Empatica Health Monitoring Platform; EmbracePlus; Empatica Care; Care Portal
Applicant
Empatica S.r.l.
Product Code
MWI · Cardiovascular
Decision Date
Oct 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Empatica Health Monitoring Platform is a wearable device and paired mobile and cloud-based software platform intended to be used by trained healthcare professionals or researchers for retrospective remote monitoring of physiologic parameters in ambulatory individuals 18 years of age and older in home-healthcare environments. As the platform does not provide real-time alerts related to variation of physiologic parameters, users should use professional judgment in assessing patient clinical stability and the appropriateness of using a monitoring platform designed for retrospective review. The device is intended for continuous data collection supporting intermittent review of the following physiological parameters: · Pulse Rate, - · Blood Oxygen Saturation under no-motion conditions, - · Respiratory Rate under no motion conditions, - · Peripheral Skin Temperature, - · Electrodermal Activity, - · Activity associated with movement during sleep The Empatica Health Monitoring Platform can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable. The Empatica Health Monitoring Platform is not intended for SpO2 monitoring in conditions of motion or low perfusion. The Empatica Health Monitoring Platform is intended for peripheral skin temperature monitoring, where monitoring temperature at the wrist is clinically indicated. The Empatica Health Monitoring Platform is not intended for Respiratory Rate monitoring in motion conditions. This device does not detect apnea and should not be used for detecting or monitoring cessation of breathing. The Empatica Health Monitoring Platform is not intended for Pulse Rate monitoring in patients with chronic cardiac arrhythmias, including atrial fibrillation and atrial/ventricular bigeminy. and is not intended to diagnose or analyze cardiac arrhythmias. The Empatica Health Monitoring Platform is not a substitute for an ECG monitor, and should not be used as the sole basis for clinical decision-making.
Device Story
Wearable wrist-worn device (EmbracePlus) collects raw physiologic data via sensors; transmits via Bluetooth Low Energy to mobile app (Care App). Care App module (EmpaDSP) computes physiologic parameters; transmits data via cellular/Wi-Fi to cloud (Empatica Cloud). Healthcare professionals/researchers access data via cloud-based portal (Care Portal) for retrospective review. Device provides no real-time alerts. Used in home-healthcare environments to support clinical decision-making; not a substitute for ECG monitoring. Benefits include continuous, non-invasive physiologic data collection for intermittent review.
Clinical Evidence
Three clinical studies (n=146) evaluated pulse rate accuracy vs. ECG reference: ≤3 bpm (no-motion), ≤5 bpm (motion). Four clinical studies (n=117) evaluated respiratory rate accuracy vs. capnography: ≤3 brpm (no-motion). Subgroup analyses confirmed performance across sex, skin tone, and clinical conditions. No adverse events reported.
Technological Characteristics
Wrist-worn wearable with MEMS accelerometer (±16g), optical PPG sensors, and high-precision temperature sensor. Connectivity: BLE to mobile, Wi-Fi/cellular to cloud. Materials compliant with ISO 10993-1. Battery: Rechargeable Li-ion. Software: EmpaDSP module for signal processing. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 80601-2-56.
Indications for Use
Indicated for ambulatory individuals 18+ years old in home-healthcare environments for retrospective remote monitoring of pulse rate, SpO2 (no-motion), respiratory rate (no-motion), peripheral skin temperature, electrodermal activity, and sleep-related movement. Contraindicated for SpO2/RR monitoring during motion, apnea detection, or pulse rate monitoring in patients with chronic cardiac arrhythmias (e.g., atrial fibrillation, bigeminy, trigeminy).
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
October 30, 2023
Empatica S.r.1. Alberto Poli, Regulatory Affairs & Quality Manager Via Stendhal, 36 Milan, 20144, Italy
Re: K230457
Trade/Device Name: Empatica Health Monitoring Platform; EmbracePlus; Empatica Care; Care Portal Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, BZQ, DQA, DRG, FLL, GZO, LEL Dated: April 3, 2023 Received: September 29, 2023
Dear Alberto Poli:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K230457
Device Name
Empatica Health Monitoring Platform; EmbracePlus; Empatica Care; Care Portal
Indications for Use (Describe)
The Empatica Health Monitoring Platform is a wearable device and paired mobile and cloud-based software platform intended to be used by trained healthcare professionals or researchers for retrospective remote monitoring of physiologic parameters in ambulatory individuals 18 years of age and older in home-healthcare environments. As the platform does not provide real-time alerts related to variation of physiologic parameters, users should use professional judgment in assessing patient clinical stability and the appropriateness of using a monitoring platform designed for retrospective review.
The device is intended for continuous data collection supporting intermittent review of the following physiological parameters:
· Pulse Rate,
- · Blood Oxygen Saturation under no-motion conditions,
- · Respiratory Rate under no motion conditions,
- · Peripheral Skin Temperature,
- · Electrodermal Activity,
- · Activity associated with movement during sleep
The Empatica Health Monitoring Platform can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable.
The Empatica Health Monitoring Platform is not intended for SpO2 monitoring in conditions of motion or low perfusion.
The Empatica Health Monitoring Platform is intended for peripheral skin temperature monitoring, where monitoring temperature at the wrist is clinically indicated.
The Empatica Health Monitoring Platform is not intended for Respiratory Rate monitoring in motion conditions. This device does not detect apnea and should not be used for detecting or monitoring cessation of breathing.
The Empatica Health Monitoring Platform is not intended for Pulse Rate monitoring in patients with chronic cardiac arrhythmias, including atrial fibrillation and atrial/ventricular bigeminy and trigeminy. and is not intended to diagnose or analyze cardiac arrhythmias. The Empatica Health Monitoring Platform is not a substitute for an ECG monitor, and should not be used as the sole basis for clinical decision-making.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
{3}------------------------------------------------
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Empatica Health Monitoring Platform
### 510(k) Summary
#### l. SUBMITTER
| Company Name | Empatica Srl |
|-----------------------------------|---------------------------------------------------------|
| Establishment Registration Number | 3012933969 |
| Contact Person | Alberto Poli, Director, Quality & Regulatory Compliance |
| Contact Person email | apo@empatica.com |
| Address | Via Stendhal, 36 - 20144, Milan, Italy |
| Telephone Number | +39 02 36165068 |
| Date prepared | September 27, 2023 |
#### II. DEVICE
Trade/Proprietary Name: Common/Usual Name:
Empatica Health Monitoring Platform Remote Patient Monitoring System
### Primary Product Code:
| Classification Regulation | Classification Name | Device Class | Product Code | Classification Panel |
|---------------------------|--------------------------------------------------------------------------|--------------|--------------|----------------------|
| 870.2300 | Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) | Class II | MWI | Cardiovascular |
#### Secondary Product Codes:
| Classificatio<br>n Regulation | Classification Name | Device<br>Class | Produc<br>t Code | Classification<br>Panel |
|-------------------------------|---------------------------------------------------------------------|-----------------|------------------|-------------------------|
| 870.2700 | Oximeter | Class II | DQA | Cardiovascular |
| 868.2375 | Monitor, Breathing Frequency | Class II | BZQ | Anesthesiology |
| 870.2910 | Transmitters and Receivers, Physiological<br>Signal, Radiofrequency | Class II | DRG | Cardiovascular |
| 882.5050 | Device, Sleep Assessment | Class II | LEL | Neurology |
| 882.1540 | Galvanic skin response measurement<br>device | Class II | GZO | Neurology |
| 880.2910 | Thermometer, Electronic, Clinical | Class II | FLL | General<br>Hospital |
#### PREDICATE DEVICES lll.
| Predicate<br>Device | Name | Submitter | Product Code(s) | 510(k) Number |
|---------------------|-------------------------------------------|----------------------|----------------------------------------|---------------|
| Primary | Current Health Monitoring<br>System Gen 2 | Spry Health,<br>Inc. | MSX<br>DQA<br>DRG<br>BZQ<br>BZG<br>FLL | K210133 |
| Secondary | Empatica Health Monitoring<br>Platform | Empatica S.r.l. | DQA<br>DRG<br>LEL<br>FLL<br>GZO | K221282 |
{5}------------------------------------------------
Empatica Health Monitoring Platform
| Predicate<br>Device | Name | Submitter | Product Code(s) | 510(k) Number |
|---------------------|-------------|----------------------|-----------------|---------------|
| Secondary | Loop System | Spry Health,<br>Inc. | DQA<br>BZQ | K181352 |
None of these predicates have been subject to a design-related recall.
#### REFERENCE DEVICE IV.
| Name | Submitter | Product Code(s) | 510(k) Number |
|-----------------|-------------------------------|-----------------|---------------|
| Everion+ System | Biofourmis Singapore Pte. Ltd | MWI, MSX, BZQ | K213863 |
#### DEVICE DESCRIPTION V.
The Empatica Health Monitoring Platform is a wearable device and software platform composed by:
- A wearable medical device called EmbracePlus,
- A mobile application running on smartphones called "Care App", ●
- A cloud-based software platform named "Care Portal". ●
The EmbracePlus is worn on the user's wrist and continuously collects raw data via specific sensors. These data are wirelessly transmitted via Bluetooth Low Energy to a paired mobile device where the Care App is up and running. The data received are analyzed by one of the Care App software modules, EmpaDSP, which computes the user physiological parameters. Based on the version of the Care App installed, the user can visualize a subset of these physiological parameters. The Care App is also responsible for transmitting, over cellular or Wi-Fi connection sensors' raw data, device information, Care App-specific information, and computed physiological parameters to the Empatica Cloud. On the Empatica Cloud, these data are stored, further analyzed, and accessible by healthcare providers or researchers via a specific cloud-based software called Care Portal.
The Empatica Health Monitoring Platform is intended for retrospective remote monitoring of physiological parameters in ambulatory adults in home-healthcare environments. It is designed to continuously collect data to support intermittent monitoring of the following physiological parameters and digital biomarkers by trained healthcare professionals or researchers: Pulse Rate (PR), Respiratory Rate (RR), blood oxygen saturation (SpO-), peripheral skin temperature (TEMP), and electrodermal activity (EDA). Activity sensors are used to detect sleep periods and to monitor the activity associated with movement during sleep.
#### INDICATION FOR USE vi.
The Empatica Health Monitoring Platform is a wearable device and paired mobile and cloud-based software platform intended to be used by trained healthcare professionals or researchers for retrospective remote monitoring of physiologic parameters in ambulatory individuals 18 years of age and older in home-healthcare environments. As the platform does not provide real-time alerts related to variation of physiologic parameters, users should use professional judgment in assessing patient clinical stability and the appropriateness of using a monitoring platform designed for retrospective review.
The device is intended for continuous data collection supporting intermittent retrospective review of the following physiological parameters:
• Pulse Rate,
- · Blood Oxygen Saturation under no-motion conditions,
- · Respiratory Rate under no motion conditions,
- · Peripheral Skin Temperature,
- · Electrodermal Activity,
- · Activity associated with movement during sleep
{6}------------------------------------------------
Empatica Health Monitoring Platform
The Empatica Health Monitoring Platform can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable.
The Empatica Health Monitoring Platform is not intended for SpO2 monitoring in conditions of motion or low perfusion.
The Empatica Health Monitoring Platform is intended for peripheral skin temperature monitoring, where monitoring temperature at the wrist is clinically indicated.
The Empatica Health Monitoring Platform is not intended for Respiratory Rate monitoring in motion conditions. This device does not detect apnea and should not be used for detecting or monitoring cessation of breathing.
The Empatica Health Monitoring Platform is not intended for Pulse Rate monitoring in patients with chronic cardiac arrhythmias, including atrial fibrillation and atrial/ventricular bigeminy, and is not intended to diagnose or analyze cardiac arrhythmias. The Empatica Health Monitoring Platform is not a substitute for an ECG monitor, and should not be used as the sole basis for clinical decision-making.
{7}------------------------------------------------
K230457
Empatica Health Monitoring Platform
#### I. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The application Enpatica Health Monitoring Platform is substantially equivate devices. The devices have similar Indications for Use, features, technology, and accuracy.
| Features | Empatica Health<br>Monitoring Platform<br>(Subject Device) | Current Health Monitoring<br>System Gen 2 (K210133) | Empatica Health Monitoring<br>Platform (K221282) | Loop System (K181352) | Analysis of differences |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Monitor, Physiological,<br>Patient (Without<br>Arrhythmia Detection Or<br>Alarms) | System, Network and<br>Communication,<br>Physiological<br>Monitors | Oximeter | Oximeter | N/A |
| Device<br>Manufacturer | Empatica S.r.l. | Current Health Ltd. | Empatica S.r.l. | Spry Health Ltd. | N/A |
| Device<br>Classification | II | ll | ll | ll | N/A |
| 510(k) number | N/A | K210133 | K221282 | K181352 | N/A |
| Primary<br>Product Code | MWI | MSX | DQA | DQA | N/A |
| Secondary<br>Product Code | DQA, BZQ, DRG, GZO, LEL,<br>FLL | FLL, DQA, BZQ, DRG,<br>BZG | DRG, GZO, LEL, FLL | BZQ | N/A |
| Intended<br>Use/Indications<br>for Use | The Empatica Health<br>Monitoring Platform is a<br>wearable device and<br>paired mobile and<br>cloud-based software<br>platform intended to be<br>used by trained<br>healthcare professionals<br>or researchers for<br>retrospective remote<br>monitoring of physiologic<br>parameters in ambulatory | The Current Wearable<br>Health Monitoring System is<br>intended for reusable<br>bedside, mobile and central<br>multi-parameter, physiologic<br>patient monitoring of adult<br>patients in professional<br>healthcare facilities, such as<br>hospitals or skilled nursing<br>facilities, or their own home.<br>It is intended for monitoring<br>of patients by trained | The Empatica Health<br>Monitoring Platform is a<br>wearable device and paired<br>mobile and cloud-based<br>software platform intended<br>to be used by trained<br>healthcare professionals or<br>researchers to remotely<br>monitor physiologic<br>parameters in ambulatory<br>individuals 18 years of age<br>and older in | The Loop System is<br>intended for adult<br>patients in the home<br>environment for<br>passive, noninvasive,<br>intermittent data<br>collection of<br>physiological<br>parameters that will<br>later be transmitted to a<br>web server for remote<br>review by a clinician. | The subject device<br>indication for use<br>includes the monitoring<br>of a subset of the<br>physiological parameters<br>monitored by the primary<br>predicate and additional<br>parameters compared<br>with the secondary<br>predicates. |
| Features | Empatica Health<br>Monitoring Platform<br>(Subject Device) | Current Health Monitoring<br>System Gen 2 (K210133) | Empatica Health Monitoring<br>Platform (K221282) | Loop System (K181352) | Analysis of differences |
| | individuals 18 years of age and older in home-healthcare environments. As the platform does not provide real-time alerts related to variation of physiologic parameters, users should use professional judgment in assessing patient clinical stability and the appropriateness of using a monitoring platform designed for retrospective review.<br>The device is intended for continuous data collection supporting intermittent retrospective review of the following physiological parameters:<br>• Pulse Rate,<br>• Blood Oxygen Saturation under no-motion conditions,<br>• Respiratory Rate under no motion conditions, | healthcare professionals.<br>The Current Wearable Health Monitoring System is intended to provide visual and audible physiologic multi-parameter alarms. The Current Wearable Health Monitoring System is intended for temperature monitoring where monitoring temperature at the upper arm is clinically indicated. The Current Wearable Health Monitoring System is intended for continuous monitoring of the following parameters in adults:<br>• Pulse rate<br>• Oxygen saturation<br>• Temperature<br>• Movement<br>The Current Wearable Health Monitoring System is intended for intermittent or spot-check monitoring, in adults, of:<br>• Respiration rate | home-healthcare environments.<br>The device supports the continuous monitoring of the following:<br>• Peripheral skin temperature,<br>• Electrodermal activity<br>• Blood Oxygen Saturation under no motion conditions<br>• Activity associated with movement during sleep<br>The Empatica Health Monitoring Platform can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable.<br>The Empatica Health Monitoring Platform is not intended for SpO2 monitoring in conditions of motion or low perfusion.<br>The Empatica Health Monitoring Platform is intended for peripheral skin temperature monitoring, | The Loop System measures and records:<br>• arterial oxygen saturation (SpO2) • heart rate (HR) • respiration rate (RR) All of these measurements are made when no motion is detected by the System. The Loop System device does not provide physiological alarms. | |
| Features | Empatica Health<br>Monitoring Platform<br>(Subject Device) | Current Health Monitoring<br>System Gen 2 (K210133) | Empatica Health Monitoring<br>Platform (K221282) | Loop System (K181352) | Analysis of differences |
| | Peripheral Skin<br>Temperature, Electrodermal Activity, Activity associated with<br>movement during sleep The Empatica Health<br>Monitoring Platform can<br>be used to analyze<br>circadian rhythms and<br>assess activity in any<br>instance where<br>quantifiable analysis of<br>physical motion is<br>desirable.<br>The Empatica Health<br>Monitoring Platform is<br>not intended for SpO2<br>monitoring in conditions<br>of motion or low<br>perfusion.<br>The Empatica Health<br>Monitoring Platform is<br>intended for peripheral<br>skin temperature<br>monitoring, where<br>monitoring temperature | Non-invasive blood<br>pressure Lung function &<br>spirometry Weight The Current Wearable<br>Health Monitoring System is<br>not intended for use in<br>high-acuity environments,<br>such as ICU or operating<br>rooms.<br>The Current Wearable<br>Health Monitoring System is<br>not intended for use on<br>acutely ill cardiac patients<br>with the potential to<br>develop life threatening<br>arrhythmias e.g. very fast<br>atrial fibrillation. For these<br>patients, they should be<br>monitored using a device<br>with continuous ECG. The<br>Current Wearable Health<br>Monitoring System is not a<br>substitute for an ECG<br>monitor.<br>The Current Wearable<br>Health Monitoring System is | where monitoring<br>temperature at the wrist is<br>clinically indicated. | | |
| Features | Empatica Health<br>Monitoring Platform<br>(Subject Device) | Current Health Monitoring<br>System Gen 2 (K210133) | Empatica Health Monitoring<br>Platform (K221282) | Loop System (K181352) | Analysis of differences |
| | at the wrist is clinically<br>indicated. | not intended for SpO2<br>monitoring in conditions of<br>high motion or<br>low perfusion | | | |
| | The Empatica Health<br>Monitoring Platform is<br>not intended for<br>Respiratory Rate<br>monitoring in motion<br>conditions. This device<br>does not detect apnea<br>and should not be used<br>for detecting or<br>monitoring cessation of<br>breathing. | | | | |
| | The Empatica Health<br>Monitoring Platform is<br>not intended for Pulse<br>Rate monitoring in<br>patients with chronic<br>cardiac arrhythmias,<br>including atrial fibrillation<br>and atrial/ventricular<br>bigeminy and trigeminy,<br>and is not intended to<br>diagnose or analyze<br>cardiac arrhythmias. The<br>Empatica Health<br>Monitoring Platform is | | | | |
| Features | Empatica Health<br>Monitoring Platform<br>(Subject Device) | Current Health Monitoring<br>System Gen 2 (K210133) | Empatica Health Monitoring<br>Platform (K221282) | Loop System (K181352) | Analysis of differences |
| | not a substitute for an<br>ECG monitor, and should<br>not be used as the sole<br>basis<br>for<br>clinical<br>decision-making. | | | | |
| Target<br>Population | Adult | Adult | Adult | Adult | The subject device and<br>the predicates are<br>identical |
| Anatomical Site | Wrist | Upper Arm | Wrist | Wrist | Clinical testing<br>demonstrated the<br>equivalence between the<br>subject device and the<br>predicates. The<br>difference in wearing<br>location<br>on the body<br>does not raise new<br>questions of<br>safety or efficacy. |
| Over the<br>Counter or Rx | Rx | Rx | Rx | Rx | The subject device and<br>the predicates are<br>identical |
| Environment | Home | Professional Healthcare<br>Facilities & Home | Home | Home | The subject device<br>includes a subgroup of<br>the predicates, hence this<br>does not raise new<br>questions of safety or<br>efficacy. |
| Features | Empatica Health<br>Monitoring Platform<br>(Subject Device) | Current Health Monitoring<br>System Gen 2 (K210133) | Empatica Health Monitoring<br>Platform (K221282) | Loop System (K181352) | Analysis of differences |
| Alarms | No…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.