United Orthopedic Knee Patient Specific Instrumentation
K230850 · Enhatch, Inc. · OOG · Dec 20, 2023 · Orthopedic
Device Facts
Record ID
K230850
Device Name
United Orthopedic Knee Patient Specific Instrumentation
Applicant
Enhatch, Inc.
Product Code
OOG · Orthopedic
Decision Date
Dec 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Anatomical segmentation of femur and tibia
Deep learning algorithms
—
—
—
—
—
—
Implant size and position selection
Optimization algorithm
—
—
—
—
Full use simulations tests using cadaver specimens
>1 (surgeons)
Indications for Use
The United Orthopedic Knee Patient Specific Instrumentation is indicated as an orthopedic instrument system to assist in the positioning of compatible total knee arthroplasty systems. It is comprised of surgical planning software (Intelligent Surgery Knee CT Segmentation Engine/Knee X-ray Segmentation Engine, and Implant Recognition Engine) intended to preoperatively plan the surgical placement of the United Orthopedics Knee implants on the basis of provided patient radiological images and 3D reconstructions of bones with identifiable anatomical landmarks, and surgical instrument components that include patient specific or customized guides fabricated on the basis of the surgical plan to precisely reference the placement of the implant components intra-operatively per the surgical plan. The United Orthopedic Knee Patient Specific Instrumentation is indicated for patients without severe bone deformities, such as a HKA greater than 15° or deformities due to prior fracture of the distal femur or proximal tibia. The instruments are intended for use with the U2 Total Knee System when the clinical evaluation complies with its cleared indications for use. The instruments are intended for single use only.
Device Story
System assists orthopedic surgeons in total knee arthroplasty (TKA) by providing patient-specific surgical guides and anatomical models. Input: patient CT or X-ray DICOM images. Software (Intelligent Surgery Knee Segmentation Engine) uses deep learning to segment femur/tibia and generate 3D models. Intelligent Surgery Knee Implant Recognition Engine uses optimization algorithms to determine implant size and position based on surgical technique. Enhatch technicians manage the SaaS preoperative planning workflow; surgeon reviews and approves the plan. Output: 3D-printed patient-specific surgical guides and anatomical models. Guides are used intra-operatively to reference implant placement. Benefits include precise implant positioning based on patient anatomy, potentially improving surgical outcomes. System is non-adaptive; algorithms are static.
Clinical Evidence
Bench testing only. Includes segmentation system testing (CT/X-ray DICOM accuracy), 3D model verification, software system testing (landmark detection accuracy), guide wear testing, and full-use cadaver specimen simulations performed by multiple surgeons to validate overall system performance and accuracy.
Technological Characteristics
System comprises software (SaaS) and 3D-printed patient-specific hardware. Software utilizes deep learning for image segmentation and optimization algorithms for implant planning. Connectivity is web-based. Hardware is single-use. Software is classified as moderate level of concern. Design controls follow ISO 14971 and IEC 62304.
Indications for Use
Indicated for adult patients undergoing total knee arthroplasty using the U2 Total Knee System. Contraindicated for patients with severe bone deformities, including HKA > 15° or deformities resulting from prior distal femur or proximal tibia fractures.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
United Orthopedic Corporation U2 Total Knee System (K051640)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
December 20, 2023
Enhatch, Inc. Monica Williams QA/RA Manager 226 Boulevard Hasbrouck Heights, New Jersey 07604
Re: K230850
Trade/Device Name: United Orthopedic Knee Patient Specific Instrumentation Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: OOG, PBF, LLZ Dated: March 24, 2023 Received: March 28, 2023
Dear Monica Williams:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Lixin Liu -S
Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No, 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
#### Indications for Use
510(k) Number (if known) K230850
Device Name
United Orthopedic Knee Patient Specific Instrumentation
#### Indications for Use (Describe)
The United Orthopedic Knee Patient Specific Instrumentation is indicated as an orthopedic instrument system to assist in the positioning of compatible total knee arthroplasty systems. It is comprised of surgical planning software (Intelligent Surgery Knee CT Segmentation Engine/Knee X-ray Segmentation Engine, and Implant Recognition Engine) intended to preoperatively plan the surgical placement of the United Orthopedics Knee implants on the basis of provided patient radiological images and 3D reconstructions of bones with identifiable anatomical landmarks, and surgical instrument components that include patient specific or customized guides fabricated on the surgical plan to precisely reference the placement of the implant components intra-operatively per the surgical plan. The United Orthopedic Knee Patient Specific Instrumentation is indicated for patients without severe bone deformities, such as a HKA greater than 15° or deformities due to prior fracture of the distal femur or proximal tibia.
The instruments are intended for use with the U2 Total Knee System when the clinical evaluation complies with its evaluation complies with its cleared indications for use. The instruments are intended for single use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) |
|--|----------------------------------------------|--|---------------------------------------------|
|--|----------------------------------------------|--|---------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (6/20) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|
{4}------------------------------------------------
## 510(k) Summary
#### 1. INTRODUCTION
This document contains the 510(k) summary for the United Orthopedic Knee Patient Specific Instrumentation. The content of this summary is based on the requirements of 21 CFR 807.92.
## 2. SUBMITTER
3.
| Name: | Enhatch, Inc. |
|-------------------------|-----------------------------------------------------------------------------------------------|
| Address: | 226 Boulevard<br>Hasbrouck Heights, NJ 07604<br>Phone: (201) 771-2034 |
| Official Correspondent: | Monica Williams<br>QA/RA Manager |
| Date Prepared: | November 20, 2023 |
| | <b>DEVICE</b> |
| Trade Name: | United Orthopedic Knee Patient Specific Instrumentation |
| Regulation Number: | 21 CFR 888.3560 |
| Regulation Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis |
| Regulatory Class: | Class II |
| Product Code: | OOG, PBE, LLZ |
#### 4. PREDICATE DEVICES
Primary predicate device:
- X-PSI Knee System, Orthosoft (K171269) .
Secondary predicate devices:
- VSP Orthopedics System, 3D Systems (K211244) ●
- United Orthopedic Corporation U2 Total Knee System (K051640) .
{5}------------------------------------------------
#### 5. DESCRIPTION OF THE DEVICE
The United Orthopedic Knee Patient Specific Instrumentation is comprised of: United Orthopedics (UO) surgical guides (hardware), anatomical models (physical replica), Intelligent Surgery Knee CT Segmentation Engine / Intelligent Surgery Knee X-ray Segmentation Engine (software), and Intelligent Surgery Knee Implant Recognition Engine (software). Enhatch is responsible for design and development of all three components of the system.
The subject device is intended to facilitate the implantation of the U2 Knee protheses under the U2 Knee System developed and distributed by United Orthopedics Corporation: U2 Total Knee System.
#### [THE SOFTWARE]
The Intelligent Surgery Knee Segmentation Engine consists of two imaging modalities, CT (Knee CT Segmentation) and X-ray (Knee X-ray Segmentation). The Intelligent Surgery Knee CT Segmentation Engine and X-ray Segmentation Engine are web applications that use deep learning algorithms to detect and extract region of interest (ROI) information (femur and tibia) from medical imaging data (DICOM). The segmentation engines generate 3D models which can be used for treatment planning of Total Knee Arthroplasty (TKA), design of surgical guides, or generation of 3D printed anatomical models.
The Intelligent Surgery Knee Implant Recognition Engine is a web application that uses an optimization algorithm as a treatment planning tool for total knee arthroplasty. It assists in selecting implant size and position, from a range of implants of a total knee implant system, using a range of run parameters based upon the TKA surgical technique of that system. The software identifies anatomical landmarks of the patient's bony anatomy and articular surface topographies to reference the position and alignment of the femoral and tibial implant components. This positioning and alignment in turn allows for design of surgical guides and of the United Orthopedic Knee Patient Specific Instrumentation .
Note, these algorithms are static and non-adaptive; they do not alter their behavior over time based on user input.
#### [THE HARDWARE]
The UO surgical guides and anatomical models are patient-specific instruments designed to facilitate the implantation of the United Orthopedics Knee protheses. The UO surgical guides are designed based on preoperative plan generated by the software Intelligent Surgery Knee Implant Recognition Engine.
{6}------------------------------------------------
### 6. INDICATIONS FOR USE
The United Orthopedic Knee Patient Specific Instrumentation is indicated as an orthopedic instrument system to assist in the positioning of compatible total knee arthroplasty systems. It is comprised of surgical planning software (Intelligent Surgery Knee CT Segmentation Engine/Knee X-ray Segmentation Engine, and Implant Recognition Engine) intended to preoperatively plan the surgical placement of the United Orthopedics Knee implants on the basis of provided patient radiological images and 3D reconstructions of bones with identifiable anatomical landmarks, and surgical instrument components that include patient specific or customized guides fabricated on the basis of the surgical plan to precisely reference the placement of the implant components intraoperatively per the surgical plan. The United Orthopedic Knee Patient Specific Instrumentation is indicated for patients without severe bone deformities, such as a HKA greater than 15° or deformities due to prior fracture of the distal femur or proximal tibia. The instruments are intended for use with the U2 Total Knee System when the clinical evaluation complies with its cleared indications for use. The instruments are intended for single use only.
## 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technical features of United Orthopedic Knee Patient Specific Instrumentation are substantially equivalent to the primary predicate, Orthosoft X-PSI Knee System (K171269), and secondary predicate devices. 3D Systems VSP Orthopedics System (K211244) and United Orthopedic Corporation U2 Total Knee System (K051640). The subject device has similar indications for use as compared to the predicates. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Includes software as a medical device component
- . Image transfer and manipulation via software to be used for 3D printing of anatomical models and guides for surgical planning
- . Visualization of radiological images (CT and X-ray), image segmentation, and 3D reconstruction of bones
- Display and measurement evaluations of anatomical landmarks that must be viewed, . accepted, or rejected by a clinician
- . For use by prespecified clinical users
The following technological differences exists between the subject and predicate devices:
{7}------------------------------------------------
- . Internal facing versus external user interface, United Orthopedic Knee Patient Specific Instrumentation software follows a SaaS preoperative planning workflow where Enhatch technicians interface directly with the software and provide end user outputs
- Segmentation edge detection versus manual outlining of patient contours, United . Orthopedic Knee Patient Specific Instrumentation has the ability to semiautomatically identify bony contours for the alignment of the final algorithmic model using deep learning algorithms whereas predicate device
- Implant sizing / placement with an optimization algorithm versus manual user controls, United Orthopedic Knee Patient Specific Instrumentation does not allow end user to perform measurements in software but uses an optimization algorithm to find the approximate location of implant size / position approved by surgeon via a preoperative plan
The following table provides details of the United Orthopedic Knee Patient Specific Instrumentation as compared to the primary and secondary predicates.
| Technological<br>Characteristics | Subject Device:<br>United Orthopedic<br>Knee Patient Specific<br>Instrumentation | Primary<br>Predicate:<br>Orthosoft X-PSI<br>Knee System<br>(K171269) | Predicate Device:<br>3D Systems VSP<br>Orthopedics System<br>(K211244) | Predicate Device:<br>United Orthopedic<br>Corporation U2<br>Total Knee System<br>(K051640) | Comparison |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use (IFU) | The United<br>Orthopedic Knee<br>Patient Specific<br>Instrumentation is<br>indicated as an<br>orthopedic instrument<br>system to assist in the<br>positioning of<br>compatible total knee<br>arthroplasty systems. It<br>is comprised of<br>surgical planning<br>software (Intelligent<br>Surgery Knee CT<br>Segmentation<br>Engine/Knee X-ray<br>Segmentation Engine,<br>and Implant<br>Recognition Engine)<br>intended to<br>preoperatively plan the<br>surgical placement of | The X-PSI Knee<br>System is indicated<br>as an orthopedic<br>instrument system<br>to assist in the<br>positioning of knee<br>replacement<br>components. It<br>involves surgical<br>planning software<br>used pre-<br>operatively to plan<br>the surgical<br>placement of the<br>components on the<br>basis of provided<br>patient radiological<br>images and 3-D<br>reconstructed<br>bones with<br>identifiable<br>placement<br>anatomical<br>landmarks, and<br>surgical instrument | The VSP®<br>Orthopedics System<br>is intended to be used<br>as a surgical<br>instrument to assist in<br>preoperative planning<br>and/or in guiding the<br>marking of bone<br>and/or in guiding<br>surgical instruments<br>in non-acute, non-<br>joint replacing<br>osteotomies for adult<br>patients in the distal<br>femur, tibia, and non-<br>sacrum pelvis. | This device is<br>indicated in knee<br>arthroplasty for<br>reduction or relief of<br>pain and/or improved<br>knee function in<br>skeletally mature<br>patients with severe<br>knee pain and<br>disability due to<br>rheumatoid arthritis,<br>osteoarthritis,<br>primary and<br>secondary traumatic<br>arthritis, polyarthritis,<br>collagen disorders,<br>avascular necrosis of<br>the femoral condyle<br>or pseudogout,<br>posttraumatic loss of<br>joint configuration,<br>particularly when | Substantially<br>Equivalent<br>All fundamental aspects<br>of the indications for<br>use are the same (e.g.<br>visualization of<br>radiological images, 3D<br>reconstruction of bones<br>and includes tools to<br>improve visualization,<br>positioning of knee<br>replacement<br>components, use of<br>clinical judgement, use<br>with fixed bearing knee<br>replacement systems) |
| the United<br>Orthopedics Knee<br>implants on the basis<br>of provided patient<br>radiological images<br>and 3D reconstructions<br>of bones with<br>identifiable anatomical<br>landmarks, and<br>surgical instrument<br>components that<br>include patient specific<br>or customized guides<br>fabricated on the basis<br>of the surgical plan to<br>precisely reference the<br>placement of the<br>implant components<br>intra-operatively per<br>the surgical plan. The<br>United Orthopedic<br>Knee Patient Specific<br>Instrumentation is<br>indicated for patients<br>without severe bone<br>deformities, such as a<br>HKA greater than 15°<br>or deformities due to<br>prior fracture of the<br>distal femur or<br>proximal tibia.<br><br>The instruments are<br>intended for use with<br>the U2 Total Knee<br>System when the<br>clinical evaluation<br>complies with its<br>cleared indications for<br>use. The instruments<br>are intended for single<br>use only. | components that<br>include patient<br>specific or<br>customized guides<br>fabricated on the<br>basis of the<br>surgical plan to<br>precisely reference<br>the placement of<br>the<br>implant<br>components intra-<br>operatively per the<br>surgical plan. The<br>X-PSI Knee<br>system is indicated<br>for patients without<br>severe bone<br>deformities, such<br>as a HKA greater<br>than 15° or<br>deformities due to<br>prior fracture of the<br>distal femur or<br>proximal tibia.<br><br>The X-PSI Knee<br>System is to be<br>used with the<br>following fixed<br>bearing knee<br>replacement<br>systems in<br>accordance with<br>their indications<br>and<br>contraindications:<br>NexGen® CR,<br>NexGen CR-Flex,<br>NexGen CR-Flex<br>Gender, NexGen<br>LPS, NexGen LPS-<br>Flex, NexGen<br>LPS-Flex Gender,<br>Persona® CR,<br>Persona PS,<br>Vanguard® CR<br>and Vanguard PS.<br>The patient<br>specific guide<br>components are<br>intended for single-<br>use only. | there is<br>patellofemoral<br>erosion, dysfunction<br>or prior patellectomy,<br>moderate valgus,<br>varus, or flexion<br>deformities. This<br>device may also be<br>indicated in the<br>salvage of previously<br>failed surgical<br>attempts if the knee<br>can be satisfactorily<br>balanced and<br>stabilized at the time<br>of surgery. This<br>device system is<br>designed for<br>cemented use only. | | | |
| Device Class | Class II | | Same | | |
| Device Code | OOG, PBF, LLZ | OOG, IWH, MBH | PBF | IWH | Substantially |
| | | LLZ | | | Equivalent |
| Patient<br>Population | Adult patients | Patients without<br>severe bone<br>deformities | Adult patients | Skeletally mature<br>patients | Substantially<br>Equivalent |
| End User | | Prespecified clinical users (orthopedic surgeons) | | | Equivalent |
| Image<br>Modality | CT, X-ray | X-ray | CT | N/A | Substantially<br>Equivalent |
| Software<br>Processing | Image segmentation,<br>implant predictability,<br>dimensioning of<br>digital representations<br>of models, generation<br>of PSI | Image<br>segmentation,<br>implant<br>predictability,<br>dimensioning of<br>digital<br>representations of<br>models, generation<br>of PSI | Image segmentation,<br>dimensioning of<br>digital<br>representations of<br>models, generation of<br>PSI | N/A | Substantially<br>Equivalent<br>Software processing<br>functionality is<br>consistent |
| Preoperative<br>Planning | Provides tool to<br>improve visualization,<br>generate 3D models<br>and perform model<br>measurements to<br>choose, size, and<br>position the implant | Provides tool to<br>improve<br>visualization,<br>generate 3D<br>models and<br>perform model<br>measurements to<br>choose, size, and<br>position the<br>implant | Provides tool to<br>improve<br>visualization,<br>generate 3D models | N/A | Substantially<br>Equivalent<br>Subject device tools for<br>preoperative planning<br>are consistent |
| Output | Patient specific<br>anatomical models and<br>guides (physical<br>outputs); and patient<br>specific surgical plans<br>and digital files<br>(digital outputs) | Patient specific<br>anatomical models<br>and guides<br>(physical outputs);<br>and patient specific<br>surgical plans and<br>digital files (digital<br>outputs) | Patient specific<br>anatomical models,<br>templates, and guides<br>(physical outputs);<br>and patient specific<br>surgical plans and<br>digital files (digital or<br>documentation<br>outputs) | N/A | Substantially<br>Equivalent<br>Device outputs are<br>consistent |
Table - Comparison Summary
{8}------------------------------------------------
# K230850 page 5 of 8
{9}------------------------------------------------
{10}------------------------------------------------
#### 8. SUMMARY OF PERFORMANCE TESTING
Safety and performance of United Orthopedic Knee Patient Specific Instrumentation has been evaluated and verified in accordance with software specifications and applicable performance standards through Software Development and Verification & Validation Procedures (ISO 14971, IEC 62304:2006 & A1:2016), User Requirements, and Federal Regulations and Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "FDA Guidance on Content of Premarket Submissions for Management of Cybersecurity in Medical Devices." The software contained in this device is considered to be of "moderate level of concern." since a failure or latent design flaw could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
United Orthopedic Knee Patient Specific Instrumentation has been evaluated through the following non-clinical testing:
- . Segmentation System Testing
- o This testing was performed to verify and validate the accuracy of the software generated segmentation masks for CT and X-ray based DICOM data. The device passed the acceptance criteria and demonstrated satisfactory performance per the intended use.
- . Model Verification Testing
- This testing was performed to verify and validate the accuracy of the software o generated 3-dimensional models from CT and X-ray based DICOM data. The device passed the acceptance criteria and demonstrated satisfactory performance per the intended use.
- Software System Testing
- This testing was performed to verify and validate the accuracy of landmark o detection, system usage and accuracy. All specimens were within the bounds of the acceptance criteria. The resulting output measurements from the system were within the bounds of the input parameters (input values produced expected output values). The device passed the acceptance criteria and demonstrated satisfactory performance per the intended use.
- Guide Wear Testing ●
- This testing was performed to verify and validate the amount of wear debris o generated from use of the surgical guides. The device passed the acceptance criteria of average weight loss and demonstrated satisfactory performance per the intended use.
{11}------------------------------------------------
- System Verification and Validation Test
- This testing was performed to verify and validate the amount overall system O performance in terms of system usage, and accuracy. Full use simulations tests using cadaver specimens were performed by multiple surgeons to verify and validate the overall system performance. The results demonstrated satisfactory performance per the intended use as in the predicate
In conclusion, United Orthopedic Knee Patient Specific Instrumentation performance testing has demonstrated the subject device's safety and effectiveness and meets its intended use statement.
#### CONCLUSION
The result of the comparison of the intended use/indications and technological characteristics of United Orthopedic Knee Patient Specific Instrumentation as compared to the primary predicate and secondary predicates demonstrate that United Orthopedic Knee Patient Specific Instrumentation is substantially equivalent to the chosen predicate devices. Additionally, the results of testing demonstrate that United Orthopedic Knee Patient Specific Instrumentation is as safe and effective as the predicate device and does not raise different questions relating to safety and/or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.