← Product Code [LYR](/productcode/LYR) · K230901

# Premier HpSA Flex (619096) (K230901)

_Meridian Bioscience, Inc. · LYR · Jul 3, 2023 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230901

## Device Facts

- **Applicant:** Meridian Bioscience, Inc.
- **Product Code:** [LYR](/productcode/LYR.md)
- **Decision Date:** Jul 3, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3110
- **Device Class:** Class 1
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K230901 · Jul 3, 2023 | Premier HpSA Flex (619096) | Meridian Bioscience, Inc. | Left-over clinical stool specimens from patients undergoing routine diagnostic testing | Retrospective clinical specimens were used to conduct a method comparison study to validate the performance of the Premier HpSA Flex assay with preserved stool specimens against an FDA-cleared comparator. | Method comparison; Retrospective clinical samples; Routine clinical care |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Method comparison study; Retrospective method comparison | Symptomatic patients with suspicions of gastroenteritis; Sample Size: 182; Number of Sites: 1 | FDA-cleared comparator assay | Positive and negative percent agreement |

## Indications for Use

The Premier HpSA Flex enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. The test is intended for use with unpreserved stool specimens or preserved stool specimens in transport media. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.

## Device Story

Premier HpSA Flex is a microwell-based enzyme immunoassay (EIA) for qualitative detection of H. pylori antigens in human stool. It modifies the predicate (K182559) by adding claims for stool specimens preserved in Cary-Blair or Culture and Sensitivity (C&S) transport media. The assay uses monoclonal anti-H. pylori capture antibodies adsorbed to microwells. Diluted patient samples and enzyme conjugate are incubated; unbound material is washed away. Substrate addition produces a color change (colorless to yellow) indicating antigen presence. Results are interpreted visually or via spectrophotometer (450 nm/630 nm). Used in clinical laboratories by technicians to aid diagnosis and monitor treatment efficacy. Benefits include objective, simple interpretation and expanded specimen handling flexibility.

## Clinical Evidence

Bench testing only. Analytical performance evaluated using contrived samples (negative stool matrix spiked with H. pylori antigen). LoD determined as 12 ng/mL for preserved stool. Reproducibility study (N=360) showed 100% agreement across sites and lots. Interference and cross-reactivity studies showed no impact from common stool substances or microorganisms. Method comparison study (N=182) against an FDA-cleared comparator showed 100% positive agreement and 98.5% negative agreement.

## Technological Characteristics

Microwell-based enzyme immunoassay (EIA). Uses monoclonal anti-H. pylori capture antibodies. Qualitative output. Visual or spectrophotometric (450 nm/630 nm) detection. Compatible with unpreserved stool or stool in Cary-Blair/C&S transport media. Reagents stored at 2–8°C. No specialized instrumentation required for visual read; generic spectrophotometer optional.

## Regulatory Identification

Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.

## Predicate Devices

- Premier Platinum HpSA PLUS ([K182559](/device/K182559.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K230901

B Applicant

Meridian Bioscience, Inc.

C Proprietary and Established Names

Premier HpSA Flex (619096)

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LYR | Class I, reserved | 21 CFR 866.3110 - Campylobacter Fetus Serological Reagents | MI - Microbiology  |

## II Submission/Device Overview:

### A Purpose for Submission:

To obtain a substantial equivalence determination for the Premier HpSA Flex by Meridian Bioscience to the predicate device. The Premier HpSA Flex assay represents a modification of the predicate device (Meridian Bioscience’s PREMIER Platinum HpSA PLUS assay: K182559)—the modification being the addition of preserved stool specimen for testing with this device.

### B Measurand:

Helicobacter pylori antigens.

### C Type of Test:

Microwell-based enzyme immunoassay with qualitative assay outputs.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The Premier HpSA Flex enzyme immunoassay (EIA) is an *in vitro* qualitative procedure for the detection of *Helicobacter pylori* antigens in human stool. The test is intended for use with unpreserved stool specimens or preserved stool specimens in transport media. Test results are intended to aid in the diagnosis of *H. pylori* infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

D Special Instrument Requirements:

Visual determination of Assay endpoint needs no special instrument. However, if the optional spectrophotometric determination is chosen, endpoint reading of the assay requires a generic spectrophotometric microplate reader capable of recording absorbance values at 450 nm and optionally at 630 nm for wavelength correction.

IV Device/System Characteristics:

A Device Description:

The Premier HpSA Flex assay represents a modification of the sponsor’s previously FDA-cleared Premier Platinum HpSA PLUS assay (K182559). The modification is the addition of a new sample type claim (i.e., human stool specimens preserved in specific transport media, namely, Cary-Blair or Culture and Sensitivity (C&amp;S) transport media) to the intended use of the previously cleared device (K182559). Both the subject and predicate devices are qualitative, *in vitro* diagnostic tests for the detection of *Helicobacter pylori* antigens present in human stool specimens.

B Principle of Operation:

The Premier HpSA Flex test is a microwell-based enzyme immunoassay that detects *H. pylori* antigens from human stool specimens. Stool specimen may be either unpreserved or preserved in Cary-Blair or Culture and Sensitivity (C&amp;S) transport media. The test utilizes a plurality (mixture) of monoclonal anti-*H. pylori* capture antibodies adsorbed to microwells. Diluted patient samples and an enzyme conjugate reagent are added to the microwells and incubated for one hour at room temperature. A wash is performed to remove unbound material. Substrate is added and incubated for 10 minutes at room temperature. Color develops by reaction with the bound enzyme, indicating the presence of *H. pylori* antigen. Stop solution is added and the results are interpreted visually or spectrophotometrically. No calculations are required, and the visible change from colorless to

K230901 - Page 2 of 14

{2}

yellow makes the interpretation of results objective and simple. The yellow color may be read with a spectrophotometer at 450 nm wavelength with or without correction at 630 nm. In addition, the HpSA test permits assessment of established or novel anti-H. pylori treatment during and post-therapy to monitor for treatment effectiveness, relapse, or eradication.

V Substantial Equivalence Information:

A Predicate Device Name(s):

PREMIER Platinum HpSA PLUS

B Predicate 510(k) Number(s):

K182559

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | Device: K230901 | Predicate: K182559  |
| --- | --- | --- |
|  Device Trade Name | Premier HpSA Flex | PREMIER Platinum HpSA PLUS  |
|  Intended Use / Indications for Use | The Premier HpSA Flex enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool.
The test is intended for use with unpreserved stool specimens or preserved stool specimens in transport media.
Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients.
Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy | The PREMIER Platinum HpSA PLUS is an in vitro diagnostic procedure for the detection of Helicobacter pylori antigens in human stool.
Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.  |
|  General Device Characteristic Similarities  |   |   |
|  Measured analyte | Helicobacter pylori antigens | Same  |
|  Antibody / Technology / Assay format | Monoclonal / Enzyme immunoassay (EIA) / Microwell-based | Same  |
|  Type of Test | Qualitative | Same  |

K230901 - Page 3 of 14

{3}

|  Device & Predicate Device(s): | Device: K230901 | Predicate: K182559  |
| --- | --- | --- |
|  Controls | Positive and negative controls included in the kit | Same  |
|  Reagent storage | Refrigerated (2–8°C) | Same  |
|  Reagent and components | Antibody Coated Microwells; Positive Control; Sample Diluent / Negative Control; Premier 20X Buffer I; Enzyme Conjugate; Premier Substrate I; Premier Stop Solution 1; Transfer pipettes; Microwell strip sealer; Wooden stick applicators | Same  |
|  Result interpretation | Visual or spectrophotometric | Same  |
|  General Device Characteristic Differences  |   |   |
|  Device & Predicate Device(s): | Device: K230901 | Predicate: K182559  |
|  Addition of specimen type | Unpreserved stool specimens and stool specimens preserved in Cary-Blair or C&S transport media. | Unpreserved (raw) stool specimens  |

VI Standards/Guidance Documents Referenced:

None

VII Performance Characteristics:

A Analytical Performance:

Analytical performance of the subject assay (Premier HpSA Flex) was evaluated using contrived samples. Briefly, a Bulk Negative Stool (BNS) matrix was prepared by pooling unpreserved stool specimens from five (5) donors or fewer into small, pooled lots. Purified and pre-quantified $H.$ pylori flagellar antigen (HpSA) was used to spike the negative stool matrix with known concentrations to create contrived positive samples. Sample matrix alone served as true negative samples. The contrived positive and negative samples were then preserved in Remel Cary Blair (“CB”; ThermoFisher, cat. #R21925) or Para-Pak Culture &amp; Sensitivity transport media (“C&amp;S”; MBI, cat. #900612) at a volumetric ratio of 1:4 (i.e., 1 part sample in negative stool matrix + 3 parts transport media, respectively).

1. Limit of Detection (LoD) / Analytical Sensitivity:

Analytical sensitivity of the subject assay was estimated in two steps, (i) by estimating the Limit of Detection (LoD) of the assay using contrived positive human stool samples preserved in either CB or C&amp;S transport media, and (ii) by determining that the performances of the assay with either medium are equivalent at near LoD antigen concentrations in samples.

K230901 - Page 4 of 14

{4}

Contrived samples (positives and negatives) were tested in the assay following instructions in labeling using three (3) kit lots and twenty (20) replicates per sample.

LoD was defined as the lowest concentration of HpSA (calculated as ng of antigen per mL of preserved stool) that produced positive results $\geq 95\%$ of the time. The LoD for the subject assay was conservatively determined to be $12\,\mathrm{ng}$ HpSA per mL of preserved stool. The previously established LoD using unpreserved (raw) stool was $4.66\,\mathrm{ng/mL}$.

Further, the equivalence between CB and C&amp;S TM was assessed at LoD and below LoD antigen concentrations in contrived samples using three (3) kit lots with 20 replicates at each of these $Lot \times Concentration$ (total of 10 testing conditions) for each transport media. CB and C&amp;S were determined to be equivalent for the preservation of specimens intended to be used with the subject assay, as shown below in Table 1.

Table 1: Equivalence of CB and C&amp;S transport media for preservation of stool specimens for use in Premier HpSA Flex assay

|   | C&S | CB | OD Difference Mean [90% CI] | EAC* (Δmean OD) | Outcome  |
| --- | --- | --- | --- | --- | --- |
|   |  Mean OD [90% CI] | Mean OD [90% CI]  |   |   |   |
|  All tested HpSA concentrations | 0.2087 [0.131–0.286] | 0.1475 [0.070–0.225] | 0.0612 [0.054–0.069] | ±0.1 | Equivalent  |
|  At LoD (12 ng/mL) | 0.2413 [0.226–0.256] | 0.2141 [0.208–0.221] | 0.0272 [0.011–0.043] | ±0.1 | Equivalent  |

* EAC: Equivalence Acceptance Criterion

2. Precision/Reproducibility:

A multi-site, blinded study was conducted at three (3) external clinical sites to evaluate the reproducibility and repeatability of the Premier HpSA Flex device. The testing panels of contrived stool samples were preserved in C&amp;S transport media. Each panel consisted of negative stool matrix alone (true negative sample, TN) or matrix spiked with known concentrations of HpSA to generate high negative (&lt;1X LoD, HN), low positive (2X LoD, LP), and moderate positive (5X LoD, MP) samples. The panels were stored at $\leq -70^{\circ}\mathrm{C}$ and shipped frozen to the study sites.

Three (3) independent sites tested the panels in a blinded fashion, with each site evaluating: ten (10) panels x four (4) members/panel x three (3) replicates/member = a total of 120 samples. In total, therefore, the study evaluated 360 samples at the three sites combined. Two (2) technicians participated in the testing at each site. Along with each panel (of 24 samples), each technician tested a positive and negative control per day, for five (5) days.

Precision assessments evaluated the variabilities of consecutive measurements of the same sample input between sites, within sites, and within run, as well as the reproducibility of measurements of each sample over time and experimental conditions. Setting the expected results to be a qualitative negative or positive, total percent agreement amongst all sites was for 360 out of 360 samples, i.e., $100\%$ (95% CI: 98.9–100%). Similarly, the total percent

K230901 - Page 5 of 14

{5}

agreement amongst all kit lots for 360 out of 360 samples, i.e.,  $100\%$  (95% CI: 98.9–100%), as summarized in Table 2.

Table 2. Reproducibility analysis of OD values (N=360) show:

|  Sample type | Site (N) | Mean OD | Assay Read out | Between Site |   | Between Run (Within Site) |   | Within Run |   | Total (Reproducibility)  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |  SD | %CV | SD | %CV | SD | %CV | SD | %CV  |
|  TN | 1 (30) | 0.0037 | NEG | N | N | 0.002 | 43.7% | 0.002 | 54.5% | 0.003 | 69.80%  |
|   |  2 (30) | 0.0054 | NEG | N | N | 0 | 0.00% | 0.002 | 29.7% | 0.002 | 29.70%  |
|   |  3 (30) | 0.0024 | NEG | N | N | 0.003 | 118% | 0.007 | 274% | 0.007 | 298%  |
|   |  ALL (90) | 0.0039 | NEG | 0.001 | 29.9% | 0.002 | 48.7% | 0.004 | 107% | 0.005 | 121%  |
|  HN | 1 (30) | 0.0056 | NEG | N | N | 0.003 | 59.5% | 0.004 | 74.7% | 0.005 | 95.50%  |
|   |  2 (30) | 0.0124 | NEG | N | N | 0.009 | 73.8% | 0.005 | 36.6% | 0.01 | 82.40%  |
|   |  3 (30) | 0.0052 | NEG | N | N | 0 | 0.00% | 0.007 | 127% | 0.007 | 127%  |
|   |  ALL (90) | 0.0077 | NEG | 0.004 | 46.0% | 0.005 | 68.90% | 0.006 | 71.0% | 0.008 | 109%  |
|  LP | 1 (30) | 0.2699 | POS | N | N | 0.04 | 14.9% | 0.08 | 28.5% | 0.09 | 32.10%  |
|   |  2 (30) | 0.3968 | POS | N | N | 0.1 | 30.5% | 0.08 | 20.6% | 0.1 | 36.80%  |
|   |  3 (30) | 0.4455 | POS | N | N | 0.1 | 28.0% | 0.07 | 16.1% | 0.1 | 32.20%  |
|   |  ALL (90) | 0.3707 | POS | 0.08 | 22.5% | 0.1 | 27.8% | 0.08 | 20.7% | 0.2 | 41.30%  |
|  MP | 1 (30) | 0.7005 | POS | N | N | 0.1 | 18.0% | 0.1 | 17.3% | 0.2 | 25.00%  |
|   |  2 (30) | 1.006 | POS | N | N | 0.2 | 16.8% | 0.2 | 15.4% | 0.2 | 22.80%  |
|   |  3 (30) | 1.0318 | POS | N | N | 0.2 | 15.8% | 0.1 | 12.6% | 0.2 | 20.20%  |
|   |  ALL (90) | 0.9128 | POS | 0.2 | 19.3% | 0.2 | 16.9% | 0.1 | 14.9% | 0.3 | 29.60%  |

Considering all data points as recommended by the FDA-recognized consensus standard CLSI EP05-A3 (2014), the Within-Run  $\%$  CV ranged  $14.9 - 107\%$  and Reproducibility (total  $\%$  CV) ranged  $29.6 - 121\%$  across all sites. As expected, the data variability was higher at lower Ag concentrations (TN and HN). However, there was no instance recorded of unexpected conversion of a presumed negative to positive or vice versa. All concentrations tested met the predefined acceptance criteria.

# 3. Linearity:

Not applicable

# 4. Analytical Specificity/Interference:

Studies were conducted to determine whether substances that may be present in human stool specimens preserved in CB or C&amp;S TM interfere with results generated by the subject assay (Premier HpSA Flex). Substances evaluated were those that may reasonably be expected to be present within patient samples and those that were determined to have the potential to affect test results.

True negative (negative stool matrix alone) and low positive (2X LoD) specimens were contrived in triplicate. The target interferent was added to each replicate. Where necessary, stock solutions of interferent substances were prepared in a diluent (sterile  $0.85\%$  saline) prior

K230901 - Page 6 of 14

{6}

to combining with stool samples; the stock concentrations were prepared such that the diluent volume did not exceed 10% of the stool matrix when combining. Specimen replicates with added interferents were preserved in C&amp;S transport medium at a ratio of 1:4 (v/v).

The same chemical or biological substances were previously evaluated for interference with the predicate assay (K182559). Table 3 below lists the potential interferents with their target concentrations (per 500 μL of patient stool) evaluated with specimens preserved in C&amp;S transport media.

Table 3. Interfering substances evaluated

|  Substance (Active Ingredient(s)) |   | Target Concentration (per 500 μL of Patient Stool)  |
| --- | --- | --- |
|  1 | TUMS | 10 mg/500 μL  |
|  2 | Mylanta (per 10 mL: Aluminum hydroxide 800 mg, Magnesium hydroxide 800 mg, Simethicone 80 mg) | 11.5 mg/500 μL  |
|  3 | Pepto-Bismol (Bismuth subsalicylate 525 mg/30 mL) | 0.44 mg/500 μL  |
|  4 | Tagamet (Cimetidine 200 mg/tablet) | 1 mg/500 μL  |
|  5 | Prilosec OTC (Omeprazole 20 mg/tablet) | 1 mg/500 μL  |
|  6 | Barium Sulfate | 25 mg/500 μL  |
|  7 | Whole Blood | 250 μL/500 μL  |
|  8 | Leukocytes (White Blood Cells) | 250 μL/500 μL  |
|  9 | Mucin | 17 mg/500 μL  |
|  10 | Hemoglobin | 62.5 mg/500 μL  |
|  11 | Stearic Acid (fecal fat) | 5.3 mg/500 μL  |
|  12 | Palmitic Acid (fecal fat) | 2.65 mg/500 μL  |
|  13 | NSAID, Ibuprofen | 0.25mg/500 μL  |

Each microwell plate of samples tested was analyzed spectrophotometrically at dual wavelength (A₄₅₀–₆₃₀nm). External controls (Positive and Negative) were included with each plate tested. All positive samples (preserved in C&amp;S transport media) tested gave the expected results indicating that none of the substances evaluated in this study showed interference in the subject assay at the concentrations tested. All negative samples were also negative.

5. Analytical Specificity/Cross-reactivity:

Studies were conducted to determine whether micro-organisms that may be present human stool specimens preserved in CB or C&amp;S transport media non-specifically reacted with the subject assay. The microorganisms (i.e., bacteria, fungus, and viruses) evaluated in this study are those that are most likely to be or are possibly present within patient stool specimens and those could potentially affect assay results.

True negative (negative stool matrix alone) and low positive (2X LoD) specimens were contrived in triplicate. Relevant bacterial organisms or fungus were tested at final concentrations of ≥1.0 × 10⁷ CFU/mL, whereas viruses were tested at ≥1.0 × 10⁵ TCID₅₀/mL. For cross-reactivity, the target micro-organisms were added to preserved true negative replicates; for microbial interference, target micro-organisms were added to preserved low positive replicates. Where necessary, microbial stock solutions were prepared in sterile 0.85% saline prior to combining with stool samples; the stock concentrations were prepared such that

K230901 - Page 7 of 14

{7}

the diluent volume did not exceed  $10\%$  of the stool matrix when combining. Specimen replicates with added micro-organisms were preserved in C&amp;S transport medium at a ratio of 1:4 (v/v).

Each microwell plate of samples tested was analyzed spectrophotometrically at dual wavelength  $(\mathrm{A}_{450 - 630\mathrm{nm}})$ . External controls (Positive and Negative) were included with each plate tested. Table 4 below lists the micro-organisms tested in preserved stool specimens to evaluate cross-reactivity and microbial interference in the performance of the subject assay. The same microbial organisms were previously evaluated for cross-reactivity and microbial interference with the predicate assay (K182559).

Table 4. List of potentially cross-reactive organisms tested.

|   | Microorganism Tested | (Strain ID)  |
| --- | --- | --- |
|  1 | Aeromonas hydrophila | ATCC 35654  |
|  2 | Bacillus subtilis | ATCC 6051  |
|  3 | Borrelia burgdorferi | N/A  |
|  4 | Campylobacter coli | ATCC 43478  |
|  5 | Campylobacter fetus | ATCC 25936  |
|  6 | Campylobacter jejuni | ATCC 33292  |
|  7 | Campylobacter lari | ATCC BAA-1060  |
|  8 | Candida albicans | ATCC 18804  |
|  9 | Citrobacter freundii | ATCC 43864  |
|  10 | Clostridium difficile | ATCC 43255  |
|  11 | Clostridium perfringens | ATCC 12915  |
|  12 | Enterobacter cloacae | ATCC 15337  |
|  13 | Enterococcus faecalis | ATCC 49532  |
|  14 | Escherichia coli | ATCC 8739  |
|  15 | Escherichia coli | ATCC 9637  |
|  16 | Escherichia coli | ATCC BAA-2196  |
|  17 | Escherichia coli O157:H7 (toxigenic) | ATCC 43895  |
|  18 | Escherichia fergusonii | ATCC 35469  |
|  19 | Escherichia hermanii | ATCC 33650  |
|  20 | Escherichia hermanii | EMDI-64  |
|  21 | Haemophilus influenzae | ATCC 9006  |
|  22 | Klebsiella pneumoniae | ATCC BAA-1900  |
|  23 | Lactococcus lactis | ATCC 49032  |
|  24 | Listeria monocytogenes | ATCC 19115  |
|  25 | Peptostreptococcus anaerobius | ATCC 27337  |
|  26 | Proteus vulgaris | ATCC 6380  |
|  27 | Pseudomonas aeruginosa | ATCC 10145  |
|  28 | Pseudomonas fluorescens | ATCC 13525  |
|  29 | Salmonella enterica subsp. enterica serovar Dublin | ATCC 15480  |
|  30 | Salmonella enterica subsp. enterica serovar Hilversum | ATCC 15784  |
|  31 | Salmonella enterica subsp. enterica serovar Typhimurium | ATCC 13311  |
|  32 | Salmonella heidelberg (Group B) | ATCC 8326  |
|  33 | Salmonella minnesota | ATCC 9700  |
|  34 | Serratia liquefaciens | ATCC 27952  |
|  35 | Serratia marcescens | ATCC 43862  |

K230901 - Page 8 of 14

{8}

|  Microorganism Tested | (Strain ID)  |
| --- | --- |
|  36 Shigella boydii | ATCC 9207  |
|  37 Shigella dysenteriae | ATCC 9361  |
|  38 Shigella flexneri | ATCC 12022  |
|  39 Shigella sonnei | ATCC 25931  |
|  40 Staphylococcus aureus | ATCC 51153  |
|  41 Staphylococcus aureus (Cowan’s) | ATCC 12598  |
|  42 Staphylococcus epidermidis | ATCC 49134  |
|  43 Yersinia enterocolitica | ATCC 23715  |
|  44 Adenovirus 41 | Tak  |
|  45 Rotavirus RV4 |   |

All spiked samples (preserved in C&amp;S transport media) were observed to be negative in the assay, indicating that none of the micro-organisms evaluated in this study showed cross-reactivity or microbial interference in the subject assay at the concentrations tested. External controls (Positive and Negative) that were included also produced the expected results.

6. Assay Reportable Range:

Not Applicable.

7. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

**Storage stability studies:**

Analytical storage stability studies were performed with a panel of contrived stool specimens preserved in CB or C&amp;S transport media to determine acceptable limits for storage and handling of such preserved specimens. For each transport media type, the panel included a true negative (TN) sample with unpreserved negative stool matrix, a moderate positive (MP) sample (5X LoD), three low positive (LP) samples (2X LoD), and a high negative (HN) sample with antigen at &lt;1X LoD. Antigen concentration for HN sample (7.5 ng/mL) was determined by testing various dilutions until testing generated results with 10–90% positivity (i.e., 1–9 positives out of 10 replicates) at the given concentration in each transport media. Prior to the study, all panel members were preserved in CB or C&amp;S transport media at 1:4 (v/v) ratio. Testing was performed following the instructions for use for the subject device. These analytical studies are listed below.

**a) Preserved specimen storage stability:**

An analytical study determined the acceptable limits for storage and handling of human stool specimens when preserved in CB or C&amp;S transport media and tested with the subject assay. This study used the contrived and preserved specimens indicated above (i.e., TN, HN, LP, and MP samples) as a “reference” panel. A separate panel of contrived “clinical” positives was prepared using archived, preserved clinical positive stool samples with pre-quantified analyte (HpSA) levels as measured by the assay. These archived preserved stool samples were serially diluted in cognitely preserved (CB or C&amp;S) negative stool matrix and tested to arrive at assay signals that were equivalent (within ± 2 SD of mean) to those of the reference panel members LP (designated “2X”) and MP (designated “5X”). Similarly, another archived preserved clinical positive specimen was serially diluted in cognitely preserved negative stool matrix to create a contrived “clinical” high negative (designated “HN”) sample for which replicates

K230901 - Page 9 of 14

{9}

generated results with  $10 - 90\%$  positivity (i.e., 1-9 assay positives out of 10 replicates in a set) in each transport media. Finally, archived clinical negative specimens were screened and preserved in CB or C&amp;S transport media to create a negative (designated "NS") "clinical" sample.

After baseline testing (time 0), each preserved sample was subjected to different storage temperatures and time conditions, and the stored sample aliquots were tested in the subject assay at designated time points ( $\sim 106$  and  $130.5$  hours for refrigerated or ambient storage; 10 and 75 days for storage in  $-20^{\circ}\mathrm{C}$  and  $-80^{\circ}\mathrm{C}$  freezers) using a minimum of 10 replicates per sample.

When subjected to different temperature conditions as indicated above and then tested in the assay, the MP, LP, and NS preserved in CB or C&amp;S transport media produced acceptable results at all timepoints and reliably demonstrated assay signal outputs in alignment with expectations. For preserved NS samples, baseline results met all acceptance criteria; prolonged storage at several timepoints resulted in the low analyte (HpSA antigen) content in these samples being variably detected by the subject immunoassay, as expected. Results from this study are shown in Table 5.

Table 5. Preserved Sample storage study results

|  Time Point | Storage temp (℃) | Sample ID | Positive in Assay | Replicates tested | Percent Positive | Positive in Assay | Replicates tested | Percent Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  C&S |   |   | CB  |   |   |
|  Base line | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |   |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 2 | 10 | 20% | 7 | 10 | 70%  |
|   |   |  NS | 10 | 10 | 0% | 0 | 10 | 0%  |
|  106-108* Hours | 2-8 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 0 | 10 | 0% | 7 | 10 | 70%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |
|   |  19-27 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |
|  130.5 Hours | 2-8 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 0 | 10 | 0% | 4 | 10 | 40%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |
|   |  19-27 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 4 | 10 | 40% | 10 | 10 | 100%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |
|  10 days | -28 to -16 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 9 | 10 | 90% | 10 | 10 | 100%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |

K230901 - Page 10 of 14

{10}

|  Time Point | Storage temp (℃) | Sample ID | Positive in Assay | Replicates tested | Percent Positive | Positive in Assay | Replicates tested | Percent Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  C&S |   |   | CB  |   |   |
|  -84 to -66 | 5X | 10 | 10 | 100% | 10 | 10 | 100% |   |
|   |  2X | 10 | 10 | 100% | 10 | 10 | 100% |   |
|   |  HN | 10 | 10 | 100% | 10 | 10 | 100% |   |
|   |  NS | 0 | 10 | 0% | 0 | 10 | 0% |   |
|  75 days | -28 to -16 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 2 | 10 | 20% | 10 | 10 | 100%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |
|   |  -84 to -66 | 5X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  2X | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  HN | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |   |  NS | 0 | 10 | 0% | 0 | 10 | 0%  |

* Note: samples were stored at 2–8°C for 108 hours and at 19–27°C for 106 hours prior to testing at these time points.

In summary, specimen stability testing of a panel of clinical specimens diluted in negative stool matrix and preserved in CB or C&amp;S transport media demonstrated reliably detectable assay signals when the preserved stool specimens were stored at ambient (19–27°C) or cold (2–8°C) temperatures for up to 120 hours or at frozen conditions (−20°C and/or −80°C) for up to 14 days. The collective data from this study supported the acceptable storage times and conditions and demonstrated analyte stability in frozen specimens, which served as the basis for supporting the use of frozen, archived specimens for the Method Comparison study.

## b) Freeze/Thaw:

An analytical study determined the acceptable limits for the freeze / thaw cycling of human stool specimens when preserved in CB or C&amp;S transport media for testing with the subject assay. The panel (Neg, HN, LP, and MP) of contrived, preserved stool samples was tested at baseline ("Time 0"), then subjected to a series of freeze/thaw cycles using a low temperature (−20°C) freezer (range: −28 to −16°C) or an ultralow (−80°C) freezer (range: −84 to −66°C) followed by ambient (19–27°C) thawing, and finally tested in the assay with 10–30 replicates per sample. For preserved Negative, LP (2X LoD), and MP (5X LoD) samples in either transport media and cycled up to 3 times through freeze/thaw cycles in low-temperature and ultralow freezers, all samples showed acceptable signal outputs aligned with expectations, and no significant trend was observed in the average ODs (A450-630) compared to baseline results. For HN samples, whereas the baseline ODs remained within the sponsor's acceptance criteria of 10–90% positive signals, when subjected to up to 3 freeze/thaw cycles, the low amount (&lt;1X LoD) of analyte present in the preserved HN samples (6.5 ng/mL of HpSA in negative stool matrix) produced only low or no signals as expected. The results are summarized in Table 6 below.

K230901 - Page 11 of 14

{11}

Table 6. Freeze thaw stability data.

|  Storage Condition | Panel Member | Positive in Assay | Replicates tested | Percent Positive | Positive in Assay | Replicates tested | Percent Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  C&S |   |   | CB  |   |   |
|  Baseline | MP | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |  LP | 30 | 30 | 100% | 30 | 30 | 100%  |
|   |  HN | 9 | 10 | 90% | 7 | 10 | 70%  |
|   |  Neg | 0 | 10 | 0% | 0 | 10 | 0%  |
|  Cycle 2 (-28 to -16°C) | MP | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |  LP | 30 | 30 | 100% | 30 | 30 | 100%  |
|   |  HN | 10 | 10 | 100% | 7 | 10 | 70%  |
|   |  Neg | 0 | 10 | 0% | 0 | 10 | 0%  |
|  Cycle 3 (-28 to -16°C) | MP | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |  LP | 30 | 30 | 100% | 30 | 30 | 100%  |
|   |  HN | 0 | 10 | 0% | 10 | 10 | 100%  |
|   |  Neg | 0 | 10 | 0% | 0 | 10 | 0%  |
|  Cycle 2 (-84 to -66°C) | MP | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |  LP | 30 | 30 | 100% | 30 | 30 | 100%  |
|   |  HN | 2 | 10 | 20% | 0 | 10 | 0%  |
|   |  Neg | 0 | 10 | 0% | 0 | 10 | 0%  |
|  Cycle 3 (-84 to -66°C) | MP | 10 | 10 | 100% | 10 | 10 | 100%  |
|   |  LP | 30 | 30 | 100% | 30 | 30 | 100%  |
|   |  HN | 0 | 10 | 0% | 0 | 10 | 0%  |
|   |  Neg | 0 | 10 | 0% | 0 | 10 | 0%  |

In summary, freeze/thaw stability testing of contrived specimens preserved in CB or C&amp;S transport media demonstrated no challenges in signal acquisition in the assay when the preserved specimens were stored frozen (in  $-20^{\circ}\mathrm{C}$  and  $-80^{\circ}\mathrm{C}$  freezers) and thawed up to a  $3^{\mathrm{rd}}$  time for testing.

## Expected Values:

Studies on the epidemiology of  $H.$  pylori have shown that this organism is present worldwide. Gastritis caused by  $H.$  pylori has been shown to correlate with age, ethnic background, family size and socioeconomic class. The prevalence of  $H.$  pylori infection in a given population can vary from  $20\%$  to  $90\%$ . In patients diagnosed with duodenal ulcers, however, it has been shown in every age group to be approximately  $80\%$ . Currently recommended eradication treatments have an efficacy rate between  $75\%$  and  $90\%$ .

The proposed subject assay "Premier HpSA Flex" assay is intended to detect the presence of  $H.$  pylori antigens in unpreserved and preserved human stool specimens. Expected values for a given population should be determined for each laboratory. The rate of positivity may vary depending on geographic location, method of specimen collection, handling and transportation, test employed and general health environment of the patient population under study.

K230901 - Page 12 of 14

{12}

8. Assay Cut-Off:

Not Applicable

B Comparison Studies:

1. Method Comparison with Comparator Device:

A method comparison study was conducted to evaluate the agreement of the subject assay (i.e., Premier HpSA Flex) with an FDA-cleared comparator assay, for use with human stool specimens preserved in CB or C&amp;S transport media. All specimens enrolled in this study were de-identified, left-over samples of liquid, semi-solid, or formed human stool, that were previously collected from symptomatic patients with suspicions of gastroenteritis and submitted to a clinical laboratory for H. pylori antigen testing under order from a prescribing healthcare provider as a part of the patients' medical care. Each specimen was sourced from one individual patient. De-identified frozen preserved clinical specimens were acceptable for use in this study since the subject device represents a modification to an existing device. Of note, originally cleared assay used fresh prospective specimens and provided support for use of frozen specimens.

From the time of receipt, the specimens were stored frozen, randomized, relabeled, and prescreened against both the current clinical diagnostic Standard of Care (SoC) and with an FDA-cleared device. A total of 182 specimens with 50 positives and 132 negatives successfully passed pre-screening (with matching SoC and pre-screen device results) and were assigned one preservative transport medium (CB or C&amp;S). The unpreserved (raw) specimens were transferred to the transport medium (N=90 in C&amp;S and 92 in CB) prior to being tested with the subject assay and the comparator.

The study was carried out at a single US site using three (3) kit lots of the subject assay (Lots 601396P167, 601396P168, and 601396P169) following the instructions in the Package Insert. In parallel, the FDA-cleared comparator assay was performed with the same preserved specimens on any given day by a separate technician following the assay's Package Insert. The results are summarized in Table 7.

Table 7. Performance comparison of Premier HpSA Flex assay and comparator device.

|   | Comparator Assay  |   |   |
| --- | --- | --- | --- |
|  Premier HpSA Flex Assay | POSITIVE | NEGATIVE | TOTAL  |
|  POSITIVE | 49* | 2 | 51  |
|  NEGATIVE | 0 | 131 | 131  |
|  TOTAL | 49 | 133 | 182  |
|   | Performance |   | Wilson-Score 95% CI  |
|  Positive Agreement | 100% (49/49) |   | [92.7–100%]  |
|  Negative Agreement | 98.5% (131/133) |   | [94.7–99.6%]  |

* One (1) sample that was positive based on SoC result and confirmed by prescreening testing gave a negative result on Comparator Method.

K230901 - Page 13 of 14

{13}

2. Matrix Comparison:
Not Applicable

C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable
2. Clinical Specificity:
Not Applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable

D Clinical Cut-Off:
Not Applicable

E Expected Values/Reference Range:
Not Applicable

VIII Proposed Labeling:
The proposed labeling included with the subject assay/device supports the finding of substantial equivalence for this device.

IX Conclusion:
The submitted information in this premarket notification (K230901) is complete and supports a substantial equivalence decision.

K230901 - Page 14 of 14

---

**Source:** [https://fda-staging.innolitics.com/device/K230901](https://fda-staging.innolitics.com/device/K230901)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
