OSCAR 15 & OSCAR 15i are a mobile fluoroscopy system designed to provide fluoroscopic and spot film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac and critical care. The system may be used for other imaging applications at the physician's discretion. OSCAR 15 and OSCAR 15i are indicated only for adult patients.
Device Story
Mobile C-arm fluoroscopy system; provides real-time fluoroscopic and spot film images for diagnostic, surgical, and interventional procedures. Inputs: X-radiation converted to visible images via Flat Panel Detector (OSCAR 15) or Image Intensifier (OSCAR 15i). Operation: X-ray tube assembly, controller, and image receptor; physician-operated in hospitals or clinics. Output: Real-time digital images displayed for clinician visualization. Clinical utility: Enables visualization of anatomical structures without film development; supports decision-making in various specialties (e.g., vascular, cardiac, orthopedic). Benefits: Improved patient accessibility via C-arm mobility; real-time guidance for procedures.
Clinical Evidence
No clinical trials; clinical evidence consists of evaluation of clinical images by a licensed radiologist to confirm sufficient diagnostic quality. Bench testing performed per IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-43, 60601-2-54, and 62366, plus 21 CFR 1020.30-32 compliance.
Technological Characteristics
Mobile C-arm fluoroscopy system. OSCAR 15: Flat Panel Detector (CsI scintillator, 145 μm pixel pitch). OSCAR 15i: Image Intensifier (9"). X-ray source: Rotating tube (0.6mm/0.3mm focal spots). Generator: High-frequency inverter, 15 kW max output. Connectivity: No wireless function. Software: Validated per FDA guidance. Standards: IEC 60601 series, 21 CFR 1020.30-32.
Indications for Use
Indicated for adult patients requiring fluoroscopic and spot film imaging during diagnostic, surgical, and interventional procedures, including cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, and critical care applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 20, 2023
Genoray Co., Ltd. % Kaitlynn Min Business Development Genoray America Inc. 147 E. Bristol Lane ORANGE, CA 92865
Re: K230787
Trade/Device Name: Oscar 15 & Oscar 15i Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA Dated: May 22, 2023 Received: July 3, 2023
Dear Kaitlynn Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radialogical Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K230787
Device Name OSCAR 15 & OSCAR 15i
### Indications for Use (Describe)
OSCAR 15 & OSCAR 15i are a mobile fluoroscopy system designed to provide fluoroscopic and spot film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, neurologic, vascular, cardiac and critical care. The system may be used for other imaging applications at the physician's discretion. OSCAR 15i are indicated only for adult patients.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
#### Exhibit 5 510(k) Summary
K230787
This 510(k) summary information is prepared in accordance with 21 CFR 807.92
- Date of Summary Preparation [21 CFR 807.92(a) (1)] 1.
: Mar. 08, 2023
### 2. Administrative Information [21 CFR 807.92(a) (1)]
| 510(k) Submitter | GENORAY Co.,Ltd<br>Address: 512, 560, Dunchon-daero, Jungwon-gu,<br>Seongnam-si, Gyeonggi-Do, Korea<br>Telephone No.: +82-31-5178-5771<br>Fax: +82-31-5178-5599<br>Contact Person: JiYeon Choi(jychoi@genoray.com) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent | GENORAY America Inc.<br>Address: 147 E. Bristol Lane, Orange, CA 92865 USA<br>Telephone No.: +1-855-436-6729<br>Fax: +1-714-786-8919<br>Contact Person: Kaitlynn Min (kaitlynn@genorayamerica.com) |
#### 3. Device Information [21 CRF 807.92(a) (2)]
| Trade / Device Name: | OSCAR 15 & OSACR 15i |
|----------------------------|------------------------------------------------------|
| Common or Usual Name: | Fluoroscopic X-ray System |
| Classification Name: | Interventional Fluoroscopic X-ray System |
| Classification Regulation: | 21 CFR 892.1650 |
| Class of Device: | Class II |
| Panel: | Radiology |
| Product Code: | OWB / Interventional Fluoroscopic X-Ray System |
| Subsequence product code: | JAA / System, X-Ray, Fluoroscopic, Image-Intensified |
#### Predicate Device Information [21 CRF 807.92(a) (3)] 4.
| Manufacturer: | GENORAY Co., Ltd |
|----------------------------|------------------------------------------------|
| Name of Device: | OSCAR 15 (K172180) |
| Common or Usual Name: | Fluoroscopic X-ray System |
| Classification Name: | Interventional Fluoroscopic X-ray System |
| Classification Regulation: | 21 CFR 892.1650 |
| Class of Device: | Class II |
| Panel: | Radiology |
| Product Code: | OWB / Interventional Fluoroscopic X-Ray System |
{4}------------------------------------------------
Subsequence product code: JAA / System, X-Ray, Fluoroscopic, Image-Intensified
#### 5. Description of the Device [21 CRF 807.92(a) (4)]
The OSCAR 15 & OSCAR 15i, C-Arm Mobile are used for providing fluoroscopic and radiographic images of patient anatomy, especially during special procedures in a hospital or medical clinics. The fluoroscopic mode of operation is very useful to the attending physician to see the images on real time without the need to develop individual films. These devices are intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification.
The OSCAR 15 & OSCAR 15i consist of the X-ray tube assembly, X-ray controller, Image receptor and some accessories with no wireless function. The difference between OSCAR 15 and OSCAR 15i is only image acquisition parts. (An Flat Panel Detector (FPD) is applied to OSCAR 15, and an Image instensifier is applied to OSCAR 15i.)
#### Indications for use (intended use) [21 CRF 807.92(a) (5)] 6.
OSCAR 15 & OSCAR 15i are a mobile fluoroscopy system designed to provide fluoroscopic and spot film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac and critical care. The system may be used for other imaging applications at the physician's discretion. OSCAR 15 and OSCAR 15i are indicated only for adult patients.
{5}------------------------------------------------
## 7. Substantial equivalence chart [21 CRF 807.92(a) (6)]
| | Proposed device | | Predicate device |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | OSCAR 15 | OSCAR 15i | OSCAR 15 |
| Manufacturer | GENORAY Co., Ltd. | GENORAY Co., Ltd. | GENORAY Co., Ltd. |
| 510(k) No. | - | - | K172180 |
| Classification<br>Name | Interventional Fluoroscopic X-Ray System | | Interventional Fluoroscopic<br>X-Ray System |
| | Image: OSCAR 15 | Image: OSCAR 15i | Image: OSCAR 15 |
| | Proposed device | | Predicate device |
| | OSCAR 15 | OSCAR 15i | OSCAR 15 |
| Indications<br>for use | OSCAR 15 & OSCAR 15i are a mobile fluoroscopy system designed to provide<br>fluoroscopic and spot film images of the patient during diagnostic, surgical and<br>interventional procedures. Examples of clinical application may include cholangiography,<br>endoscopy, urologic, orthopedic, neurologic, vascular, cardiac and critical care. The system<br>may be used for other imaging applications at the physician's discretion. OSCAR 15 and<br>OSCAR 15i are indicated only for adult patients. | OSCAR 15 & OSCAR 15i are a mobile fluoroscopy system designed to provide<br>fluoroscopic and spot film images of the patient during diagnostic, surgical and<br>interventional procedures. Examples of clinical application may include cholangiography,<br>endoscopy, urologic, orthopedic, neurologic, vascular, cardiac and critical care. The system<br>may be used for other imaging applications at the physician's discretion. OSCAR 15 and<br>OSCAR 15i are indicated only for adult patients. | OSCAR 15 is a mobile fluoroscopy system<br>designed to provide fluoroscopic and spot<br>film images of the patient during<br>diagnostic, surgical and interventional<br>procedures. Examples of clinical<br>application may include cholangiography,<br>endoscopy, urologic, orthopedic,<br>neurologic, vascular, cardiac and criticalcare. The system may be used for other<br>imaging applications at the physician's<br>discretion.. |
| Max. output<br>power | 15 kW | 15 kW | 15 kW |
| X-ray Tube | Rotating tube | Rotating tube | Rotating tube |
| | Large : 0.6 mm<br>Small : 0.3 mm | Large : 0.6 mm<br>Small : 0.3 mm | Large : 0.6 mm<br>Small : 0.3 mm |
| | High Frequency Inverter | High Frequency Inverter | High Frequency Inverter |
| | Proposed device | | Predicate device |
| | OSCAR 15 | OSCAR 15i | OSCAR 15 |
| | [Radiography] | [Radiography] | [Radiography] |
| | • Output kV, mAs<br>: 120kV @ 100mAs (max.) | • Output kV, mAs<br>: 120kV @ 100mAs (max.) | • Output kV, mAs<br>: 120kV @ 100mAs (max.) |
| | • kV range<br>: 40kV ~ 120kV, Step: 1kV | • kV range<br>: 40kV ~ 120kV, Step: 1kV | • kV range<br>: 40kV ~ 120kV, Step: 1kV |
| | • mAs range<br>: 0.4mAs ~ 100mAs, 24 Steps<br>(Snapshot: 1.0mA ~ 50.0mA) | • mAs range<br>: 0.4mAs ~ 100mAs, 24 Steps<br>(Snapshot: 1.0mA ~ 50.0mA) | • mAs range<br>: 0.4mAs ~ 100mAs, 24 Steps |
| Generator | [Fluoroscopy] | [Fluoroscopy] | [Fluoroscopy] |
| | • Output kV, mA<br>: 120kV, 4mA (max.) | • Output kV, mA<br>: 120kV, 4mA (max.) | • Output kV, mA<br>: 120kV, 4mA (max.) |
| | • kV range<br>: 40kV ~ 120kV, step: 1kV | • kV range<br>: 40kV ~ 120kV, step: 1kV | • kV range<br>: 40kV ~ 120kV, step: 1kV |
| | • mA range<br>- Small focus :<br>0.2mA ~ 6.0mA, Continuous mode<br>2.0mA ~ 10.0mA, Pulsed mode<br>- Large focus :<br>1.0mA ~ 20.0mA, Continuous mode<br>3.0mA ~ 50.0mA, Pulsed mode | • mA range<br>- Small focus :<br>0.2mA ~ 6.0mA, Continuous mode<br>2.0mA ~ 10.0mA, Pulsed mode<br>- Large focus:<br>1.0mA ~ 20.0mA, Continuous mode<br>3.0mA ~ 50.0mA, Pulsed mode | • mA range<br>- Small focus :<br>0.2mA ~ 6.0mA, Continuous mode<br>1.0mA ~ 10.0mA, Pulsed mode<br>- Large focus :<br>1.0mA ~ 20.0mA, Continuous mode<br>1.0mA ~ 48.0mA, Pulsed mode |
| | Proposed device | | Predicate device |
| | OSCAR 15 | OSCAR 15i | OSCAR 15 |
| | FXDD-1212GA (Flat Panel Detector) | E5830SD-P4A (Image Intensifier) | DualRay-Q (Flat Panel Detector) |
| Detector | • Active image area : 296.96 x 296.96 mm | • Size of incident surface : 230 mm (9") | • Active image area : 260 x 256 mm |
| | • Maximum resolution: 3.4 lp/mm | • Conversion factor<br>29 (cd/m2)/(uGy/s)<br>250 (cd/m2)/(uGy/s) | • Central Resolution: 4.6 lp/mm |
| | • Type : TFT | • Rated of contrast<br>30:1 (10% Area Contrast)<br>19:1(10mm Dia. Contrast) | • Type : CMOS |
| | • Pixel sampling resolution : 16 bits | • DQE: 65% | • Pixel sampling resolution : 14 bits |
| | • Pixel pitch : 145 μm | • Resolution<br>9" - 52 (Lp/cm typ.)<br>6" - 58 (Lp/cm typ.)<br>4.5" - 68 (Lp/cm typ.) | • Pixel pitch: 100 μm |
| | • MTF: 78% (0.5 lp/mm) | • Out image diameter : 25 ±0.5 mm | • MTF: 56% |
| | • DQE: 60% (0.5 lp/mm) | Previously cleared under K211780. | • DQE: 59% |
| | • Scintillator : CsI | | • Scintillator: CsI |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | Previously cleared under K180473. | | |
| Dimensions | SID : 1000 mm | SID : 1000 mm | SID : 1000 mm |
| | C-arc Panning : ±12.5° | C-arc Panning : ±12.5° | C-arc Panning : ±12.5° |
| | C-arc Orbital angle : 150° | C-arc Orbital angle : 135° | C-arc Orbital angle : 155° |
| | C-arc up/down range : 500 mm | C-arc up/down range : 500 mm | C-arc up/down range : 500 mm |
| | C-arc back/forward range : 200 mm | C-arc back/forward range : 200 mm | C-arc back/forward range : 200 mm |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
When compared to the predicate device (K172180), the proposed devices presented in this submission has the identical
- Indications for use
- Maximum output power
- X-ray source
The differences between proposed device and predicate device (K172180) are change of detector, addition of New model OSCAR 15i.
### A new model, OSCAR 15i has been added to the predicate device (K172180).
The difference between OSCAR 15i and predicate device (K172180) is image acquisition parts. An Flat Panel Detector (FPD) is applied to OSCAR 15, and an Image instensifier is applied to OSCAR 15i. Brief description of technology and conversion principle of whole detector chain through an Inage intensifier is below. Total conversion is that x-ray signal irradiated to the surface of the IIT is converted into a pixel value of a digital image in proportion to its anount. That of conversion method described by several steps by the IT' and the CCD camera attached o IT output window. FPDs designed specifically for thoroscopic use and dose efficiency generally superior to the II systems that they replace, except at the lowest fluoro dose levels. Advantages include excellent image uniformity, no veiling glar or vigneting, and small, thin physical size for improved patient accessibility. Since a detective is applied to the proposed device, the performance related to image quality is also different. The image performance was evaluated according to the IEC standard through performance bench testing denonstrated that these differences do not railer and effectiveness in companson with the predicate device. Furthermore, clinical images have been evaluated by a licensed radits confirmed the sufficient diagnostic quality to provide accurate information.
{10}------------------------------------------------
### 8. Safety, EMC and Performance data comparison to Predicate [21 CRF 807.92(b)]
OSCAR 15 & OSCAR 15i have been successfully completed verification and validation testing per GENORAY quality system as well engineering bench testing in support of this submission. The system has been tested and is compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-43, IEC 60601-2-54, IEC 62366. Also, OSCAR 15 &…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.