K230727 · Novadaq Technologies Ulc (A Part of Stryker) · IZI · Jun 5, 2023 · Radiology
Device Facts
Record ID
K230727
Device Name
SPY Portable Handheld Imaging (SPY-PHI) System
Applicant
Novadaq Technologies Ulc (A Part of Stryker)
Product Code
IZI · Radiology
Decision Date
Jun 5, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1600
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Visible white light and near-infrared illumination during open-field surgical procedures.
Device Story
SPY-PHI is a handheld, open-field surgical imaging system providing real-time white-light and near-infrared (NIR) fluorescence visualization. System comprises a handheld imager head, camera cable, and video processor/illuminator. During surgery, surgeon administers SPY AGENT GREEN (indocyanine green); imager captures reflected light, rejecting NIR excitation light via internal filters. CMOS sensor converts optical data to digital signals, processed by the video processor for display on a monitor. Modes include white light, overlay, contrast (SPY), and color-segmented fluorescence. Used by surgeons to assess blood flow, tissue perfusion, and lymphatic structures. Output aids intraoperative decision-making by delineating vessels and lymph nodes, potentially improving surgical precision and patient outcomes.
Clinical Evidence
FILM-B prospective, multi-center, open-label study (n=152) in early-stage breast cancer patients. Compared SPY-PHI/SPY AGENT GREEN against Tc99m radioactive colloid/gamma probe for sentinel lymph node biopsy. Primary endpoint: proportion of lymph nodes identified. Results: 89% identified by SPY-PHI vs 66% by Tc99m (difference 23%, 95% CI 3.67%-9.48%, p<0.0001). Secondary endpoints: 95% of subjects had at least one lymph node detected by both methods; 99% of lymph nodes identified by SPY-PHI were visualized via lymphatic vessel mapping. No serious adverse events reported.
Technological Characteristics
Open-field imaging system; CMOS image sensor; RGB LED light source. NIR excitation at 805 nm; white light 400-700 nm. 1080p resolution at 60 Hz. Connectivity via camera cable to video processor. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60825-1 (Class 3R laser safety).
Indications for Use
Indicated for adult and pediatric patients (1 month+) for intraoperative fluorescence angiography (with IV SPY AGENT GREEN) in vascular, gastrointestinal, organ transplant, and plastic/micro/reconstructive surgeries. Indicated for lymphatic system visualization (with interstitial SPY AGENT GREEN) and lymphatic mapping/lymph node delineation in adults with breast cancer (with intradermal SPY AGENT GREEN).
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SPY-PHI System with SPY-PHI Fluorescence Assessment Software (K202244)
SPY Portable Handheld Imaging (SPY-PHI) System (K200737)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5, 2023
Novadaq Technologies ULC (a part of Stryker) Cara Cahill Manager, Regulatory Affairs 8329 Eastlake Drive, Unit 101 Burnaby, British Columbia V5A 4W2 Canada
Re: K230727
Trade/Device Name: SPY Portable Handheld Imaging (SPY-PHI) System Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic X-Ray System Regulatory Class: Class II Product Code: IZI Dated: March 15, 2023 Received: March 16, 2023
Dear Cara Cahill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Carr -S
Jessica Carr, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230727
#### Device Name
SPY Portable Handheld Imaging (SPY-PHI) System
#### Indications for Use (Describe)
Upon intravenous administration of SPY AGENT GREEN (indocyanine green for injection, USP), the SPY-PH System is used with SPY AGENT GREEN to perform intraoperative fluorescence angiography. The SPY-PHI System is indicated for use in adult and pediatric patients one month of age and older.
The SPY-PHI System is indicated for fluorescence imaging of blood flow and tissue perfusion before, and after. vascular, gastrointestinal, organ transplant, and plastic, micro- and reconstructive surgical procedures.
Upon interstitial administration of SPY AGENT GREEN, the SPY-PHI System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Upon intradermal administration of SPY AGENT GREEN, the SPY-PHI System is indicated for fluorescence imaging of lymph nodes and delineation of lymphatic vessels during lymphatic mapping in adults with breast cancer for which this procedure is a component of intraoperative management.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| <div style="text-align:left;"> <span style="padding-right:5px">☒</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="text-align:left;"> <span style="padding-right:5px">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in all lowercase letters.
## 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c)
## Submitter:
| Applicant | Novadaq Technologies ULC (A Part of Stryker)<br>8329 Eastlake Drive, Unit 101<br>Burnaby, British Columbia V5A 4W2, Canada |
|----------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Cara Cahill<br>Manager, Regulatory Affairs<br>Phone: (408) 754-2473<br>Email: cara.cahill@stryker.com |
| Date Prepared | May 10, 2023 |
#### Subject Device:
| Name of Device | SPY Portable Handheld Imaging (SPY-PHI) System |
|----------------------|------------------------------------------------|
| Common or Usual Name | Fluorescence Angiographic System |
| Classification Name | Angiographic X-Ray System, 21 CFR 892.1600 |
| Regulatory Class | Class II |
| Product Code | IZI |
| 510(k) Review Panel: | Radiology |
## Predicate Device:
| Predicate Devices | Manufacturer |
|----------------------------------------------------------------------------|-------------------------------------------------|
| SPY-PHI System with SPY-PHI Fluorescence Assessment<br>Software (K202244)* | Novadaq Technologies ULC<br>(A Part of Stryker) |
| SPY Portable Handheld Imaging (SPY-PHI) System<br>(K200737) | Novadaq Technologies ULC<br>(A Part of Stryker) |
*primary predicate
## Device Description:
The SPY-PHI System is a real-time white-light and near-infrared illumination/ fluorescence imaging system used during open-field surgical procedures. Near-infrared illumination is used for fluorescence imaging using SPY AGENT® GREEN for the visual assessment of blood flow, tissue perfusion and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. It consists of the SPY Portable Handheld Imager/ imaging head with an integrated light guide and camera cable and the Video Processor/ Illuminator. Additionally, SPY-QP Fluorescence Assessment Software is provided as an optional upgrade to the SPY-PHI System that enables relative quantification of NIR fluorescence.
{4}------------------------------------------------
# Indications for Use:
| Item | Subject Device | Predicate Device | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| | SPY Portable Handheld Imaging (SPY-PHI)<br>System | SPY Portable Handheld Imaging<br>(SPY-PHI) System | |
| Intended Use | Visible white light and near-infrared illumination<br>during open-field surgical procedures. | Same as subject device | |
| Indications for Use | Upon intravenous administration of SPY AGENT<br>GREEN (indocyanine green for injection, USP), the<br>SPY-PHI System is used with SPY AGENT GREEN<br>to perform intraoperative fluorescence angiography.<br>The SPY-PHI System is indicated for use in adult and<br>pediatric patients one month of age and older.<br>The SPY-PHI System is indicated for fluorescence<br>imaging of blood flow and tissue perfusion before,<br>during, and after: vascular, gastrointestinal, organ<br>transplant, and plastic, micro- and reconstructive<br>surgical procedures.<br>Upon interstitial administration of SPY AGENT<br>GREEN, the SPY-PHI System is used to perform<br>intraoperative fluorescence imaging and visualization<br>of the lymphatic system, including lymphatic vessels<br>and lymph nodes.<br>Upon intradermal administration of SPY AGENT<br>GREEN, the SPY-PHI System is indicated for<br>fluorescence imaging of lymph nodes and delineation<br>of lymphatic vessels during lymphatic mapping in<br>adults with breast cancer for which this procedure is a<br>component of intraoperative management. | Upon intravenous administration<br>of SPY AGENT GREEN<br>(indocyanine green for injection,<br>USP) the SPY-PHI System is used<br>with SPY AGENT GREEN to<br>perform intraoperative<br>fluorescence angiography. The<br>SPY-PHI System is indicated for<br>use in adult and pediatric patients<br>one month of age and older.<br>The SPY-PHI System is indicated<br>for fluorescence imaging of blood<br>flow and tissue perfusion before,<br>during, and after: vascular,<br>gastrointestinal, organ transplant,<br>and plastic, micro- and<br>reconstructive surgical procedures.<br>Upon interstitial administration of<br>SPY AGENT GREEN, the SPY-<br>PHI System is used to perform<br>intraoperative fluorescence<br>imaging and visualization of the<br>lymphatic system, including<br>lymphatic vessels and lymph<br>nodes. | |
| Feature | Subject Device | Predicate Device | |
| Feature | SPY Portable Handheld Imaging (SPY-PHI) System | SPY Portable Handheld Imaging (SPY-PHI) System | |
| Manufacturer | Novadaq Technologies ULC. (a part of Stryker) | Same as subject device | |
| Submission Reference | Current Submission | K202244, K200737 | |
| Imaging System Type | Open-Field | Same as subject device | |
| Imaging Modes | White Light<br>Near-Infrared:<br>- Overlay Mode1<br>- SPY Mode (Contrast) 1<br>- SPY Color Segmented Fluorescence (CSF) | Same as subject device | |
| Principles of Operation | Via Imager, light is transmitted through the Light Cable and projected to illuminate the surgical site/ area of interest. The light emitted and reflected from the surgical site/ area of interest (aside from NIR excitation light, which is blocked by an internal rejection filter) is acquired by the Imager, projected onto the image sensor and converted to a digital signal. The digital signal is transmitted via the Camera Cable to the VPI which receives the optical image information, processes the information, and outputs it to digital video. | Same as subject device | |
| System Components | SPY-PHI Imager<br>Video Processor/ Illuminator<br>SPY-QP Fluorescence Assessment Software | Same as subject device | |
| Safety Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60825-1 | Same as subject device | |
| Optical<br>Imaging<br>Specifications | Dynamic Range | The user shall be able to visualize SPY AGENT GREEN in physiology applications. The response on the system monitor to the minimum clinically relevant concentration of SPY AGENT GREEN shall be at least 6.9(ΔE), and the system response to the maximum clinically relevant concentration shall be at least twice that at low concentrations. | Same as subject device |
| | Localization | The user shall be able to visualize SPY AGENT GREEN in anatomy applications. The response on the system monitor shall be at least 10.35 (ΔΕ) under clinically relevant conditions. | Same as subject device |
| SPY-PHI Imager | Image Sensor: CMOS image sensor | Same as subject device | |
| Video Processor/ Illuminator | Light Source/ RGB LEDs | Same as subject device | |
| Feature | Subject Device | | Predicate Device |
| | SPY Portable Handheld Imaging (SPY- | | SPY Portable Handheld |
| | PHI) System | | Imaging (SPY-PHI) System |
| | Laser Safety | Class 3R | Same as subject device |
| | Class | | |
| | Wavelengths | White Light: | Same as subject device |
| | | 400 nm – 700 nm | |
| | | Near-infrared: 805 nm | |
| | Imaging | Digital | Same as subject device |
| | Processing/ | | |
| | Video Output | | |
| | Resolution | 1080p (1920x1080) | Same as subject device |
| | Frame Rate | 60 Hz | Same as subject device |
| Imaging Agent | | SPY AGENT GREEN (indocyanine green for | Same as subject device |
| | injection, USP) | | |
{5}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device:
{6}------------------------------------------------
1 SPY Mode (Contrast) and Overlay Mode are recommended for use in breast lymphatic mapping.
{7}------------------------------------------------
## Performance Testing:
| Test | Method | Result |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Electromagnetic<br>Compatibility | In accordance with IEC 60601-1-2:2014, Medical electrical<br>equipment, Part 1. General requirements for safety, 2. Collateral<br>standard: Electromagnetic compatibility - Requirements and tests | PASS |
| Electrical Safety | In accordance with:<br>• IEC 60601-1:2005+A1:2012, Medical electrical equipment, Part 1.<br>General requirements for safety and essential performance; &<br>applicable national deviations<br>• IEC 60601-1-6:2020-07, Medical electrical equipment - Part 1-6:<br>General requirements for basic safety and essential performance | PASS |
| Laser Safety | In accordance with IEC 60825-1:2014, Safety of laser products - Part<br>1: Equipment classification and requirements | PASS |
| Usability | In accordance with IEC62366-1:2015, Medical devices -- Part 1:<br>Application of usability engineering to medical devices | PASS |
| Performance Testing - Bench | In accordance with design input specifications including optical<br>imaging performance specifications | PASS |
| Software | In accordance with IEC 62304:2006, Medical device software -<br>Software life cycle processes | PASS |
Note: The device is not patient contacting. Therefore, biocompatibility does not apply.
## Performance Testing - Clinical:
The Novadaq Technologies ULC-sponsored FILM-B Clinical Study was performed under an IDE (G170109), in accordance with 21 CFR 812.
FILM-B was a single-arm, prospective. multi-center, open-label study in subjects with early stage breast cancers with clinically negative nodal status and who were scheduled for surgery that included clinically indicated sentinel lymph node biopsy with technetium 99m radioactive colloid (Tc99m). Subjects received Tc99m the day before or the day of surgery and SPY AGENT GREEN was injected into the identified breast of subjects at the beginning of the operative procedure. All subjects received the injection into a single breast. A total of 152 subjects were enrolled in the study. Lymphatic mapping was performed intraoperatively using the SPY-PHI System and a handheld gamma probe, or the surgeon's visual and palpation examination. The resected tissues were evaluated by histopathology to confirm presence of lymph nodes.
The FILM-B study design resulted in unbiased, quality, reliable and repeatable data for use in the evaluation of the effectiveness and safety of SPY AGENT GREEN and SPY-PHI in identifying lymph nodes during lymphatic mapping and sentinel lymph node biopsy in subjects with earlystage breast ca…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.