← Product Code [NEU](/productcode/NEU) · K230836

# SchurSign Tissue Marker (K230836)

_Surgmark GmbH · NEU · Jan 23, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230836

## Device Facts

- **Applicant:** Surgmark GmbH
- **Product Code:** [NEU](/productcode/NEU.md)
- **Decision Date:** Jan 23, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The SchurSign Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.

## Device Story

SchurSign Tissue Marker is a sterile, single-use implantable device used to mark soft tissue locations during open, percutaneous, or endoscopic surgical procedures. The device consists of a chitosan-based marker containing Barium Sulfate, preloaded into a rigid-shaft delivery system (cannula with handle, push rod, plunger, and end cap). The clinician inserts the delivery system into the target tissue and deploys the marker by depressing the plunger. The marker is visible via X-ray, mammography, ultrasound, and MRI. By providing a permanent radiographic landmark, the device assists clinicians in localizing specific tissue sites for subsequent surgical or diagnostic interventions, potentially improving surgical precision and patient outcomes.

## Clinical Evidence

Bench testing and animal studies provided. Ex vivo ultrasound imaging in chicken breast confirmed performance equivalence to the predicate. An in vivo swine study evaluated localization via X-ray and ultrasound, supported by histopathology, demonstrating performance equivalent to the predicate. Biocompatibility testing included cytotoxicity (ISO 10993-5), acute/subacute systemic toxicity (ISO 10993-11), sensitization (ISO 10993-10), irritation (ISO 10993-23), implantation (ISO 10993-6), pyrogenicity (USP-NF 2021), genotoxicity (ISO 10993-3), and chemical characterization (ISO 10993-18). All tests met acceptance criteria.

## Technological Characteristics

Marker material: Chitosan with Barium Sulfate. Delivery system: Stainless steel cannula (14-10 gauge, 12 cm length). Form factor: Cylindrical marker (1.5-3.0 mm diameter, 5-10 mm length). Sterilization: Ethylene oxide (EO). Imaging: X-ray, mammography, ultrasound, MRI. Single-patient use, preloaded delivery system.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Beacon ([K130763](/device/K130763.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

January 23, 2024

SurgMark GmbH Christine König CEO Maria-Louisen-Straße 122 Hamburg, 22301 Germany

Re: K230836

Trade/Device Name: SchurSign Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: April 27, 2023 Received: December 12, 2023

Dear Christine König:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of 

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tek N. Digitally signed by Lamichhane Tek N. Lamichhane -S
Date: 2024.01.23 -5 17:29:36 -05'00' Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230836

Device Name SchurSign Tissue Marker

Indications for Use (Describe) Under supervision of a healthcare professional

· The SchurSign Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.

Type of Use (Select one or both, as applicable)

| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-.052z" fill-rule="evenodd"></path> </svg>     Prescription Use (Part 21 CER 801 Subpart D)    </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div>                                                                                         |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Date Prepared: January 22, 2024

I. SUBMITTER

# Submitter of 510(k):

SurgMark GmbH Maria-Louisen-Straße 122 D-22301 Hamburg Germany +49 177 4075795 / info@surgmark.com www.surgmark.com

Contact Person: Dr. Christine König Christine.koenig@surgmark.com

# II. DEVICE

Name of Device: SchurSign Tissue Marker

Common or Usual Name: Implantable Clip

Classification Panel: General and Plastic Surgery

Classification of the device: Class II, 21 CFR 878.4300

Product Code: NEU

III PREDICATE DEVICE

Primary Predicate: Beacon (K130763) Manufactured by Scion Medical Technologies, LLC

## IV DESCRIPTION

SchurSign Tissue Marker consists of a radiographic soft tissue marker and a delivery system. SchurSign is a sterile, single patient use, discrete marker that is visible on standard radiographs (x-ray, mammography) as well as ultrasound, and Magnetic Resonance Imaging (MRI).

The proposed SchurSign Tissue Marker is placed into soft tissue during open, percutaneous, or endoscopic procedures to mark a surgical location.

{5}------------------------------------------------

The proposed SchurSign Tissue Marker is comprised of chitosan filled with Barium Sulfate.

The proposed SchurSign Tissue Marker delivery system is a distal delivery needle tip, rigid shaft, sterile, and single patient use preloaded delivery system incorporating the SchurSign Tissue Marker.

The delivery system consists of a cannula with a handle, a push rod with a plunger, and an end cap. The tissue marker is retained within the delivery system until placement is desired, where it is delivered through the end port by fully depressing the plunger into the handle. The SchurSign Tissue Marker delivery system is used to place the SchurSign Tissue Marker into soft tissue during open, percutaneous, or endoscopic procedures to radiographically mark a surgical location. The delivery system device has a bevelled 12 cm / 14 to 10 gauge needle with 1 cm depth marks and a plunger.

SchurSign is available in seven different sizes:

| Model            | Diameter (mm) | Length (mm) |
|------------------|---------------|-------------|
| SchurSign 1.5-5  | 1.5           | 5           |
| SchurSign 1.5-8  | 1.5           | 8           |
| SchurSign 2.0-5  | 2.0           | 5           |
| SchurSign 2.0-8  | 2.0           | 8           |
| SchurSign 2.0-10 | 2.0           | 10          |
| SchurSign 2.5-10 | 2.5           | 10          |
| SchurSign 3.0-10 | 3.0           | 10          |

## V. INDICATIONS FOR USE

Under supervision of a healthcare professional (Rx)

- SchurSign Tissue Marker is indicated for use to radiographically mark soft ● tissue during a surgical procedure or for future surgical procedures.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

SchurSign has substantially equivalent indications to the Beacon Tissue Marker (K130763) in that they are indicated for use to mark soft tissue during a surgical procedure or for future surgical procedures.

SchurSign and Beacon are both made of polymeric material containing barium sulfate and have a cylindrical shape. Moreover, they are also sterilized by ethylene oxide.

Both SchurSign and Beacon have the same delivery system, which is a distal delivery needle tip, rigid shaft, sterile, and single patient use preloaded delivery system incorporating the marker. The delivery system consists of a cannula with a handle, a push rod with a plunger, and an end cap. The tissue marker is retained within the delivery system until placement is desired, where it is delivered through the end port by fully depressing the plunger into the handle. The SchurSign Tissue Marker delivery system is used to place the SchurSign Tissue Marker into soft tissue

{6}------------------------------------------------

during open, percutaneous, or endoscopic procedures to radiographically mark a surgical location.

The following table compares the SchurSign device to the predicate device with respect to intended use, material characteristics, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

| Parameter                                                | Device                                                                                                                                                                       | Primary Predicate<br>Device                                                                                                                                               | Equivalence |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Trade name                                               | SchurSign Tissue<br>Marker                                                                                                                                                   | Beacon Tissue<br>Marker                                                                                                                                                   | -           |
| Company Name                                             | SurgMark GmbH                                                                                                                                                                | Scion Medical<br>Technologies, LLC                                                                                                                                        | -           |
| 510(k) #                                                 | K230836                                                                                                                                                                      | K130763                                                                                                                                                                   | -           |
| Product Code                                             | NEU                                                                                                                                                                          | NEU                                                                                                                                                                       | Same        |
| Indications For<br>Use                                   | The SchurSign<br>Tissue Marker is<br>indicated for use to<br>radiographically<br>mark soft tissue<br>during a surgical<br>procedure or for<br>future surgical<br>procedures. | The Beacon Tissue<br>Marker is indicated<br>for use to<br>radiographically<br>mark soft tissue<br>during a surgical<br>procedure or for<br>future surgical<br>procedures. | Same        |
| Type of<br>polymeric<br>material                         | Chitosan                                                                                                                                                                     | Oxford Performance<br>Materials (OPM)<br>OXPEKK-IG200                                                                                                                     | Different   |
| Physical form                                            | Cylinder                                                                                                                                                                     | Cylinder                                                                                                                                                                  | Same        |
| Dimensions of<br>tissue marker                           | Length: 5-10 mm,<br>Diameter: 1.5-3.0<br>mm                                                                                                                                  | Length: 5 mm,<br>Diameter: 1.5 mm                                                                                                                                         | Different   |
| Is the marker for<br>single use?                         | Yes                                                                                                                                                                          | Yes                                                                                                                                                                       | Same        |
| X-ray contrast<br>agent<br>incorporated in<br>the marker | Barium Sulfate                                                                                                                                                               | Barium Sulfate                                                                                                                                                            | Same        |
| Sterilization<br>Method                                  | EO                                                                                                                                                                           | EO                                                                                                                                                                        | Same        |
| Imaging modality                                         | X-ray,<br>mammography,<br>ultrasound, and<br>Magnetic<br>Resonance Imaging                                                                                                   | X-ray,<br>mammography,<br>ultrasound, and<br>Magnetic<br>Resonance Imaging                                                                                                | Same        |
| Type of delivery<br>system                               | Distal delivery<br>needle tip, rigid                                                                                                                                         | Distal delivery<br>needle tip, rigid                                                                                                                                      | Same        |

# DEVICE COMPARISON CHART

{7}------------------------------------------------

|                                                                                | shaft and sterile.                                                        | shaft and sterile.                                                        |           |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------|
| Dimensions of<br>canula                                                        | Length: 12 cm<br>Diameter: 14 - 10<br>gauge                               | Length: 12 cm<br>Diameter: 14 gauge                                       | Different |
| Components of<br>the delivery<br>system                                        | Cannula with a<br>handle, a push rod<br>with a plunger, and<br>an end cap | Cannula with a<br>handle, a push rod<br>with a plunger, and<br>an end cap | Same      |
| Material of<br>canula                                                          | Stainless steel                                                           | Stainless steel                                                           | Same      |
| Is the tissue<br>marker already<br>preloaded in the<br>delivery system?        | Yes                                                                       | Yes                                                                       | Same      |
| Are the tissue<br>marker and the<br>delivery system<br>sterilized<br>together? | Yes                                                                       | Yes                                                                       | Same      |

The only difference between SchurSign and Beacon is the polymeric composition of the tissue marker and that SchurSign comes in different sizes. To prove that SchurSign is as safe and efficacious as its predicate device, a series of tests were done as shown below.

## VII. PERFORMANCE DATA

In vitro - Performance data were provided in support of the substantial equivalence determination. The characteristics of the SchurSign Tissue Marker were substantially equivalent to the predicate device. SchurSign is equivalent to Beacon in an ex vivo test of ultrasound imaging of chicken breast.

In vivo - A swine study was done where x-ray and ultrasound were used to localize SchurSign. Histopathology was also conducted. The animal study showed that the performance of SchurSign was equivalent to its predicate device.

## Biocompatibility testing

#### In vitro and in vivo tests:

To demonstrate that SchurSign is biocompatible, the following tests were done in accordance with ISO standards:

- 1) Cytotoxicity according to ISO 10993-5
- 2) Acute systemic toxicity according to ISO 10993-11
- 3) Sensitization according to ISO 10993-10
- 4) Irritation according to the ISO 10993-23

{8}------------------------------------------------

- 5) Subacute toxicity according to ISO 10993-11
- 6) Implantation according to ISO 10993-6
- 7) Pyrogenicity according to USP-NF 2021
- 8) Genotoxicity according to ISO 10993-3

9) Chemical characterization - according to ISO ISO 10993-18

Chronic toxicity and carcinogenicity endpoints were addressed through chemical characterization and toxicological risk assessment. It was concluded that neither the components nor potential leachables of SchurSign are of risk for the patient.

| Test                                    | Result                                                                                                                                                                                                                      |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity                            | The test article extract showed no cytotoxic potential<br>to L-929 mouse fibroblast cells.                                                                                                                                  |
| Acute systemic toxicity                 | There was no mortality or evidence of systemic<br>toxicity from the extracts injected into mice.                                                                                                                            |
| Sensitization                           | The topical application of the SC and SO extracts<br>did not induce delayed sensitization in the guinea<br>pig. The test article was not considered a sensitizer.                                                           |
| Irritation/Intracutaneous<br>reactivity | The test article met the requirements of the test<br>since the difference between each test extract<br>overall mean score and corresponding control blank<br>overall mean score was 0.0 for the SC and SO test<br>extracts. |
| Subacute toxicity                       | There was no evidence of systemic toxicity from the<br>test article 4 weeks following subcutaneous<br>implantation in the rat.                                                                                              |
| Implantation                            | The macroscopic reaction was not significant as<br>compared to the negative control article.<br>Microscopically, the test article caused a moderate<br>reaction as compared to the negative control article.                |
| Pyrogenicity                            | The test article met the requirements of the US<br>Pharmacopoeia and was judged as non-pyrogenic.                                                                                                                           |
| Genotoxicity                            | The SC and DMSO test article extracts were<br>considered to be non-mutagenic in the bacterial<br>reverse mutation study.                                                                                                    |

## VIII. CONCLUSIONS

The SchurSign Tissue Marker to be distributed by SurgMark GmbH is substantially equivalent in design and function to Beacon Tissue Marker (K130763) manufactured by Scion Medical Technologies, LLC. The subject device has the same intended use and similar characteristics and functional properties as the predicate device. Moreover, documentation supplied in this submission demonstrates that any differences in their technological characteristics or materials do not raise any new questions of safety or performance, and that the non-clinical data for the device supports the safety of the device.

---

**Source:** [https://fda-staging.innolitics.com/device/K230836](https://fda-staging.innolitics.com/device/K230836)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
