Light curing one-component opaquer to mask framework structures. Signum opaque F is the opaquer component for veneering materials and denture base materials (heat/cold curing denture acrylics) offered by Kulzer.
Device Story
Signum opaque F is a light-curing, one-component resin paste used by dental professionals to mask framework structures (metal or zirconia) for crowns, bridges, and partial dentures. The device is supplied in a 3ml syringe with a plunger and closing cap. The clinician dispenses the paste onto a mixing block, applies it to the framework using a disposable brush in multiple uniform, semi-overlapping layers, and polymerizes each layer individually using a light-curing unit. The opaquer serves as a base layer for subsequent veneering or denture base materials. By masking the underlying framework, the device improves the aesthetic outcome of the final dental restoration. It is intended for extraoral use only.
Clinical Evidence
No clinical data. Evaluation is based on internal data and scientific literature regarding veneering composites. Bench testing included shear bond strength on non-precious metal, precious metal, zirconium dioxide, and PEEK frameworks, as well as thermal shelf-life testing.
Technological Characteristics
Light-curing, one-component resin paste. Packaged in 3ml syringe with plunger and closing tip. Applied via disposable brush and polymerized via light-curing. Compatible with metal, zirconia, and PEEK frameworks. Conforms to applicable dental material standards.
Indications for Use
Indicated for extraoral color-masking of framework structures (metal or zirconia) for crowns, bridges, and partial dentures using veneering materials and denture base materials.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2023
Kulzer, LLC David Vincent Director, QA/RA 4315 S. Lafayette Blvd South Bend, Indiana 46614
Re: K230644
Trade/Device Name: Signum opaque F Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: August 4. 2023 Received: August 4, 2023
Dear David Vincent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K230644
Device Name
Signum opaque F
Indications for Use (Describe)
Light curing one-component opaquer to mask framework structures. Signum opaque F is the opaquer component for veneering materials and denture base materials (heat/cold curing denture acrylics) offered by Kulzer.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) #: | K230644 | 510(k) Summary | Prepared on: | 2023-02-24 |
|-----------|---------|----------------|--------------|------------|
|-----------|---------|----------------|--------------|------------|
Contact Details
| Applicant Name | Kulzer, LLC | 21 CFR 807.92(a)(1) |
|-----------------------------|----------------------------------------------------------|---------------------|
| Applicant Address | 4315 S. Lafayette Blvd South Bend IN 46614 United States | |
| Applicant Contact Telephone | (574) 299-5421 | |
| Applicant Contact | Mr. David Vincent | |
| Applicant Contact Email | David.Vincent@kulzer-dental.com | |
Device Name
| Device Trade Name | Signum opaque F | 21 CFR 807.92(a)(2) |
|---------------------|------------------------------|---------------------|
| Common Name | Tooth shade resin material | |
| Classification Name | Material, Tooth Shade, Resin | |
| Regulation Number | 872.3690 | |
| Product Code | EBF | |
Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | 21 CFR 807.92(a)(3) |
|-------------|----------------------------------------------------------|--------------|---------------------|
| K152373 | VITA VM LC FLOW | EBF | |
Device Description Summary
21 CFR 807.92(a)(4)Signum opaque F is a light-curing, one component opaquer used for extraoral color-masking of framework structures for crowns, bridges, and/or partial dentures. The veneering composites are processed by dental restoration by veneering framework-supported restorations, such as metal or zirconia frameworks.
Signum opaque F is a paste stored in a syringe featuring a plunger and a closing cap. The product is dispensed and is then applied with a disposable brush until the scaffolding frame is completely covered. After polymerization, the crown or bridge is inserted into the oral cavity, adapted and connected to the remaining residual teeth or functionally adapted.
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
### Intended Use/Indications for Use
Light curing one-component opaquer to mask framework structures. Signum opaque F is the opaquer component for veneering materials and denture base materials (heat/cold curing denture acrylics) offered by Kulzer.
## Indications for Use Comparison
Our submission device, Signum opaque F and the predicate device, VITA VM LC Flow (K152373) have the same indications for use. They both are used in conjunction with bonding systems and provide a layer of resin type material to act as an opaquer is to mask framework structures. They both have the indication for extraoral use with veneering materials. VITA VM LC Flow (K152373) also has individualization of acrylic teeth, long term temporaries, and reproduction of gingival areas. While the verbiage of the indications for use differs and the scope of the indications for use of they are both indicated for extraoral use and have the indication to act as an opague laver to mask the underlying metal frameworks.
{4}------------------------------------------------
Our indication for use for Signum opaque F is listed as one of the indications listed for the predicate device, VITA VM LC Flow (K152373), They are both used as an opaquer for color-masking of framework structures.
### Technological Comparison
21 CFR 807.92(a)(6)
Signum opaque F and the predicate device, VITA VM LC Flow (K152373) are dispensed in a paste form. Signum opaque F's primary packaging is a 3ml syringe with plunger and closing tip is used for dispensing of the product. Signum opaque F is placed on the mixing block and stired thoroughly. The opaque is applied with a suitable, disposable brush in several, uniform, semioverlapping layers until the scaffolding frame is completely covered. Each layer must be polymerized individually. Vita VM LC Flow is also packaged in a syringe with a plunger to extruded, Vita VM LC recommends using a metal spatula to adjust thixotropically (their consistency can be changed - from firmer to softer - by slightly pressing onto them with an instrument). VITA VM LC Flow is then applied in layers and polymerized.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Mechanical testing was performed in regard to testing shear bond strength. This testing was performed on various metal frameworks (i.e. non-precious metal, precious metal, zirconium dioxide, PEEK (high performance polymers)).
Thermal testing was performed in the shelf-life report, utilizing testing methods at differing temperatures.
All nonclinical testing was performed with conformity to applicable standards.
This clinical evaluation is based on internal data and scientific literature currently available about veneering composites and comparable medical devices. Veneering composites of Kulzer are marketed purpose without any hithout any hint for negative benefit-risk ratio. The final individualized medical devices, prepared in a conventional process, is well-established. Therefore, clinical investigation with veneering composites c&b is not necessary and the route for this clinical evaluation is literature based.
In regard to the conforming standards of both Signum opaque F and the predicate device VITA VM LC Flow, the same standards are referenced. Therefore, it can be determined that the testing methods and results of both devices are in alignment and substantial equivalence can be concluded.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.