K230593 · Jiangsu Haize Medical Scientific Development Co., Ltd. · GEI · May 30, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230593
Device Name
Victan
Applicant
Jiangsu Haize Medical Scientific Development Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
May 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Victan is a sterile, single-patient-use electrosurgical accessory intended to conduct electrosurgical current for cutting and coagulation of tissue and/or to provide suction and irrigation functions to the surgical site.
Device Story
Victan is a sterile, single-patient-use electrosurgical accessory; functions as a conduit for high-frequency (HF) electrical current from an external generator to tissue for cutting and coagulation; provides integrated suction and irrigation capabilities via connection to a negative pressure drainage device. Device components include electrode, external tube, irrigation/suction tubes, control buttons, rotator, handle, and connectors. Operated by clinicians in surgical settings; requires HF generator with nominal frequency between 300 kHz and 3 MHz. Output is controlled by the surgeon to perform tissue manipulation and site maintenance. Benefits include combined electrosurgical and fluid management functionality in a single sterile device.
Clinical Evidence
No clinical data or animal studies were provided. Substantial equivalence is supported by bench testing, including appearance, dimension, leakage, pressure resistance, continuity, irrigation/suction flow, and tensile strength. Biocompatibility was evaluated per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity). Electrical safety and EMC testing were performed per IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2.
Technological Characteristics
Monopolar electrosurgical accessory; stainless steel electrode; pistol or trumpet-style handle; EO sterilized (3-year shelf life). Operates with HF generators (300 kHz–3 MHz). Complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2. No software or algorithm components.
Indications for Use
Indicated for use as a sterile, single-patient-use electrosurgical accessory for cutting and coagulation of tissue and/or providing suction and irrigation functions at the surgical site.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 30, 2023
Jiangsu Haize Medical Scientific Development Co., Ltd. Vincy Gao RA No.99, Furongzhongsan Road, Xishan Economical Development Zone Wuxi, Jiangsu 214000 China
Re: K230593
Trade/Device Name: Victan Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 9, 2023 Received: May 9, 2023
Dear Vincy Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Mark Trumbore -S. The date of the signature is 2023.05.30. The time of the signature is 10:01:30 -04'00'.
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230593
Device Name Victan
Indications for Use (Describe)
Victan is a sterile, single-patient-use electrosurgical accessory intended to conduct electrosurgical current for cutting and coagulation of tissue and/or to provide suction and irrigation functions to the surgical site.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) Summary
### Single Use Surgical Electrocoagulator
| Compiled by: | Vincy Gao |
|--------------|--------------|
| Date: | 2023.01.02 |
| Reviewed by: | Zhang Rong |
| Date: | 2023.01.16 |
| Approved by: | Qian Jianmin |
| Date: | 2023.02.15 |
### Jiangsu Haize Medical Scientific Development Co.,Ltd
{4}------------------------------------------------
### 510(k) Summary
#### 1. Submitter Information
| Applicant/Submitter: | Jiangsu Haize Medical Scientific Development Co.,Ltd. |
|----------------------|-------------------------------------------------------|
| | No.99, Furongzhongsan Road, Xishan Economical |
| | Development Zone, Wuxi, Jiangsu, China |
| | Phone: 086-0510-82833301 |
| | Fax: 086-0510-82833301 |
| Contact Person: | Vincy Gao |
| | chenll@haizemed.net |
#### 2. Device
| Trade Name | Victan |
|---------------------|-------------------------------------------------------------|
| Common Name | Single Use Surgical Electrocoagulator |
| Classification Name | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number | 21 CFR 878.4400 |
| Device Class | Class II |
| Product Code | GEI |
| 510(K) Number | K230593 |
#### 3. Predicate Device
| Manufacturer | Product Name | 510(K) Number |
|--------------------------|--------------|---------------|
| Sejong Medical Co., Ltd. | LAP-iX | K173112 |
#### 4. Device Description
The Single Use Surgical Electrocoagulator is a sterile, single-patient-use electrosurgical accessory, and deliver electrical current from high-frequency generator to the surgical site. At the same time ,
{5}------------------------------------------------
the product can also provides suction or irrigation by connecting to a negative pressure drainage device. The proposed product consists of Electrode, External tube, Irrigation and suction tube, External tube control button, Rotator, Irrigation switch, Suction switch, Handle, Suction connecting tube, Suction connector, Irrigation connector, Cable, Irrigation connecting tube, Outer tube.
Single Use Surgical Electrocoagulator should be operated with HF generators having a nominal frequency of the HF voltage between 300 kHz and 3 MHz. Lower frequencies may cause nerve stimulation and thus uncontrolled convulsion. In case of higher frequencies, user/patient safety cannot be guaranteed because insulation may heat up so that dielectric strength is no longer given. All equipment conforming to this specs can be used.
The product was sterilized by EO. The valid period of sterilization was 3 years.
#### 5. Indications for Use
Single Use Surgical Electrocoagulator is a sterile, single-patient-use electrosurgical accessory intended to conduct electrosurgical current for cutting and coagulation of tissue and/or to provide suction and irrigation functions to the surgical site.
| Item | Proposed Device | Predicate Device | Equivalence |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Trade Name | Victan | LAP-iX | / |
| Manufacturer | Jiangsu Haize Medical<br>Scientific Development<br>Co.,Ltd. | Sejong Medical Co., Ltd. | / |
| Classification Name | Electrosurgical, cutting &<br>coagulation device &<br>accessories | Electrosurgical, cutting &<br>coagulation device &<br>accessories | Same |
| 510(K) Number | K230593 | K173112 | / |
| Device Class | Class II | Class II | Same |
| Product Code | GEI | GEI | Same |
| Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 | Same |
| Indications for Use | Single Use Surgical<br>Electrocoagulator is a<br>sterile, single-patient-use<br>electrosurgical accessory<br>intended to conduct<br>electrosurgical current<br>for cutting and<br>coagulation of tissue<br>and/or to provide suction<br>and irrigation functions | Single Use Surgical<br>Electrocoagulator is a<br>sterile, single-patient-use<br>electrosurgical accessory<br>intended to conduct<br>electrosurgical current<br>for cutting and<br>coagulation of tissue<br>and/or to provide suction<br>and irrigation functions | Same |
#### 6. Comparison of Technological Characteristics with the Predicate Device
{6}------------------------------------------------
| | to the surgical site. | to the surgical site. | |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Energy Type | Radiofrequency | Radiofrequency | Same |
| Electrode Type | Monopolar | Monopolar | Same |
| End-tip Type | Angle, Spade, Hook, Needle | L-hook,Spoon | Different<br>(After the assessment,<br>End-tip type does not<br>affect electrical safety or<br>performance.) |
| Handle Type | Pistol Type, Trumpet Type | Pistol Type, Trumpet Type | Same |
| Length | Function Tube<br>Length:190/335/435mm | Tip Length:37/48cm | Different<br>(After the assessment,<br>End-tip type does not<br>affect electrical safety or<br>performance.) |
| Materials(Electrode) | Stainless steel | Stainless steel | Same |
| Sterilization | EO | EO | Same |
| Performance Test | Appearance, Dimension,<br>Leakage, Pressure<br>Resistance, Continuity,<br>Irrigation and Suction<br>Flow, Tensile Strength,<br>Electrical safety and<br>electromagnetic<br>compatibility (EMC) | Nominal size, Leakage,<br>Continuity,<br>Dielectric Strength Test<br>of Electrode cable, HF<br>Leakage current Test of<br>Electrode<br>cable, and Tensile<br>Strength Test of<br>Electrode cable | Equivalent<br>(The actual test results<br>are the same.) |
#### 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 7.1. Biocompatibility
The device has been evaluated for its biological safety according to ISO10993-1 Biological Evaluation of Medical Devices - Part 1:Evaluation and Testing Within a Risk Management Process. Following endpoints have assessed during the evaluation:
- * Cytotoxicity
- * Sensitization
- * Intracutaneous Reactivity
- * Acute Systemic Toxicity
- * Material-Mediated Pyrogenicity
{7}------------------------------------------------
#### 7.2. Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Single Use Surgical Electrocoagulator. The subject device complies with relevant clauses of the IEC 60601-1, IEC-60601-2-2 standards for safety and the IEC 60601-1-2 for EMC.
#### 7.3. Sterilization
The subject device and the predicate device are sterilized via Ethylene Oxide; both devices are sterilized to the same sterility assurance level.
#### 7.4. Shelf-life
Accelerated Aging testing confirmed that the expiration date of subject device is 3 year from the manufacturing date.
#### 7.5. Bench Testing
Haize Medical conducted bench testing to ensure that the design of Single Use Electrocoagulator applicable the specifications and that the product is safety and effective in use.
Test included:
- Appearance
- Dimension
- Leakage ●
- Pressure Resistance
- Continuity
- Irrigation and Suction Flow
- Tensile Strength
#### 7.6. Animal Study
This premarket notification does not rely on preclinical animal testing to demonstrate substantial equivalence.
#### 7.7. Clinical Study
This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence.
#### 8. Substantial Equivalence
The subject and predicate device have the same fundamental technology and indications for use. The performance, dimensions and materials of the subject device are similar to the predicate device. Although there some minor differences with each product, these differences between the subject device and the predicate device do not raise new or different questions of safety and efficacy. There is no new technology and no difference that would raise new or different questions of safety or efficacy.
{8}------------------------------------------------
#### 9. Conclusion
The subject device are identical in intend use, indications and technology compared to the predicate device. The results of the test demonstrate that the subject devices do not raise new questions of safety and effectiveness and are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.