K230492 · Avitus Orthopaedics, Inc. · FMF · Jun 26, 2023 · General Hospital
Device Facts
Record ID
K230492
Device Name
Avitus® Precision Autograft Delivery System
Applicant
Avitus Orthopaedics, Inc.
Product Code
FMF · General Hospital
Decision Date
Jun 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Avitus® Precision Autograft Delivery System is intended to be used for the delivery of autograft or hydrated allograft bone graft material to an orthopaedic surgical site.
Device Story
Manual surgical instrument for delivery of bone autograft or hydrated allograft to orthopaedic surgical sites. System includes reloadable graft cartridges, bone loading scoop, bone packer, straight/angled delivery cannulas, and threaded delivery plunger. Clinician loads graft material into cartridge using scoop/packer; attaches cannula; inserts plunger. Two-stage deployment: plunger depressed to move graft into cannula; plunger knob twisted to engage threads and deliver controlled volume of graft. Used in clinical orthopaedic settings. Provides precise, small-volume graft delivery compared to predicate. Benefits include controlled, accurate placement of bone graft material during surgery.
Clinical Evidence
No clinical testing was performed. Non-clinical bench testing using cadaveric bone graft material confirmed device reliability, structural integrity, and successful delivery performance.
Indicated for the delivery of autograft or hydrated allograft bone graft material to an orthopaedic surgical site in patients requiring bone grafting procedures.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
June 26, 2023
Avitus Orthopaedics, Inc. % Robert Mclain Sr. Quality and Regulatory Affairs Consultant Keystone Regulatory Services, LLC 324 E. Main Street Leola, Pennsylvania 17540
Re: K230492
Trade/Device Name: Avitus® Precision Autograft Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: May 4, 2023 Received: May 4, 2023
Dear Robert Mclain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark | Digitally signed by<br>Mark Trumbore -S |
|-------------|-----------------------------------------|
| Trumbore -S | Date: 2023.06.26<br>15:57:56 -04'00' |
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230492
Device Name Avitus® Precision Autograft Delivery System
Indications for Use (Describe)
The Avitus® Precision Autograft Delivery System is intended to be used for the delivery of autograft or hydrated allograft bone graft material to an orthopaedic surgical site.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K230492
Avitus Orthopaedics, Inc. Avitus Precision Autograft Delivery System
> Robert McLain Sr. Quality and Regulatory Affairs Consultant Keystone Regulatory Services, LLC 717-656-9656 rob.mclain@keystoneregulatory.com
> > June 12, 2023
{4}------------------------------------------------
# Section 1
# 510(k) Summary
#### Submission Owner and Correspondent 1.1
## Submission Owner
Avitus Orthopaedics, Inc. 6 Armstrong Rd., Suite 107 Shelton, CT 06484 USA Contact: Maxim Budyansky Phone: 860-372-7213 Email: maxim@avitusortho.com
# Submission Correspondent
Keystone Regulatory Services, LLC 342 E. Main Street, Suite 207 Leola, PA 17540 Contact: Robert McLain Phone: 717-656-9656 E-Mail: rob.mclain@keystoneregulatory.com
#### 1.2 Date Summary Prepared
June 12, 2023
#### Device Trade Name 1.3
Avitus Precision Autograft Delivery System
#### 1.4 Device Common Name
Piston syringe.
{5}------------------------------------------------
#### Device Classification Name 1.5
Syringe, Piston Classified as Class 2 at 21 CFR 880.5860, product code FMF.
#### 1.6 Predicate Device
The Spinal Surgical Strategies, LLC Bi-Portal Bone Graft Delivery Device cleared under K142661.
#### 1.7 Description of the Device
The Avitus Precision Autograft Delivery System is designed to deliver bone autograft or hydrated allograft material to an orthopaedic surgical site. The device delivers approximately up to 0.5cc of bone graft material per reloadable cartridge. Each system comes with the following:
- · 2 x Graft Cartridge Features a graft channel, latching lid, and connection points to the Delivery Cannula and Threaded Delivery Plunger.
- · 1 x Bone Loading Scoop Accessory Used to add graft to the Graft Cartridge channel.
- · 1 x Bone Packer Accessory Used to compress graft into the Graft Cartridge channel.
- · 1 x Straight Delivery Cannula Features a hub, which connects to a loaded Graft Cartridge, and straight-tipped cannula, through which graft material is delivered to the orthopaedic surgical site.
- · 1 x Angled Delivery Cannula Features a hub, which connects a loaded Graft Cartridge, and angle-tipped cannula, through which graft material is delivered laterally to the orthopaedic surgical site.
- · 1 x Threaded Delivery Plunger Features a plunger rod, knob, and threads. During graft delivery, the rod is depressed into the Graft Cartridge until the Plunger's threads engage with the Cartridge. This action transfers graft material into the Cannula. With the threads engaged, the user twists the Plunger knob to deliver a controlled volume of graft to the orthopaedic surgical site.
#### Indications for Use 1.8
The Avitus Precision Autograft Delivery System is intended to be used for the delivery of autograft or hydrated allograft bone graft material to an orthopaedic surgical site.
#### 1.9 Technological Characteristics
Table 1.1 compares the technological characteristics of the proposed Avitus Precision Autograft Delivery System and the predicate Bi-Portal Bone Graft Delivery Device.
{6}------------------------------------------------
| Feature | Proposed Device - Avitus Precision<br>Autograft Delivery System | Predicate Device -<br>Bi-Portal Bone Graft<br>Delivery Device<br>(K142661) |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Product Code | FMF; Syringe, Piston | FMF; Syringe, Piston |
| Classification<br>and Regulation | 880.5860 | 880.5860 |
| Indication for<br>Use | Intended to be used for the delivery of bone<br>autograft or hydrated allograft. | Intended to be used for<br>the delivery of bone<br>autograft or hydrated<br>allograft. |
| Materials | Biocompatible Polycarbonate (Graft Cartridge,<br>Cannula, Plunger, Pack tool, Bone Scoop<br>Accessory<br>Biocompatible Stainless Steel (Cannula and<br>Plunger) | Biocompatible Plastics<br>(Entire Device) |
| Mechanism of<br>Operation | Graft material dispersed from device tip by<br>depressing the plunger. Plunger is threaded and<br>is twisted clockwise to disperse graft material. | Graft material dispersed<br>from device tip by<br>depressing the plunger. |
| Recommended<br>Graft Volume | Approximately up to 0.5cc per Cartridge | Up to 4.0cc in Barrel |
| Bone Graft Exit<br>Hole<br>Configuration | Single cannula endhole; 3.4mm diameter | Two sideholes; 5.6mm x<br>14.8mm |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide |
Table 1.1: Device Comparison
This section contains a comparison of technological characteristics of the Avitus Precision Autograft Delivery System and the predicate Bi-Portal Bone Graft Delivery Device and, where there are differences, provides an explanation of why there is no negative impact on substantial equivalence.
### Materials
Both the Avitus Precision Autograft Delivery System and the predicate K142661 Bi-Portal Bone Graft Delivery Device are made of biocompatible plastic materials. All elements of the Avitus Precision Autograft Delivery System are composed of polycarbonate with the sole exception of the cannula and plunger components, which contain biocompatible stainless steel in addition to polycarbonate. The polycarbonate and stainless steel materials which make up the Avitus Precision Autograft Delivery System are identical to the polycarbonate and stainless steel materials which make up the Avitus Orthopaedics, Inc. Bone Harvester devices (K152474). Similarities in Avitus Precision Autograft Delivery System and Avitus Bone Harvester device scale, manufacturing methods, and intended clinical use environment indicate that materials will not negatively impact the overall use and function of the new device. Biocompatibility has been demonstrated with the uti
{7}------------------------------------------------
lization of identical materials and processes to the materials found in the Avitus Orthopaedics. Inc. Bone Harvester devices, there is no negative impact on product safety or effectiveness.
## Mechanism of Operation
The predicate K142661 Bi-Portal Bone Graft Delivery Device deploys bone autograft or hydrated allograft when the user depresses the Threaded Delivery Plunger into the Cannula. The Avitus Precision Autograft Delivery System deploys bone autograft or hydrated allograft using a two stage deployment process. First, user depresses the Threaded Delivery Plunger into the Graft Cartridge until the Plunger's back-end threads engage with the Graft Cartridge, moving bone graft material into the Cannula. Next, the user turns the Threaded Delivery Plunger to deploy the desired volume of bone graft material from the Cannula. Despite this additional feature, which provides the user with precise control over the deployed volume of bone graft, the Avitus Precision Autograft Delivery System and the predicate K142661 Bi-Portal Bone Graft Delivery Device share basic mechanism of action. The addition of the threaded mechanism in the Avitus Precision Autograft Delivery System does not negatively impact product safety or effectiveness.
## Recommended Graft Volume
The predicate K142661 Bi-Portal Bone Graft Delivery Device can accommodate up to 4.0 cc of hydrated allograft or autograft material in its syringe barrel. The Avitus Precision Autograft Delivery System can accommodate approximately up to 0.5 cc of hydrated allograft or autograft material in each reloadable Graft Cartridge. If more graft is required at the delivery site, the user replaces the Graft Cartridge with another loaded Graft Cartridge. This is repeated as many times as needed by opening the lid of the Graft Cartridge and reloading it after bone delivery. The Avitus Precision Autograft Delivery System and the predicate K142661 Bi-Portal Bone Graft Delivery Device are intended to deliver different volumes of graft. There is no negative impact of recommended graft volume on product safety or effectiveness because the Avitus Precision Autograft Delivery System will be utilized by clinicians who want to deliver smaller, precise volumes of graft.
## Bone Graft Exit Hole Configuration
The predicate K142661 Bi-Portal Bone Graft Delivery Device deploys bone autograft or hydrated allograft through bilateral sideholes. Each sidehole measures 14.8 mm x 5.6 mm. The Avitus Precision Autograft Delivery System deploys bone autograft through a single endhole in the case of the Straight Delivery Cannula and through a single sidehole in the case of the Angled Delivery Cannula. Both Straight and Angled Delivery Cannulas have an exit hole diameter of 3.4 mm. This difference, combined with the difference in recommended graft volume, indicates that the devices are designed to deliver different volumes of graft material, however, there is no difference between the proposed and predicate devices in terms of function and intended purpose. The Avitus Precision Autograft Delivery System effectively delivers graft material. This has been demonstrated in performance testing. There is no negative impact of the difference in bone graft exit hole configuration on product safety or effectiveness.
#### 1.10 Non-Clinical Testing
Avitus Orthopaedics, Inc. performed worst-case non-clinical testing in which the Avitus Precision Autograft Delivery System was used to deliver cadaveric bone graft material. Testing found that the Avitus Precision Autograft Delivery System meets the reliability and confidence requirements for successful delivery and for structural integrity during normal use.
{8}------------------------------------------------
#### 1.11 Biocompatibility
The previously-cleared Avitus Bone Harvester and the Avitus Precision Autograft Delivery System are both external communicating devices with limited tissue/bone/dentin contact (<24 hours). Per ISO 10993-1:2018 and FDA Guidance Use of International Standard ISO 10993-1, the following biocompatibility endpoints are appropriate for both devices:
- · Cytotoxicity
- · Sensitization
- · Irritation
- · Acute Systemic Toxicity
- · Material-Mediated Pyrogenicity
Per Attachment D of the aforementioned FDA guidance, the results of biocompatibility testing of the Avitus Bone Harvester may be applied to the subject device, considering that the material formulation, body contact, and sterilization of direct and indirect contacting components are identical and that the geometry and manufacturing processes (noting that no new plasticizers, fillers, additives, cleaning agents, mold release agents have been added to the subject device manufacturing process) of the Avitus Bone Harvester represent worst-case biocompatibility conditions.
#### Clinical Testing 1.12
No clinical testing was performed in association with this submission.
#### 1.13 Conclusions
The results of the comparison of design, materials, intended use and technological characteristics demonstrate that the device is as safe and effective as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.