← Product Code [FED](/productcode/FED) · K230748

# Disposable Ureteral Access Sheath (K230748)

_Youcare Technology Co.,Ltd. (Wuhan) · FED · Aug 2, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230748

## Device Facts

- **Applicant:** Youcare Technology Co.,Ltd. (Wuhan)
- **Product Code:** [FED](/productcode/FED.md)
- **Decision Date:** Aug 2, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Disposable Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract

## Device Story

Disposable Ureteral Access Sheath functions as a conduit for endoscopic urological procedures; facilitates passage of endoscopes and surgical instruments into urinary tract. Device consists of sheath tube, sheath tube base, dilator, and center base of dilator; includes fiber image channel, irrigation channel, and equipment channel interfaces. Operated by physicians in clinical settings. Procedure involves inserting guide wire into ureter; advancing dilator/sheath assembly over guide wire; connecting negative pressure aspirator to irrigation channel; confirming position via endoscope; removing dilator; and inserting surgical instruments. Provides clear visualization and access for urological interventions; benefits patient by enabling minimally invasive access to urinary tract.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-5, -10, -11. Performance testing included bending resistance, friction coefficients, dimensional analysis, flow testing, patency, connection integrity, peak tensile force, and Luer taper testing. Sterilization validated per ISO 11135 and ISO 11737. Shelf life and package integrity validated per ASTM F1980, D3078, F1929, and F88.

## Technological Characteristics

Materials: Sheath tube (Pebax6333, SUS 304, PTFE, PAM), base (ABS, PC), dilator (LDPE), obturator (PC, silica gel). Sensing/Actuation: Mechanical conduit. Energy: None. Dimensions: 12Fr ID; 35cm/45cm lengths. Connectivity: None. Sterilization: EO. Software: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Well Lead Ureteral Access Sheath ([K151084](/device/K151084.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

August 2, 2023

YouCare Technology Co., Ltd. (Wuhan) Bing Hu Manager Tangxunhu North Street Wuhan, Hubei 430000 China

Re: K230748

> Trade/Device Name: Disposable Ureteral Access Sheath Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FED Dated: July 5, 2023 Received: July 5, 2023

Dear Bing Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230748

Device Name Disposable Ureteral Access Sheath

Indications for Use (Describe)

The Disposable Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5 - 510(k) Summary

Date of Summary Preparation: 07/16/2023

# 1. Submitter's Identifications

Submitter's Name: YouCare Technology Co., Ltd.(Wuhan). Address: Great Wall Science and Technology Park, Tangxunhu North Street, East Lake Development Zone Wuhan, 430223 Hubei P.R. China. Contact Person: Bing Hu Contact Email: hb@youcaretech.com Telephone: +86-27-87926396-830

# 2. Correspondent's Identifications

Correspondent's Name: YouCare Technology Co., Ltd.(Wuhan). Address: Great Wall Science and Technology Park, Tangxunhu North Street, East Lake Development Zone Wuhan, 430223 Hubei P.R. China. Contact Person: Bing Hu Contact Email: hb@youcaretech.com Telephone: +86-27-87926396-830

# 3. Name of the Device

Product Name: Disposable Ureteral Access Sheath Trade Name: Disposable Ureteral Access Sheath Model: YC-Videoaid-sheath-A YC-Videoaid-sheath-B Classification Name: Endoscope and accessories Regulation Number: 21 CFR§ 876.1500 Product Code: FED Device Classification: Class II

### 4. The Predicate Devices

K151084

Well Lead Ureteral Access Sheath

This predicate has not been subject to a design-related recall. No reference devices were used in this submission

### 5. Device Description

The product is composed of sheath tube, sheath tube base, dilator and center base of dilator. The sheath tube base contains a backwater interface. The center base of dilator consists of upper cover of dilator base, lower cover of dilator base , fiber image channel interface F, irrigation channel interface I, equipment channel interface E, obturator, two-way water valve, needle free joint. There is no guide wire. The product is sterilized by EO, it should be sterile and only for

1 / 6

{4}------------------------------------------------

single use.

# 6. Indications for use

The Disposable Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract.

### 7. Operating instructions

8. Substantial Equivalence

1. Check the package, it is forbidden to use when sterile barrier is damaged.

2. Open the package, and make sure that the product structure is complete without missing of product structure, and it is firmly connected between two-way water valve and irrigation channel, obturator and fiber image channel, needle free joint and equipment channel

3. Insert the guide wire (diameter no more than 0.035 inch or 0.89mm) into ureter to the predetermined length, and then build the channel.

4. The endoscope is inserted through the obturator into the fiber image channel until the objective lens of endoscope reaches the end of the dilator. Connect the negative pressure aspirator to irrigation channel.

5. After flushing and lubricating the sheath tube with normal saline, take away the needle free joint, insert the guide wire into the channel (marked with a notch) on the top of dilator which until it extends out of the equipment channel interface E, then reinstall the needle free joint after threading guide wire out of the needle free joint ..

6. Push the dilator/sheath along guide wire, open the two-way water valve and water irrigation system to ensure the view of the endoscope is clear. Confirm the position of dilator/sheath through the endoscope.

7. Fix the sheath in place, at the same time, loosen the joint and remove the dilator.

8. Import the required endoscope or other surgical instruments.

|                             | New device                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate device                                                                                                                                                                                                         | Comparison  |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| "K"NUMBERS                  | /                                                                                                                                                                                                                                                                                                                                                                                                                                        | K151084                                                                                                                                                                                                                  | /           |
| Manufacturer                | YouCare Technology Co.,<br>Ltd.(Wuhan).                                                                                                                                                                                                                                                                                                                                                                                                  | Well Lead Medical Co. Ltd                                                                                                                                                                                                | /           |
| Regulation #                | 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                 | 876.1500                                                                                                                                                                                                                 | Same        |
| Product Code                | FED                                                                                                                                                                                                                                                                                                                                                                                                                                      | FED                                                                                                                                                                                                                      | Same        |
| Classification<br>name      | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                                                | Endoscope and accessories                                                                                                                                                                                                | Same        |
| Intended use                | It is intended to establish a<br>channel during urologic surgical<br>procedures, facilitating the<br>passage of endoscopes and other<br>instruments into the urinary<br>tract.                                                                                                                                                                                                                                                           | The Well Lead ureteral access<br>sheath is used to establish a<br>conduit during endoscopic<br>urological procedures facilitating<br>the passage of endoscopes and<br>other instruments into the urinary<br>tract.       | Same        |
| Regulatory Class            | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                 | Class II                                                                                                                                                                                                                 | Same        |
| Sterility                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                      | Same        |
| Sterilization<br>Method     | EO                                                                                                                                                                                                                                                                                                                                                                                                                                       | EO                                                                                                                                                                                                                       | Same        |
| Single Use                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                      | Same        |
| Sheath ID                   | 12Fr                                                                                                                                                                                                                                                                                                                                                                                                                                     | 10Fr, 12Fr, 14Fr                                                                                                                                                                                                         | Different 1 |
| Sheath length               | 45cm, 35cm                                                                                                                                                                                                                                                                                                                                                                                                                               | 13cm, 20cm, 28cm, 35cm, 45cm,<br>55cm                                                                                                                                                                                    | Different 2 |
| Primary<br>structure        | The product is composed of sheath<br>tube, sheath tube base, dilator and<br>center base of dilator. The sheath<br>tube base contains a backwater<br>interface. The center base of dilator<br>consists of upper cover of dilator<br>base, lower cover of dilator base ,<br>fiber image channel interface F,<br>irrigation channel interface I,<br>equipment channel interface E,<br>obturator, two-way water valve,<br>needle free joint. | The Well Lead Ureteral Access<br>Sheath is comprised of three<br>components: sheath, dilator and<br>connector. The outer surface of<br>the sheath has a hydrophilic<br>coating.                                          | Different 3 |
| Materials                   | sheath tube: Pebax6333 SA01<br>MED, SUS 304, PTFE, PAM<br>sheath tube base: ABS, PC<br>(backwater interface: PC)<br>dilator: LDPE                                                                                                                                                                                                                                                                                                        | Sheath Tube: PTFE, Nylon (PA),<br>stainless steel<br>Outer surface: Nylon (PA)<br>Inner surface: PTFE<br>Dilator: Polyethylene (PE)                                                                                      | Different 4 |
|                             | center base of dilator                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                          |             |
|                             | upper cover of the dilator base:<br>ABS                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                          |             |
|                             | lower cover of the dilator base:<br>ABS                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                          |             |
|                             | fiber image channel interface F:<br>PC                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                          |             |
|                             | irrigation channel interface I: PC<br>equipment channel interface E:<br>PC                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                          |             |
|                             | obturator: PC, silica gel                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                          |             |
|                             | two-way water valve: PC, POM                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                          |             |
|                             | needle-free joint: PC, silica gel                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                          |             |
| Package                     | Single-use EO sterilized pouch with<br>one device per pouch                                                                                                                                                                                                                                                                                                                                                                              | Single-use EO sterilized pouch with<br>one device per pouch                                                                                                                                                              | Same        |
| Biocompatible               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                      | Same        |
| Shelf Life                  | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                  | 3 years                                                                                                                                                                                                                  | Same        |
| Bending<br>Resistance       | Bending any section of the<br>dilator, sheath tube or the<br>assembly of sheath tube and<br>dilator into a ring with a radius<br>of 5cm for 1min, there should be<br>no crease, crack or other<br>undesirable phenomena.                                                                                                                                                                                                                 | Bending any section of the<br>dilator, sheath tube or the<br>assembly of sheath tube and<br>dilator into a ring with a radius<br>of 5cm for 1min, there should<br>be no crease, crack or other<br>undesirable phenomena. | Same        |
| Coefficients of<br>Friction | When the sheath tube is tested<br>for friction, the friction<br>coefficient shall not exceed 0.03.                                                                                                                                                                                                                                                                                                                                       | When the sheath tube is tested<br>for friction, the friction<br>coefficient shall not exceed<br>0.03.                                                                                                                    | Same        |
| Peak tensile<br>force       | The peak tensile force of the<br>sheath tube and dilator should<br>not be less than 15N. And the<br>peak tensile force of the junction<br>between sheath tube and sheath<br>tube base, the dilator and center<br>base of dilator also shall meet<br>the requirement                                                                                                                                                                      | The Dilator to Hub Tensile<br>Strength>15N.                                                                                                                                                                              | Same        |

# Comparison of technology characteristics

{5}------------------------------------------------

# YouCare Technology Co., Ltd.(Wuhan).

{6}------------------------------------------------

# YouCare Technology Co., Ltd.(Wuhan).

{7}------------------------------------------------

### 9. Substantial equivalence discussion:

### Different 1 -sheath ID

This difference is in sheath ID. Different sheath inner diameter device will be selected by physician per patient's condition. Moreover, a small diameter difference has little effect on the product safety performance

# Different 2 -Sheath length

This difference is in sheath length. Different sheath length device will be selected by physician per patient's condition, the length difference did not affect the safety and efficacy of the product.

# Different 3 - Primary structure

The sheath base of the new device is similar to the connector of the predict product. The New device has three instrument channels --- image channel, irrigation channel and equipment channel, it does not raise different questions of safety and effectiveness for patient, it only increases the operation requirements for doctors.

# Different 4 - Materials

The materials of sheath tube and dilator are different. The new device is 304 Stainless Steel and Pebax6333 and Low-Density Polyethylene(LDPE), the materials of predicate device is Stainless Steel、Nylon(PA)and Polyethylene. Pebax6333 is a block polymer composed of nylon and polyether parts, which has the properties of nylon. Although the materials are different, the performance of the product is not affected, and both passed the biological test.

# Substantial Equivalence discussion:

The proposed devices and the predicated device have classification information, similar materials, same specifications, same performance effectiveness. The Model of Proposed Device is included in the Model of Predicate device. The Indications for use of proposed device is same as the predicated device. the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.

# 10. Non-Clinical Tests Performed:

Non-clinical Testing: A series of preclinical tests were performed to assess the safety and performance of the Disposable Ureteral Access Sheath including biocompatibility physical and mechanical performance.

The following biocompatibility and safety tests were conducted in compliance with ISO standards:

# Biocompatibility

- . Cytotoxicity Test: ISO10993-5
- . Sensitization: ISO 10993-10
- . Intracutaneous reactivity:1SO10993-10

{8}------------------------------------------------

- Material mediated pyrogenicity: ISO10993-11
- Acute systemic toxicity: ISO10993-11 ●

# Ethylene Oxide Sterilization Validation:

- ISO 11135:2014
- ISO 11737-1:2018
- ISO11737-2:2019
- ISO 10993-7: 2008

# Shelf life and Package Validation:

- ASTM F1980-2016 .
- ASTM D3078-02-2021
- ASTMF 1929-15
- . ASTM F88/F88M-15

# 11. Performance testing

The following performance testing was conducted:

- . Bending resistance
- Coefficients of Friction
- Determining the Dimensions
- Flow test
- Patency
- . Connection
- Peak tensile force
- . Luer taper

### 12. Clinical Test Conclusion

No clinical study is included in this submission.

### 13. Conclusion:

Comparison to the predicate devices for Intended Use, structure and physical, it shows that our products are as safe, as effective and performs as well as the predicate devices.

--- End of this section ---

---

**Source:** [https://fda-staging.innolitics.com/device/K230748](https://fda-staging.innolitics.com/device/K230748)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).