reusable | Reusable<br>For single patient use | Reusable<br>For single patient use | Reusable<br>For single patient use | |
| Characteristic | New Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | Comparison |
| Device<br>Name/Model | Ova+ | Livia | Perfect EMS | |
| 510(k) | K230926 | K183110 | K200694 | |
| Manufacturer | TensCare Ltd | LifeCare Ltd. | TensCare Ltd | |
| Power Source(s)1 | 3.7V Lithium ion<br>battery(rechargeable) | 3.7V Lithium ion<br>battery(rechargeable) | 2 x AA alkaline battery | Internal power<br>source<br>Substantially<br>Equivalent. |
| Method of Line<br>current Isolation | Not possible to<br>connect patient lead<br>and charger at same<br>time - use same<br>socket | Output is electrically<br>disabled when<br>connect to charger, by<br>means of<br>microprocessor<br>charging circuit | No connection to Line<br>Current | Substantially<br>Equivalent. |
| Patient Leakage<br>Current<br>- Normal Condition<br>(μΑ) | Battery<br>powered<br>(< 10μA) | Battery<br>powered<br>(< 10μA) | Battery<br>powered<br>(< 10μA) | Identical |
| Patient Leakage<br>Current<br>- Single Fault<br>Condition (μΑ) | Battery<br>powered<br>(< 50μA) | Battery<br>powered<br>(< 50μA) | Battery<br>powered<br>(< 50μA) | Identical |
| Average DC current<br>through Electrodes<br>when device is on<br>but<br>no pulses are being<br>applied (μΑ) | 0 µA | 0 µA | 0 µA | Identical with<br>primary<br>predicate |
| Number of Output<br>Modes | 4<br>1 Dysmenorrhea &<br>3 Pain | 1 Dysmenorrhea | 1 Pain | Substantially<br>equivalent |
| Number of Output<br>Channels | 1 | 1 | 2 | Identical with<br>primary<br>predicate |
| Regulated Current<br>or Voltage? | Voltage | Current | Voltage | Substantially<br>Equivalent. |
| Software/Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Yes | Identical |
| Automatic No-Load<br>Trip? | No | Yes | Yes | Substantially<br>Equivalent. |
| Automatic Overload<br>Trip? | No | Yes | Yes | Substantially<br>Equivalent. |
| Automatic Shut Off? | No | No | Yes | Identical |
| Characteristic | New Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | Comparison |
| Device<br>Name/Model | Ova+ | Livia | Perfect EMS | |
| User Override<br>Control? | Yes | Yes | Yes | Substantially<br>Equivalent. |
| Indicator Display:<br>-On/Off status<br>-Low battery<br>-Voltage/Current<br>level<br>-Time to cut-off | Yes<br>Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes<br>Yes | Identical |
| Timer Range<br>(minutes) | No Timer | The Livia has no<br>internal timer | 20-40 | Identical with<br>primary<br>predicate |
| Compliance with<br>Voluntary<br>Standards? | IEC 60601-1,<br>IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11<br>IEC 62304<br>ISO 14971<br>IEC 60601-1-6<br>IEC 62366-1 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10,<br>ISO 10993-1, -5 and -<br>10 | IEC 60601-1:2005<br>IEC 60601-1-2<br>IEC 60601-2-10 IEC<br>60601-1-11<br>IEC 62304<br>ISO 14971 | Substantially<br>Equivalent. |
| Compliance with 21<br>CFR 898? | Yes | Yes | Yes | Identical |
| Weight g | 19.5 | 37 | 146.5 | Similar |
| Dimensions (mm) | 65 x 38 x 10 | 55X 55 X 20 | 66×136×30.7 | Similar |
| Housing Materials<br>and Construction | Silicone, ABS<br>plastics | ABS plastics | ABS plastics | Substantially<br>Equivalent. |
| Sterilization | N/A | N/A | N/A | Identical |
| Parameter | New Device | Primary Predicate<br>Device | Secondary<br>Predicate<br>Device | Comparison |
| Device<br>Name/Model | Ova+ | Livia | Perfect EMS<br>OTC | |
| 510(k) | K230926 | K183110 | K200694 | |
| Manufacturer | TensCare Ltd | LifeCare Ltd. | TensCare Ltd | |
| Waveform | Biphasic,<br>Equal duration…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.