The Compact Wearable Pump is intended to express and collect breastmilk from lactating women for the purpose of feeding collected breastmilk to a baby. The Compact Wearable Pump is intended for a single user.
Device Story
Compact Wearable Pump; battery-powered breast pump for single-user, home-use. Device consists of pump body, breast flange, flange insert, diaphragm, manifold disk, and collection cup. User wears device inside bra; operates via press-button interface on main body. Microcontroller-based system controls vacuum levels and cycle speeds; stimulation mode (8 levels) and expression mode (8 levels, 3 cycle styles). Device creates seal around nipple; applies/releases suction to express milk into collection container. Backflow protection membrane prevents milk ingress into vacuum system. Powered by internal 3.8V lithium-ion battery; charged via USB-C. LED indicators provide status. Healthcare provider does not operate; patient self-manages. Output is expressed breastmilk for infant feeding; device facilitates lactation management for nursing mothers.
Clinical Evidence
Bench testing only. Biocompatibility (ISO 10993-5, 10993-10), electrical safety (ANSI/AAMI ES60601-1, IEC 62133-2, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and performance verification (vacuum levels, backflow protection, use life, battery functionality) were performed. No clinical data provided.
Technological Characteristics
Powered breast pump; 3.8V 1300mAh Li-ion battery; USB-C charging. Microcontroller-based control. Materials compliant with 21 CFR 174-179. Vacuum range 36-280 mmHg; cycling 0.53-3.04 cycles/sec. Wearable form factor. Backflow protection membrane. Non-sterile. No wireless/mobile app connectivity.
Indications for Use
Indicated for lactating women to express and collect breastmilk for infant feeding. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Lansinoh Smartpump 3.0 Double Electric Breast Pump (K222726)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2023
Lansinoh Laboratories, Inc. Lindsay Ewers Director of Quality Assurance 99 Canal Center Plaza. Suite 550 Alexandria, VA 22314
Re: K230469
> Trade/Device Name: Compact Wearable Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: May 31, 2023 Received: May 31, 2023
Dear Lindsay Ewers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230469
Device Name Compact Wearable Pump
Indications for Use (Describe)
The Compact Wearable Pump is intended to express and collect breastmilk from lactating women for the purpose of feeding collected breastmilk to a baby. The Compact Wearable Pump is intended for a single user.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K230469
#### 1. Submitter Information
| Applicant: | Lansinoh Laboratories Inc. |
|------------|----------------------------------------------------------|
| Contact: | Lindsay Ewers |
| Phone: | (727) 542-3743 |
| Email: | lewers@lansinoh.com |
| Address: | 99 Canal Center Plaza, Suite 550<br>Alexandria, VA 22314 |
#### 2. Correspondent Information
| Contact: | Lindsay Ewers |
|----------|-------------------------------|
| | Director of Quality Assurance |
| Firm: | Lansinoh Laboratories |
#### 3. Date prepared: June 29, 2023
#### 4. Device Information
| Device Name: | Compact Wearable Pump |
|--------------------|-----------------------------|
| Common Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered Breast Pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Regulatory Class: | Class II |
#### 5. Predicate Device Information
| Device Name: | Lansinoh Smartpump 3.0 Double Electric Breast Pump |
|----------------|----------------------------------------------------|
| 510(k) Number: | K222726 |
| Manufacturer: | Lansinoh Laboratories |
The predicate device has not been subject to a design-related recall.
## 6. Device Description
The Compact Wearable Pump is a battery-powered breast pump which consists of the following components: pump body, breast flange, flange insert, diaphragm, manifold disk, collection cup, charging cord, and carrying bag. The subject device is capable of single pumping (using both packaged pumps independently). It is provided non-sterile and can be re-used by a single user. It does not incorporate off-theshelf (OTS) software or wireless technology/mobile app software functionality. The subject device cannot be operated while plugged into AC power.
The Compact Wearable Pump is designed to work in a user's bra. The main body includes a press-button user interface, pump body, and LED indicator lights. The user interface allows the user to switch from stimulation to expression mode and control the vacuum levels within those modes. Stimulation mode consists of 8 vacuum levels and fixed cycle speeds when used. Expression mode includes 8 vacuum levels with three different cycle speed configurations that vary duration between suction and let down for further customization by the user.
{4}------------------------------------------------
The Compact Wearable Pump is capable of providing vacuum levels from 36-159 mmHg with cycling rates of 1.67-3.04 cycles per second in stimulation mode and vacuum levels from 78-280 mmHg with cycling rates from 0.53-1.99 cycles per second dependent on which expression mode is used. The Compact Wearable Pump is charged with a 5 V DC USB type C adaptor and powered by a 3.8 V, 1300mAh internal rechargeable lithium-ion battery.
The breast pump expresses breast milk by creating a seal around the nipple using a flange and applying and releasing suction to the milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.
All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 174-179.
# 7. Indications for Use
The Compact Wearable Pump is intended to express and collect breastmilk from lactating women for the purpose of feeding collected breastmilk to a baby. The Compact Wearable Pump is intended for a single user.
### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
| | Compact Wearable<br>Pump<br>K230469<br>Subject Device | Smartpump 3.0 Double<br>Electric Breast Pump<br>K222726<br>Predicate Device | Comparison |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Product Code | HGX | HGX | Same |
| Regulation Number | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Regulatory Class | Class II | Class II | Same |
| Patient Population | Lactating Women | Lactating Women | Same |
| Indications for Use | The Compact Wearable Pump is<br>intended to express and collect<br>breastmilk from lactating<br>women for the purpose of<br>feeding collected breastmilk to a<br>baby. The Compact Wearable<br>Pump is intended for a single<br>user. | The Lansinoh Smartpump 3.0<br>Double Electric Breast Pump is<br>intended to express and collect<br>the breastmilk of a nursing<br>woman for the purpose of<br>feeding collected breastmilk to a<br>baby. The pump is intended for a<br>single user. | Same Intended Use |
| Pump Options | Single or Double | Single or Double | Same |
| Cycling control<br>mechanism | Microcontroller | Microcontroller | Same |
| Backflow Protection | Yes | Yes | Same |
| Mobile Application | No | No | Same |
| Indicators | Yes, LED indicators | Yes, LCD display and LED<br>indicators | Similar |
| Single User | Yes | Yes | Same |
# Table 1: Comparator Table for Subject and Predicate Devices
{5}------------------------------------------------
| Media separation<br>(backflow<br>protection) | Yes | Yes | Same |
|----------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Expression pattern | 2-Phase | 2-Phase | Same |
| Expression modes | 3 styles | 3 styles | Same |
| Suction levels<br>(Stimulation) | 36-159 mmHg | 78-170 mmHg | Different. Differences in massage<br>suction levels do not raise<br>different questions of safety and<br>effectiveness |
| Suction levels<br>(Expression) | 78-280 mmHg | 119-280 mmHg | Different. Differences in<br>pumping suction levels do not<br>raise different questions of safety<br>and effectiveness |
| Cycles per Second<br>(stimulation) | 1.67-3.04 | 1.61-2.56 | Different. Differences in<br>available cycle speed do not raise<br>different questions of safety and<br>effectiveness. |
| Cycles per Second<br>(Expression) | Mode 1: 0.74-1.99<br>Mode 2: 0.64-1.72<br>Mode 3: 0.53-1.32 | 0.61-1.74 | Different. Differences in<br>available cycle speed do not raise<br>different questions of safety and<br>effectiveness. |
| Suction levels | 8 vacuum levels | 8 vacuum levels | Same |
| User Interface | On-Off switch, vacuum<br>adjustment, Pump style switch | On-Off switch, vacuum<br>adjustment, double/single<br>pumping, Pump style switch | Different. Differences in<br>available suction levels do not<br>raise different questions of safety<br>and effectiveness. |
| Adjustable Suction<br>Levels | Yes | Yes | Same |
| Design | Wearable Milk Collector and<br>Flange | Tabletop Milk Collector and<br>Flange | Different. Differences in overall<br>design do not, on their own, raise<br>different questions of safety and<br>effectiveness. |
| Power Supply | 3.8 V, 1300 mAh Lithium-Ion<br>Polymer battery | AC adapter or 7.4 V, 1500<br>mAh Lithium-Ion Polymer<br>battery | Different. Differences in<br>available power supply do not<br>raise different questions of safety<br>and effectiveness as performance<br>is independently assessed for<br>each available power<br>configuration. |
The indications for use of the subject and predicate device are similar, and they have the same intended use (i.e., for collection of breast milk from the breasts of lactating women).
The subject and predicate devices have similar technological features, including expression mode options, control mechanism, backflow protection, and device indicators. However, as shown in the table above, there are technological differences between the subject and predicate device, including different vacuum and cycle specifications, power supply, wear configurations, and vacuum levels. The different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
### 9. Summary of Non-Clinical Performance Testing
## Biocompatibility
Biocompatibility studies on the patient-contacting components of the subject device, including irritation, cytotoxicity, and sensitization testing were performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1. "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process", as follows:
- . Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010)
- . Irritation (ISO 10993-10:2010)
The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing.
### Electrical Safety
Testing was conducted in accordance with ANSI/AAMI ES60601- 1:2005/A2:2010 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
#### Electromagnetic Compatibility
Testing was conducted in accordance with IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
### Software
Software was evaluated for a moderate level of concern as recommended in the 2005 FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
### Performance Testing
Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:
- Vacuum level verification testing at each mode/cycle demonstrated that the devices . meet mode/cycle specifications.
- . Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
- Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life.
- Battery performance testing was conducted to demonstrate that the battery remains functional . during its stated battery use-life.
- Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
#### 10. Conclusion
The results of the performance testing described above demonstrate that the Compact Wearable Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.