← Product Code [MQV](/productcode/MQV) · K230486

# Cove Strip (K230486)

_SeaSpine Orthopedics Corporation · MQV · Aug 21, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230486

## Device Facts

- **Applicant:** SeaSpine Orthopedics Corporation
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Aug 21, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Cove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cove Strip resorbs and is replaced with bone during process. Cove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis.

## Device Story

Cove Strip is a bone void filler implant composed of 70:30 Beta-Tricalcium Phosphate (B-TCP) and Hydroxyapatite (HA) ceramic granules within a Type-1 collagen scaffold. It functions as an osteoconductive substrate to support new bone formation. Used by surgeons in clinical settings, the device is mixed with autograft in a 1:1 ratio and implanted into bony voids or gaps in the posterolateral spine or pelvis. The implant resorbs over time and is replaced by the patient's own bone during the healing process. It is provided sterile for single use.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing (biocompatibility per ISO 10993-1, bacterial endotoxin per USP<85>/<161> and AAMI ST72) and an in vivo animal study assessing resorption, remodeling, and fusion rates compared to a predicate device, including local effects after implantation per ISO 10993-6:2016.

## Technological Characteristics

Composition: 70:30 Beta-Tricalcium Phosphate (B-TCP) and Hydroxyapatite (HA) ceramic granules with Type-1 collagen scaffold. Sterilization: Ethylene Oxide (ISO 11135) to SAL 10^-6. Biocompatibility: ISO 10993-1. Endotoxin: AAMI ST72, USP<85>, USP<161> (limit ≤20 EU/Device). Form factor: Strip, provided in double blister tray.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- MASTERGRAFT Putty ([K140375](/device/K140375.md))
- Integra MOZAIK™ Bone Regeneration Matrix — Strip ([K063124](/device/K063124.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

August 21, 2023

SeaSpine Orthopedics Corporation Cindy Toyama Regulatory Affairs Specialist 2 Goodyear Irvine, California 92618

Re: K230486

Trade/Device Name: Cove Strip Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MOV Dated: July 22, 2023 Received: July 24, 2023

Dear Cindy Toyama:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image shows the name "Jesse Muir-S" in a large, bold, sans-serif font. The text is black and stands out against a white background. There is a faint, light blue watermark in the background, partially obscuring the text.

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230486

Device Name Cove Strip

Indications for Use (Describe)

Cove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cove Strip resorbs and is replaced with bone during process. Cove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

### K230486

### Contact Details

| Applicant Name: | SeaSpine Orthopedics Corporation            |
|-----------------|---------------------------------------------|
| Address:        | 5770 Armada Drive, Carlsbad CA              |
| Phone number:   | (949) 855-7175                              |
| Fax number:     | (760) 683-6874                              |
| Contact Person: | Cindy Toyama, Regulatory Affairs Specialist |
| Date Prepared:  | June 21, 2023                               |

Device Name

| Device/Trade Name:   | Cove Strip                                            |
|----------------------|-------------------------------------------------------|
| Common Name:         | Bone Void Filler                                      |
| Classification Name: | Filler, Bone Void, Calcium Compound (21 CFR 888.3045) |
| Class:               | II                                                    |
| Product Code:        | MQV                                                   |

### Legally Marketed Predicate Device

| 510(k) Number               | Product Code | Trade Name                                           | Manufacturer                         |  |
|-----------------------------|--------------|------------------------------------------------------|--------------------------------------|--|
| Primary Predicate Device    |              |                                                      |                                      |  |
| K140375                     | MQV          | MASTERGRAFT Putty                                    | Medtronic Sofamor<br>Danek USA, Inc. |  |
| Additional Predicate Device |              |                                                      |                                      |  |
| K063124                     | MQV          | Integra MOZAIK™ Bone<br>Regeneration Matrix — Strip* | Integra LifeSciences<br>Corporation  |  |

*Since the most recently cleared 510(k) for the predicate device, Integra MOZAK™ Osteoconductive Scaffold – Strip (K063124), under contractual agreements between IsoTis OrthoBiologics, Inc. and Integra LifeSciences Corporation, the device has been rebranded under a new trade name for IsoTis OrthoBiologics, Inc., and will herein be referred to as "IsoTis Mozaik Strip". IsoTis Mozaik Strip is legally manufactured by IsoTis OrthoBiologics, Inc.

{4}------------------------------------------------

#### Device Description

Cove Strip is comprised of a ceramic granule consisting of a ratio of 70:30 Beta-Tricalcium Phosphate (B-TCP): Hydroxyapatite (HA) combined with highly purified Type-1 collagen. The collagen scaffold contains ceramic granules throughout, providing osteoconductive substrates that support new bone formation. Cove Strip combined with autograft in a 1:1 ratio, is intended to be placed in the posterolateral spine. The implant is provided sterile and non-pyrogenic for single use in a double blister tray configuration.

#### Indications for Use

Cove Strip is an implant intended to fill bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Cove Strip resorbs and is replaced with bone during the healing process. Cove Strip must be used with autograft as a bone graft extender in the posterolateral spine and pelvis.

#### Summary of Technological Characteristics

The subject device is identical or similar to the cited predicate devices in regard to intended use/indications for use, device description, technological characteristics (i.e. design, materials, manufacturing, labeling, sterility, etc.), and non-clinical performance (i.e., in vivo performance (animal) study).

#### Summary of Non-Clinical Testing to Support Substantial Equivalence

Non-clinical testing performed on the subject device includes tests for biocompatibility and bacterial endotoxin to establish safety. Biocompatibility was performed in accordance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Bacterial endotoxin testing complies with AAMI ST72 Bacterial Endotoxins – Test Methods, Routine Monitoring, and Alternatives to Batch Testing, USP<85> Bacterial Endotoxin Test, and USP<161>, Medical Devices – Bacterial Endotoxin and Pyrogen Tests, and has been validated ensure a BET limit of ≤20 EU/Device.

Equivalency was established with a predicate device consisting of identical product sizing and packaging that allowed us to adopt the sterilization validation. Sterilization complies with ISO 11135, Sterilization of health care products-Ethylene Oxide-Requirements to ensure a sterility assurance level (SAL) of 10-6.

An in vivo (animal) study for safety and performance demonstrated comparable resorption, remodeling and rates of fusion when compared to a legally marked predicate. The study employed various analyses and endpoints were assessed at several time points. The subject device was also assessed per ISO 10993-6:2016 Biological evaluation of medical devices – Part 6: Tests for local effects after implantation.

#### Clinical Testing

Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data.

{5}------------------------------------------------

### Conclusions

The submitted data demonstrate that the subject device is substantially equivalent to the cited legally marketed predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K230486](https://fda-staging.innolitics.com/device/K230486)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
