K230499 · Livsmed, Inc. · GEI · Apr 26, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230499
Device Name
ArtiSential Laparoscopic Instruments-Electrodes
Applicant
Livsmed, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.
Device Story
Single-use, sterile, invasive laparoscopic instrument; used in conjunction with electrosurgical units. Device features jaw, 8mm diameter shaft, and grip with control ring. Operator (surgeon) manipulates grip to open/close jaw, articulate jaw (pitch/yaw ±80°), and rotate jaw 360°. Device delivers high-frequency radiofrequency (RF) energy from electrosurgical unit to tissue via bipolar electrodes; bioimpedance generates heat for tissue coagulation and dissection. Used in OR; provides surgeon with precise tissue manipulation and electrosurgical control. Benefits include improved access and control during minimally invasive procedures.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993-1), electrical safety (IEC 60601 series), and performance testing (appearance, dimensions, operational articulation, tensile strength, feedthrough resistance). Thermal effects on porcine tissue (liver, kidney, muscle) were evaluated via histological analysis. All tests passed.
Indicated for electrosurgical coagulation, dissection, and grasping of tissue during laparoscopic and general surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 26, 2023
LivsMed Inc. Dong Wook Lee QMR (Quality Management Representative) #304, D-dong, 700, Pangyo-ro, Bundang-gu Seongnam-si. Gyeonggi-do 13516 Korea, South
Re: K230499
Trade/Device Name: ArtiSential Laparoscopic Instruments-Electrodes, Bipolar series (four versions, ABF01series, ABD01 series, ABD02 series and ABD04 series)
Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: February 24, 2023 Received: February 24, 2023
Dear Dong Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for LIV'SMED. The logo is in a dark blue color, with the apostrophe in LIV'SMED being a lighter blue color. The font is sans-serif and the letters are all capitalized.
# 3. Indications for Use Statement
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K230499
#### Device Name
ArtiSential Laparoscopic Instruments-Electrodes, Bipolar series (four versions, ABF01 series, ABD02 series and ABD04 series)
#### Indications for Use (Describe)
Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{4}------------------------------------------------
# 510(k) Summary
## 1. General Information
| Applicant/Submitter: LivsMed Inc. | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | #304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516<br>Seongnam-si, Gyeonggi-do, Republic of Korea<br>Tel) +82-70-4282-7652<br>Fax) +82-31-706-3211 |
| Contact Person: | Dong Wook Lee / QMR<br>(Quality Management Representative) |
| Address: | #304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516<br>Seongnam-si, Gyeonggi-do, Republic of Korea<br>Tel) +82-70-7709-4993<br>Fax) +82-31-706-3211<br>Email) dongwook.livsmed@gmail.com |
| Preparation Date: | 02-24-2023 |
### 2. Device Name and Code
| Device Trade Name | ArtiSential Laparoscopic Instruments-Electrodes,<br>Bipolar Series (four versions, ABF01 series, ABD01<br>series, ABD02 series and ABD04 series) |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Electrosurgical Instruments |
| Classification Name | Electrosurgical, cutting & coagulation & accessories |
| Product Code | GEI |
| Regulation Number | 21 CFR 878.4400 |
| Classification | Class II |
| Review Panel | General & Plastic Surgery |
## 3. Predicate Devices
ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are substantially equivalent to the following devices
{5}------------------------------------------------
Table 3.1 Predicate device 1
| Applicant | Device Name | 510(k) Number |
|--------------|--------------------------------------------|---------------|
| LivsMed Inc. | ArtiSential Bipolar<br>Fenestrated Forceps | K200875 |
### Table 3.2 Predicate device 2
| Applicant | Device Name | 510(k) Number |
|--------------|--------------------------------------------|---------------|
| LivsMed Inc. | ArtiSential Bipolar<br>Fenestrated Forceps | K220384 |
### 4. Device Description
The ArtiSential Laparoscopic Instruments – Electrodes, Bipolar series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are sterile, single-use, invasive instruments that used in laparoscopic surgery. There are four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series. Three versions are same except for jaw shape. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system.
### 5. Indications for use
- 5.1 Indications for use
Indications for use include electrosurgical coagulation, dissection, and grasping of tissue during the performance of laparoscopic and general surgical procedures.
### 6. Technical Characteristics in Comparison to Predicate Devices
| | Proposed device | Predicate Device 1 | Predicate Device 2 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K)<br>Number | In process | K200875 | K220384 |
| Manufacture | LivsMed, Inc. | LivsMed, Inc. | LivsMed, Inc. |
| Device Name | ArtiSential Laparoscopic<br>Instruments-Electrodes | ArtiSential Laparoscopic<br>Instruments-Electrodes | ArtiSential Laparoscopic<br>Instruments-Electrodes |
| Clearance Date | N/A | 05-21-2020 | 02-24-2022 |
| Classification /<br>Regulation | Class 2 / 878.4400 | Class 2 / 878.4400 | Class 2 / 878.4400 |
| Product Code | GEI | GEI | GEI |
| Intended for | Prescription Use | Prescription Use | Prescription Use |
| Indications for<br>Use | Electrosurgical<br>coagulation, dissection,<br>and grasping of tissue<br>during the performance<br>of laparoscopic and<br>general surgical<br>procedures. | Electrosurgical<br>coagulation, dissection,<br>and grasping of tissue<br>during the performance<br>of laparoscopic and<br>general surgical<br>procedures. | Electrosurgical<br>coagulation, dissection,<br>and grasping of tissue<br>during the performance<br>of laparoscopic and<br>general surgical<br>procedures. |
| Principles of | This product is a single- | This product is a single- | This product is a single- |
| operation | use instrument used in | use instrument used in | use instrument used in |
| | electrosurgical units to | electrosurgical units to | electrosurgical units to |
| | hold soft tissues or | hold soft tissues or | hold soft tissues or |
| | coagulate and make an | coagulate and make an | coagulate and make an |
| | incision (tissue | incision (tissue | incision (tissue |
| | dissection) during | dissection) during | dissection) during |
| | general laparoscopic | general laparoscopic | general laparoscopic |
| | surgery, which uses the | surgery, which uses the | surgery, which uses the |
| | principle of applying | principle of applying | principle of applying |
| | high-frequency currents | high-frequency currents | high-frequency currents |
| | from the electrode to the | from the electrode to the | from the electrode to the |
| | human body to generate | human body to generate | human body to generate |
| | heat by bioimpedance | heat by bioimpedance | heat by bioimpedance |
| | when radio frequency | when radio frequency | when radio frequency |
| | (RF) energy from the | (RF) energy from the | (RF) energy from the |
| | electrosurgical unit | electrosurgical unit | electrosurgical unit |
| | applies an electric | applies an electric | applies an electric |
| | current to the electrode | current to the electrode | current to the electrode |
| | part, and using the | part, and using the | part, and using the |
| | generated heat to incise | generated heat to incise | generated heat to incise |
| | cellular tissues and cause | cellular tissues and cause | cellular tissues and cause |
| | coagulation. | coagulation. | coagulation. |
| | It is composed of a jaw, | It is composed of a jaw, | It is composed of a jaw, |
| | Φ8 diameter shaft, grip | Φ8 diameter shaft, grip | Φ8 diameter shaft, grip |
| | (including a control | (including a control | (including a control |
| | ring), and electrosurgical | ring), and electrosurgical | ring), and electrosurgical |
| | unit connection | unit connection | unit connection |
| | electrode connector. | electrode connector. | electrode connector. |
| | During a procedure with | During a procedure with | During a procedure with |
| | this product, the jaw | this product, the jaw | this product, the jaw |
| | opens if the control ring | opens if the control ring | opens if the control ring |
| | opens, and jaw closes if | opens, and jaw closes if | opens, and jaw closes if |
| | the control ring closes. | the control ring closes. | the control ring closes. |
| | In addition, the jaw is | In addition, the jaw is | In addition, the jaw is |
| | also bent up, down, left | also bent up, down, left | also bent up, down, left |
| | and right within a range | and right within a range | and right within a range |
| | of ±80° or more by | of ±80° or more by | of ±80° or more by |
| | moving the grip up, | moving the grip up, | moving the grip up, |
| | down, left and right, and<br>the jaw can also turn | down, left and right, and<br>the jaw can also turn | down, left and right, and<br>the jaw can also turn |
| | 360° when rotating the | 360° when rotating the | 360° when rotating the |
| | grip. | grip. | grip. |
| Energy Type | Radiofrequency | Radiofrequency | Radiofrequency |
| Electrode type | Bipolar | Bipolar | Bipolar |
| (monopolar or | | | |
| bipolar) | | | |
| Physical | - Shaft diameter: 8mm | - Shaft diameter: 8mm | - Shaft diameter: 8mm |
| dimensions | - Shaft Length: 250mm, | - Shaft Length: 250mm, | - Shaft Length: 250mm, |
| and design | 380mm, 450mm | 380mm, 450mm | 380mm, 450mm |
| (size, length) | | | |
| Rated voltage | 200Vp | 200Vp | 200Vp |
| Materials | Stainless steel | Stainless steel | Stainless steel |
| (electrode) | | | |
| Materials<br>(insulation) | Polyetherimide | Polyetherimide | Polyetherimide |
| | | | |
| Materials<br>(Shaft) | Glass fiber | Glass fiber | Glass fiber |
| | | | |
| Articulating<br>feature | Pitch:±80° or more,<br>Yaw:±80° or more and<br>Open-Close | Pitch:±80° or more,<br>Yaw:±80° or more and<br>Open-Close | Pitch:±80° or more,<br>Yaw:±80° or more and<br>Open-Close |
| Tip rotation | 360° | 360° | 360° |
| Sterilization | EO | EO | EO |
Table 6.1 Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
### 7. Performance Data
### 7.1 Biocompatibility
The device has been evaluated for its biological safety according to ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1 : Evaluation and Testing Within a Risk Management Process". Following endpoints have been assessed during the evaluation:
- Cytotoxicity -
- -Intracutaneous reactivity
- -Skin Sensitization
- Acute systemic toxicity -
- -Pyrogenicity
### 7.2 Electrical Safety
The ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) have been tested according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18 and IEC 60601-2-2. The test setup included:
The device had passed all performed tests.
#### 7.3 Sterilization
ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series. ABD01 series, ABD02 series and ABD04 series) are provided sterile, intended to be single-use. This product is EO-Sterilization in accordance with ISO-11135.
- 7.4 Shelf life
The proposed expiration date is 3 years from the manufacturing date. The real-time testing will be performed to confirm the shelf-life for 3 years
#### 7.5 Performance test
The device had passed all performed tests.
{8}------------------------------------------------
| Test clause and specification | Test requirement | Results- Remarks |
|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| 1. Appearance<br>: Implement the visual<br>inspection for surface of the<br>device | There should be no defects in<br>the appearance of the product<br>and there should be no<br>problem in use. | No crack, stain or no<br>substances on the surface<br>of the product |
| 2. Dimension<br>: Measure by ruler and vernier<br>calipers | It shall be within $\pm$ 5% of the<br>indicated value of the<br>dimensional term. | Pass<br>Refer to [Test result] on<br>9-50 page at attachment<br>12 |
| 3. Operational test<br>: Manipulating the grip and<br>control ring, and measure the<br>angle at bending and rotation<br>by goniometer. | The jaw must be smoothly<br>opened and closed and free<br>from jamming, the jaw and<br>hub can be bent up, down,<br>left, and right a range of<br>above $\pm$ 80° and are capable<br>of 360° rotation. | The jaw and hub are bent<br>up, down, left and right<br>within above 80° and can<br>rotate 360°. |
| 4. Tensile strength<br>: Hold the jaw and shaft<br>connections respectively and<br>apply a force of 20 N using<br>Push pull gauge. | The jaw and shaft<br>connections shall not be<br>damaged from pulling of 20<br>N. | No damage to the<br>connection when applying<br>a force of 20N |
| 5. Feedthrough test<br>: Electrical conduction<br>between the electrode tip and<br>the connector is tested using a<br>DMM (digital multi meter). | Electricity should be<br>transmitted between the<br>electrode tip and the<br>connector. | The resistance value<br>between the electrode tip<br>and the connector is less<br>than 1Ω |
#### 7.6 Thermal effect
Thermal effects on tissue were also tested. A histological analysis was performed on thermal effect to porcine tissues(liver, kidney and abdominal muscle) through an electrosurgical device.
Based on these performance characteristics, the results demonstrate that the performance requirements were met, the device performs as intended and that the subject device has substantially equivalent performance characteristics to the predicate devices.
### 8. Substantial Equivalence
ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) indication for use is same to the predicate device 1 (K200875) and predicate device 2 (K220384). The energy type, electrode type, sterilization as well as physical characteristics are the same. Although there are some minor differences with each product, these differences between the ArtiSential Bipolar Series and the predicate device do not raise new or different questions of safety and efficacy. There is no new technology and no difference that would raise new or different questions of safety or efficacy.
{9}------------------------------------------------
### 9. Conclusions
In conclusion, the comparison carried out covers all products, models, sizes, and the entire intended purpose of the device under evaluation. The subject device which is the ArtiSential Laparoscopic Instruments-Electrodes, Bipolar Series(four versions, ABF01 series, ABD01 series, ABD02 series and ABD04 series) are same to the predicate device in principles of operation, technological characteristics, as well as performance characteristics. The testing was conducted to evaluate the performance of subject device in comparison to the predicate device. Results of validation and verification activities in design control that include testing/certification to designated standards and performance testing of the devices has demonstrated substantial equivalence of the subject device to the predicate in terms of safety and effectiveness for requested intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.