K230859 · Wenzhou Kangshun Medical Devices Co.,Ltd · DXQ · Jun 28, 2023 · Cardiovascular
Device Facts
Record ID
K230859
Device Name
Aneroid Sphygmomanometer
Applicant
Wenzhou Kangshun Medical Devices Co.,Ltd
Product Code
DXQ · Cardiovascular
Decision Date
Jun 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The aneroid sphygmomanometer is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on children or adults. The device is intended to a patient and manually inflated along with a manual method for detecting Korotkoff sounds. There are a variety of stethoscopes can be an optioned accessory,depending on the model.
Device Story
Manual non-invasive aneroid sphygmomanometer; consists of inflation cuff, manual inflation bulb, and aneroid manometer gauge. Used in home, hospital, or clinical settings by medical professionals or patients. Operator wraps cuff around patient's upper arm (or leg for children); manually inflates cuff using bulb; listens for Korotkoff sounds via stethoscope (optional accessory). Manometer displays pressure readings (0-300 mmHg). Device provides manual blood pressure measurement; aids clinicians in patient assessment and monitoring.
Clinical Evidence
No clinical data. Bench testing only. Performance validated per ISO 81060-1:2007. Biocompatibility validated per ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2021, and ISO 10993-23:2021 (cytotoxicity, skin sensitization, and irritation tests).
Technological Characteristics
Manual aneroid sphygmomanometer. Materials: Zinc alloy/aluminum (manometer), PVC (tubing/bulb), Nylon (cuff). Measurement range: 0-300 mmHg; 2 mmHg interval. Accuracy: +/- 3 mmHg. Non-electronic/mechanical operation. Biocompatible per ISO 10993. Performance per ISO 81060-1.
Indications for Use
Indicated for measurement of systolic and diastolic blood pressure in children and adults. Intended for use by medical professionals or in home settings. Requires manual inflation and manual detection of Korotkoff sounds.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 28, 2023
Wenzhou Kangshun Medical Devices Co.,Ltd Eric Shi OA&OC Manager No. 706 Yanyun Road,Lingkun Street,Oujiangkou Industrial Cluster District Wenzhou City, Zhejiang 325011 China
Re: K230859
Trade/Device Name: Aneroid Sphygmomanometer Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: March 29, 2023 Received: March 29, 2023
Dear Eric Shi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Hetal B. Patel -S
for
Robert Kazmierski Acting Assistant Director Division of Cardiac Electrophysiology Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230859
Device Name Non-Automated Blood Pressure Meter
#### Indications for Use (Describe)
The aneroid sphygmomanometer is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on children or adults. The device is intended to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
There are a variety of stethoscopes can be an optioned accessory,depending on the model.
Type of Use (Select one or both, as applicable)
| <span> <span style="text-decoration:overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="text-decoration:overline;">X</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Sponsor: Wenzhou Kangshun Medical Devices Co.,Ltd. Subject Device: Aneroid Sphygmomanometer 510(k) number: K230859 File No.: Chapter 6 510(k) Summary
# 510(k) Summary
#### 1.Submitter Information
Company Name:Wenzhou Kangshun Medical Devices Co.,Ltd. Address:No.706 Yanyun Road,Lingkun Street,Oujiangkou Industrial Cluster District,Wenzhou City,Zhejiang 325011,China. Phone:+86-13857771939 Contact Person (including title):Eric Shi (QA&QC Manager) E-mail:eric-shi@isosh.com Subject Device Information Traditional Type of 510(k): Common Name: Non-Automated Blood Pressure Meter Trade Name: Aneroid Sphygmomanometer Regulation Name: Blood Pressure Cuff Product Code: DXQ,LDE Regulation Number: 21 CFR 870.1120 Regulation Class: 2 2.Predicate Device Information Common Name: WX non-Automated Blood Pressure Meter Regulation Name: Blood Pressure Cuff 510(k) number: K212416 Product Code: DXQ Regulation Number: 21 CFR 870.1120 Regulation Class: 2
# 3.Device Description
The product is a manual non-invasive aneroid sphygmomanometer which respectively uses an inflation cuff wrapped around the upper arm.
{4}------------------------------------------------
The cuff is inflated and deflated by a manual inflation bulb. Besides a manometer (Aneroid gauge), the accessories of MODEL Ks-1222,Ks-1621,Ks-3201,Ks-2203 also includes one selected cuff, one inflation bulb, instruction manual and a vinyl bag.
It is conjunction with Stethoscope when use.
### 4.Indications for Use
The product is intended to be used by medical professionals or in the home for the measurement of
systolic and diastolic pressure on children or adults. The device is intended to be manually attached
to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
There are a variety of stethoscopes can be an optioned accessory,depending on the model.
# 5.Summary of Comparison and Technological Characteristics
#### Table 1 - General Comparison
| Elements of<br>Comparison | Subject Device | Predicate Device | Comment |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Product Name | Non-Automated Blood<br>Pressure Meter | WX non-Automated Blood Pressure<br>Meter | -- |
| General Comparison | | | |
| 510(k) Number | K230859 | K212416 | N/A |
| Device Classification | Blood Pressure Cuff | Blood Pressure Cuff | Same |
| Regulation Number | 21 CFR 870.1120 | 21 CFR 870.1120 | Same |
| Product Code | DXQ,LDE | DXQ | Same |
| Indications for Use | The product is intended to be<br>used by medical professionals<br>or in the home for the<br>measurement of systolic and<br>diastolic pressure on children<br>or adults.The device is<br>intended to be manually<br>attached to a patient and<br>manually inflated along with a | WX non-Automated Blood Pressure<br>Meter is intended to be used by<br>medical professionals or in the<br>home for the measurement of<br>systolic and diastolic pressure on<br>children or adults.<br>The device is intended to be<br>manually attached to a patient and<br>manually inflated along with a | Same |
| Elements of<br>Comparison | Subject Device | Predicate Device | Comment |
| | manual method for detecting<br>Korotkoff sounds.<br>There are a variety of<br>stethoscopes can be an<br>optioned accessory,depending<br>on the model. | manual method for detecting<br>Korotkoff sounds.<br>There are a variety of stethoscopes<br>can be an optioned accessory,<br>depending on the model.<br>The device is applicable to all<br>patients at least 2 years of age. | |
| Model | Ks-1222,Ks-1621,Ks-3201,<br>Ks-2201,Ks-2203 | WX02,WX0201, WX0202,<br>WX0203 | |
| Over-TheCounter Use | Yes | Yes | |
| Where used | Home, Hospital, heathcare<br>facility, ambulance etc. | Home, Hospital, heathcare facility,<br>ambulance etc. | Same |
| Target population | Children or adults | Infants, children, young adults, and<br>adults | Different 1 |
| Anatomical sites | Upper Arm (leg for child) | Upper Arm (leg for child) | Same |
| Measurement Method | Auscultatory Korotkoff sounds<br>Method | Auscultatory Korotkoff sounds<br>Method | Same |
| Inflation | Manual | Manual | Same |
| Deflation | Manual | Manual | Same |
| Display | Aneroid Manometer | Aneroid Manometer | Same |
| The monitor scale | 0 to 300 mmHg with a<br>minimum interval of 2 mmHg | From 0 to 300 mmHg with a<br>minimum interval of 2 mmHg | Same |
| Design of blood<br>pressure meter | The device comprises tubing<br>attached to a soft inelastic cuff<br>with an integrated inflatable<br>bladder that is wrapped | The device comprises tubing<br>attached to a soft inelastic cuff with<br>an integrated inflatable bladder that<br>is wrapped around the patient's limb | Same |
| Elements of<br>Comparison | Subject Device | Predicate Device | Comment |
| | around the patient's limb and<br>secured by hook and loop<br>closure. | and secured by hook and loop<br>closure. | |
| Design<br>of Stethoscope | Three types of option:<br>Single head<br>Dual head<br>Sprague Rappaport | Three types of option:<br>Single head<br>Dual head<br>Sprague Rappaport | Same |
| Materials | The manometer:<br>Zinc Alloy or Aluminum<br>materials.<br>The tubing, inflation bulb:<br>PVC.<br>Cuff:Nylon cloth. | The manometer:<br>Zinc Alloy or Aluminum materials.<br>The tubing,inflation bulb:<br>PVC or nature latex<br>Cuff:<br>Nylon cloth or cotton cloth for<br>outside layer.<br>Cuff bladder:<br>PVC or nature latex | Same |
| Accuracy | Pressure:<br>+/- 3 mmHg of reading | Pressure:<br>+/- 3 mmHg of reading | Same |
| Compatibility<br>with<br>environment | It can be used from 50°F to<br>104°F (10°C to 40°C) and 15%<br>~ 85%RH humidity | It can be used from 50°F to<br>104°F(10°C to 40°C) and 15% ~<br>85%RH humidity | Same |
| Cuff Size | Ks-1222:500mm*140mm<br>Ks-1621:520mm*140mm<br>Ks-2201:630mm*175mm<br>Ks-2203:720mm*210mm<br>Ks-3201:380mm*110mm | 20"x 5.5" (510mm*140mm)<br>21.7"x 6.3" (550mm*160mm)<br>24.4"x 6.9" (620mm*175mm)<br>28.3"x 8.3" (720mm*210mm)<br>13.4"x 4.15" (340mm*105mm) | Different 2 |
| Cuff Circumference | Fits arm circumferences | Fits arm circumferences 220mm to | Same |
| Elements of<br>Comparison | Subject Device | Predicate Device | Comment |
| | 220mm to 440 mm. The<br>standard cuff should be<br>available for use in measuring<br>a child's leg blood pressure<br>and for children with larger<br>arms | 440 mm. The standard cuff should<br>be available for use in measuring a<br>child's leg blood pressure and for<br>children with larger arms | |
| Cuff Color | Blue | Blue, Pink, Black | Same |
| Contents<br>(with<br>accessories) | Aneroid gauge, Arm Cuff, Inflation Bulb, Vinyl storage<br>pouch and Instruction Manual,<br>Stethoscope (option) | Aneroid gauge, Arm Cuff, Inflation<br>Bulb, Vinyl storage pouch and<br>Instruction Manual, Stethoscope<br>(option) | Same |
| Biocompatibility | Conforms to the requirements<br>of ISO 10993 series<br>standards. | Biocompatible as requirement of<br>ISO 10993-1 ISO 10993-5 ISO<br>10993-10 | Same |
| Performance | compatible as requirement of<br>ISO 81060-1 | compatible as requirement of ISO<br>81060-1 | Same |
{5}------------------------------------------------
Wenzhou Kangshun Medical Devices Co.,Ltd. Sponsor: Subject Device: Aneroid Sphygmomanometer 510(k) number: K230859 File No.: Chapter 6 510(k) Summary
{6}------------------------------------------------
Wenzhou Kangshun Medical Devices Co.,Ltd. Sponsor: Subject Device: Aneroid Sphygmomanometer 510(k) number: K230859 File No.: Chapter 6 510(k) Summary
{7}------------------------------------------------
Sponsor: Wenzhou Kangshun Medical Devices Co.,Ltd. Subject Device: Aneroid Sphygmomanometer 510(k) number: K230859 File No.: Chapter 6 510(k) Summary
Different 1: The target population of the product is for children or adults and the target population of the predicate device is for Infants, children, young adults, and adults.
Different 2: The cuff size of Ks-1222,Ks-1621,Ks-2201 is shorter than predicate device.The cuff size of Ks-2203 is equivalent that predicate device.The cuff size of Ks-3201 is more than predicate device
# 6.Non-clinical Tests Performed on the Proposed Device
The company has performed non-clinical performance testing based on its risk assessment utilizing Failure Mode Effect Analysis (FMEA).
Following Quality System processes,required testing was conducted to validate the cumulative
modifications made to the subject devices.
{8}------------------------------------------------
| Sponsor: | Wenzhou Kangshun Medical Devices Co.,Ltd |
|-----------------|------------------------------------------|
| Subject Device: | Aneroid Sphygmomanometer |
| 510(k) number: | K230859 |
| File No.: | Chapter 6 510(k) Summary |
Substantial Equivalence is being supported with full performance testing representing the current devices currently marketed.
#### Performance test
• ISO 81060-1:2007 Non-invasive sphygmomanometersPart 1: Requirements and test methods for
non-automated measurement type
### Biocompatibility test
Testing was conducted according to relevant sections of ISO 10993:2019 Biological evaluation of
medical devices.
Biocompatibility testing included:
- MTT cytotoxicity test
- Skin Sensitization Test
- Skin Irritation Test
- ISO 10993-1:2018 Biological evaluation of medical devices-Part 1: Evaluation and testing within a
risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices-Part 10: Tests for skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices-Part 23: Tests for irritation
# 7.Clinical Testing
Not Applicable
# 8.Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(k)
submission K212416,the product is as safe,as effective.
#### 9.Summary Prepared Date
22 June 2023
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.