K230599 · Recensmedical, Inc. · GEH · Jun 22, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230599
Device Name
TargetCool™
Applicant
Recensmedical, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jun 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
TargetCool™ (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch). TargetCool™ (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures - Molluscum Contagiosum - Skin Tags - Actinic Keratosis - Lentigo - Verruca Plana - Verruca Vulgaris - Verruca Lesions - Genital Lesions - Seborrheic Keratosis
Device Story
Handheld cryosurgical device delivering controlled cooling or freezing to skin tissue. Inputs: CO2 gas cartridge and saline solution (for boosting mode). Operation: Thermoelectric cooling and cryogen spray. Cooling/Boosting modes reduce nerve conduction velocity to increase pain threshold; Freezing mode applies CO2 (-79°C) for tissue necrosis. Device features LCD display for real-time skin temperature, set temperature, and status; includes safety alarm/light if skin temperature drops below -1°C for >1 second. Used by clinicians in clinical settings. Output: Controlled thermal application to target area. Benefits: Pain relief, reduction of inflammation/hematoma, and precise surgical destruction of skin lesions.
Clinical Evidence
Clinical publications and data provided to demonstrate safety and effectiveness for pediatric patients. Bench testing confirmed compliance with ISO 13485, IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC). Biocompatibility of patient-contacting polycarbonate guard established via predicate equivalence.
Technological Characteristics
Handheld device; polycarbonate guard (patient-contacting); CO2 gas cartridge; thermoelectric cooling system; LCD interface; safety alarm (audible/visual) for < -1°C skin temperature. Sterilization not specified. Connectivity: None (standalone).
Indications for Use
Indicated for patients requiring temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains, or minor sports injuries, and as an adjunct to rehabilitative treatment. Also indicated for surgical destruction of target tissue (Molluscum Contagiosum, Skin Tags, Actinic Keratosis, Lentigo, Verruca Plana, Verruca Vulgaris, Verruca Lesions, Genital Lesions, Seborrheic Keratosis) via cryoablation.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2023
RecensMedical, Inc. Lee Yeonui Regulatory Affairs Manager #507, #908, SK V1 center, 830 Dongtansunhwan-daero Hwaseong-si, Gyeonggi-do 18468 Korea. South
Re: K230599
Trade/Device Name: TargetCool™ Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH, MLY Dated: Mav 26, 2023 Received: May 26, 2023
Dear Lee Yeonui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Mark Trumbore -S" in a large font on the left side of the image. On the right side of the image, it says "Digitally signed by Mark Trumbore -S" and "Date: 2023.06.22 11:57:21 -04'00'". The text on the right side of the image is smaller than the text on the left side of the image.
Mark Trumbore Assistant Director, THT4A1: Robotically-Assisted Surgical Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230599
Device Name TargetCool™
#### Indications for Use (Describe)
TargetCool™ (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).
TargetCool™ (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures
- Molluscum Contagiosum
- Skin Tags
- Actinic Keratosis
- Lentigo
- Verruca Plana
- Verruca Vulgaris
- Verruca Lesions
- Genital Lesions
- Seborrheic Keratosis
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## 1. ADMINISTRATIVE INFORMATION
| Manufacturer Name | RecensMedical, Inc. |
|---------------------------------|----------------------------------------------------------------------------------|
| | #507, #908, SK V1 center, 830 Dongtansunhwan<br>daero, Hwaseong-si, Gyeonggi-do, |
| | Republic of Korea |
| Phone: | +82 31 8043 3064 |
| Fax: | +82 31 630 2092 |
| Date of the summary preparation | March 01, 2023 |
| Contact person: | Yeonui Lee / Regulatory Affairs Manager<br>yui.lee@recensmedical.com |
### 2. DEVICE NAME AND CLASSIFICATION
| Trade name: | TargetCool™ |
|-----------------------------|-----------------------------------|
| Common name: | Cryosurgical Device |
| Classification name: | Cryosurgical Unit and Accessories |
| Classification Regulations: | 21 CFR 878.4350 / 21 CFR 878.4810 |
| Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
| Product code: | GEH, MLY |
### 3. PRIMARY PREDICATE DEVICE
| 510(k) Number: | K203481 |
|-----------------------------|-----------------------------------|
| Trade name: | CryoVIVE |
| Classification name: | Cryosurgical Unit and Accessories |
| Classification Regulations: | 21 CFR 878.4350 |
| Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
| Product code: | GEH, MLY |
# 4. SECOND PRIDICATE DEVICE
510(k) Number: Trade name: Classification name:
K221234 TargetCool™ Cryosurgical Unit and Accessories
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "TargetCool" with the word "Target" in a bright blue color and the word "Cool" in a light gray color. The letters are in a sans-serif font. There is a trademark symbol in superscript next to the word "Cool".
| Classification Regulations: | 21 CFR 878.4350 |
|-----------------------------|---------------------------|
| Class: | Class 11 |
| Classification Panel: | General & Plastic Surgery |
| Product code: | GEH, MLY |
#### 5. INDICATIONS FOR USE
TargetCool™ (Cooling mode, Boosting) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch).
TargetCool™ (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures
- Molluscum Contagiosum
- Skin Tags
- Actinic Keratosis
- Lentigo
- Verruca Plana
- Verruca Vulgaris
- Verruca Lesions
- Genital Lesions
- Seborrheic Keratosis
### 6. DEVICE DESCRIPTION
TargetCool™ is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool™ device consists of a main device, nozzle, a guard, a filter, and a cartridge.
The main device produces controlled cooling based on thermoelectric cooling, which controls the temperature of the targeted area. TargetCool™ displays the skin temperature measured in real time, the set cooling temperature and the device status through the LCD display. Also, if the measured temperature is below -1 ℃ and lasts for more than 1 second, the status light blinks in blue with a beep sound.
The principle of pain relief through skin cooling is as the skin cools by spraying the cryogen onto the skin, the nerve conduction velocity (NCV) of the skin decreases, increasing the pain threshold (PTH) and pain tolerance (PT). Through this mechanism, pain in the skin is relieved. The advances represented by TargetCool™ lie in the fact that the device executes this process in a precise, controlled, and rapid way.
{5}------------------------------------------------
### 7. PERFORMANCE DATA
The Company's Performance Data for TargetCool™ is as follows:
#### Bench Testing
TargetCool™ complies with all applicable standards, including ISO 13485:2016, IEC 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility.
Biocompatibility (ISO 10993) was also performed to demonstrate conformance with established industry standards. The only patient-contacting material on TargetCool™ is the guard, which is comprised of polycarbonate. The polycarbonate is the exact same material used in CryoVIVE, which was FDA-cleared under K203481. Therefore, the biocompatibility test is not applicable for the TargetCool™.
The device hazard analysis was completed and risk-control implemented to mitigate identified hazards. The testing results support that all the specifications have met the acceptance criteria of each module and interaction of processes.
TargetCool™ passed all testing and supports the claims of substantial equivalence and safe operation.
#### Clinical Testing
Clinical publications and clinical data demonstrated the safety and effectiveness of the TargetCoolTM for treatment of pediatric patient.
#### 8. SUBSTANTIAL EQUIVALENCE
The comparison chart below provides evidence to support the equivalence determination between TargetCool™ and the predicate device (K220674) with respect to intended use, technological characteristics and principles of operation. TargetCool™ shares the same indications for use, device operation, technical and functional capabilities, and therefore is substantially equivalent to the predicate device.
| Product Name | TargetCool TM<br>(Subject Device) | TargetCool TM<br>(K221234) | Comparison |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for Use / Intended Use | TargetCoolTM is indicated for the<br>temporary reduction of pain, swelling,<br>inflammation, and hematoma from<br>minor surgical procedures, minor<br>sprains or other minor sports injuries,<br>and as an adjunct to rehabilitative<br>treatment (e.g., intermittent cold with<br>stretch). | TargetCoolTM is indicated for the<br>temporary reduction of pain, swelling,<br>inflammation, and hematoma from<br>minor surgical procedures, minor<br>sprains or other minor sports injuries,<br>and as an adjunct to rehabilitative<br>treatment (e.g., intermittent cold with<br>stretch). | Same |
| Component | Main system, Control button, LCD,<br>Cooling nozzle, Freezing nozzle,<br>Cooling guard, filter, CO2 cartridge<br>and Boosting accessories (Boosting<br>nozzle, a Boosting guard, and a | Main system, Trigger, LCD Display,<br>Cooling-Nozzle, Guard, Filter, CO2<br>Cartridge | Difference |
#### [TargetCool™-Cooling mode, Boosting mode]
{6}------------------------------------------------
| | Boosting Container). | | | |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------|
| Mechanism of<br>Action | The unit blows very low-temperature<br>gas and cold-water at temperature and<br>time settings, onto the desired<br>treatment area | | The unit blows very low-temperature<br>gas at temperature and time settings,<br>onto the desired treatment area | Similar |
| Mode type | Cooling Mode, Boosting Mode | | Cooling mode | Difference |
| Cryogen Type | CO2, Cold-water (Saline solution) | | CO2 | Same |
| Temperature | 2-4 °C within 5 sec | | 2-4 °C within 5 sec | Same |
| Treatment<br>Duration | Cooling Mode | Continuous (0 sec)<br>PL (2 sec pre-set)<br>1~60sec | 0~60 sec | Similar |
| | Boosting Mode | 3~4 min | | |
| Gas Volume | 65g cartridge | | 65g cartridge | Same |
| Safety feature | Alarm and status light blinking if the<br>temperature of the skin is determined to<br>be less than -1 °C for 1 second. | | Alarm and status light blinking if the<br>temperature of the skin is determined to<br>be less than -1 °C for 1 second. | Same |
The Cooling mode & Boosting mode of TargetCool™ is equivalent in intended use, principles of operation, and performance temperature to TargetCool™ (K221234) and raises no new issues of safety or effectiveness.
The differences are the Boosting mode and the Boosting accessories for the Boosting mode.
In Boosting mode, saline cooled by CO2 gas is sprayed onto the skin. The saline solution is used for the purpose of keeping the skin temperature constant at 2-4℃ for 3-4 minutes and for safety purposes to prevent the skin temperature from dropping below freezing.
Cooling mode uses CO2 gas and Boosting mode uses cold-water, but the temperature of Cooling mode and Boosting mode is the same (2~4℃). Therefore, the difference does not raise new questions of safety and effectiveness
| Product Name | TargetCool™<br>(Subject Device) | CryoVIVE<br>(K203481) | Comparison |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Indications for<br>Use<br>/ Intended Use | TargetCool™ are intended for the<br>surgical destruction of target tissue by<br>applying cryogenic gases at extreme<br>low temperatures<br>- Molluscum Contagiosum<br>- Skin Tags<br>- Actinic Keratosis<br>- Lentigo<br>- Verruca Plana<br>- Verruca Vulgaris<br>- Verruca Lesions<br>- Genital Lesions<br>- Seborrheic Keratosis | The CryoVIVE are intended for the<br>surgical destruction of target tissue by<br>applying cryogenic gases at extreme<br>low temperatures<br>- Molluscum Contagiosum<br>- Skin Tags<br>- Actinic Keratosis<br>- Lentigo<br>- Verruca Plana<br>- Verruca Vulgaris<br>- Verruca Lesions<br>- Genital Lesions<br>- Seborrheic Keratosis | Same |
| | Component | Main system, Control button, LCD,<br>Cooling nozzle, Freezing nozzle,<br>Cooling guard, filter, CO2 cartridge<br>and Boosting accessories (Boosting<br>Nozzle, a Boosting guard, and a | Main system, Control button, LCD,<br>Nozzle (Cooling and Freezing), Guide<br>tip, filter, CO2 cartridge |
[TargetCoolTM-Freezing mode]
{7}------------------------------------------------
| | Boosting Container). | | |
|------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|------|
| Mechanism of<br>Action | Cryogen, CO2 is delivered to the<br>treatment site to effect cellular<br>destruction | Cryogen, CO2 is delivered to the<br>treatment site to effect cellular<br>destruction | Same |
| Temperature | CO2 (-79°C) | CO2 (-79°C) | Same |
| Gas dispensing<br>rate | 0.578 g/sec | 0.578 g/sec | Same |
| Gas Volume | 65g cartridge | 65g cartridge | Same |
| Tissue damage | Cell necrosis occurs only inside the ice<br>ball | Cell necrosis occurs only inside the ice<br>ball | Same |
Cryoablation is the fundamental technological principle for the freezing mode of the subject device, TargetCool™, and the predicate devices (CryoVIVE).
The Freezing mode of TargetCool™ is equivalent in intended use, principles of operation, and performance temperature to TargetCool™ (K221234) and raises no new issues of safety or effectiveness. The only difference is that the freezing nozzle's length is shortened and the guard is not inserted during the procedure in the freezing mode. Because of this the device can be placed closer to the lesion, so that it can be applied to smaller lesion sizes. However, these differences do not significantly affect safety and/or effectiveness.
#### 8. CONCLUSION
TargetCool™ and the legally marketed predicate devices have the same intended use, Indications for Use statement and the technological characteristics. While the technological characteristics differ between the two systems, the differences are minor. Performance testing data established that the TargetCool™ is safe and effective as the legally marked predicate devices and that the TargetCool™ does not raise any different questions of safety and effectiveness than the predicate.
On this basis and in accordance with 21 CFR\$ 807.100(b), TargetCool™ is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.