# Zeta Cranial Navigation System (K230661)

_Zeta Surgical · HAW · Sep 8, 2023 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230661

## Device Facts

- **Applicant:** Zeta Surgical
- **Product Code:** [HAW](/productcode/HAW.md)
- **Decision Date:** Sep 8, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Zeta Cranial Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is only indicated for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites.

## Device Story

Zeta Cranial Navigation System is a stereotaxic, image-guided planning and intraoperative guidance system for cranial procedures. It takes 3D preoperative CT/MRI scans (DICOM format) as input. The system uses structured light and machine vision to perform automatic, markerless, pinless patient registration. It tracks wireless, unpowered surgical instruments via optical technology. The system displays the instrument's position relative to the 3D scan on a monitor, providing real-time guidance for surgeons. It is used in operating rooms and interventional suites by neurosurgeons and staff. The output assists in precise instrument positioning, potentially improving surgical accuracy and workflow efficiency by eliminating the need for rigid patient fixation. The system includes a cart, sensor head, positioning arms, and software.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including accuracy testing under simulated clinical conditions (varying surgical illumination, dynamic patient motion, and worst-case environmental factors), design validation, and human factors/usability testing.

## Technological Characteristics

System components: cart, sensor head, positioning arms, monitor, and software. Tracking: optical. Registration: structured light and machine vision (automatic, markerless, pinless). Imaging: 3D CT/MRI (max 2mm slice thickness). Connectivity: standalone (no network). User interface: non-sterile touchscreen with sterile cover. Software: Major level of concern. Sterilization: reusable, non-sterile, cleaning instructions provided.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- 7D Surgical System Cranial Biopsy and Ventricular Catheter Placement Application ([K192945](/device/K192945.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 8, 2023

Zeta Surgical Roman Stolyarov, Ph.D. Vice President of Product 2 Park Plaza. Suite 400 Boston, Massachusetts 02116

Re: K230661

Trade/Device Name: Zeta Cranial Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: August 9, 2023 Received: August 9, 2023

Dear Dr. Stolyarov:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows a digital signature. The name "Adam D. Pierce -S" is written on the left side of the image. On the right side of the image, it says "Digitally signed by Adam D. Pierce -S" and "Date: 2023.09.08 19:01:29-04'00'".

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230661

Device Name Zeta Cranial Navigation System

### Indications for Use (Describe)

The Zeta Cranial Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is only indicated for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites.

| Type of Use (Select one or both, as applicable)            |                                                           |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Zeta Cranial Navigation System is provided below.

#### SUBMITTER 1.

| Applicant:                | Zeta Surgical Inc.<br>115 Kingston Street, Floor 2<br>Boston, MA 02111                                                                                                               |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                  | Roman Stolyarov, PhD<br>Vice President of Product<br>Zeta Surgical Inc.<br>115 Kingston Street, Floor 2<br>Boston, MA 02111<br>+1 (214) 284-7990<br>roman.stolyarov@zetasurgical.com |
| Submission Correspondent: | Roman Stolyarov, PhD<br>Vice President of Product<br>Zeta Surgical Inc.<br>115 Kingston Street, Floor 2<br>Boston, MA 02111<br>+1 (214) 284-7990<br>roman.stolyarov@zetasurgical.com |
| Date Prepared:            | September 7, 2023                                                                                                                                                                    |

#### 2. DEVICE

| Device Trade Name:  | Zeta Cranial Navigation System          |
|---------------------|-----------------------------------------|
| Device Common Name: | Neurological Stereotaxic Instrument     |
| Classification Name | Stereotaxic instrument, 21 CFR 882.4560 |
| Regulatory Class:   | Class II                                |
| Product Code:       | HAW                                     |

#### PREDICATE DEVICE 3.

Predicate Device: 7D Surgical System Cranial Biopsy and Ventricular Catheter Placement Application [K192945]

{4}------------------------------------------------

#### DEVICE DESCRIPTION 4.

The Zeta Cranial Navigation System is a stereotaxic, image guided planning and intraoperative guidance system enabling computer-assisted cranial interventional procedures. The system assists surgeons with the precise positioning of surgical instruments relative to patient anatomy by displaying the position of navigated surgical instruments relative to 3D preoperative medical scans.

#### 5. INDICATIONS FOR USE

The Zeta Cranial Navigation System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The device is indicated only for cranial surgery where reference to a rigid anatomical structure can be identified, does not require rigid fixation of the patient, and does not require fixation of a navigated instrument guide to the patient. The system is intended to be used in operating rooms and in less acute surgical settings such as interventional procedure suites.

#### SUBSTANTIAL EQUIVALENCE 6.

# Comparison of Indications

| Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Zeta Cranial Navigation System is a stereotaxic<br>image guidance system intended for the spatial<br>positioning and orientation of neurosurgical<br>instruments used by surgeons. The device is indicated<br>only for cranial surgery where reference to a rigid<br>anatomical structure can be identified, does not<br>require rigid fixation of the patient, and does not<br>require fixation of a navigated instrument guide to the<br>patient. The system is intended to be used in operating<br>rooms and in less acute surgical settings such as<br>interventional procedure suites. | The 7D Surgical System is a stereotaxic image<br>guidance system intended for the spatial positioning<br>and orientation of neurosurgical instruments used by<br>surgeons. The system is also intended to be used as the<br>primary surgical luminaire during image guided<br>surgery. The device is indicated for cranial surgery<br>where reference to a rigid anatomical structure can be<br>identified. |

# Technological Comparisons

The table below compares the key technological features of the subject device to the predicate device (7D Surgical System Cranial Biopsy and Ventricular Catheter Placement Application, K192945).

| Table 1: | Technological Comparison |
|----------|--------------------------|
|----------|--------------------------|

|                                                 | Proposed Device<br>(K230661)                                                                                                              | Predicate Device<br>(K192945)                                                                                                             |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                   | K230661                                                                                                                                   | K192945                                                                                                                                   |
| Applicant                                       | Zeta Surgical                                                                                                                             | 7D Surgical                                                                                                                               |
|                                                 | Proposed Device<br>(K230661)                                                                                                              | Predicate Device<br>(K192945)                                                                                                             |
| Device Name                                     | Zeta Cranial Navigation System                                                                                                            | 7D Surgical System Cranial Biopsy and<br>Ventricular Catheter Placement Application                                                       |
| Classification Regulation                       | 21 CFR 882.4560                                                                                                                           | 21 CFR 882.4560                                                                                                                           |
| Product Code                                    | HAW                                                                                                                                       | HAW                                                                                                                                       |
| Classification                                  | Class II                                                                                                                                  | Class II                                                                                                                                  |
| Intended User                                   | Neurosurgeons and neurosurgical procedure<br>staff                                                                                        | Neurosurgeons and neurosurgical procedure<br>staff                                                                                        |
| Intended Use<br>Environment                     | Neurosurgical operating room and facilities<br>equipped for interventional cranial procedures                                             | Neurosurgical operating room and facilities<br>equipped for interventional cranial procedures                                             |
| Anatomical Site                                 | Head                                                                                                                                      | Head                                                                                                                                      |
| Principal of Operation                          | Preoperative image upload<br>• Surgical planning<br>• Patient registration<br>• Instrument guidance                                       | Preoperative image upload<br>• Surgical planning<br>• Patient registration<br>• Instrument guidance                                       |
| Technology                                      |                                                                                                                                           |                                                                                                                                           |
| Imaging Modalities                              | 3D CT / MRI (max slice thickness 2mm)                                                                                                     | 3D CT / MRI (max slice thickness 2mm)                                                                                                     |
| Data Format                                     | DICOM                                                                                                                                     | DICOM                                                                                                                                     |
| CD-ROM Input                                    | Yes                                                                                                                                       | Yes                                                                                                                                       |
| USB Input                                       | Yes                                                                                                                                       | Yes                                                                                                                                       |
| DICOM Compliance                                | Yes                                                                                                                                       | Yes                                                                                                                                       |
| Instrument Tracking<br>Technology               | Optical                                                                                                                                   | Optical                                                                                                                                   |
| Registration Technology                         | Structured light and machine vision                                                                                                       | Structured light and machine vision                                                                                                       |
| Major System<br>Components                      | Cart<br>• Sensor head<br>• Sensor head positioning arm<br>• Monitor<br>• Monitor positioning arm<br>• Tracked instruments<br>• Software   | Cart<br>• Sensor head<br>• Sensor head positioning arm<br>• Monitor<br>• Monitor positioning arm<br>• Tracked instruments<br>• Software   |
| Workflow Components                             | Upload<br>• Segmentation<br>• Planning<br>• Staging/Positioning<br>• Registration<br>• Instrument Calibration<br>• Navigation             | Upload<br>• Segmentation<br>• Planning<br>• Instrument Calibration<br>• Registration (includes Staging/Positioning)<br>• Navigation       |
|                                                 | Proposed Device<br>(K230661)                                                                                                              | Predicate Device<br>(K192945)                                                                                                             |
| Planning Features                               | Multiple target point selection, multiple<br>instrument selection                                                                         | Multiple target point selection, multiple<br>instrument selection                                                                         |
| Navigation Features                             | Target point projection, instrument rendering,<br>instrument extended trajectory, physical<br>distance measurement, multiple perspectives | Target point projection, instrument rendering,<br>instrument extended trajectory, physical<br>distance measurement, multiple perspectives |
| Image Guided                                    | Yes                                                                                                                                       | Yes                                                                                                                                       |
| Real-time Display of the<br>Instrument Position | Yes                                                                                                                                       | Yes                                                                                                                                       |
| Provide Guidance for<br>Surgical Instruments    | Yes                                                                                                                                       | Yes                                                                                                                                       |
| Instrument Technology                           | Wireless and unpowered                                                                                                                    | Wireless and unpowered                                                                                                                    |
| Instrument<br>Compatibility                     | Brainlab Disposable Stylet                                                                                                                | 7D instruments                                                                                                                            |
| Initial Registration<br>Method                  | Automatic                                                                                                                                 | Manual (requires user defined features)                                                                                                   |
| Continuous Registration<br>Method               | Automatic and pinless                                                                                                                     | Patient is pinned                                                                                                                         |
| Markerless Registration                         | Yes                                                                                                                                       | Yes                                                                                                                                       |
| Network Connectivity                            | No                                                                                                                                        | Yes                                                                                                                                       |
| User Interface                                  | Non-sterile touchscreen monitor that is covered<br>with a transparent sterile cover during the<br>procedure                               | Non-sterile workstation with foot pedal<br>controls for use during the procedure                                                          |
| Performance                                     |                                                                                                                                           |                                                                                                                                           |
| Maximum Tracking<br>Speed (cm/s)                | 0.88                                                                                                                                      | N/A - pinned system                                                                                                                       |
| Effective Navigation<br>Latency (c)             | 0.17                                                                                                                                      | Unknown                                                                                                                                   |

{5}------------------------------------------------

{6}------------------------------------------------

#### 7. PERFORMANCE DATA

# Biocompatibility Testing

There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.

# Electrical safety and electromagnetic compatibility (EMC)

The Zeta Cranial Navigation System was tested in accordance with the following standards:

- IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical . electrical equipment: Part 1: General requirements for basic safety and essential

{7}------------------------------------------------

performance including US deviations, with the exception of Clause 11.7 regarding biocompatibility. The device passed all tests.

- IEC 62304:2006+ Amd 1:2015. Medical device software Software life cycle . processes. The device passed all tests.
- IEC 60601-1-2:2014+A1:2021, Medical electrical equipment Part 1-2: General ● requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances – Requirements and tests. The device passed all tests.
- IEC 60601-1-6, Edition 3.2 2020-07, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. The device passed all tests.

# Software Verification and Validation Testing

Software verification and validation testing was conducted, and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device was considered a Major level of concern.

Cybersecurity documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.

## Sterilization, Cleaning, and Shelf Life

## Sterilization and Cleaning

The device is reusable, provided non-sterile, and is not sterile when used. Cleaning instructions are provided in the labeling.

## Shelf-Life

-

Shelf-life is not applicable due to the low likelihood of time-dependent product degradation.

# Bench Testing

The following bench testing was performed to demonstrate substantial equivalence:

- Accuracy testing under different conditions, including:
	- Simulated clinical procedures using virtual targets that can be located only with o the guidance system
	- Surgical illumination levels O
	- Dynamic patient motion O
	- о Other worst-case physical and environmental conditions
- Design validation testing -
- Human factors testing, following the FDA Guidance Document, "Applying Human -Factors and Usability Engineering to Medical Devices"

{8}------------------------------------------------

# Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

# Clinical Data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

#### 8. CONCLUSION

The results of testing described above demonstrate that the Zeta Cranial Navigation System is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K230661](https://fda-staging.innolitics.com/device/K230661)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).