K230663 · Jeisys Medical, Inc. · GEI · Aug 14, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230663
Device Name
Density
Applicant
Jeisys Medical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
DENSITY is a software-controlled radiofrequency (RF) electrosurgical device operating at 6.78 MHz. It delivers RF energy to target tissue via a handpiece with an electrode tip to induce electrothermal reactions, resulting in tissue coagulation and hemostasis. The device utilizes a 'reverse thermal gradient' principle, heating tissue to 65–75°C while cooling the epidermis with cryogen to facilitate collagen fiber denaturation and contraction. Components include a main unit, handpiece, six electrode tips (monopolar and bipolar), foot switch, and neutral electrode pad. Used in clinical settings by healthcare professionals, the device output is managed via a color touch panel. The system provides controlled thermal energy to achieve clinical outcomes in dermatologic and general surgical procedures.
Clinical Evidence
No clinical performance testing was performed. Evidence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), high-frequency surgical equipment requirements (IEC 60601-2-2), usability engineering (IEC 60601-1-6, IEC 62366), software verification/validation (IEC 62304), and biocompatibility (ISO 10993 series).
Technological Characteristics
RF electrosurgical device; 6.78 MHz operating frequency; 400 W max power. Features monopolar and bipolar electrode tips. Cooling system uses 1234ze cryogen with solenoid valve control. User interface via color touch panel. Software-controlled. Biocompatibility per ISO 10993. Safety/performance per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 62304, and IEC 62366.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
August 14, 2023
Jeisys Medical Inc. % Sanghwa Myung Regulatory Affairs Consultant E&m D-1474, 230, Simin-daero, Dongan-gu Seoul, Gyeonggi-do 14067 Korea, South
Re: K230663
Trade/Device Name: Density Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: July 14, 2023 Received: July 14, 2023
Dear Sanghwa Myung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed Mark by Mark Trumbore - Trumbore - S Date: 2023.08.14 S 14:21:44 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 5. Indication for use
See next pages
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230663
Device Name DENSITY
Indications for Use (Describe)
DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 6. 510(k) Summary
## K230663
This summary of 510(k) Safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
Submitter: Jeisys Medical Incorporated 307 Daeryung Techno Town 8th Gamasan-ro 96, Geumcheon-Gu, Seoul, 08501, South Korea Telephone: +82-70-7435-4930 Fax: 82-10-2843-3547 E-mail: brkim@jeisys.com
- Contact Person: E&M Regulatory Affair/Sang Hwa Myung Telephone: +80-70-7807-0550 FAX: +82-31-388-9263 Cellphone: +82-10-4952-6638 E-mail: mshenmc@gmail.com Primary Contact: Sanghwa Myung
Date 510(k) summary Prepared: July 14, 2023
#### Trade Name: DENSITY
Common Name: Electrosurgical Device Classification: II Product Code: GEI Regulation Numbers: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[ Review Panel: General & Plastic Surgery (ODE)
#### Description of Device:
The DENSITY generates radiofrequency (RF) energy by means of high RF at 6.78MHz. The RF energy is delivered through the skin into the target tissue via a handpiece equipped with an electrode tip. As the RF energy passes through the tissue, it generates an electrothermal reaction which is capable of coagulating the tissue. In detail, the main body uses the "reverse thermal gradient" principle to deliver RF energy, and thus, heat is generated and selective coaqulation occurs while cooling the epidermis, resulting in denaturation and contraction of collagen fibers.
The DENSITY is an RF (radiofrequency), software-controlled electrosurgical device used for electrocoagulation of soft tissue and hemostasis.
The DENSITY consists of the following components:
•Electrosurgical Unit - Main body
{5}------------------------------------------------
- •Handpiece
- •6 different electrode tips
- •Neutral electrode pad and neutral electrode pad cable, cleared under K201685
- •Foot switch
- •Power cord
Indication for use:
DENSITY indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
#### Predicate Device:
| Predicate Devices | K170758<br>Applicant: Solta Medical, Inc<br>Trade/Device Name: Thermage FLX System<br>Regulation Number: 21 CFR 878.4400<br>Regulation Name: Electrosurgical Cutting and Coagulation<br>Device and Accessories<br>Regulatory Class: Class II<br>Product Code: GEI, ISA |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### Substantial Equivalence:
Comparison table is as follows.
#### Table 1: Substantial equivalence comparison
| | Proposed Device | Predicate Device |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K170758 |
| Manufacturer | Jeisys | So;ta Medical, Inc |
| Trade/Device Name | DENSITY | Thermage FLX System |
| Classification Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Electrosurgical cutting and<br>coagulation device and<br>accessories |
| Indication for use | DENSITY indicated for use<br>in dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. | The radiofrequency-energy<br>only delivery components<br>of the Thermage FLX<br>System are indicated for<br>use in: • Dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis; |
| Output energy type | Radio Frequency | Radio Frequency |
| User interface | Color Touch Panel | Color Touch Panel |
| Operating Frequency | 6.78MHz | 6.78MHz |
| Max Power | 400 W | 400W |
| Electrode tip | 3 types (0.25 cm², 4 cm²) | 4 types (0.25 cm², 3.0 cm², |
{6}------------------------------------------------
| | 16cm²) | 4 cm², 16cm²) |
|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Coolant solution | Cryogen | Cryogen |
| Temperature range | 65~75°C | 65~75°C |
| RF Time | 50 ~ 800 ms | 50 ~ 1000 ms |
| Impedance | 75 - 400 Ω | 75 - 400 Ω |
| Communication to<br>generator | LCD touch screen | LCD touch screen |
| Style of electrode tip | Monopolar, Bipolar | Monopolar |
| Number of active<br>electrodes per applicator | Number of output patterns<br>- I-Tip(Around eye) I, F-<br>Tip(Face): 1ea<br>- I-Tip II(Around eye), F-<br>Tip II(Face): 2ea<br>- B-Tip(Body), B-Tip(Body)<br>II: 4ea | Number of output patterns<br>- EYE Tip 0.25: 1ea<br>- TOTAL Tip 3.0: 1ea<br>- NEW TOTAL TIP 4.0: 1ea<br>- BODY TIP 16.0: 4ea |
| Coolant control<br>parameters<br>(valve type; valve<br>activation, power supply,<br>solution) | Type of gas: 1234ze<br>Method of gas control:<br>Handpiece solenoid<br>control operation<br>Mouth valve specifications:<br>12V<br>Solenoid valve<br>specifications:<br>1.8V 24.8 BAR | Type of gas: 1234ze<br>Method of gas control:<br>Handpiece solenoid control<br>operation<br>Mouth valve specifications:<br>12V<br>Solenoid valve<br>specifications:<br>1.8V 24.8 BAR |
#### - Discussion
- Electrosurgical Unit:
- Operating frequency of 6.78 MHz: same as for the predicate device. ●
- Max power, same as for the predicate device.
- . Electrode tip: our device's tip types are including the predicate device.
- Temperature range: same as for the predicate device.
- o Impedance: same as for the predicate device.
- Coolant solution: same as the reference device.
- 0 Style of electrode tip: Subject device have two type of electrode tip that is monopolar and bipolar. Predicate device is monopolar type. This mode difference should not affect performance or safety.
- 0 Different #1 Subject device and predicate device are divided into types of TIP depending on the area to be applied. This difference does not affect performance or safety.
- 0 Coolant control parameters: DENSITY and predicate devices have the same parameters, but the exact technical specifications are unknown.
There are no significant differences between DENSITY and the predicate devices (K170758) that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics
{7}------------------------------------------------
#### Non-clinical Performance Data:
- -. Basic safety and essential performance of the DENSITY is evaluated in accordance with IEC 60601-1:2005/AMDI:2012
- -. Effect to the device by electromagnetic disturbances is evaluated in accordance with the FDA-recognized consensus standard, IEC 60601-1-2:2014.
- -. Medical electrical equipment Part 2 -2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, IEC 60601-2-2
- -. General Requirements for Basic Safety and Essential Performance Collateral Standard: Usability is evaluated in accordance with the FDA-recognized consensus standard, IEC 60601-1-6:2013.
- -. The software for DENSITY is verified and validated in accordance with its moderate level of concern. Software life cycle processes are evaluated according to the FDA-recognized consensus standard, IEC 62304:2006.
- -. Application of usability engineering to medical devices is evaluated in accordance with IEC 62366:2008 based on Human Factor Engineering
- -. Biocompatibility of DENSITY is documented in the reference of ISO 10993-1:2009, ISO 10993-5:2009, and 10993-10:2010.
#### Risk Management
A risk analysis was conducted based on ISO 14971:2012 Medical devices – Application of risk management to medical devices
#### Clinical Data:
No clinical performance testing was performed.
#### Conclusion
Subject device is substantially equivalent to its predicate devices with same indication for use and technological characteristics. Any minor differences in the human interface and accessories design do not raise any new types of safety and effectiveness issues, as verified by performance testing. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that we conclude that substantially equivalent with predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.