K230631 · Argon Medical Devices · DYB · May 22, 2023 · Cardiovascular
Device Facts
Record ID
K230631
Device Name
Kodiak™ Dual Port Coaxial Introducer Kit
Applicant
Argon Medical Devices
Product Code
DYB · Cardiovascular
Decision Date
May 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Kodiak™ Dual Port Coaxial Introducer Kit is indicated to introduce therapeutic or diagnostic devices into the vasculature.
Device Story
Kodiak™ Dual Port Coaxial Introducer Kit facilitates percutaneous access for therapeutic or diagnostic devices. System comprises 16F/14F braided reinforced sheaths with fluoropolymer coating, radiopaque marker band, over-molded hub with dual hemostasis valves, side port with stopcock, removable Y-connector, 14F dilator, 16ga blunt flushing needle, and high-pressure Luer adapter. Used in clinical settings by physicians; inserted over a pre-placed guidewire under imaging guidance. Provides a conduit for device delivery into the vasculature. Single-use device.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including system leak, tensile strength, tip strength, radiopacity, kink testing, dimensional analysis, functional fit, simulated use, Luer testing, power injection, tip buckling, particulate, and packaging integrity. Biocompatibility testing performed per ISO 10993 standards, including cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis, thromboresistance, PTT, complement activation, and platelet/leukocyte counts.
Technological Characteristics
Coaxial 16F/14F braided reinforced sheaths with fluoropolymer coating; radiopaque marker band; over-molded hub with dual hemostasis valves; side port with stopcock; removable Y-connector; 14F dilator; 16ga blunt flushing needle; high-pressure Luer adapter. Compatible with 0.038" guidewires. Single-use. Materials and design are similar to predicate. Sterilization method not specified.
Indications for Use
Indicated for patients requiring the introduction of therapeutic or diagnostic devices into the vasculature.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 22, 2023
Argon Medical Devices Pratusha Kudumula Senior Regulatory Affairs Specialist 1445 Flat Creek Road Athens, Texas 75751
Re: K230631
Trade/Device Name: Kodiak™ Dual Port Coaxial Introducer Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: March 2, 2023 Received: March 7, 2023
Dear Pratusha Kudumula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Misti E. Malone -S
Misti Malone, PhD Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K230631
Device Name Kodiak™ Dual Port Coaxial Introducer Kit
Indications for Use (Describe)
The Kodiak™ Dual Port Coaxial Introducer Kit is indicated to introduce therapeutic or diagnostic devices into the vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Date prepared | May 5th, 2023 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Argon Medical Devices, Inc.<br>1445 Flat Creek Road<br>Athens, Texas 75751 USA<br>Facility Registration number:1625425 |
| Contact person | Pratusha Kudumula<br>Sr. Regulatory Affairs Specialist<br>309-706-9488 |
| Device Name | Kodiak™ Dual Port Coaxial Introducer Kit |
| Trade Name | Kodiak™ Dual Port Coaxial Introducer Kit |
| Common name | Catheter Introducer |
| Regulation name | Catheter Introducer |
| Classification number | 21 CFR 870. 1340 |
| Primary product code | DYB |
| Regulatory class | II |
| Predicate device | Performer Introducer (K171999) Cook, Inc. |
| Description of the<br>Device: | The Kodiak™ Dual Port Coaxial Introducer Kit is intended to introduce<br>therapeutic or diagnostic devices into the vasculature. The Kit Model<br>381416000 contains a coaxial 16F and 14F braided reinforced sheaths with<br>a fluoropolymer coating, an embedded radiopaque marker band, and an<br>over molded hub assembly containing two hemostasis valves and side<br>port with stopcock.<br>It also contains a removable Y-connector with two 7F (ID) ports with<br>hemostasis valves, a 14F dilator, a 16ga blunt flushing needle, and a high-<br>pressure Luer adapter. After percutaneous access is established and a<br>working guidewire is placed, the introducer system is inserted over the<br>guidewire and advanced under imaging. Therapeutic or diagnostic devices<br>are inserted through the system. The Kodiak™ Dual Port Coaxial Introducer<br>Kit is a single use device. |
## 510(k) Summary
The Kodiak™ Dual Port Coaxial Introducer Kit is intended to introduce Indication for Use: therapeutic or diagnostic devices into the vasculature.
{4}------------------------------------------------
### Technological Characteristics:
A comparison of the technological characteristics of the subject device and the predicate devices shows the Kodiak™ Dual Port Coaxial Introducer Kit to be substantially equivalent to the current marketed predicate devices.
Equivalence is based upon the product performance, design and intended use. The Kodiak™ Dual Port Coaxial Introducer Kit and the predicate devices have similar materials of construction, dimensional specifications, design, and sterilization process.
| SUBJECT DEVICE | PRIMARY PREDICATE<br>DEVICE |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Kodiak™ Dual Port Coaxial<br>Introducer Kit | Cook Performer Introducer |
| Principle of Operation | |
| After percutaneous access is<br>established and a working<br>guidewire is placed, the introducer<br>system is inserted over the<br>guidewire and advanced under<br>imaging. Therapeutic or diagnostic<br>devices are inserted through the<br>system. | After access is established and a<br>working guidewire is placed, the<br>introducer system is inserted over<br>the guidewire and advanced<br>under imaging. Therapeutic or<br>diagnostic devices are inserted<br>through the system. |
| Device Description | |
| The system consists of the<br>following components:<br>• 16F x 45cm outer sheath<br>• 14F x 53cm inner sheath<br>• 14F x 61cm dilator<br>• 7F / 7F Y-connector<br>• 16ga blunt flushing needle<br>• High pressure Luer adapter | Performer® Introducers are<br>designed to perform as a<br>• Introducer sheath 6F-18F;<br>13cm – 85cm<br>• 6F - 18F dilators of varying<br>lengths |
| Guidewire Compatibility | |
| 0.038" | 0.035" or 0.038" depending on<br>introducer size |
Performance Tests (Non-Clinical):
No performance standards have been established under section 514, performance standards, of the Food, Drug and Cosmetic Act for these devices. A series of testing was conducted in accordance with protocols based on requirements outlined in guidance and industry standards and the
{5}------------------------------------------------
testing below was shown to meet the acceptance criteria that was determined to demonstrate substantial equivalence.
The following tests were performed under the specified testing parameters to support the Kodiak™ Dual Port Coaxial Introducer Kit substantial equivalence.
Performance Testing, including:
- System Leak ●
- . Tensile Strength
- Tip Strength ●
- Radiopacity ●
- Kink Testing .
- Dimensional ●
- . Functional Fit
- Simulated Use ●
- Luer Testing .
- Power Injection Testing ●
- Tip Buckling ●
- Particulate ●
- Packaging Integrity .
Biocompatibility Testing, including:
- . Cytotoxicity (ISO 10993-5)
- Sensitization (ISO10993-10) ●
- Irritation Intracutaneous Reactivity (ISO10993-23) ●
- Acute Systemic Toxicity (ISO10993-11) .
- . Pyrogenicity (ISO 10993-11)
- . Hemolysis Direct/Indirect (ISO 10993-4)
- . Thromboresistance (ISO 10993-4)
- Partial thromboplastin time (ISO 10993-4) ●
- Complement Activation (ISO 10993-4) ●
- Platelet and Leukocyte Count (ISO 10993-4) ●
| Substantial<br>Equivalence: | Based on the Indication for Use, design, and safety and performance<br>testing, the Kodiak™ Dual Port Coaxial Introducer Kit meets the<br>requirements for its intended use and is substantially equivalent to the<br>predicate device. |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The results of all testing demonstrate that the Kodiak™ Dual Port Coaxial<br>Introducer Kit is substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.