← Product Code [GXY](/productcode/GXY) · K230533

# Motive™ Electrode (K230533)

_Cymedica Orthopedics, Inc. · GXY · Mar 22, 2023 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230533

## Device Facts

- **Applicant:** Cymedica Orthopedics, Inc.
- **Product Code:** [GXY](/productcode/GXY.md)
- **Decision Date:** Mar 22, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1320
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Motive™ Electrode is intended for use as a disposable, conductive, and adhesive interface between the user's skin and an electrical muscle stimulator. Motive Electrode is intended for adults of 22 years of age and older.

## Device Story

Motive Electrode serves as disposable, conductive, adhesive interface between user skin and neuromuscular electrical stimulator; worn on thigh to cover quadriceps. Device consists of multi-layer sheet with three medical-grade hydrogel pads (two rectangular, one circular) on PET backing. Silver ink traces provide electrical conductivity from magnetic connector housing to gel pads. Used by patients for multiple applications; non-sterile. Electrical impulses delivered through skin to target muscles. Benefits include facilitating electrical stimulation therapy for muscle stimulation. No clinical data; bench testing confirms electrical impedance, current density, and power density performance.

## Clinical Evidence

No clinical data. Bench testing only. Verification included electrical/mechanical testing, impedance measurements, maximum current/power density analysis, and shelf-life testing. Biocompatibility testing performed per ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2021 (irritation/sensitization).

## Technological Characteristics

Multi-layer electrode sheet; PET substrate; conductive silver ink traces; medical-grade hydrogel pads. Dimensions: 15.6 x 21.8 cm total; 124.5 cm² total conductive area. Magnetic pogo pin connector. Non-sterile. 6-month shelf life. Biocompatibility per ISO 10993-5 and ISO 10993-10.

## Regulatory Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

## Predicate Devices

- ASCEND Electrode, NeuroMetrix, Inc. ([K140586](/device/K140586.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

March 22, 2023

CyMedica Orthopedics, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K230533

Trade/Device Name: Motive™ Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: February 27, 2023 Received: February 27, 2023

## Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Tushar Bansal -S

for Heather Dean, Ph.D Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230533

Device Name Motive(TM) Electrode

Indications for Use (Describe)

The Motive™ Electrode is intended for use as a disposable, conductive, and adhesive interface between the user's skin and an electrical muscle stimulator.

Motive Electrode is intended for adults of 22 years of age and older.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Motive™ Electrode CyMedica Orthopedics, Inc.

#### 1 Regulatory Information

| 1.1 Trade/Proprietary Name:     | Motive™ Electrode                                                                                 |  |
|---------------------------------|---------------------------------------------------------------------------------------------------|--|
| 1.2 Common Name:                | Electrode, Cutaneous                                                                              |  |
| 1.3 Regulation Names & Numbers: | Cutaneous Electrode, 21 CFR 882.1320                                                              |  |
|                                 | Product Codes:<br>Code- GXY; Electrode, Cutaneous                                                 |  |
| 1.4 Classification:             | Class II                                                                                          |  |
| 1.5 Manufacturer Name:          | CyMedica Orthopedics, Inc.                                                                        |  |
|                                 | 2120 East 6t Street, Suite 8<br>Tempe, AZ 85288<br>Telephone (480) 664-1282<br>FAX (866) 296-2772 |  |

#### 1.6 Predicate Device Information:

Trade name - ASCEND Electrode, NeuroMetrix, Inc. Common name - Electrode, Cutaneous Code - GXY; Electrode, Cutaneous Classification name - Cutaneous Electrode, 21 CFR 882.1320 510(k) Number – K140586

These devices are reviewed by the Neuromodulation and Physical Medicine Devices (DHT5B).

#### Submission Information 2

## Submission Number:

Date:

Contact:

February 15, 2023

Kereshmeh Shahriari 2120 East 6th Street, Suite 8 Tempe, AZ 85288 kereshmeh@cymedicaortho.com 

{4}------------------------------------------------

Telephone (480) 664-1282 FAX (866) 296-2772

#### 3 Intended Use/Indications for Use

The Motive™ Electrode is intended for use as a disposable, conductive, and adhesive interface between the user's skin and an electrical muscle stimulator. Motive Electrode is intended for adults of 22 years of age and older.

#### 4 Device Description

The Motive Electrode provides an electrically conductive interface between a neuromuscular electrical stimulator (muscle stimulator) and a user's skin. The Motive Electrode is provided nonsterile, is designed and intended for single patient use only, multiple application use, and is disposable. The electrode sheet is to be worn on the thigh covering the quadriceps muscles to apply electrical stimulation therapy when connected to a muscle stimulator. The electrodes are conductive pads that contact the skin. The electrical stimulation impulses are delivered through the electrodes on the skin in direct proximity to the muscles to be stimulated.

The Motive Electrode consists of multiple layers. The skin-contacting layer of the electrode sheet is comprised of three separate medical grade, self-adhering, biocompatible, and conductive hydrogel (gel) pads. These gel pads are placed on a single polyester or polyethylene terephthalate (PET) sheet. These gel pads include two rectangular 2 inch x 4 inch (5.1 x 10.2 cm) and one circular. 2.165 inch diameter (5.5 cm) pads.

The outermost layer of the electrode contains a magnetic connector housing for electrical connection to the muscle stimulator. Silver ink printed conductive electrical traces on the PET film are utilized to provide electrical conductivity from the muscle stimulator to the gel pads. When not in use, the electrode sheet is covered by a paper-backed release liner.

#### Summary of Non-Clinical/ Bench Studies 5

Verification of the Motive Electrode include electrical and mechanical tests to show that it meets its target specifications over a range of operating and storage conditions. Verification and performance testing further demonstrate that it meets user needs as reflected in the functional specification.

To demonstrate the safety, the Motive device was tested per the following standards:

- ISO 10993-5: 2009, Biological evaluation of medical devices- Part 5: Tests for in vitro . cytotoxicity
- ISO 10993-10: 2010, Biological evaluation of medical devices- Part 10: Tests for ● irritation and skin sensitization

#### 6 Biocompatibility Testing

The patient contacting materials in the Motive™ Electrode include a conductive hydrogel used in the form of gel pads. The hydrogel is placed on a Polyethylene Terephthalate (PET) backing liner. PET is a known biocompatible material. The hydrogel is medical grade, self-adhering, and biocompatible. The gel pads (hydrogel) are skin contacting. A biocompatibility test was conducted on the patient contacting materials per the following standards. The test results were acceptable, and the materials were considered biocompatible.

- ISO 10993-5: 2009, Biological evaluation of medical devices- Part 5: Tests for in vitro .

{5}------------------------------------------------

cytotoxicity

- ISO 10993-10: 2021, Biological evaluation of medical devices- Part 10: Tests for skin irritation and sensitization
#### 7 Shelf-Life

The Motive Electrode has a 6-month shelf life. Accelerated aging shelf-life testing was performed to ensure that the Motive Electrode performs as intended over the course of its labeled shelf-life.

#### 8 Summary of Non-Clinical/ Bench Studies

Verification of the Motive Electrode included electrical and mechanical tests. The electrical tests demonstrated the electrodes impedance measurements were acceptable and electrodes performed effectively and safely in delivering transmitting electrical stimulation to the user's tissue. The electrical tests also demonstrated that the Motive Electrode's Maximum Current Density and Maximum Average Power Density were acceptable and similar to the predicate device, ASCEND Electrode.

Mechanical testing demonstrated that the electrodes maintained the mechanical integrity for different lavers and components of the electrode sheet.

Visual evaluation demonstrated that the Motive Electrodes were intact and interfaced with the muscle stimulator.

Verification and performance testing further demonstrate that the Motive Electrodes meet user's needs as reflected in the product specification.

#### 9 Comparison of Technological Characteristics with the Predicate Device

The predicate device for Motive Electrode is ASCEND Electrode manufactured by NeuroMetrix, Inc., 510(k) NUMBER K140586. The Motive Electrode and ASCEND Electrode have the same intended use. Both are intended as disposable. single-patient for multiple application use electrodes with an adhesive conductive hydrogel interface to deliver treatment from an electrical stimulator to the User's skin. Both are multi-layer electrode sheet designs housing multiple individual electrodes in a pre-defined geometry. Both are substantially equivalent in composition and material, consisting of flex silver pads or traces printed onto a plastic film base sheet.

The primary difference between the two devices is that the Motive Electrode consists of three electrodes compared to the ASCEND Electrode's four electrodes. The shape and surface area of the Motive Electrodes are within range of the predicate ASCEND Electrodes. The individual predicate ASCEND Electrodes contain a smaller surface area (36 by 46 mm, surface area of 1656 mm²). In contrast, the smallest Motive Electrode contains a larger surface area (50.8 mm diameter circle, surface area of 2026 mm²).

Table 1 compares design details of the Motive Electrode to the predicate ASCEND Electrode:

{6}------------------------------------------------

## Table 1: Motive Electrode and predicate ASCEND Electrode comparison

|                                     | Subject Device<br>Motive Electrode<br>CyMedica Orthopedics,<br>Inc.<br>510(k), K230533                                                                                    | Predicate Device<br>ASCEND Electrode,<br>NeuroMetrix, Inc.<br>510(k), K140586                                                                                                   | Comparison                                                                                                  |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| General Characteristics             |                                                                                                                                                                           |                                                                                                                                                                                 |                                                                                                             |
| Intended Use                        | The Motive™ Electrode is<br>intended for use as a<br>disposable, conductive, and<br>adhesive interface between the<br>user's skin and an electrical<br>muscle stimulator. | The ASCEND Electrode is<br>intended for use as disposable,<br>conductive, adhesive interface<br>between the user's skin and a<br>transcutaneous electrical nerve<br>stimulator. | N/A                                                                                                         |
| Number of<br>Electrodes             | 3, two rectangular and one<br>circular                                                                                                                                    | Not publicly available                                                                                                                                                          | Any differences do not raise<br>any issue in equivalency of<br>both as it relates to safety or<br>efficacy. |
| Sterile                             | Non-sterile                                                                                                                                                               | Non-sterile                                                                                                                                                                     | N/A                                                                                                         |
| Single-Patient<br>Use               | Yes                                                                                                                                                                       | Yes                                                                                                                                                                             | N/A                                                                                                         |
| Number of<br>applications           | Multiple                                                                                                                                                                  | Not publicly available                                                                                                                                                          | N/A                                                                                                         |
| Shelf life                          | 6 months                                                                                                                                                                  | Not publicly available                                                                                                                                                          | N/A                                                                                                         |
| Technical Characteristics           |                                                                                                                                                                           |                                                                                                                                                                                 |                                                                                                             |
| Substrate                           | Polyethylene terephthalate<br>(PET) film                                                                                                                                  | Not publicly available                                                                                                                                                          | N/A                                                                                                         |
| Conductor                           | Conductive silver ink traces                                                                                                                                              | Not publicly available                                                                                                                                                          | N/A                                                                                                         |
| Conductive<br>interface             | Medical grade conductive<br>hydrogel                                                                                                                                      | Not publicly available                                                                                                                                                          | N/A                                                                                                         |
| Connector                           | Pogo pin connector array                                                                                                                                                  | Not publicly available                                                                                                                                                          | Any differences do not raise<br>any issue in equivalency of<br>both as it relates to safety or<br>efficacy. |
| Overall<br>Dimensions               | 15.6 x 21.8 cm, total area 211<br>cm², includes three electrodes                                                                                                          | Not publicly available                                                                                                                                                          | Any differences do not raise<br>any issue in equivalency of<br>both as it relates to safety or<br>efficacy. |
| Conductive<br>Surface<br>Dimensions | Two rectangular pads 5.1 x<br>10.2 cm and one circular pad<br>5.1 cm diameter, total area<br>124.5 cm²                                                                    | Not publicly available                                                                                                                                                          | Any differences do not raise<br>any issue in equivalency of<br>both as it relates to safety or<br>efficacy. |

#### 10 Conclusion

The summary includes the conclusions drawn from the nonclinical tests (discussed above) that demonstrate that the Motive Electrode is as safe, as effective, and performs as well as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K230533](https://fda-staging.innolitics.com/device/K230533)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).