K230892 · Aju Pharm Co., Ltd. · MAI · Jun 9, 2023 · Orthopedic
Device Facts
Record ID
K230892
Device Name
Fixone Biocomposite Anchor
Applicant
Aju Pharm Co., Ltd.
Product Code
MAI · Orthopedic
Decision Date
Jun 9, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction; Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair; Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, lliotibial Band Tenodesis; Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. The following indications are for the S and SL type anchors only: Hand Wrist. Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Hip: Capsular repair, Acetabular Labral repair
Device Story
Fixone Biocomposite Anchor system; used for reattaching soft tissue to bone. Components: biodegradable anchor (PLGA/β-TCP) pre-loaded with nonabsorbable UHMWPE suture on insertion driver. Types: screw-in (B, N, BN), knotless (Kc, KcN), and push-in (S, SL). Operated manually by surgeons in clinical settings. Anchor implanted into bone; suture secures soft tissue. Provides secure fixation during healing; anchor degrades over time. Benefits: facilitates soft tissue repair in various orthopedic procedures.
Clinical Evidence
Bench testing only. Included biocompatibility, mechanical testing, EO residues, shelf-life, pyrogenicity, and endotoxin monitoring. No clinical data presented.
Technological Characteristics
Materials: PLGA (70%) and β-TCP (30%). Suture: Non-absorbable UHMWPE. Sterilization: EO gas per ISO 11135:2014. Configuration: Cannulated anchor with eyelet, pre-loaded on manual insertion driver. Mechanical principle: Screw-in or push-in fixation. Shelf-life: 5 years.
Indications for Use
Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, knee, elbow, hand/wrist, and hip procedures. Specific procedures include rotator cuff, Bankart, SLAP, biceps, AC separation, deltoid, capsular, ligament, and tendon repairs/reconstructions. S and SL types are indicated for hand/wrist and hip procedures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
June 9, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Aju Pharm Co., Ltd. % Peter Chung President Plus Global 300. Atwood street Pittsburgh, Pennsylvania 15213
# Re: K230892
Trade/Device Name: Fixone Biocomposite Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: March 31, 2023 Received: March 31, 2023
# Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sara S. Thompson -S
For
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230892
Device Name Fixone Biocomposite Anchor
The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, lliotibial Band Tenodesis;
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
The following indications are for the S and SL type anchors only:
Hand Wrist. Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Hip: Capsular repair, Acetabular Labral repair
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
### 1. Applicant
- 1) Company : AJU Pharm Co.,Ltd.
- 2) Address : A-207, 697, Pangyo-ro, Seongnam-si, Gyeonggi-do, Korea
- 3) Tel : 82-31-765-4420
- 4) Fax : 82-31-602-7818
- Preparation date : Mar. 31, 2023 5)
- ୧) Contact person : Peter Chung, 412-512-8802
- 7) Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
- 8) Submission date : June 8, 2023
- 9) Submission type : Special
- 10) Prior related submissions : K171299, K192484, K192032, K203523
- 11) Submission Number : K230892
### 2. Device Information
- Trade name : Fixone Biocomposite Anchor 1)
- Common name : Fastener, Fixation, Biodegradable, Soft Tissue 2)
- 3) Regulation name : Single/multiple component metallic bone fixation appliances and accessories.
- 4) Product code : MAI
- 5) Regulation number : 888.3030
- 6) Class of device : Class II
- 7) Panel : Orthopedic
### 3. The legally marketed device to which we are claiming equivalence
K203523, Aju Pharm Co., Ltd. / Fixone hybrid knotless anchor (Only for UHMWPE suture)
### Reference devices
K171299, Aju Pharm Co., Ltd. / Fixone Biocomposite Anchor
K192484, Aju Pharm Co., Ltd. / Fixone Biocomposite Anchor (Special submission of K171299)
K192032, Aju Pharm Co., Ltd. / Fixone Biocomposite Small Anchor
### 4. Device description
The Fixone Biocomposite Anchor is intended for reattaching soft tissue to bone with sutures. The anchor is manufactured from biodegradable materials (PLGA copolymer and β-TCP). A nonabsorbable UHMWPE suture is inserted into the anchor. The anchor is implanted using a provided driver.
B,N,BN type(Knotted suture anchors) are 'screw-in' anchors used alone or in combination with Kc,KcN type(Knotless suture anchors) for soft-tissue repair.
Kc,KcN type(Knotless suture anchor) are used for a soft-tissue repair in combination with the B,N,BN type.
SL type 'push-in' anchors that are ideal for soft-tissue repair in narrow or deep areas.
S type are 'push-in' anchors that are ideal for soft-tissue repair in the foot, ankle, hand and wrist
The Fixone Biocomposite Anchor consists of cannulated anchors with an eyelet. They are pre-loaded on an insertion device. The Fixone Biocomposite Anchor is intended to provide secure reattachment of the soft tissue to bone.
Devices are provided sterile. Single use only.
{4}------------------------------------------------
#### 5. Intended Use :
#### The Fixone Biocomposite Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee and elbow in the following procedures:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction;
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, lliotibial Band Tenodesis;
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
#### The following indications are for the S and SL type anchors only:
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers
Hip: Capsular repair, Acetabular Labral repair
#### 6. Predicate device comparison table
| Manufacturer | AJU Pharm Co.,Ltd. | AJU Pharm Co.,Ltd. | AJU Pharm Co.,Ltd. | AJU Pharm Co.,Ltd. | Remark |
|--------------|--------------------|--------------------|--------------------|--------------------|--------|
| 510(k) No. | K171299 | K192032 | K192484 | K213008 | N/A |
{5}------------------------------------------------
| Indication for<br>use | The Fixone Biocomposite<br>Anchors are intended for<br>fixation of suture (soft<br>tissue) to bone in the<br>shoulder, foot/ankle,<br>knee and elbow in the<br>following procedures:<br>Shoulder: Rotator Cuff<br>Repair, Bankart Repair,<br>SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-<br>Clavicular Separation<br>Repair, Deltoid Repair,<br>Capsular Shift or<br>Capsulolabral<br>Reconstruction;<br>Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles<br>Tendon Repair;<br>Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament<br>Repair, Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis;<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or<br>Radial Collateral Ligament<br>Reconstruction. | The Fixone Biocomposite<br>Small Anchor are intended<br>to be used for suture (soft<br>tissue) fixation to bone in<br>the foot, ankle, knee,<br>hand, wrist, elbow,<br>shoulder, and hip.<br>Shoulder: Rotator Cuff<br>Repair, Bankart Repair,<br>SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-<br>Clavicular Separation<br>Repair, Deltoid Repair,<br>Capsular Shift or<br>Capsulolabral<br>Reconstruction<br>Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles<br>Tendon Repair, Metatarsal<br>Ligament Repair, Hallux<br>Valgus reconstruction,<br>digital tendon transfers,<br>Mid-foot reconstruction<br>Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament Repair,<br>Patellar Tendon Repair,<br>Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis<br>Hand/Wrist: Scapholunate<br>Ligament Reconstruction,<br>Carpal Ligament<br>Reconstruction,<br>Repair/Reconstruction of<br>collateral ligaments,<br>Repair of Flexor and<br>Extensor Tendons at the<br>PIP, DIP, and MCP joints<br>for all digits, digital tendon<br>transfers<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or<br>Radial Collateral Ligament<br>Reconstruction<br>Hip: Capsular repair,<br>Acetabular Labral repair | The Fixone Biocomposite<br>Anchors are intended for<br>fixation of suture (soft<br>tissue) to bone in the<br>shoulder, foot/ankle,<br>knee and elbow in the<br>following procedures:<br>Shoulder: Rotator Cuff<br>Repair, Bankart Repair,<br>SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-<br>Clavicular Separation<br>Repair, Deltoid Repair,<br>Capsular Shift or<br>Capsulolabral<br>Reconstruction;<br>Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles<br>Tendon Repair;<br>Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament Repair,<br>Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis;<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or<br>Radial Collateral Ligament<br>Reconstruction. | The Fixone Biocomposite<br>Anchors are intended for<br>fixation of suture (soft<br>tissue) to bone in the<br>shoulder, foot/ankle,<br>knee and elbow in the<br>following procedures:<br>Shoulder: Rotator Cuff<br>Repair, Bankart Repair,<br>SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-<br>Clavicular Separation<br>Repair, Deltoid Repair,<br>Capsular Shift or<br>Capsulolabral<br>Reconstruction;<br>Foot/Ankle: Lateral<br>Stabilization, Medial<br>Stabilization, Achilles<br>Tendon Repair;<br>Knee: Medial Collateral<br>Ligament Repair, Lateral<br>Collateral Ligament Repair,<br>Posterior Oblique<br>Ligament Repair, Iliotibial<br>Band Tenodesis;<br>Elbow: Biceps Tendon<br>Reattachment, Ulnar or<br>Radial Collateral Ligament<br>Reconstruction.<br>The following indications<br>are for the S and SL type<br>anchors only:<br>Hand/Wrist: Scapholunate<br>Ligament Reconstruction,<br>Carpal Ligament<br>Reconstruction,<br>Repair/Reconstruction of<br>collateral ligaments,<br>Repair of Flexor and<br>Extensor Tendons at the<br>PIP, DIP, and MCP joints<br>for all digits, digital tendon<br>transfers<br>Hip: Capsular repair,<br>Acetabular Labral repair | Similar |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Classification<br>name | Fastener, Fixation,<br>Biodegradable, Soft Tissue | Fastener, Fixation,<br>Biodegradable, Soft Tissue | Fastener, Fixation,<br>Biodegradable, Soft Tissue | Fastener, Fixation,<br>Biodegradable, Soft Tissue | Same |
| Trade name | Fixone Biocomposite<br>Anchor | Fixone Biocomposite Small<br>Anchor | Fixone Biocomposite<br>Anchor | Fixone Biocomposite<br>Anchor | Same |
{6}------------------------------------------------
| Manufacturer | AJU Pharm Co.,Ltd. | AJU Pharm Co.,Ltd. | AJU Pharm Co.,Ltd. | AJU Pharm Co.,Ltd. | Remark |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------|
| Model/type | 80 model codes including<br>BAB-45001a | 9 model codes including<br>SAB-30001a | 161 model codes including<br>BAB-45001a | 211 model codes including<br>BAB-45002ah | N/A |
| Product<br>configuration | Driver | Driver | Driver | Driver | |
| | Anchor | Anchor | Anchor | Anchor | Same |
| | Suture | Suture | Suture | Suture | |
| Material | PLGA(70%) + B-TCP(30%) | PLGA(70%) + B-TCP(30%) | PLGA(70%) + B-TCP(30%) | PLGA(70%) + B-TCP(30%) | Same |
| Anchor | | | | | |
| Outside<br>diameter | 4.5mm/4.75mm/5.5mm/5<br>.75mm/6.5mm | 3.0mm | 3.0mm/4.5mm/4.75mm<br>5.5mm/5.75mm/6.5mm | 3.0mm/4.5mm/4.75mm<br>5.5mm/5.75mm/6.5mm | Similar |
| Length of<br>anchor | 14.6mm/15mm/15.7mm/<br>16mm/16.8mm | 11.9mm | 11.9mm/14.6mm/15mm/<br>15.7mm/16mm/16.5mm/<br>16.8mm | 11.9mm/14.6mm/15mm/<br>15.7mm/16mm/16.5mm/<br>16.8mm/17mm/17.7mm | Similar |
| Fixation<br>mechanism | Anchor holds soft tissue<br>and suture. Anchor and<br>suture could endure the<br>tension of moving during<br>degradation of anchor. | Anchor holds soft tissue<br>and suture. Anchor and<br>suture could endure the<br>tension of moving during<br>degradation of anchor. | Anchor holds soft tissue<br>and suture. Anchor and<br>suture could endure the<br>tension of moving during<br>degradation of anchor. | Anchor holds soft tissue<br>and suture. Anchor and<br>suture could endure the<br>tension of moving during<br>degradation of anchor. | Same |
| Inner diameter | Same as subject | Same as subject | Same as subject | Same as predicate | Same |
| Degradation | Average 15% to 4 weeks | Average 15% to 4 weeks | Average 15% to 4 weeks | Average 15% to 4 weeks | Same |
| Suture | | | | | |
| Absorbable | Non-absorbable | Non-absorbable | Non-absorbable | Non-absorbable | Same |
| Suture<br>diameter | 0.50~0.599 (USP size 2) | 0.35~0.399 (USP size 0) | 0.50~0.599 (USP size 2) | 0.35~0.399 (USP size 0)<br>0.50~0.599 (USP size 2) | Similar |
| Material | Polyethylene | Polyethylene | Polyethylene | Polyethylene (UHMWPE) | Same |
| Colorant | Blue / Green / Violet | Blue / Green / Violet | Blue / Green / Violet | Blue / Green / Violet /<br>Black | Similar |
| Sterilization | EO Gas sterilization<br>According to ISO 11135:<br>2014 | EO Gas sterilization<br>According to ISO 11135:<br>2014 | EO Gas sterilization<br>According to ISO 11135:<br>2014 | EO Gas sterilization<br>According to ISO 11135:<br>2014 | Same |
| Biodegradable | No | No | No | No | Same |
| Principle of<br>operation | Manual | Manual | Manual | Manual | Same |
| Shelf-life | 5 years | 5 years | 5 years | 5 years | Same |
| Inserter<br>configuration | Shaft that holds the<br>anchor from inside with<br>suture | Shaft that holds the<br>anchor from inside with<br>suture | Shaft that holds the<br>anchor from inside with<br>suture | Shaft that holds the<br>anchor from inside with<br>suture | Same |
K192032, Fixone Biocomposite Small Anchor could be used in smaller surgery spot like hand, wrist and hip. Differences of intended use was just come from device size.
The differences between predicate and subject device which could effect the performance of device is thread and suture.
Thread – Added new performance test data of dual-threaded anchors. KcN type is from the Kc type. KcN type anchor has same design with Kc type anchor except only the thread. BN type anchor. The eyelet part of BN type is different with the eyelet of N type anchor. BN type has same design with N type anchor except only the eyelet design.
Suture – Added new performance test data of only UHMWPE suture. (Previously cleared in K203523)
These differences could be explained with new test report and also could be explained by predicate devices because all of these predicate devices were made with same manufacturing process of subject device.
#### 7. Performance data:
Bench test were performed. Bench testing included biocompatibility, mechanical testing including EO residues, shelf-life testing, pyrogenicity testing and endotoxin monitoring. The tests demonstrated that the device performs in a substantially equivalent manner to the prior related devices.
#### 8. Conclusion
The device is investigated for function to compare the operation of function between Fixone Biocomposite Anchor and prior related devices.
{7}------------------------------------------------
Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.
Therefore, it is concluded that Fixone Biocomposite Anchor is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.