Syngo Carbon Space is a software intended to display medical data and to support the review and analysis of medical images by trained medical professionals.
Device Story
Syngo Carbon Space is a software-only medical image management and processing system installed on standard IT hardware. It operates via a client-server architecture, where the server processes and renders data from connected modalities, and the client provides the user interface for interactive viewing and analysis. It supports DICOM and non-DICOM objects, including images, videos, and clinical documents. Used by trained medical professionals in clinical environments (e.g., Radiology, Cardiology), it enables image review, post-processing, and result distribution. The device provides tools for measurement, annotation, and segmentation (e.g., distance, angle, ROI, organ segmentation). Output is displayed on certified medical monitors to support clinical decision-making. It does not perform automated diagnostic interpretation (CAD). The system integrates with RIS and third-party reporting systems via FHIR/HL7. It benefits patients by facilitating efficient access to medical images and clinical data across the healthcare enterprise.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical performance testing, including verification and validation of software specifications and hazard analysis. Previous phantom and reader study results for the integrated Lesion Quantification Algorithm were evaluated and deemed applicable with minor design mitigations.
Technological Characteristics
Software-only medical device; client-server architecture. Supports DICOM (2016a) and non-DICOM (HL7, FHIR R4) communication. Runs on standard IT hardware (Windows/Linux). Features include MPR, MIP, MinIP, VRT, and various measurement/segmentation tools. Cybersecurity controls include user access control, audit trails, network/firewall control, and system hardening. Software level of concern: Moderate.
Indications for Use
Indicated for display, rendering, and post-processing of medical data (mostly medical images) by trained medical professionals in healthcare institutions (e.g., Radiology, Nuclear Medicine, Cardiology). Diagnostic Workspace is for display, rendering, and post-processing; Physician Access is for display and rendering. Not for diagnosis of digital mammography images, not for use as a sole basis for clinical decisions, and Physician Access is not for diagnostic use on mobile devices in the USA.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Siemens Healthcare GmBH % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
March 21, 2023
Re: K230561
Trade/Device Name: Syngo Carbon Space VA30A Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: II Product Code: LLZ Dated: February 28, 2023 Received: February 28, 2023
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K230561
Device Name Syngo Carbon Space (VA30A)
#### Indications for Use (Describe)
Syngo Carbon Space is a software intended to display medical data and to support the review and analysis of medical images by trained medical professionals.
Syngo Carbon Space "Diagnostic Workspace" is indicated for display, rendering, post-processing of medical data (mostly medical images) within healthcare institutions, for example, in the field of Radiology, Nuclear Medicine and Cardiology.
Syngo Carbon Space "Physician Access" is indicated for display and rendering of medical data within healthcare institutions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots. Below the logo is the date "December 09, 2022".
# Traditional 510(k) Summary
# K230561
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### 1. Submitter:
Siemens Healthcare GmbH Henkestrasse 127 91052 Erlangen Germany
#### 2. Establishment Registration Number: 3004977335
#### 3. Contact Person:
Mr. Vijay Ramadas Regulatory Affairs Manager Siemens Healthcare GmbH, Siemensstr. 3 91301 Forchheim Germany
| E-mail: | vijay.ramadas@siemens-healthineers.com |
|------------|----------------------------------------|
| Telephone: | +49 (172) 4324369 |
| Fax: | +49 (9191) 18-4404 |
#### 4. Device Name and Classification
| Device/Trade Name: | Syngo Carbon Space (VA30A) |
|------------------------|------------------------------------------------|
| Classification Panel: | Radiology Devices |
| Classification Number: | 21 CFR 892.2050 |
| Classification Name: | Medical Image Management and Processing System |
| Device Class: | Class II |
| Product Code: | LLZ |
#### 5. Predicate Device(s):
#### Main Predicate Device:
| Device/Trade Name: | Syngo Carbon Space (VA20A) |
|------------------------|------------------------------------------------|
| K Number | K213665 |
| Classification Panel: | Radiology Devices |
| Classification Number: | 21 CFR 892.2050 |
| Classification Name: | Medical Image Management and Processing System |
| Device Class: | Class II |
| Product Code: | LLZ |
{4}------------------------------------------------
### Reference Predicate Device 1:
| Device/Trade Name: | syngo.via (Version VB40A) |
|------------------------|---------------------------------------------|
| K Number | K191040 |
| Classification Panel: | Radiology Devices |
| Classification Number: | 21 CFR 892.2050 |
| Classification Name: | Picture archiving and communications system |
| Device Class: | Class II |
| Product Code: | LLZ |
### Reference Predicate Device 2:
| Device/Trade Name: | syngo.plaza |
|------------------------|---------------------------------------------|
| K Number | K180563 |
| Classification Panel: | Radiology Devices |
| Classification Number: | 21 CFR 892.2050 |
| Classification Name: | Picture archiving and communications system |
| Device Class: | Class II |
| Product Code: | LLZ |
### 6. Device Description:
Syngo Carbon Space is a software only medical device which is intended to be installed on recommended common IT Hardware. The hardware is not seen as part of the medical device. Syngo Carbon Space is intended to support reviews and analysis of medical images by trained medical practitioners. The software is used in Radiology for reading images and throughout the healthcare institutions for image & result distribution.
Syngo Carbon Space is a medical device, provided in two variants/options.
- Diagnostic Workspace (Fat/Thick Client) -
- -Physician Access (Thin/Web Client)
In any scenario, both the options can be installed/run on the same machine and be used simultaneously.
Syngo Carbon Space Diagnostic Workspace provides a reading workspace for Radiology that supports display of medical image data & documents and connects intelligent work tools (diagnostic and non-diagnostic software elements) to enable easy access to the data needed, easy access to external tools and creation of actionable results.
Syngo Carbon Space Physician Access provides a zero-footprint web application for enterprise-wide viewing of DICOM, non-DICOM, multimedia data and clinical documents to facilitate image and result distribution in the healthcare institution.
Since Syngo Carbon Space is a software only product, shelf-life is not applicable because of low likelihood of time-dependent product degradation. Hence performance data is not needed to establish maintenance of device performance characteristics over the shelf-life period
{5}------------------------------------------------
Syngo Carbon Space is a software only medical device, which is delivered by download only option to be installed on common IT hardware. This hardware must fulfil the defined requirements. Any hardware platform that complies to the specified minimum hardware and software requirements and with successful installation verification and validation activities can be supported. The hardware itself is not seen as part of the medical device and therefore not in the scope of the 510(k) submission.
Syngo Carbon Space provides tools and features to cover the radiological tasks reading images and support reporting through third party tools. It supports DICOM and Non-DICOM objects. In a comprehensive imaging suite, Syngo Carbon Space interoperates with a Radiology Information System (RIS) to enable customer specific workflows.
Syngo Carbon Space is based on a client-server architecture. The server processes and renders the data from the connected modalities. The server provides central services including image processing and temporary storage. The client provides the user interface for interactive image viewing and processing and can be installed and started on each workplace that has a network connection to the server.
Syngo Carbon Space supports various monitor setups and can be adapted to a range of image types by connecting different monitor types.
The subject device and the predicate device share fundamental scientific technology. This device description holds true for the subject device and the predicate device.
### 7. Intended Use:
Syngo Carbon Space is a software intended to display medical data and to support the review and analysis of medical images by trained medical professionals.
#### 8. Indications for Use:
Syngo Carbon Space is a software intended to display medical data and to support the review and analysis of medical images by trained medical professionals.
Syngo Carbon Space "Diagnostic Workspace" is indicated for display, rendering, postprocessing of medical data (mostly medical images) within healthcare institutions, for example, in the field of Radiology, Nuclear Medicine and Cardiology.
Syngo Carbon Space "Physician Access" is indicated for display and rendering of medical data within healthcare institutions.
{6}------------------------------------------------
# 9. Summary of Differences between the Subject Device and the Predicate Device:
The differences between the subject device described in this premarket notification (Highlighted) and the following comparison table:
| Specification | Subject Device | Predicate Device | Comparison | Impact to Safety &<br>Effectiveness |
|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name<br>and version | Syngo Carbon Space VA30A | Syngo Carbon Space VA20A<br>(K213665) | New version of the<br>product | NA |
| Manufacturer | Siemens Healthcare GmbH | Siemens Healthcare GmbH | Same | NA |
| Indications for<br>use | Syngo Carbon Space is a software<br>intended to display medical data and to<br>support the review and analysis of<br>medical images by trained medical<br>professionals.<br><br>Syngo Carbon Space "Diagnostic<br>Workspace" is indicated for display,<br>rendering, post-processing of medical<br>data (mostly medical images) within<br>healthcare institutions, for example, in<br>the field of Radiology, Nuclear<br>Medicine and Cardiology.<br><br>Syngo Carbon Space "Physician<br>Access" is indicated for display and<br>rendering of medical data within<br>healthcare institutions. | Syngo Carbon Space is a software<br>intended to display medical data and to<br>support the review and analysis of<br>medical images by trained medical<br>professionals.<br><br>Syngo Carbon Space "Diagnostic<br>Workspace" is indicated for display,<br>rendering, post-processing of medical<br>data (mostly medical images) within<br>healthcare institutions, for example, in<br>the field of Radiology, Nuclear<br>Medicine and Cardiology.<br><br>Syngo Carbon Space "Physician<br>Access" is indicated for display and<br>rendering of medical data within<br>healthcare institutions. | Same | NA |
| Specification | Subject Device | Predicate Device | Comparison | Impact to Safety &<br>Effectiveness |
| Contraindicatio<br>ns | Syngo Carbon Space "Diagnostic<br>Workspace" is not intended for<br>diagnosis of digital mammography<br>images.<br>Syngo Carbon Space "Diagnostic<br>Workspace" is not intended to be used<br>as a sole basis for clinical decisions.<br><br>Syngo Carbon Space "Physician<br>Access" is not intended for diagnosis<br>of digital mammography images.<br>Syngo Carbon Space "Physician<br>Access" is not intended to be used for<br>diagnostic purpose on mobile devices<br>in the United States of America<br>(USA).<br>Syngo Carbon Space "Physician<br>Access" is not intended to be used as<br>a sole basis for clinical decisions. | Syngo Carbon Space "Diagnostic<br>Workspace" is not intended for<br>diagnosis of digital mammography<br>images.<br>Syngo Carbon Space "Diagnostic<br>Workspace" is not intended to be used<br>as a sole basis for clinical decisions.<br><br>Syngo Carbon Space "Physician<br>Access" is not intended for diagnosis<br>of digital mammography images.<br>Syngo Carbon Space "Physician<br>Access" is not intended to be used for<br>diagnostic purpose on mobile devices<br>in the United States of America<br>(USA).<br>Syngo Carbon Space "Physician<br>Access" is not intended to be used as a<br>sole basis for clinical decisions. | Same | NA |
| Software<br>architecture | Syngo Carbon Space is based on a<br>client-server architecture | Syngo Carbon Space is based on a<br>client-server architecture | Same | NA |
| Image<br>communication | Standard network protocols like<br>TCP/IP and standard communication<br>protocol including DICOM (2016a)<br>and non-DICOM objects. Supports<br>interfacing with HL7 (v2.5 / v2.3.1 /<br>v2.3 / FHIR R4) | Standard network protocols like<br>TCP/IP and standard communication<br>protocol including DICOM (2016a)<br>and non-DICOM objects. Supports<br>interfacing with HL7 (v2.5 / v2.3.1 /<br>v2.3 / FHIR R4) | Same | NA |
| Specification | Subject Device | Predicate Device | Comparison | Impact to Safety &<br>Effectiveness |
| Image display<br>algorithms | MPR: MPR, MPR Thick, MPR/MPR* MIP: MIP, MIP Thin MinIP View VRT*: Plain VRT, Adapt VRT, VRT Thin, Cinematic VRT Fused View * Invert Image * available in in Diagnostic<br>Workspace only | MPR: MPR, MPR Thick, MPR/MPR* MIP: MIP, MIP Thin MinIP View VRT*: Plain VRT, Adapt VRT, VRT Thin, Cinematic VRT Fused View * Invert Image * available in in Diagnostic<br>Workspace only | Same | NA |
| Specification | Subject Device | Predicate Device | Comparison | Impact to Safety &<br>Effectiveness |
| Measurement,<br>Evaluation/Inte<br>rpretation<br>Tools | • Distance (Distance line, Distance<br>Polyline)<br>• Angle and Angle on stack*<br>• 2D ROI (Circle, Freehand,<br>Polygonal, Auto Contour) *<br>• 3D VOI (Sphere, Freehand) *<br>• Pixel Lens<br>• Ranges (Parallel, Radial, Radial<br>Sliced, Curved, Spine) *<br>• Lesion Quantification*<br>• Assisted Perpendicular Tool*<br>• Automatic Organ Segmentation*<br>• Interactive Tissue Segmentation*<br>• Time Curve, Time ROI*<br>• SUV Measurement*<br>• Automatic Anatomy Labeling<br>(rib, spine) *<br>• Next Study/Previous<br>Study/Nearline study*<br>• Change Geometry*<br>• Snapshot<br>• CT Lung Change*<br>• MR General Reading*<br>• Alpha Blending*<br>* Available in in Diagnostic<br>Workspace only | • Distance (Distance line, Distance<br>Polyline)<br>• Angle<br>• 2D ROI (Circle, Freehand,<br>Polygonal, Auto Contour) *<br>• 3D VOI (Sphere, Freehand) *<br>• Pixel Lens<br>• Ranges (Parallel, Radial, Radial<br>Sliced, Curved, Spine) *<br>• Lesion Quantification*<br>• Assisted Perpendicular Tool*<br>• Automatic Organ Segmentation*<br>• Interactive Tissue Segmentation*<br>• Time Curve, Time ROI*<br>• SUV Measurement*<br>• Automatic Anatomy Labeling (rib,<br>spine) *<br>• Next Study/ Previous Study*<br>* available in in Diagnostic<br>Workspace only | The tool set in the<br>subject device is<br>enhanced.<br>New basic reading tools<br>were added to the<br>subject device. | This differences<br>between the<br>predicate device and<br>the subject device<br>does not impact the<br>safety and<br>effectiveness of the<br>subject device as the<br>necessary measures<br>taken |
| Supported<br>objects<br>for display | DICOM image object display: | DICOM image object display: | Same | NA |
| | • CR Image | • CR Image | | |
| | • CT Image | • CT Image | | |
| | • DX Image | • DX Image | | |
| | • ES Image | • ES Image | | |
| | • GM Image | • GM Image | | |
| | • MG Image | • MG Image | | |
| | • MR Image | • MR Image | | |
| | • NM Image | • NM Image | | |
| | • PET Image | • PET Image | | |
| | • OP/OPT Image | • OP/OPT Image | | |
| | • RF Image | • RF Image | | |
| | • RT IMAGE | • RT IMAGE | | |
| | • SM (WSI) | • SM (WSI) | | |
| | • XA Image | • XA Image | | |
| | • US Image | • US Image | | |
| | • Secondary capture objects | • Secondary capture objects | | |
| | DICOM non-image object display: | DICOM non-image object display: | | |
| | • ECG | • ECG | | |
| | • Encapsulated PDF | • Encapsulated PDF | | |
| | • PR | • PR | | |
| | Non-DICOM file display: | Non-DICOM file display: | | |
| | • Images: BMP, GIF, JPEG<br>(JFIF), JPEG 2000, JPEG-LE,<br>JPEG-LS, PCX, PNG, PNM,<br>TIFF, WBMP | • Images: BMP, GIF, JPEG (JFIF),<br>JPEG 2000, JPEG-LE, JPEG-LS,<br>PCX, PNG, PNM, TIFF, WBMP | | |
| | • Video: FLV, H.264, H.265,<br>INDEO2, INDEO3, INDEO4,<br>MPEG1, MPEG2, MPEG4 | • Video: FLV, H.264, H.265,<br>INDEO2, INDEO3, INDEO4,<br>MPEG1, MPEG2, MPEG4, VP8,<br>VP9, WMV1, WMV2, WMV3 | | |
| Specification | Subject Device…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.