APOLLO ADVANTAGE Nitrile Examination Gloves

K230633 · Dell Corning Corporation · LZA · Jun 5, 2023 · General Hospital

Device Facts

Record IDK230633
Device NameAPOLLO ADVANTAGE Nitrile Examination Gloves
ApplicantDell Corning Corporation
Product CodeLZA · General Hospital
Decision DateJun 5, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

APOLLO ADVANTAGE Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

Device Story

APOLLO ADVANTAGE Nitrile Examination Gloves are non-sterile, powder-free, blue nitrile rubber gloves. Used in medical settings by clinicians to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. Performance verified through bench testing against ASTM standards for physical dimensions, tensile strength, elongation, and watertightness. Biocompatibility confirmed via ISO 10993 testing for irritation, sensitization, and systemic toxicity.

Clinical Evidence

No clinical data. Bench testing only. Compliance with ASTM D6319 (physical dimensions, tensile strength, elongation), ASTM D5151 (watertightness), ASTM D6124 (residual powder), and ISO 10993 (biocompatibility: irritation, sensitization, systemic toxicity) demonstrated.

Technological Characteristics

Material: Nitrile rubber. Form factor: Disposable, non-sterile, powder-free, blue gloves. Sizes: S, M, L, XL. Standards: ASTM D6319 (dimensions/properties), ASTM D5151 (holes), ASTM D6124 (powder), ISO 10993-10/11 (biocompatibility).

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hands to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font. June 5, 2023 Dell Corning Corporation % Yiren Qiao General Manager Dell Corning Asia Co., Ltd 12E, 7th Floor, Bank of China Tower, Qingshan District Wuhan, Hubei 430000 China Re: K230633 Trade/Device Name: APOLLO ADVANTAGE Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: March 3, 2023 Received: March 7, 2023 Dear Yiren Qiao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Bifeng Qian -S Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. # 510(k) Number (if known) K230633 Device Name APOLLO ADVANTAGE Nitrile Examination Gloves Indications for Use (Describe) APOLLO ADVANTAGE Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) 区 Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Dow Corning Corporation. The logo features the words "Dow" and "Corning" stacked on top of each other in a bold, sans-serif font. The word "Dow" is in a lighter gray color, while "Corning" is in a darker blue color. Below the two words is the word "corporation" in a smaller, lighter font. ## 510(k) Summary ## K230633 This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. #### 5.1 Submitter | Submitted by: | Dell Corning Corporation | |-----------------|-----------------------------------------------------------------------------------------| | Address: | 575 John Dodd Road, Spartanburg, South Carolina 29303, USA. | | Contact Person: | Yiren Qiao<br>General Manager | | | | | | Dell Corning Asia Co., Ltd | | Address: | 12E, 7th Floor, Bank of China Tower, Qingshan District,<br>Wuhan, Hubei 430000<br>China | | Phone: | +86-15871756987 | | Email: | kinginsky@126.com | | Date Prepared: | May 18, 2023 | #### 5.2 Device | Device Name: | APOLLO ADVANTAGE Nitrile Examination Glove | |----------------------|--------------------------------------------| | Classification Name: | Polymer Patient Examination Glove | | Regulatory Class: | I | | Regulation Number: | 21 CFR 880.6250 | | Regulation Name: | Non-powdered patient examination glove | | Product Code: | LZA | | Size: | S, M, L, XL | #### 5.3 Predicate Device | Device Name: | Nitrile Examination Glove, K213739 | | | |----------------------|----------------------------------------|--|--| | Manufacturer: | Hangzhou Runheng Medical Co., Ltd | | | | Classification Name: | Polymer Patient Examination Glove | | | | Regulatory Class: | I | | | | Regulation Number: | 21 CFR 880.6250 | | | | Regulation Name: | Non-powdered patient examination glove | | | | Product Code: | LZA | | | | Size: | XS, S, M, L, XL | | | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Dell Corning Corporation. The word "Dell" is in a gray, italicized font and is stacked on top of the word "Corning" which is in a bold, blue font. Below the two words is the word "corporation" in a smaller, blue font. #### 5.4 Device Description APOLLO ADVANTAGE Nitrile Examination Gloves are made of Nitrile rubber and are blue in color. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151 and ASTM D6319. The proposed device is non-sterile and powder free. #### 5.5 Indication for Use: APOLLO ADVANTAGE Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. | Table 5-1 General Comparison | | | | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Item | Proposed Device | Predicate | Remark | | | APOLLO<br>ADVANTAGE Nitrile<br>Examination Gloves | Device(K213739)<br>Nitrile Examination<br>Glove | | | Product Code | LZA | LZA | Same | | Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same | | Classification | I | I | Same | | Indication for Use | APOLLO<br>ADVANTAGE Nitrile<br>Examination Gloves are<br>disposable devices<br>intended for medical<br>purposes that are worn<br>on the examiner's hands<br>to prevent<br>contamination between<br>patient and examiner. | Nitrile Examination<br>Gloves are disposable<br>devices intended for<br>medical purpose that<br>are worn on the<br>examiner's hand to<br>prevent contamination<br>between patient and<br>examiner. | Same | | Material | Nitrile | Nitrile | Same | | Powered or Powered<br>Free | Powered Free | Powered Free | Same | | Color | Blue | Blue | Same | #### 5.6 Comparison of Technological Characteristics | Proposed Device | Standard | Designation | Size | | | | Tolerance | |---------------------------------------------------------|---------------|-------------|------|----|-----|-----|-----------| | APOLLO<br>ADVANTAGE<br>Nitrile<br>Examination<br>Gloves | ASTM<br>D6319 | Length,mm | S | M | L | XL | min | | | | Width, mm | 80 | 95 | 110 | 120 | ±10 | | Thickness,mm | | | | | | | | | Finger | | | 0.05 | | | | min | | Palm | | | 0.05 | | | | min | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Dell Corning Corporation. The word "Dell" is in a light gray color and is stacked on top of the word "Corning" which is in black. Below the word "Corning" is the word "Corporation" which is in a light blue color. | Predicate | Standard | Designation | Size | | | | | Tolerance | |------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------|------|----|----|-----|-----|-----------| | Device(K213739)<br>Nitrile<br>Examination<br>Glove<br>ASTM D6319 | ASTM<br>D6319 | Length,mm | XS* | S | M | L | XL | min | | | | Width,mm | 70 | 80 | 95 | 110 | 120 | ±10 | | | | Thickness, mm | | | | | | | | | | Finger | 0.05 | | | | min | | | | | Palm | 0.05 | | | | min | | | Remark | Similar<br>*Difference discussion: No size XS is included in our<br>submission. But all sizes we submitted met the specification of<br>ASMT D6319. | | | | | | | | - Table 5-3 Performance Comparison | | Item | | APOLLO<br>ADVANTAGE<br>Nitrile Examination<br>Gloves<br>(Proposed Device) | Nitrile Examination<br>Glove(Predicate<br>Device, K213739) | Remark | |------------------------|----------------------------------------------------|------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------------|---------| | Colorant | | | Blue | Blue | Same | | Physical<br>Properties | Before<br>Aging<br>Comply<br>with<br>ASTM<br>D6319 | Tensile<br>Strength | 14MPa,min | 14MPa,min | Same | | | Comply<br>with<br>ASTM<br>D6319 | Ultimate<br>Elongation | 500%min | 500%min | Same | | | After<br>Aging<br>Comply<br>with<br>ASTM<br>D6319 | Tensile<br>Strength | 14Mpa, min | 14Mpa, min | Same | | | Comply<br>with<br>ASTM<br>D6319 | Ultimate<br>Elongation | 400% min | 400% min | Same | | Freedom from Holes | | | Free from holes when<br>tested in accordance<br>with ASTM D5151<br>AQL=2.5 | Free from holes<br>when tested in<br>accordance with<br>ASTM D5151<br>AQL=2.5 | Same | | Powder Content | | | 0.14mg-0.21mg | <2mg/glove | Similar | {6}------------------------------------------------ | Table 5-4 Biocompatibility Testing Comparison | | | | | |-----------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------| | Item | APOLLO<br>ADVANTAGE<br>Nitrile Examination<br>Gloves<br>(Proposed Device) | Nitrile Examination<br>Glove (Predicate<br>Device, K213739) | Remark | | | Material | Nitrile | Nitrile | Same | | | Biocompatibility | Irritation<br>ISO 10993-10:<br>2010 | The test<br>article has no skin<br>irritation under the<br>conditions of the<br>study. | Not an irritant under<br>the conditions of the<br>study | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | Sensitization<br>ISO 10993-10:<br>2010 | The test article<br>showed no evidence<br>of causing delayed<br>dermal contact<br>sensitization under<br>the conditions of the<br>study. | Not a sensitizer<br>under the conditions<br>of the study | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | Systemic<br>toxicity<br>ISO 10993-<br>11:2017 | The test article<br>showed no evidence<br>of causing acute<br>system toxicity<br>under the conditions<br>of the study. | Device extracts do<br>not pose a systemic<br>toxicity concern<br>under the conditions<br>of the study. | Same | | | | | | | | | | | | | | | | | | | | | | | | | #### Table 5-4 Biocompatibility Testing Comparison #### 5.7 Non-clinical Performance Data Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity. ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves. ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. | Test Method | Purpose | Acceptance Criteria | Results | |-------------|--------------------------|----------------------------------------|-------------------------------------------------------------| | ASTM D6319 | Physical Dimensions Test | Length(mm):<br>S:≥220;<br>M/L/XL:≥230; | Length(mm):<br>S:221-238mm;<br>M:232-251mm;<br>L:232-243mm; | Table 5-5 Summary of non-clinical performance testing {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Dow Corning Corporation. The word "Dow" is in a light gray color and is stacked on top of the word "Corning" which is in a dark blue color. Below the word "Corning" is the word "corporation" in a smaller font and a light blue color. | | | Width(mm): | | | XL:234-250mm; | |--------------|-------------------------------------------------|-----------------------------------------------|-------------------------------------------------------------------|----------------|------------------------------------------------------------------------------------------------------------------------------------| | | | S:80±10; | | | | | | | M:95±10; | | | Width(mm): | | | | L:110±10; | | | S:80-89mm; | | | | XL:120±10 | | | M:94-104mm; | | | | | | | L:102-111mm;; | | | | | | | XL:111-122mm; | | | | Thickness(mm): | | Thickness(mm): | | | | | Finger:≥0.05 | | Finger: | | | | | Palm:≥0.05 | | | S:0.11-0.15mm | | | | | | M:0.11-0.15mm | | | | | | | L:0.11-0.14mm | | | | | | | XL:0.11-0.16mm | | | | | | | | | | | | | | Palm: | | | | | | | Finger:0.08mm | | | | | | | S:0.08mm | | | | | | | M:0.08mm | | | | | | | | L:0.08mm | | | | | | XL:0.08mm | | | ASTM D5151 | Watertightness Test for<br>Detection of Holes | | Meet the requirements of<br>ASTM D5151 AQL 2.5 | | 0 glove water leakage<br>found | | ASTM D6124 | Powder Content | Meet the requirements of<br>ASTM D6124 <2.0mg | | | 0.14-0.21mg | | ASTM D6319 | Physical Properties | Before<br>Aging | Tensile<br>Strength | ≥14Mpa | 34.1-39.0Mpa | | | | | Ultimate<br>Elongation | ≥500% | 520.808%-637.854% | | | | After<br>Aging | Tensile<br>Strength | ≥14Mpa | 32.0-37.0Mpa | | | | | Ultimate<br>Elongation | ≥500% | 463.000-562.586% | | ISO 10993-11 | Biocompatibility:<br>Acute Systemic<br>Toxicity | | Non-acute systemic toxicity<br>under the conditions of the study. | | The test article showed<br>no evidence of causing<br>acute system toxicity<br>under the conditions of<br>the study. | | ISO 10993-10 | Biocompatibility:<br>Skin Irritation | | Non-irritating under the<br>conditions of the study. | | The test article has no<br>skin irritation under the<br>conditions of the study. | | ISO 10993-10 | Biocompatibility:<br>Dermal Sensitization | | Non-sensitizing under the<br>conditions of the study. | | The test article showed<br>no evidence of causing<br>delayed dermal contact<br>sensitization under the<br>conditions of the study. | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Dell Corning Corporation. The word "Dell" is written in a stylized font in a light gray color. Below "Dell" is the word "Corning" in a bold, dark blue font. Underneath "Corning" is the word "corporation" in a smaller, regular font, also in dark blue. | | | | conditions of the study. | |--|--|--|--------------------------| |--|--|--|--------------------------| #### 5.8 Clinical Test Data Not Applicable #### 5.9 Conclusion The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicated device Nitrile Examination Glove by Hangzhou Runheng Medical Co., Ltd, cleared under 510(k) K213739.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...