The Zimed Distal Medial Tibial Plate and Screw System is indicated for fixation of complex extra-articular and intraarticular distal medial tibial fractures and osteotomies.
Device Story
Implantable bone fixation system; consists of metallic plates and screws; used for anatomical alignment and splinting of complex extra-articular and intra-articular distal medial tibial fractures and osteotomies; facilitates early patient mobility; intended for long-term bone contact; supplied non-sterile; utilized by orthopedic surgeons in clinical/surgical settings; provides mechanical stabilization to support bone healing.
Clinical Evidence
Bench testing only. Performance testing conducted in accordance with ASTM F382, ASTM F136, and ASTM F543 standards. Compliance with FDA guidance for orthopedic non-spinal metallic bone screws and fracture fixation plates demonstrated through mechanical performance testing.
Technological Characteristics
Metallic bone fixation plates and screws. Materials conform to ASTM F136. Design features include anatomical alignment and locking capability. Mechanical performance tested per ASTM F382 (plates) and ASTM F543 (screws).
Indications for Use
Indicated for patients requiring fixation of complex extra-articular and intra-articular distal medial tibial fractures and osteotomies.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282)
Submission Summary (Full Text)
{0}------------------------------------------------
January 4, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.
Zimed Medikal % Barry Sands President and Founder RQMIS, Inc. 110 Haverhill Road Suite 524 Amesbury, Massachusetts 01913
Re: K230507
Trade/Device Name: Zimed Distal Medial Tibial Plate and Screw System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: December 5, 2023 Received: December 6, 2023
Dear Barry Sands:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
### K230507 - Barry Sands
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
## Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230507
Device Name
Zimed Distal Medial Tibial Plate and Screw System
Indications for Use (Describe)
The Zimed Distal Medial Tibial Plate and Screw System is indicated for fixation of complex extra-articular and intraarticular distal medial tibial fractures and osteotomies.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K230507
| 510(K) SUMMARY | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter information | |
| Date Prepared | January 2, 2024 |
| Applicant/Manufacturer | Zimed Medikal Sanayi ve Ticaret Limited Sirketi<br>Aydinlar Mah.03070 Cad.No:04 Sehitkamil Gaziantep<br>Turkiye<br>Gaziantep, Sehitkamil, 27580, Turkiye |
| Primary Contact | Barry Sands<br>President and Founder, RQMIS<br>Phone: (978) 358-7307<br>Email: regulatorysubmissions@rqmis.com |
## Device information
| Trade Name of Subject Device | Zimed Distal Medial Tibial Plate and Screw System |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Plate, Fixation, Bone; Screw, Fixation, Bone |
| Device class | Class II |
| Product Code | HWC, HRS (primary) |
| Regulation Number | 888.3030 Single/multiple component metallic bone<br>Fixation appliances and accessories (Primary)<br>888.3040 Smooth or threaded metallic bone fixation<br>fastener |
| Predicate Devices: | Primary predicate: Synthes (USA) LCP Distal Tibia<br>Plate (K013248)<br>Reference predicate: HEMC BRAND Locking Bone Plates<br>and Screws Osteosynthesis Plating System, HEMC BRAND<br>of DHS Plating system (K222282) |
{4}------------------------------------------------
## Device Description
Zimed Distal Medial Tibial Plate and Screw System is used for complex extra-articular and intraarticular distal tibial fractures, and osteotomies surgeries specifically designed for the anatomical alignment of the broken bones and to provide splinting with the ability to ensure the early mobility of the patient. Zimed plate and screw system is implantable device in long term contact with bones.
Zimed Distal Medial Tibial Plate and Screw System will be supplied non-sterile.
### Indications for Use
The Zimed Distal Medial Tibial Plate and Screw System is indicated for fixation of complex extraarticular and intra-articular distal medial tibial fractures and osteotomies.
### Intended Use – Comparison of the subject device with Primary & Reference Predicate Devices
The subject device, Zimed Distal Medial Tibial Plate and Screw System, has similar intended use and indication for use to the primary predicate Synthes (USA) LCP Distal Tibia Plate (K013248), and reference predicate, HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282)
## Technological Characteristics - Comparison with the primary predicate, Synthes (USA) LCP Distal Tibia Plate (K013248) and reference predicate, HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282)
The technological characteristics comparison demonstrated that the subject device, the Plate and Screw System are substantially equivalent in intended uses, designs, materials, and operational principles to the previously cleared predicate devices, respectively.
#### Basis of Substantial Equivalence
The substantial equivalence of the subject device was determined per the FDA guidance document, "The 510(k) program: evaluating substantial equivalence in premarket Notification [510(k)]." The analysis of the technological characteristics, which include design, material composition and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2) (ii)(A), demonstrated that the Plate and Screw System are substantially equivalent to the identified predicate devices, respectively.
Zimed Distal Medial Tibial Plate and Screw System are substantially equivalent to the respective predicates. Comparison to the predicates does not raise different questions of safety and effectiveness. The performance testing, device comparisons, and feature analysis
{5}------------------------------------------------
demonstrate that the subject device is substantially equivalent to the identified predicate devices, respectively.
#### Performance Testing
The performance tests on the subject device were conducted in accordance with ASTM F382, ASTM F136 and ASTM F543 standards. Performance test reports confirmed that the subject device, Zimed Distal Medial Tibial Plate and Screw System, and the reference predicate, HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, HEMC BRAND of DHS Plating system (K222282) meet ASTM F136 and ASTM F543 and used the same test methodology as the required specifications. We utilized the FDA guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway" for the screw testing, and the FDA guidance, "Orthopedic Fracture Fixation Plates - Performance Criteria for Safety and Performance Based Pathway" for the plate testing. Therefore, the subject device is substantially equivalent to the respective predicates' devices identified throughout this submission.
#### Substantial Equivalence Conclusion
Zimed Distal Medial Tibial Plate and Screw System has similar intended use, material comparison, design, operational principle, and equivalent technological characteristics to the respective predicate devices. The non-clinical performance test reports support the safety, effectiveness and performance of the subject device and demonstrate that any difference in technological characteristics and material composition do not raise any new questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.