Modius Sleep is a non-invasive, home-use neurostimulation device to treat chronic insomnia in adults aged 22 and older.
Device Story
Modius Sleep is a battery-powered, non-invasive transdermal neurostimulation device for home use. It delivers low-level electrical energy (up to 1mA) to the vestibular nerve via two self-adhesive electrode pads placed on the skin behind the ears (mastoid processes). The device uses electrical vestibular nerve stimulation (VeNS) to treat chronic insomnia. Users adjust stimulation intensity via buttons on the device enclosure. The device is operated by the patient; it is not possible to recharge the battery while the device is in use. The device provides a 30-minute treatment session. Healthcare providers use the device output (patient-reported improvement in insomnia symptoms) to monitor treatment progress. The device benefits patients by providing a drug-free therapy for chronic insomnia.
Clinical Evidence
A 4-week, multi-site, randomized, double-blind, sham-controlled trial (NCT04452981) enrolled 126 participants (mean age 41, 67% female) with chronic insomnia. Primary endpoint was change in Insomnia Severity Index (ISI) score from baseline to 4 weeks. The Modius Sleep group showed a statistically significant greater reduction in ISI score compared to the sham control group (complete case analysis: -5.80 vs -3.52, p<0.001). Adverse events were minor, self-resolving, and comparable between groups (12% in Modius Sleep vs 8% in sham).
Technological Characteristics
Plastic enclosure with PCBA (microcontroller, transformer driver, EEPROM). Delivers symmetrical biphasic rectangular pulses (0.25Hz, 1mA max) via two self-adhesive electrodes. Powered by 3.75V Lithium Polymer battery. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Connectivity: micro-USB for charging only.
Indications for Use
Indicated for the treatment of chronic insomnia in adults aged 22 and older.
Regulatory Classification
Identification
A cranial electrotherapy stimulator is a prescription device that applies electrical current that is not intended to induce a seizure to a patient's head to treat psychiatric conditions.
Special Controls
In combination with the general controls of the FD&C Act, the virtual reality behavioral therapy device for pain relief is subject to the following special controls:
*Classification.* (1) Class II (special controls) when intended to treat insomnia and/or anxiety. The special controls for this device are:(i) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device to treat insomnia and/or anxiety.
(ii) Components of the device that come into human contact must be demonstrated to be biocompatible.
(iii) The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC) in its intended use environment.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The technical parameters of the device, including waveform, output mode, pulse duration, frequency, train delivery, maximum charge, and energy, must be fully characterized and verified.
(vi) The labeling for the device must include the following:
(A) The intended use population and the intended use environment;
(B) A warning that patients should be monitored by their physician for signs of worsening;
(C) A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur;
(D) A warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur;
(E) A detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(F) Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions. This information must be based on the results of clinical studies for the device;
(G) A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and
(H) Information on validated methods for reprocessing any reusable components between uses.
(vii) Cranial electrotherapy stimulator devices marketed prior to the effective date of this reclassification must have an amendment submitted to the previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
(2) Class III (premarket approval) when intended to treat depression.
(c) Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 19, 2020, for any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 19, 2020, been found to be substantially equivalent to any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 27, 2023
Neurovalens Limited Jason McKeown, MD CEO 8 Carmagrim Road Portglenone. BT44 8BP United Kingdom
Re: K230826
Trade/Device Name: Modius Sleep Regulation Number: 21 CFR 882.5800 Regulation Name: Cranial electrotherapy stimulator Regulatory Class: Class II Product Code: QJQ Dated: March 21, 2023 Received: March 24, 2023
Dear Dr. McKeown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Pamela D. Scott -S
Pamela Scott, MS Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230826
Device Name
Modius Sleep
Indications for Use (Describe)
Modius Sleep is a non-invasive, home-use neurostimulation device to treat chronic insomnia in adults aged 22 and older.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# EUROVALENS
### Modius Sleep
Traditional 510(k)
Version: 01
Page: 1 of 30
## 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
#### Submitter l.
Neurovalens Limited 8 Carmagrim Road Portglenone County Antrim BT44 8BP
United Kingdom
### Contact person/ official correspondence:
Dr Jason McKeown CEO Neurovalens Ltd. Tel: +44 (0) 7412508707 Email: Jason.mckeown@neurovalens.com
### Date Prepared
27th October 2023
#### II. Device
Name of Device: Modius Sleep Common or usual name: Electrical Vestibular Nerve Stimulator (VeNS) Classification name: Cranial Electrotherapy Stimulator (CES) Regulatory Class: II Product Code: Requested Regulation number: 882.5800 Classification panel: Neurology Special Controls: Modius Sleep complies with the special controls as set in FDA 21 CFR 882.5800 for a Class II, Cranial Electrical Stimulation (CES) device.
{5}------------------------------------------------
# ROVALENS
### Modius Sleep
Traditional 510(k)
Page: 2 of 30
## 510(k) Summary
Version: 01
#### Predicate devices lll.
Primary predicate device CES Ultra, K062284 Secondary predicate device Alpha-Stim® CS, K903014
No reference devices were used in this submission.
#### Device description IV.
Modius Sleep is a low-risk non-invasive transdermal neurostimulation device to treat chronic insomnia. The proposed mechanism of the device is through a technology known as electrical vestibular nerve stimulation (VeNS).
It consists of a battery-powered device designed to transcutaneously deliver low-level electrical energy (up to 1mA) to the skin behind the ears, over the mastoid processes. The delivery of this neurostimulation is through two self-adhesive electrode pads. These pads are placed on the skin behind each ear (mastoid area). The intensity of the electric pulse can be adjusted up or down by the user. When turned on, the device delivers a small electrical impulse which stimulates the vestibular nerve. Adjustments to the stimulation level may be made using the up and down buttons on the device, which are located just above the power button. The device can also be paused by pressing the power button twice. When finished the user removes the device and disposes of the electrode pads after each use. When the device is not being used it can be charged through a micro-USB cable. For safety reasons, it is not possible to recharge the battery while the device is in use in stimulation mode.
{6}------------------------------------------------
# OVALENS
## Modius Sleep Traditional 510(k)
Version: 01
Page: 3 of 30
## 510(k) Summary
The key components of the Modius Sleep include the Modius Sleep device (plastic enclosure and printed circuit board assembly (PCBA), stimulation pads (K210448 and K132588), skin cleansing wipes (K121655) and charging accessories. The PCBA consists of a microcontroller, USB connector (charging only), transformer driver, IO expander and EEPROM memory. The embedded software within the device manages overall functionality of the device from Stimulation Control, Power Management, and user interaction (Indication LED and Audio tones).
Biocompatibility testing in accordance with the endpoints of ISO 10993-1 to be addressed for a surface medical device, with prolonged: > 24 hours less than 30 days contact duration, demonstrated good biocompatibility for all components of the Modius Sleep device.
#### Device Usage
Daily sessions of Modius Sleep has been shown to improve insomnia when assessed at 4 weeks.
After using Modius Sleep there are usually no physical limitations imposed so most users can resume normal activities immediately. Some users may have a response that affect their ability to perform potentially hazardous tasks, such as operating a motor vehicle or heavy machinery for up to several hours after treatment. However, no significant lasting side effects have been reported.
{7}------------------------------------------------
# EUROVALENS
## Modius Sleep
Traditional 510(k)
Version: 01
Page: 4 of 30
## 510(k) Summary
#### Indications for use V.
Modius Sleep is a non-invasive, home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older.
### VI. Comparison of the technological characteristics with the predicate device
Modius Sleep is substantially equivalent with respect to indications for use, stimulation parameters (i.e., current levels, frequencies, pulse width and amplitude) and electrode placement to the predicate devices CES Ultra (cleared by K062284), and Alpha-Stim® CS (cleared by K903014).
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "NEUROVALENS" in all capital letters. The font is a sans-serif font, and the color is a dark teal. The letters are evenly spaced and the word is centered.
Modius Sleep
Traditional 510(k)
Page: 5 of 30
## 510(k) Summary
Version: 01
#### Table 1. Summary of substantial equivalence
| Property | Proposed Device | Primary Predicate | Secondary Predicate | Comment |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Device | Modius Sleep | CES Ultra | Alpha-Stim® CS | |
| Indications for Use | Modius Sleep is a non-invasive, home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older. | The CES Ultra is indicated for the treatment of insomnia, depression or anxiety | The Alpha-Stim® 100 is a precision medical instrument used for the management of pain, anxiety, depression, and/or insomnia. | All devices are indicated for the treatment of insomnia. |
| Target Population | Adults 22 and older | Adults | Adults, and children under adult supervision | N/A |
| Environment | Home | Home | Home | N/A |
| Waveform | Symmetrical Biphasic Rectangular Wave | Symmetrical Biphasic Rectangular Wave | Symmetrical Biphasic Rectangular Wave | Identical |
| Current Intensity Range | $0\u03bc$ A - 1000 $\u03bc$ A | $0\u03bc$ A - 1500 $\u03bc$ A continually adjustable | $0\u03bc$ A - 500 $\u03bc$ A | Modius Sleep and the CES Ultra have near identical current intensity ranges |
| Pulse Width Range | 1s | 2ms | 250ms - 1s | Pulse width varies depending on frequency range |
| Number of electrodes | Two | Two | Two | Identical |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "NEUROVALENS" in a sans-serif font. The word is written in a dark teal color. The letters are all capitalized and evenly spaced.
Modius Sleep
Traditional 510(k)
Page: 6 of 30
## 510(k) Summary
Version: 01
| Property | Proposed Device | Primary Predicate | Secondary Predicate | Comment |
|---------------------|---------------------------------|-------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Electrode placement | Mastoid | Mastoid or earlobes | Head - earlobes | The Modius Sleep electrodes are placed on<br>mastoids. CES Ultra also supports mastoid only<br>electrode placement. |
| Power Source | 3.75 Lithium Polymer<br>Battery | 9V Alkaline Battery | 2 x AAA NiMH | |
| Frequency | 0.25Hz | 100Hz | 0.5Hz, 1.5Hz, 100Hz | |
| Treatment Range | 30 min | 30min, 60min or<br>continuous | 10min, 20min, 60 min or<br>continuous | Modius Sleep maximum treatment time is within<br>the range of the predicates |
| Unit Controls | Built into the device | Built into the device | Built into the device | Identical |
| Dimensions | 16.5cm x 15.1cm x<br>6.6cm | 13.5cm x 6.4cm x<br>3.3cm | 9.8cm x 6.3cm x 2cm | Modius Sleep is larger than the predicate<br>devices |
| Enclosure | Plastic | Plastic | Plastic | Identical |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image displays the logo of a company called NEUROVALENS. Below the logo, the text "Version: 01" is visible. The text "510(k) Summary" is at the bottom of the image.
### Table 2. Safety Information Comparison
| Property | Proposed Device | Primary Predicate | Secondary Predicate | Comment | |
|-------------------|--------------------------------|----------------------------------------------|-------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|
| | Modius Sleep | CES Ultra | Alpha-Stim® CS | | |
| Electrical Safety | Complies with IEC<br>60601-1 | Complies with IEC<br>60601 | Complies with<br>IEC 60601 | The proposed device and predicate devices are<br>identical | |
| EMC | Complies with IEC<br>60601-1-2 | Complies with IEC<br>60601-1-2 | Complies with IEC<br>60601-1-2 | The proposed device and predicate devices are<br>identical | |
| Software | Level of<br>Concern | Moderate | Moderate | Moderate | The proposed device and predicate devices are<br>identical |
| | Verification &<br>Validation | Complies with<br>FDA Guidance<br>Requirement | Complies with FDA<br>Guidance Requirement | Complies with FDA<br>Guidance Requirement | |
| Biocompatibility | Complies with ISO<br>10993 | Complies with ISO<br>10993 | Complies with<br>ISO 10993 | The proposed device and predicate devices are<br>identical | |
Modius Sleep
Traditional 510(k)
Page: 7 of 30
{11}------------------------------------------------
| NEUROVALENS | Modius Sleep |
|----------------|--------------------|
| | Traditional 510(k) |
| Version: 01 | Page: 8 of 30 |
| 510(k) Summary | |
Table 3. Technical Information for Modius Sleep Output Parameters
| | Modius Sleep | CES Ultra | Alpha-Stim CS | |
|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|----------------------|----------------------------------|----------------------------|
| 510(k) Number | K230826 | K062284 | K903014 | |
| Device Name | Modius Sleep | CES ultra | Alpha-Stim CS | |
| Manufacturer | Neurovalens Ltd | Neuro-Fitness LLC | Electromedical Products Inc. | |
| Power Source† | 3.75 Lithium Polymer<br>Battery | 9V Alkaline Battery | 2 x AAA NiMH | |
| -Method of Line Current Isolation | DC:DC transformer | N/A Non Rechargeable | N/A Non-Rechargeable | |
| - Patient Leakage Current (as per ANSI/AAMI 60601-1) | 0μΑ | Unknown | Unknown | |
| -Normal Condition (μΑ) | 0-1000μΑ | 0-1500μΑ | 0-500μΑ | |
| -Single Fault Condition (μΑ) | 2.3μA | Unknown | 1.3mA | |
| Average DC current through the electrodes when the device is on but no pulses are being delivered (µA) | 0μΑ | Unknown | Unknown | |
| Number of output channels | 1 | 1 | 1 | |
| -If more than one channel, is the stimulus delivered to each channel synchronous or alternating between each channel? | N/A | N/A | N/A | |
| -If more than one channel, describe method of channel isolation | N/A | N/A | N/A | |
| Software/Firmware/Microprocessor Control | Yes | Yes | Yes | |
| Automatic Overload Trip? | No | No | No | |
| Automatic Shut Off? | Yes | Yes | Yes | |
| | | Modius Sleep | CES Ultra | Alpha-Stim CS |
| User Override Control? | | Yes | Yes | Yes |
| Mode or Program Name | | Stimulation mode | Stimulation mode | Stimulation/Treatment Mode |
| | Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | Biphasic |
| | Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular | Rectangular |
| | Maximum Output Voltage (volts) (+/-1%) | 500mV @500 Ω | Unknown | @500 Ω |
| | | 2 V @ 2 kΩ | Unknown | @ 2 kΩ |
| | | 10V @10 kΩ | Unknown | @10 kΩ |
| | Maximum Output Current (specify units) (+/-2%) | 1000μA @500 Ω | Unknown | 500μA @ 500 Ω |
| | | 1000μA @ 2 kΩ | 1.5mA @ 2kΩ | 500μA @ 2 kΩ |
| | | 1000μA @10 kΩ | Unknown | 500μA @10 kΩ |
| | Duration of primary (depolarizing) phase† (msec) | 4000<br>(4 seconds) | 10ms | 2000ms, 1000ms, 10ms |
| | Pulse Duration†† (msec) | 2000<br>(2 second) | 5ms | 1000ms, 500ms, 5ms |
| | Frequency†† (Hz) [or Rate†† (pps)] | 0.25Hz | 100Hz | 0.5Hz, 1Hz, 100Hz |
| | For interferential modes only: Beat Frequency† (Hz) | | | |
| | For multiphasic waveforms Symmetrical phases? only: | | | |
| | Phase Duration† (include units),<br>(state range, if applicable), (both phases, if asymmetrical) | | Not Applicable for these devices | |
| | | Modius Sleep | CES Ultra | Alpha-Stim CS |
| Net Charge (microcoulombs (mC) per pulse) (If zero, state method of achieving zero net charge.) | | 1000 μC @2kΩ | 0.75μC @500 Ω | 300μC, 150 μC, 1.5 µC |
| Maximum Phase Charge, (μC) | | 1000μC @2kΩ | 0.75μC@ 2 kΩ | 500µC @ 500Ω |
| Maximum Current Density,††† (mA/cm², r.m.s.) | | 0.5mA /cm²<br>@500 Ω | 1.91mA/cm²<br>@500 Ω | ~1.2mA/cm² @ 500Ω |
| Maximum Average Current (average absolute value), mA | | 1mA @2kΩ | 3.07mA/cm² @<br>2 kΩ | Unknown |
| Maximum Average Power Density,††† (W/cm²), (using smallest electrode conductive surface area) | | 0.99mW/cm² @2 kΩ | 9.19mW/cm² @2 kΩ | Unknown |
| Burst Mode†††† (i.e., pulse trains): | (a) Pulses per burst | 2 | 2 | 2, 2, 2 |
| | (b) Bursts per second | 0.25 | 100 | 0.5, 1, 100 |
| | (c) Burst duration (seconds) | 4 | 0.01 (10ms) | 2, 1, 0.01s |
| | (d) Duty Cycle: Line (b) x Line (c) | 0.5 (50%) | 0.2(20%) | 0.5 (50%) |
| ON Time (seconds) | | 2 sec | 10ms | 1sec, 0.5sec, 10ms |
| OFF Time (seconds) | | 2 sec | 40ms | 1sec, 0.5sec, 10ms |
{12}------------------------------------------------
# NEUROVALENS
Traditional 510(k)
Page: 9 of 30
## 510(k) Summary
Version: 01
{13}------------------------------------------------
# NEUROVALENS
Traditional 510(k)
Page: 10 of 30
## 510(k) Summary
Version: 01
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "NEUROVALENS" in a bold, sans-serif font. The letters are a dark teal color. The word is displayed prominently, taking up a significant portion of the image.
Traditional 510(k)
Page: 11 of 30
## 510(k) Summary
Version: 01
### Table 4. Comparison of Device Outputs
| Output value | Modius Sleep | CES Ultra | Alpha Stim CS |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------|
| Output value at | Provide for all load conditions. | Provide for all load conditions. | Provide for all load conditions. |
| 500Ω, 2kΩ and 10kΩ | 1mA at all loads | 1.5mA at all loads | 0.5mA at all loads |
| Load conditions | 0.25Hz | 100Hz | 0.5Hz |
| Electrode<br>surface | Conductive Area (16mm diameter) | Ear clips | 1.5cm Diameter |
| area in cm2 | 2.0cm² | Diameter 7.9mm<br>0.5cm² Area | 1.77cm² |
| Current density | 500Ω<br>0.5mA/cm², | 500Ω<br>1.91mA/cm² | 500Ω<br>1.2mA/cm² |
| | 2kΩ<br>0.5mA/cm² | 2kΩ<br>3.07mA/cm² | 2kΩ<br>Unknown |
| | 10kΩ<br>0.5mA/cm² | 10kΩ<br>Unknown | 10kΩ<br>Unknown |
| Charge density | 500Ω<br>497.63µC/cm² – Per Pulse | 500Ω<br>Unknown | 500Ω<br>Max 500uC |
| | 2kΩ<br>497.63µC/cm² – Per Pulse | 2kΩ<br>Unknown | 2kΩ<br>Max 500uC |
| | 10kΩ<br>497.63µC/cm² – Per Pulse | 10kΩ<br>Unknown | 10kΩ<br>Max 500uC |
| Power density | 500Ω,<br>0.25mW/cm²<br>2KΩ<br>0.99mW/cm²<br>10KΩ<br>4.97mW/cm² | 500Ω,<br>Unknown<br>2KΩ<br>9.19ma/cm²<br>10KΩ<br>Unknown | 500Ω,<br>Unknown<br>2KΩ<br>Unknown<br>10KΩ<br>Unknown |
| Max phase charge<br>(pulse width x peak<br>current) | 500Ω,<br>1000 μC – Per Pulse<br>2KΩ<br>1000 μC – Per Pulse<br>10KΩ<br>1000 μC – Per Pulse | 500Ω,<br>Unknown<br>2KΩ<br>0.75 μC<br>10KΩ<br>Unknown | 500Ω,<br>500 μC<br>2KΩ<br>Unknown<br>10KΩ<br>Unknown |
| Max phase charge<br>density (pulse width<br>x peak current) /<br>electrode surface<br>area | 500Ω<br>497.63μC/cm² – Per Pulse<br>2KΩ<br>497.63μC/cm² – Per Pulse<br>10KΩ<br>497.63μC/cm² – Per Pulse | 500Ω<br>Unknown<br>2KΩ<br>Unknown<br>10KΩ<br>Unknown | 500Ω<br>Unknown<br>2KΩ<br>Unknown<br>10KΩ<br>Unknown |
| Max average power<br>density (Duty cycle x<br>peak current)^2 x | 500Ω,<br>0.12mW/cm² | 500Ω,<br>Unknown | 500Ω,<br>Unknown |
| (load Ω) ÷ electrode<br>surface area) | 2kΩ<br>0.5mW/cm²<br>10kΩ<br>2.49mW/cm² | 2kΩ<br>9.19mW/cm²<br>10kΩ<br>Unknown | 2kΩ<br>Unknown<br>10kΩ<br>Unknown |
| Electrode number | 2 Electrodes | 2 Electrodes | 2 Electrodes |
{15}------------------------------------------------
# NEUROVALENS
Traditional 510(k)
Page: 12 of 30
## 510(k) Summary
Version: 01
{16}------------------------------------------------
| NEUROVALENS | Modius Sleep Traditional 510(k) |
|----------------|---------------------------------------------|
| | Version: 01 |
| | Page: 13 of 30 |
| 510(k) Summary | |
{17}------------------------------------------------
# NEUROVALENS
### Modius Sleep
Traditional 510(k) Version: 01
Page: 14 of 30
## 510(k) Summary
#### Performance Data VII.
#### Non-clinical tests
The Modius Sleep was evaluated for its safety and effectiveness based on the following testing:
#### Table 5: Non-clinical tests
| Test Name | Test Description | Results |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device Ship/Transport<br>Testing | Ensure device, enclosed in the selected shipping container,<br>meets ASTM D4169 specifications. | Passed |
| Biocompatibility Testing | Testing and analysis of the Modius Sleep device has<br>demonstrated compliance to ISO 10993-1: Biological<br>evaluation of medical devices - Guidance | Passed |
| Electrical Safety | The Modius Sleep device was tested to confirm that it met<br>the applicable standards for electrical safety (IEC 60601-1) | Passed |
| EMC | The Modius Sleep device was tested to confirm that it met<br>the applicable standards for electromagnetic compatibility<br>(EMC) (IEC 60601-1-2) | Passed |
| Usability Testing | Modius Sleep was assessed with regards to usability for<br>compliance with IEC 62366 - Medical devices - Application<br>of usability engineering to medical devices | Passed |
| Maximum Current &<br>Phase Charge Testing | The Maximum Current & Phase Charge Testing was to detail<br>the Max Current / Power output of the system under different<br>load conditions and determine the Max Phase Charge, Max<br>Current Density and Max Power Density of the Modius Sleep<br>device. | Passed |
| System Verification and<br>Validation Testing | The system verification and validation testing was performed<br>to verify the hardware and firmware of the Modius Sleep<br>Device | Passed |
{18}------------------------------------------------
Version: 01
Page: 15 of 30
## 510(k) Summary
#### Labeling
The labeling for the device included the intended use population and the intended use environment; warning that patients should be monitored by their physician for signs of worsening; A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur; a warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur; a detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device; Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions based on the results of clinical studies for the device; A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and Information on validated methods for reprocessing any reusable components between uses.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The software for this device was considered as a "moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
#### Animal Studies
No animal studies were conducted as part of submission to prove substantial equivalence.
#### Clinical Studies
{19}------------------------------------------------
Version: 01
Page: 16 of 30
## 510(k) Summary
CES devices collectively demonstrate a class effect of CES for treating anxiety and/or insomnia. However, it cannot be concluded, based on available information alone, that specific CES devices or stimulation parameters are effective for treating anxiety and/or insomnia. As such, individuals using this device should work with the prescribing medical provider to determine the best treatment settings to use. As such, please see below for a general summary of pertinent clinical literature that has been published using various combinations of CES devices, stimulation settings, and electrode positions.
### General CES Summary (January 1, 1970 to November 8, 2022)
As of November 2022, there were a total of 23 studies that evaluated the effectiveness of CES on insomnia. Of the 11 randomized control trials (RCT), some reported greater reduction in insomnia symptoms in the CES group compared to placebo12345, while others reported no significant differences in measures of insomnia symptoms between the 2 groups®73,9;10;11 .Among the 8 observational studies, CES treatment was associated with less frequent12 and less intense13 sleep disturbances, improved soundness of sleep14, less difficulty falling asleep14;15;16 and feeling more rested15 or improved mood14 in the morning. Several observational studies reported no impact of CES on insomnia1718 or reported an effect that did not persist after the first week19. A meta-analysis of 14 RCTs indicated that CES versus sham treatment had no impact on insomnia20. Lastly, 3 reviews evaluated the effectiveness of CES on insomnia. One review found 23 studies, 6 of which were RCTs of which only 2 studied participants with insomnia21. Of these two RCTs, only 1 showed improved patient-reported sleep latency and objective changes in sleep, while the other did not show any change21. The other 2 reviews concluded that studies evaluating the effectiveness of CES on insomnia were inconclusive22 or inconsistent²3. In studies that reported improvement in insomnia with use of CES, the reported stimulation parameters, electrode placement, and treatment schedule varied widely and were only evaluated in a small number of combinations.
Please note that not all combinations of parameters provided below have been studied for safety and effectiveness; therefore, individuals using this device should work with the prescribing medical provider to determine the best treatment settings to use and should not apply electrodes to locations that are not indicated in the specific product's labeling. Please
{20}------------------------------------------------
# IEUROVALENS
## Modius Sleep
Traditional 510(k)
Version: 01
Page: 17 of 30
## 510(k) Summary
also note that not all CES devices are capable of providing the same stimulation parameters or combination of stimulation parameters.
Please see information below on the 23 studies (January 1, 1970 to November 8, 2022) noted above that evaluated the effectiveness of CES on insomnia:
### Electrode placement:
- . Over orbits and mastoids
- Over eyelid and mastoids .
- Over orbits and occiput .
- . Earlobes
#### Waveform:
- . Square
- Rectangular, monophasic .
#### Frequency:
- . 100 Hz
- 350 Hz
- 0.5 Hz ●
- 14 Hz ●
#### Pulse Width:
- 0.7 ms .
- 1 ms
- 2 ms .
Maximum current:
- 0.01-1.9mA
Treatment schedule:
- 30 minutes, 5 times a week for 1-2 weeks ●
{21}------------------------------------------------
Modius Sleep
Traditional 510(k)
Version: 01
Page: 18 of 30
## 510(k) Summary
- 30-minute sessions twice a day ●
- single 20-30 minute session
- . 15 minutes daily for 24 days
- 30 minutes for 1-7 days .
#### Clinical Testing of Modius Sleep
Clinical testing of Modius Sleep included a pivotal randomized, double blind sham controlled clinical trial to evaluate the safety and efficacy of the Modius Sleep Electrical Vestibular Stimulation (VeNS) device, for the treatment of chronic insomnia.
#### Study design
A 4-week pivotal randomized controlled trial with two groups 1) Modius Sleep device group and 2) Sham (placebo) control group was used to evaluate the safety and efficacy of the Modius Sleep device as a treatment for chronic insomnia.
The study was a multi-site, randomised, double blinded, sham-controlled trial conducted from May 2022 to January 2023. The study has been registered at https://clinicaltrials.gov (ClinicalTrials.gov Identifier: NCT04452981). Research sites included University of Ulster, United Kingdom (UK), and Hong Kong Polytechnic University, Hong Kong (HK). The study protocol was approved at the UK site by Wales Research Ethics Committee 7 (Ref: 22/WA/0022) and at the HK site by the Human Subjects Ethics Sub-Committee, Hong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Kong Polytechnic University (Ref: HSEARS20220320001). Informed consent was obtained from all participants.
Adults with an Insomnia Severity Index (ISI) score of 15 or higher were recruited using online platforms and randomly allocation ratio) to receive the Modius Sleep device (intervention group) or the sham device (active control group). Both groups were advised to use their allocated devices for 30 minutes per day prior to going to sleep for a four-week duration. The devices are identical in appearance and therefore neither the participant nor the clinical study team knew whether the Modius Sleep device or sham device was allocated. Primary and secondary outcomes were captured at baseline and at 4 weeks.
18
{22}------------------------------------------------
Version: 01
Page: 19 of 30
## 510(k) Summary
126 participants (aged 19-67 years) with moderate to severe insomnia assessed by the insomnia severity index (ISI) were enrolled. While the ISI is a useful screening tool for insomnia, it does not cover all components of The AASM (American Academy of Sleep Medicine) International Classification of Sleep Disorders (ICSD), and The Diagnostic and Statistical Manual of Mental Disorders, Fifth Edition (DSM-V) that are used for diagnosis of chronic insomnia. For this reason, further information was collected by an online survey to establish that the participants had chronic insomnia at the time of enrolment as defined by ICSD and DSM-V. Specifically, information was collected regarding the duration and frequency of their sleep disturbance and associated daytime symptoms, confirmation of adequate opportunity and conditions to sleep, and confirmation of no other co-existing conditions. This further information provided confirmation that the participants had Chronic Insomnia at the time of enrolment in the Modius Sleep trial.
For the duration of the study, participants agreed not to take prescribed or over the counter sleeping medication; not to undergo any extreme lifestyle changes that could have impacted their sleep quality; not to use sleep trackers; not to travel to different time zones and to maintain a familiar sleeping routine.
#### Primary Effectiveness Endpoint
The efficacy of the Modius Sleep device was quantified by change in the Insomnia Severity Index (ISI) score24 at Week 0 and Week 4 between the Modius Sleep group and sham control group.
#### Primary Safety Endpoint
An evaluation of the safety of the Modius Sleep device was quantified by the occurrence in adverse events between the Modius Sleep group and sham control group.
#### Results
A total of 149 eligible participants were recruited and randomised to the Modius Sleep group (n=75) or the sham control group (n=74). Twenty-three participants withdrew from the study
{23}------------------------------------------------
# NEUROVALENS
## Modius Sleep
Traditional 510(k)
Version: 01
Page: 20 of 30
## 510(k) Summary
before the 4-week visit. A complete case analysis was undertaken on all participants (n=126) with available data at baseline and at the 4-week visit. The mean age of the sample (67% female) was 41 years (age range: 19 to 67 years).
### Table 6: Participant accountability
| Stage | Modius Sleep Device | Sham control device | Total |
|--------------------------------------------|---------------------|---------------------|-------|
| Enrollment | 75 | 74 | 149 |
| Treatment | 75 | 74 | 149 |
| Primary safety endpoint<br>analysis | 75 | 74 | 149 |
| Primary effectiveness<br>endpoint analysis | 61 | 65 | 126 |
| Secondary endpoint<br>analysis | 61 | 65 | 126 |
{24}------------------------------------------------
# ROVALENS
## Modius Sleep
Traditional 510(k)
Page: 21 of 30
## 510(k) Summary
Version: 01
Table 7: The effect of Modius Sleep on Insomnia Severity Index (Primary outcome all participants)
| | Modius Sleep | Sham control | Between-group difference mean (95% CI);<br>P valueb | | |
|---------------------------------------------------|-----------------------------------|-----------------------------------|-----------------------------------------------------|----------------------------------|----------------------------------|
| | Mean (SD) | Mean (SD) | Model 1 | Model 2 | Model 3 |
| ITT (MI) | | | | | |
| Baseline | 19.77 (0.48)c | 18.36 (0.52)c | | | |
| 4 weeks | 13.99 (0.65)c | 14.67 (0.78)c | -2.08<br>(-3.93, -0.24)<br>0.028 | -2.13<br>(-4.34, 0.07)<br>0.058 | -1.93<br>(-4.02, 0.16)<br>0.070 |
| Within-group<br>mean change<br>(95% CI); P valuea | -5.78<br>(-6.93, -4.63)<br><0.001 | -3.70<br>(-4.99, -2.41)<br><0.001 | | | |
| ITT (LOCF) | | | | | |
| Baseline | 19.77 (4.14) | 18.36 (4.51) | | | |
| 4 weeks | 14.82 (5.12) | 15.22 (6.11) | -1.80<br>(-3.24, -0.35)<br>0.015 | -1.88<br>(-3.68, -0.07)<br>0.042 | -1.76<br>(-3.62, 0.10)<br>0.064 |
| Within-group<br>mean change<br>(95% CI); P valuea | -4.95<br>(-5.90, -3.99)<br><0.001 | -3.15<br>(-4.25, -2.05)<br><0.001 | | | |
| Complete case | | | | | |
| Baseline | 19.72 (4.03) | 18.34 (4.69) | | | |
| 4 weeks | 13.92 (4.67) | 14.82 (6.30) | -2.28<br>(-3.85, -0.71)<br>0.005 | -2.49<br>(-4.45, -0.53)<br>0.013 | -2.40<br>(-4.36, -0.44)<br>0.017 |
| Within-group<br>mean change<br>(95% CI); P valuea | -5.80<br>(-6.79, -4.81)<br><0.001 | -3.52<br>(-4.74, -2.30)<br><0.001 | | | |
ª Within-group mean change in ISI scores from baseline to week 4 analysed using paired sample t-test. "Between-group difference in ISI scores (dependent variable) from baseline to week 4 analysed using ANCOVA. Model 1 includes treatment group as the independent variable; model 2 includes treatment group-by-site interaction; and model 3 includes the covariates in model 2 as well as sex (Male vs Female), age continuous, and race (White vs Asian vs Mixed). Standard error in brackets. Intention-to-treat analysis, ITT (Multiple Imputation, MI and Last observation carried forward, LOCF) includes all participants randomised and who had baseline ISI scores (Modius Sleep, n=73 and sham control n=74). Complete case analysis includes participants who had baseline and 4-week ISI scores (Modius Sleep n=61 and sham control n=65).
{25}------------------------------------------------
Version: 01
Page: 22 of 30
## 510(k) Summary
#### Outcomes
The effect of the Modius Sleep device on change in ISI score from baseline to the 4-week visit is shown in Table 7 above. The results show that the Modius Sleep group had a greater and statistically significant reduction in ISI score compared with the sham control group.
AE reports were minor and all self- resolving including, headaches, nystagmus (eye movements), low mood, and mild tinnitus. Rate of AE reporting was 12% (9 participants out of 75) in the Modius Sleep group vs 8% (6 participants out of 74) in the Sham group. There were no device related SAE's reported during the study. One non-device related SAE was reported during the study. Further information below.
#### Effectiveness
The effect of the Modius Sleep device on change in ISI score from baseline to the 4-week visit is shown in Table 7. The ITT (MI and LOCF) and complete case analyses both demonstrated that both randomization groups reduced their ISI scores; however, the Modius Sleep group had a significantly greater reduction in ISI score (model 1). When site and treatment groupby-site interaction were added as covariates (model 2), the greater reduction in ISI score observed in the Modius Sleep group remained significant for the LOCF ITT and complete case analyses and approached statistical significance for the MI ITT analysis. Model 3 included the covariates from model 2 and demographic variables including age, sex, and race. The greater reduction in ISI score observed in the Modius Sleep group remained statistically significant for the complete case analysis and approached statistical significance for both ITT (MI and LOCF) analyses.
#### Discussion (Effectiveness)
To determine if the treatment effect found in model 1 was homogeneous across the research sites (HK and UK), model 2 additionally included the variables site and treatment-by-site interaction. The statistically significant treatment effect remained for the LOCF ITT and complete case approaches as described above. Furthermore, the treatment-by-site interaction was not significant in the ITT (MI & LOCF) or complete case models, which indicates that the treatment effects can be combined across the sites to create an overall estimate of treatment effect. The treatment effect, however, did not remain significant for the MI ITT approach. This
{26}------------------------------------------------
Version: 01
Page: 23 of 30
## 510(k) Summary
approach accounts for uncertainty and is less bias than the single imputation method (LOCF); however, it provides the most conservate estimate of treatment effect.
To investigate consistency of the treatment effect across subgroups, model 3 included the covariates sex (Male vs Female), continuous age, and race (White vs Asian vs Mixed). Results showed that the covariates added in the model were not significant in the ITT (MI or LOCF) or complete case models which indicates the treatment effect is consistent across these subgroups. As mentioned above, the statistically significant treatment effect remained for the complete case analysis and only approached statistical significance for both ITT (MI and LOCF) analyses. An ITT analysis includes all the participants who were randomized and analysed according to the treatment group they were assigned. The benefit of this type of analysis is that it produces unbiased estimates of treatment effects, provides the greatest generalisability, and preserves the sample size. The treatment effects, however, are usually conservative because it includes people who have withdrawn from the study and those who did not comply with treatment. The complete case analysis includes participants who completed the study and had data collected at baseline and at the 4 week follow up visit. This type of analysis provides information on how the treatment (Modius Sleep) performs in an ideal situation, i.e., includes only people who adhered to the intervention. The addition of the covariates in model 3 may have also reduced the power to detect a significant treatment effect in the ITT analyses.
#### Safety
A total of 22 Adverse Events were reported during the 4-week study (16 reported by the Modius Sleep group vs. 6 reported by the sham control group), these included headaches and nausea, which are known side effects of cranial electrical stimulation. One finding of a Serious Adverse Event (minor cerebrovascular accident) was reported during follow up with a participant who had been previously withdrawn. Assessment of the participant by the hospitalbased medical team reported previously undiagnosed hypertension (high blood pressure) which is believed to have been the cause of this event, and therefore this was not classified as a device related SAE. The patient's hypertension was subsequently treated with no further events reported.
{27}------------------------------------------------
Version: 01
Page: 24 of 30
## 510(k) Summary
The adverse events were all minor, caused a minimal discomfort to participants, and all selfresolved during the period of the study. Clinical and research studies report the possibility of mild adverse effects including skin irritation, discomfort from the stimulation, sensation of feeling off-balance, dizziness, nystagmus (eye movements), blurred or flashing vision, headache, nausea, and tinnitus.
#### Table 8: Number (%) of Adverse Events (Safety Data Analysis) according to Randomisation Group (all participants, n=149)
| | | Total<br>n (%) | Modius<br>Sleep<br>n (%) | Sham<br>Control<br>n (%) |
|----------------------------|------------------------------|----------------|--------------------------|--------------------------|
| Nervous System Disorders | Headache/migraine | 7 (4.7) | 6 (4.0) | 1 (0.7) |
| Eye Disorders | Flashes in peripheral vision | 2 (1.3) | 2 (1.3) | 0 (0) |
| | Shadow in peripheral vision | 1 (0.7) | 1 (0.7) | 0 (0) |
| | Tingling in eye | 1 (0.7) | 1 (0.7) | 0 (0) |
| Ear disorders | Ear pain | 2 (1.3) | 0 (0) | 2 (1.3) |
| | Tinnitus | 2 (1.3) | 1 (0.7) | 1 (0.7) |
| | Itching in ear | 1 (0.7) | 0 (0) | 1 (0.7) |
| Mood disorders | Low Mood | 2 (1.3) | 2 (1.3) | 0 (0) |
| Mouth/dental disorders | Metal fillings pulsing | 1 (0.7) | 0 (0) | 1 (0.7) |
| | Grinding teeth | 1 (0.7) | 1 (0.7) | 0 (0) |
| Gastrointestinal Disorders | Nausea | 1 (0.7) | 1 (0.7) | 0 (0) |
| Other | Tingling in arm | 1 (0.7) | 1 (0.7) | 0 (0) |
Adverse Events are grouped by Body System and Reported Term and sorted by descending frequency of reported event. Includes withdrawals. Participants could have reported more than one event.
The AE's reported in this study are listed in the table above. There was no significant difference between the AE's reported in both groups. All AE's were self-limiting and resolved when the device usage was stopped.
{28}------------------------------------------------
Version: 01
Page: 25 of 30
## 510(k) Summary
The risk of these adverse effects occurring is usually mitigated by reducing or stopping the stimulation. If a user does experience these adverse effects, they should stop stimulation by turning off or removing the device.
The following warnings have been included in the labelling:
- . To mitigate the risk of dizziness, the device must be worn when seated, and start stimulation at a low level and increase gradually to the highest comfortable level. If you do experience dizziness, stop stimulation by turning off or removing the device. If dizziness persists, please consult your physician.
- . To mitigate the risk of headaches, the device must be worn when seated, and start stimulation at a low level and increase gradually to the highest comfortable level. If you do experience a headache, stop stimulation by turning off or removing the device. If the headache persists, please consult your physician.
- . The Modius Sleep device should only be used as prescribed by a trained physician.
### Generalizability of study data
The Modius Sleep trial consisted of two clinical sites, with one based in the United Kingdom (UK) and the other in Hong Kong (HK). It should be noted that the UK site received ethical approval to recruit remotely, and additionally this extended to include remote recruitment across The Republic of Ireland (ROI). Therefore, three populations were included in the Modius Sleep study, with two of these populations being recruited nationwide (as opposed to the close proximity of the clinical site).
UK and HK participants met the same definition of insomnia and had the same ISI inclusion criteria of 15 or greater. The baseline ISI scores reported for the Modius Sleep study are in
1
{29}------------------------------------------------
Version: 01
Page: 26 of 30
## 510(k) Summary
line with literature, and comparable across the two sites (HK 19.11 and UK 18.82). An Insomnia study involving Hong Kong participants, that utilized the ISI to measure sleep improvement, has reported comparable baseline ISI scores to the US. Yin et al.25 conducted a randomised controlled trial (RCT) to evaluate the efficacy and safety of acupuncture treatment for primary insomnia. The study enrolled patients from Shanghai Municipal Hospital of Traditional Chinese Medicine, and the primary outcome was change in ISI score, and the intention to treat population demonstrated a baseline ISI score of 17. Ong et al.20, evaluated the efficacy of mindfulness meditation for the treatment of chronic insomnia recruiting patient from Rush University, US. The average baseline ISI score for the three treatment groups was 16.8. Another trial evaluating mindfulness in Los Angeles reported an average baseline ISI score of 17.8.27
Furthermore, a recent population-based survey of adults in Hong Kong estimated the prevalence of insomnia to be 20.7%.20 This was twofold higher than the insomnia estimate (11.9%) from a 2002 survey, using a comparable measure29. Approximately 10% of the adult population suffers from an insomnia disorder and another 20% experiences occasional insomnia symptoms30. Data from The National Health Interview Survey reported 14.5% of adults had trouble falling asleep and 17.8% of adults had trouble staying asleep. As such, the prevalence of insomnia in Hong Kong is comparable to both global and US estimates. Additionally, a study comparing the both the prevalence of insomnia according to different diagnostic measures, and between Hong Kong and the United States by adopting a similar methodology used by the America Insomnia Survey (AIS), showed that the prevalence of insomnia disorder was similar for Hong Kong and the United States.31
An important consideration regarding generalizability for this body of evidence is that because we were seeking to treat individuals prior to their use of other interventions (e.g., drugs), the remaining potential confounding factor would be if there were differences in the presentation and diagnosis of the disease based on race/ethnicity. We were unable to find any evidence to suggest that such a difference exists. We also note that when verifying the presence of insomnia at each site as described above, the same criteria were used in both cases. Even if
{30}------------------------------------------------
Version: 01
Page: 27 of 30
## 510(k) Summary
both sites were within either the UK or Hong Kong, it would have been entirely possible for the baseline mean to vary between those two sites and yet not affect overall generalizability.
### Study Limitations
As with all clinical trials, confounding factors can make it difficult to interpret how well the intervention has worked. Therefore, to ensure a robust dataset with a low number of variables, several conditions and medications were excluded in this clinical trial.
- . Persons currently taking over the counter or prescribed medications for sleep, beta blockers, or who regularly use antihistamines.
- . Persons with a history of stroke, or severe injury to the brain.
- Persons with an active significant mental health disorder or cognitive impairment. ●
- . Persons with any disease which may cause vestibular neuropathy.
- . Persons with myelodysplastic syndromes, or history of malignancy in the past 12months.
### Conclusion
A complete case analysis of this pivotal study reported that participants using the Modius Sleep device daily demonstrated a statistically significant improvement in their insomnia at 4 weeks.
#### Safety and Effectiveness Conclusion VIII.
Based on the clinical performance as documented in the pivotal clinical study, Modius Sleep is a low-risk, non-invasive, drug-free therapy, that provides a statistically significant benefit at 4 weeks when used to treat chronic insomnia.
Based on this clinical testing and technical comparison, it can be determined that Modius Sleep device is substantially equivalent to the listed predicate devices.
{31}------------------------------------------------
# IEUROVALENS
## Modius Sleep Traditional 510(k)
Version: 01
Page: 28 of 30
## 510(k) Summary
#### References
- Electrosleep: A double-blind clinical study. Rosenthal, S H. 2, 1972, Biological 1 Psychiatry, Vol. 4, pp. 179-185.
- 2 Efficiency of transcranial electrostimulation on anxiety and insomnia symptoms during a washout period in depressed patients: A double-blind study. Philip, P, et al. 5, Mar 1, 1991, Biol Psychiatry, Vol. 29, pp. 451-456.
- 3 Electrosleep therapy: A controlled double blind study. Feighner, J P, Brown, S L and Olivier, J E. 2, 1973, Journal of Nervous and Mental Disease, Vol. 157, pp. 121- 128.
- 4 The treatment of insomnia through the use of electrosleep: An EEG study. Weiss, M F. 2, 1973, Journal of Nervous and Mental Disease, Vol. 157, pp. 108-120.
- 5 The effect of a single cranial electrotherapy stimulation on multiple stress measures. Heffernan, M. 1995, The Townsend Letter for Doctors and Patients, Vol. 147, pp. 60-64.
- 6 A clinical trial of electrosleep therapy with a psychiatric inpatient sample. Levitt, E A, James, McI N and Flavell, P. 4. 1975, Australian and New Zealand Journal of Psychiatry, Vol. 9, pp. 287-290
- 7 The management of chronic hysteria: A review and double-blind trial of electrosleep and other relaxation methods. Scallet, A, Cloninger, R and Othmer, E. 6, 1976, Diseases of the Nervous System, Vol. 37, pp. 347-353.
- Electrosleep therapy: A double-blind trial. Hearst. E D. et al. 4, 1974, Archives of 8 General Psychiatry, Vol. 30, pp. 463-466.
- റ A comparison of relaxation techniques with electrosleep therapy for chronic, sleeponset insomnia: A sleep-EEG study. Coursey, R D, et al. 1, 1980, Biofeedback & Self Regulation, Vol. 5, pp. 57-73.
- 10 Cranial Electrotherapy Stimulation to Improve the Physiology and Psychology Response, Response-Ability, and Sleep Efficiency in Athletes with Poor Sleep Quality. Chang, W D, et al. 4, Feb 9, 2022, Int J Environ Res Public Health, Vol. 19.
- 11 Cranial electrostimulation improves slow wave sleep in collegiate population: a polysomnographic study. Aseem, A, Chaudhry, N and Hussain, M E. 1, Jan-Mar 2022, Sleep Sci, Vol. 15, pp. 88-94.
- 12 Electrosleep: A preliminary communication. Rosenthal, S H and Wulfsohn, N L. 2. 1970, Journal of Nervous and Mental Disease, Vol. 151, pp. 146-151.
- 13 An exploratory investigation of CES as an employee stress management technique. Matteson, M T and Ivancevich, John M. 1986, Journal of Health and Human Resource Administration, Vol. 9, pp. 93-109.
- 14 Studies of electrosleep on normal adults, insomniacs, and hypertensive patients. Nagata, K, et al. 3-4, 1981, Tokushima Journal of Experimental Medicine, Vol. 28, pp. 69-83.
{32}------------------------------------------------
# EUROVALENS
### Modius Sleep Traditional 510(k)
Version: 01
Page: 29 of 30
## 510(k) Summary
- 15 The effects of electrosleep on insomnia revisited. Cartwright, R D and Weiss, M F. 2, 1975, Journal of Nervous and Mental Disease, Vol. 161, pp. 134-137.
- 16 Quantitative EEG analysis of electrosleep using analog frequency analyzer and digital computer methods. Itil, T, et al. 6, 1972, Diseases of the Nervous System, Vol. 33, pp. 376-381
- 17 Does electrosleep induce natural sleep? Empson, J J A. 6, 1973, Electroencep…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.