K230488 · Walz Elektronik GmbH · FFK · Aug 31, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K230488
Device Name
EL27-Compact; Sterile EHL-Probes
Applicant
Walz Elektronik GmbH
Product Code
FFK · Gastroenterology, Urology
Decision Date
Aug 31, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EL27-Compact is a device to be used together with EHL-probes from Walz Elektronik GmbH for disintegration of concernments in the upper and lower urinary tract and in gastroenterology, particularly in the bile duct.
Device Story
Table-top electrohydraulic lithotripter; used with disposable EHL probes inserted through endoscope working channel. Physician places probe tip near stone in saline environment; activates via footswitch. Device delivers high-voltage electric pulses to probe tip; generates high-pressure shockwaves to fragment stones. Used in hospital environments by physicians. Output is mechanical shockwave energy; clinical benefit is stone disintegration. System includes generator, power cord, footswitch, and probe cable.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via visual inspection, dimensional tolerance checks, endoscopic insertion testing, shockwave measurement, probe tip withdrawal force, and electrode consumption. Comparative testing on plasterboard confirmed consistent stone-fragmentation effects across intensity settings. Electrical safety and EMC compliance verified per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Table-top electrohydraulic generator; 3 intensity settings (Low, Medium, High); footswitch-controlled. Disposable EHL probes (epoxy, stainless steel, copper, polyimide, PVDF, polyolefin) sterilized via ethylene oxide. Complies with IEC 60601-1 and IEC 60601-1-2. Proprietary connection between generator and probe.
Indications for Use
Indicated for disintegration of concrements (stones) in the upper and lower urinary tract and in gastroenterology, particularly in the bile duct, for patients requiring endoscopic lithotripsy.
Regulatory Classification
Identification
An electrohydraulic lithotriptor is an AC-powered device used to fragment urinary bladder stones. It consists of a high voltage source connected by a cable to a bipolar electrode that is introduced into the urinary bladder through a cystoscope. The electrode is held against the stone in a water-filled bladder and repeated electrical discharges between the two poles of the electrode cause electrohydraulic shock waves which disintegrate the stone.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notifications for Intracorporeal Lithotripters.”
1.9 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION (K914514)
3 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION (K914515)
9 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION (K914516)
1.9 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION (K913955)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
August 31, 2023
Walz Elektronik GmbH Bernd Vollmer CEO Walddorfer Strasse 40 Rohrdorf, Baden-Wuertemberg 72229 Germany
Re: K230488 Trade/Device Name: EL27-Compact Regulation Number: 21 CFR§ 876.4480 Regulation Name: Electrohydraulic Lithotriptor Regulatory Class: II Product Code: FFK Dated: August 3, 2023 Received: August 3, 2023
Dear Bernd Vollmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino -S
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K230488
Device Name EL27-Compact
Indications for Use (Describe)
The EL27-Compact is a device to be used together with EHL-probes from Walz Elektronik GmbH for disintegration of concernments in the upper and lower urinary tract and in gastroenterology, particularly in the bile duct.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Image: @ logo | 5 - 510(k) Summary | K230488<br>Date.:23.05.2022<br>Rev:1.0<br>Page: 5-3 of 5-7 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| General | | |
| 510(k) Submitter: | Walz Elektronik GmbH<br>Walddorfer Straße 40<br>72229 Rohrdorf, Germany<br>Phone: (+49) 7452 2020 | |
| Contact Person: | Bernd Vollmer, CEO<br>Walz Elektronik GmbH<br>Walddorfer Straße 40<br>72229 Rohrdorf, Germany<br>E-Mail Address: vollmer@walz-el.de | |
| Official Correspondent: | German American Chamber of Commerce, Inc.<br>United States<br>80 Pine Street, 24st Floor<br>New York, 10005<br>Phone Number: (212) 974-8830<br>Emergency Phone: (212) 974-8830<br>Fax Number: (212) 974-8867<br>E-Mail Address: legalservices@gaccny.com | |
| Date Prepared: | 21.02.2023 | |
# Predicate Devices
| Predicate Device | 510(k) No. |
|-------------------------------------------------|------------|
| AUTOLITH TOUCH | K130368 |
| AUTOLITH URO-TOUCH | K130368 |
| 1.9 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION | K914514 |
| 3 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION | K914515 |
| 9 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION | K914516 |
| 1.9 FRENCH LITHOTRIPTER ELECTRODE, MODIFICATION | K913955 |
{4}------------------------------------------------
## Device Description
(a
The EL27-Compact is a device, designed to be used with Walz Elektronik GmbH disposable EHL-Probes for the fragmentation of concrements in the gastroenterology and urology.
A physician inserts and pushes the disposable EHL probe through the working channel of an endoscope and places the tip of the probe in front of the stone surrounded with saline solution. With pressing the footswitch, the device provides electric pulse energy in form of a high-voltage pulse. This pulse generates high pressure shockwaves that fragment hard stones.
The EL27-Compact is a table-top device and consists of the generator itself, the power cord, the footswitch, a probe cable for connection between the device and the EHL-probes.
Regulation Name: Regulation Number: Regulatory Class: Product Codes: Review Panel: Trade Name: Generic/Common Name: Lithotripter, Electro-Hydraulic 21 CFR. 876.4480 Class II FFK Gastroenterology/Urology EL27-Compact Lithotripter
{5}------------------------------------------------
### Indications for use
The EL27-Compact is a device to be used together with EHL-probes from Walz Elektronik GmbH for disintegration of concernments in the upper and lower urinary tract and in gastroenterology, particularly in the bile duct.
# Comparison of Technological Characteristics
### Generator
ರ
- -Both devices, the predicate and the subject of this 510(k) submission, are used in the field of urology and gastroenterology for fragmentation of concrements.
- -Both devices are table-top units that connect over a probe connecting cable to a disposable probe that, is used in the working channel of an endoscope. The physician inserts and pushes the disposable EHL probe through the working channel of a compatible endoscope and places the tip of the probe in front of the stone surrounded with saline solution (0,9% isotonic saline solution). With pressing the footswitch, the device provides electric pulse energy in form of a high-voltage pulse. This pulse generates high pressure shockwaves that fragment hard stones.
- Both devices have a selection of 3 intensities (Low, Medium, High) -
- -Both devices' outputs are controlled over a footswitch
- -Both devices are used in hospital environment
- -Both devices are made of commonly used materials for building electronic devices
#### Disposable EHL Probes
- -Both probe types have a proprietary connection to the generator
- -Both probe types are sterile and for single-use
- Both probe types are made of similar materials like epoxy, stainless steel, copper, -Polyimid, Polyvinylidenfluorid or Polyolefin
- -Both probe types have the tip of the probe as the patient contacting part
- -Both probe types are sterilized with ethylene oxide
{6}------------------------------------------------
## Summary of Performance Testing
The following performance testing was conducted in substantial equivalence determination.
### Biocompatibility Evaluation
ರ
The EL27-Compact does not contain components that come directly or indirectly in patient contact. Biocompatibility testing according to ISO 10993-1 for these components is not required. Biocompatibility evaluation of the patient contacting EHL-Probes were successfully conducted previously according to ISO 10993-1 Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing. There have been no changes that have impacted the established biocompatibility.
- 1. Biological Safety Evaluation of EHL Probes
- 2. In vitro Cytotoxicity testing
- 3. Testing for contact allergens
- 4. Testing for tissue compatibility
### Bench Testing
These comprehensive validation bench tests support equivalence to the predicate device. Testing confirmed that comparable effects could be achieved for applicable modes of operation.
For the EHL-Probe, the following tests were performed:
- -Visual inspection of the appearance of the probes
- -Measuring the dimensions to check if tolerances are okay
- -Insertion (not P4,5/3000/f) of probes into an Endoscope
- -Measuring the shockwaves
- Measuring withdrawing force of the tip of the probes -
- -Measuring the electrode consumption.
For the EL27, the following test was performed:
The Test performed was to set up the 3 devices and connect them to 3Fr. Probes. Choosing the different intensities on the devices, pulses have been emitted on a piece of plasterboard. The distance of the probe to the plasterboard have been kept the same overall tests. The resulting holes in the plasterboard, which vary in depth caused by the chosen intensity, can be compared to another.
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC performance testing have been confirmed to be in compliance with the relevant requirements as noted below.
{7}------------------------------------------------
- IEC60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance
- IEC 60601-1:2005+AMD1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
# Software Verification and Validation Testing
Software testing has been performed and documented accordingly to the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
### Substantial Equivalence
The predicate device is composed of a generator, a probe cable, a foot switch and a power cable. The indications for use, principles of operation, fundamental technology of the generator are identical to the predicate device. It has full compatibility with the existing device and so that users can select from the entire instrument line-up. The indicated patient population and procedures are also identical to the predicate devices. However, functions and principles of operation are identical to the predicate device.
## Conclusion
In summary, the EHL-Probes, the EL27-Compact and its accessories are at least substantially equivalent to the predicate devices and present no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.