C660G/C810G LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.
Device Story
JUSHA-C660G/C660/C810G LCD monitors are high-resolution display systems for medical imaging. Inputs include digital image data from medical imaging modalities (X-ray, MRI) via DisplayPort, DVI, or Type-C interfaces. The device transforms these inputs into high-luminance, grayscale-calibrated visual outputs using built-in DICOM standard Look-Up Tables (LUTs). Features include ambient brightness adaptation, real-time DICOM automatic calibration, full-screen brightness equalization, and presence induction to optimize display performance. Used in clinical environments by trained medical practitioners to view diagnostic images; the output assists clinicians in interpreting medical images for diagnosis. The device benefits patients by providing high-fidelity, consistent, and calibrated image visualization for accurate clinical decision-making.
Clinical Evidence
No clinical data or animal studies were required or provided. Substantial equivalence is supported by bench testing, including measurement of luminance response, luminance/chromaticity non-uniformity (per TG18 guidelines), temporal response, and luminance stability, alongside electrical safety and EMC testing.
Indicated for trained medical practitioners to display and view digital images for the diagnosis of X-ray or MRI. Not indicated for the display of mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nanjing Jusha Display Technology Co., Ltd Donny Lee Certification Engineer 8A, Block 1. Nanjing International Service Outsourcing Mansion, No. 301. Hanzhongmen street NANJING, JIANGSU 210036 CHINA
June 9, 2023
### Re: K230728
Trade/Device Name: JUSHA-C810G/C810G LCD Monitor, JUSHA-C660/JUSHA- C660G/C660/C660G LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: March 16, 2023 Received: March 16, 2023
Dear Donny Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230728
#### Device Name
JUSHA-C810G/C810G LCD Monitor, JUSHA-C660/JUSHA-C660G/C660/C660G LCD Monitor
Indications for Use (Describe)
C660G/C810G LCD Monitor is intended to be used in displaying digital images diagnosis of X-ray or MRI.etc.by trained medical practitioners. The device does not support the display of mammography images for diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# (K230728)
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | Mar 16, 2023 |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Nanjing Jusha Display Technology Co., Ltd<br>Add: 8A, Block 1. Nanjing International Service Outsourcing<br>Mansion, No. 301, Hanzhongmen street, Nanjing City, Jiangsu<br>Province, 210036 China. |
| Contact Person: | Donny Lee<br>Certification Engineer<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783273<br>lidongdong@jusha.com.cn |
| Device Trade Name: | JUSHA-C810G/C810G LCD Monitor, JUSHA-C660/JUSHA-<br>C660G/C660/C660G LCD Monitor |
| Common/Usual Name: | 6MP/8MP Color LCD Monitor |
| Classification Name: | Medical Image Management and Processing System 21CFR 892.2050 |
| Product Code: | PGY |
| Predicate Device(s): | C630G, K222121 |
| Device Description: | JUSHA-C660G/JUSHA-C660/C660G/C660 LCD Monitor is the<br>display system with high resolution (3280×2048), high luminance (800<br>cd/m2), and 16-bit grayscale (65536 level), built-in DICOM standard<br>LUT. In particular, C660G has ambient brightness adaptation inside,<br>on top of which C660G has real-time DICOM automatic calibration,<br>full-screen brightness equalization and presence induction system,<br>therefore this display automatically adjust according to different<br>requirements to achieve the best results.<br><br>The product is consisted of the following components:<br>- the display with stand<br>- a graphic card |
| | - agraphic card driver CD |
| | - an AC power cable |
| | - an external power supply |
| | - a Type-C cable |
| | - Three DP cables |
| | - a USB cable |
| | JUSHA-C810G/ C810G LCD Monitor is the display system with high resolution (3840×2160), high luminance (800 cd/m2), and 14-bit grayscale (16384 grayscale), built-in DICOM standard LUT. In particular, C810G has ambient brightness adaptation inside, on top of which C810G has real-time DICOM automatic calibration, full-screen brightness equalization and presence induction system, therefore this display automatically adjust according to different requirements to achieve the best results. |
| | The product is consisted of the following components: |
| | - the display with stand |
| | - a graphic card |
| | - agraphic card driver CD |
| | - an AC power cable |
| | - an external power supply |
| | - a DVI cable |
| | - two Mini DP switch to DP cable |
| | - a USB cable |
| | |
| | The LCD Monitors are designed, tested, and will be manufactured in accordance with both mandatory and voluntary standards: |
| | |
| | 1. IEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI<br>ES60601-1:2005+A1:2012+C1:2009+A2:2010,<br>CAN/CSA C22.2 NO.60601-1:14, Medical equipment<br>medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance.<br>2. IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015,<br>Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic disturbances -<br>Requirements and tests. |
| Intended Use: | C660G/C810G LCD Monitor is intended to be used in displaying and<br>viewing digital images for diagnosis of X-ray or MRI, etc. by trained<br>medical practitioners. The device does not support the display of<br>mammography images for diagnosis. |
| Technology: | JUSHA-C660G/JUSHA-C660/C660G/C660 LCD Monitor is the<br>display system with high resolution (3280×2048), high luminance (800<br>cd/m2), and 16-bit grayscale (65536 level), built-in DICOM standard<br>LUT. In particular, C660G has ambient brightness adaptation inside,<br>on top of which C660G has real-time DICOM automatic calibration,<br>full-screen brightness equalization and presence induction system,<br>therefore this display automatically adjust according to different<br>requirements to achieve the best results.<br>JUSHA-C810G/ C810G LCD Monitor is the display system with high<br>resolution (3840×2160), high luminance (800 cd/m2), and 14-bit<br>grayscale (16384 grayscale), built-in DICOM standard LUT. In<br>particular, C810G has ambient brightness adaptation inside, on top of<br>which C810G has real-time DICOM automatic calibration, full-screen<br>brightness equalization and presence induction system, therefore this<br>display automatically adjust according to different requirements<br>to<br>achieve the best results. |
| Determination of Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor(C660G,C810G) complies with voluntary standards<br>as following:<br>1 IEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI ES60601-<br>1:2005+A1:2012+C1:2009+A2:2010, CAN/CSA C22.2<br>NO.60601-1:14, Medical equipment medical electrical equipment<br>- Part 1: General requirements for basic safety and essential<br>performance. |
| 2 | IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015, Medical<br>electrical equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral standard:<br>Electromagnetic disturbances - Requirements and tests |
| | C660G/C810G LCD Monitor are substantially equivalent to C630<br>LCD Monitor. They have equivalent characteristics andfunctions<br>according to comparison table, pleaserefer to 12. Product Comparison |
| | The following quality assurance measures were applied to the<br>development of the system: |
| | • Risk Analysis |
| | • Requirements Reviews |
| | • Design Reviews |
| | • Raw materials verification |
| | • Testing on unit level (Module verification) |
| | • Integration testing (System verification) |
| | • Final acceptance testing (Validation) |
| | • Performance testing (Verification) |
| | • Safety testing (Verification) |
| | Summary of Clinical Tests: |
| | The subject of this premarket submission, LCD Monitor, did not<br>require clinical studies to support substantial equivalence. |
| | The proposed device is Substantially Equivalent (SE) to the predicate<br>device which is US legally market device. Therefore, the subject<br>device is determined as safe and effectiveness. |
| Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the<br>C660G/C810G LCD Monitor to be as safe, as effective, and<br>performance is substantially equivalent to the predicate device(s) |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
# 12.1 Product Comparison
This comparison identifies the similarities and differences of the proposed C660G/C810G LCD Monitor device to the legally marketed predicate C630 LCD Monitor device to which substantial equivalency is claimed.
{7}------------------------------------------------
| Attributes | Predicate Device | Proposed Device | | Discussion of Differences |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Performance/Specifications | | | | |
| Product | C630G LCD<br>Monitor | C660G LCD<br>Monitor | C810G LCD<br>Monitor | |
| 510(k) Number | K222121 | K230728 | | |
| Screen<br>technology | 30 inches, Color-TFT LCD Panel | 30inches,Color-TFT LCD Panel | 31.5inches,<br>Color-TFT LCD<br>Panel | C660G and<br>C630G are<br>same, C810G is<br>larger than<br>C630G. |
| Viewing angle<br>(H, V) | Horizontal<br>178°, Vertical<br>178° | Horizontal<br>178°, Vertical 178° | Horizontal<br>178°, Vertical<br>178° | Same |
| Resolution | 3280x2048 | 3280x2048 | 3840x2160 | C660G and<br>C630G are<br>same, C810G is<br>better than<br>C630G. |
| Display area | 645.5 (H) mm ×<br>403 (V)mm | 645.5 (H) mm ×<br>409.3 (V)mm | 697.31(H) mm×<br>392.23(V)mm | C660G and<br>C630G are<br>same, C810G is<br>larger than<br>C630G. |
| Contrast Ratio | 1000:1 | 2000:1 | 1300:1 | C660G and<br>C810G are larger<br>than C630G. |
| DICOM<br>calibrated<br>luminance | 500cd/m² | 800 cd/m² | 1000cd/m² | C660G and<br>C810G are larger<br>than C630G. |
| Pixel Pitch | 0.197mm×<br>0.197mm | 0.197mm×<br>0.197mm | 0.1816mm×<br>0.1816mm | C660G and<br>C630G are<br>same, C810G is<br>better than<br>C630G. |
| Backlight | LED | LED | LED | Same. |
| Attributes | Predicate Device | Proposed Device | Discussion of Differences | |
| Product | C630G LCD<br>Monitor | C660G LCD<br>Monitor | C810G LCD<br>Monitor | |
| 510(k) Number | K222121 | K230728 | | |
| DICOM LUT | 16-bit:65536 | 16-bit:65536 | 14-bit:16384 | This parameter is an output<br>range.C810G is smaller than<br>them, But the three products<br>can only output 10bit images. |
| Luminance<br>calibration | Built in<br>calibration<br>sensor provided | Built in calibration<br>sensor provided | Built in<br>calibration<br>sensor provided | Same. |
| Video Signal Input | | | | |
| Input signals | DisplayPort 1.2a<br>DVI | DisplayPort 1.2a<br>Type-C<br>DVI | DisplayPort 1.2a<br>DVI | The difference<br>only shows that<br>they have<br>different<br>input,has nothing<br>to do with the<br>display function. |
| Input<br>terminational | DisplayPort×2<br>DVI×1 | DisplayPort×2<br>Type-C<br>DVI×1 | DisplayPort×2<br>DVI×1 | The difference<br>only shows that<br>they have<br>different<br>input,has nothing<br>to do with the<br>display function. |
| Output signals | DisplayPort 1.2a | DisplayPort 1.2a | NA | C660G and<br>C630G are<br>same, C810G<br>has no output<br>interface. |
| Output<br>Terminational | DisplayPort×1 | DisplayPort×1…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.