TENS & EMS Device (Model: LY-ET-01, LY-ET-02, LY-ET-04): - To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. - To stimulate healthy muscles in order to improve and facilitate muscle performance. - To temporarily increase local blood circulation in healthy muscles.
Device Story
TENS & EMS Device provides electrical pulse stimulation to treatment areas via electrode pads; intended for OTC pain management and muscle stimulation. Device features 1-3 channels (model dependent) and 3-5 stimulation modes. User operates device via ON/OFF knob, mode selection button, intensity dial/buttons, and time selection button. Device transforms electrical energy into pulsed, symmetric, biphasic waveforms; output parameters (frequency, intensity, pulse duration) are controlled by internal firmware. Healthcare providers or patients use device to manage chronic/intractable pain or muscle soreness; output affects nerve/muscle tissue to provide symptomatic relief, improve muscle performance, and increase local blood circulation. Benefits include non-pharmacological pain management and muscle conditioning.
Clinical Evidence
No clinical data. Bench testing only. Device performance validated via electrical safety (IEC 60601-1, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), biocompatibility (ISO 10993-1, -5, -10), and usability (IEC 62366-1, IEC 60601-1-6).
Technological Characteristics
Microprocessor-controlled TENS/EMS stimulator; DC 3.0V (2x AAA) power source. Delivers pulsed, symmetric, biphasic rectangular waveforms with interphase interval. Features 1 channel; intensity levels (5-15); frequency 1-136 Hz; pulse duration 200µs. Materials compliant with ISO 10993. Type BF applied part. Connectivity: None (standalone). Sterilization: Not specified.
Indications for Use
Indicated for temporary relief of pain associated with sore/aching muscles in back, neck, upper/lower extremities due to exercise or household strain; symptomatic relief/management of chronic, intractable pain; relief of pain associated with arthritis; stimulation of healthy muscles to improve/facilitate performance; and temporary increase of local blood circulation in healthy muscles.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
May 22, 2023
Jiangxi Royall Smart Technology Co., Ltd. % Cassie Lee Manager Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road Huangpu District Guangzhou, Guangdong China
Re: K230443
Trade/Device Name: TENS & EMS Device (LY-ET-01, LY-ET-02, LY-ET-04) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX, NYN Dated: February 21, 2023 Received: February 21, 2023
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Robert Kang -S
for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230443
Device Name
TENS & EMS Device (Model: LY-ET-01, LY-ET-02, LY-ET-04)
Indications for Use (Describe)
TENS & EMS Device (Model: LY-ET-01, LY-ET-02, LY-ET-04):
- To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
- To stimulate healthy muscles in order to improve and facilitate muscle performance .
- To temporarily increase local blood circulation in healthy muscles.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of K230443
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 801.92.
#### 1. Submitter's Information
Sponsor Name: Jiangxi Royall Smart Technology Co., Ltd. Establishment Registration Number: Applying Address: Workshop 3#, Shangyou Lefeng Technology Co., Ltd., Huangbu Town, Ganzhou City, Jiangxi Province Postal code: 341200 Contact Person (including title): Shunzhou Yang (Head of Quality Division) Tel: +86 13421375258 Fax: / E-mail: shunzhou.yang@lefengmotors.com
#### Application Correspondent:
Contact Person: Ms. Cassie Lee Share Info (Guangzhou) Medical Consultant Ltd. Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Tel: +86 20 8266 2446 Email: regulatory@share-info.com
#### 2. Date of the summary prepared: May 18, 2023
#### 3. Subject Device Information
Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Trade Name: TENS & EMS Device Model Name: LY-ET-01, LY-ET-02, LY-ET-04 510(K) Number: K230443 Review Panel: Neurology Product Code: Primary product code: NUH Secondary product code: NGX, NYN Regulation Number: 882.5890 Regulatory Class: II
{4}------------------------------------------------
#### 4. Predicate Device Information
Predicate Device: Sponsor: Guangzhou Xinbo Electronic Co., Ltd. Trade Name: Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I Common Name: TENS, EMS, Stimulator for pain relief Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter 510(K) Number: K163611 Review Panel: Neurology Product Code: Primary product code: NUH Secondary product code: NGX, NYN Requlation Number: 21 CRF 882.5890
Requlation Class: II
#### Reference device:
Sponsor: Counter Scientific Development (GZ) Ltd Trade/Device Name: Pain Therapy System, Model PTS-II Common Name: TENS, EMS, Stimulator for pain relief Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter 510(K) Number: K150277 Review Panel: Neurology Product Code: Primary product code: NUH Secondary product code: NGX, NYN GZJ Regulation Number: 21 CRF 882.5890 Regulatory Class: II
#### 5. Device Description
TENS & EMS Device is a pain management device with a combination of TENS and EMS therapy technology.
It can stimulate healthy muscles in order to improve and facilitate muscle performance, and temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve.
TENS & EMS Device (Model: LY-ET-01, LY-ET-04) has 3 modes (Model LY-ET-02 has 5 modes) and one channel, which provide electrical pulse stimulation through the electrode pads to the treatment area. The
{5}------------------------------------------------
TENS & EMS Device has the operating elements of ON/OFF knob, Mode Selection button, Intensity adjustment Dial (or Intensity Increase button and Intensity decrease button) and Time Selection button which is user-friendly and easy to control.
The device is equipped with electrode pads and electrode wire is used to connect the pads to the main unit. All the accessories can only be purchased by a local distributor.
#### 6. Intended Use / Indications for Use
TENS & EMS Device (Model: LY-ET-01, LY-ET-02, LY-ET-04):
- To be used for temporary relief of pain associated with sore and aching muscles in the upper and । lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
- To stimulate healthy muscles in order to improve and facilitate muscle performance. -
- -To temporarily increase local blood circulation in healthy muscles.
#### 7. Comparison to predicate device and conclusion
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Reference device | Remark | |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Company | Jiangxi Royall Smart<br>Technology Co., Ltd. | Guangzhou Xinbo<br>Electronic Co., Ltd. | Counter Scientific<br>Development (GZ) Ltd | -- | |
| Trade Name | TENS & EMS Device | Pain Therapy Device, | Pain Therapy System | -- | |
| 510(k) Number | K230443 | K163611 | K150277 | -- | |
| Classification<br>Name | Transcutaneous<br>Electrical Nerve<br>Stimulator For Pain<br>Relief, Over The<br>Counter | Transcutaneous<br>Electrical Nerve<br>Stimulator For Pain<br>Relief, Over The<br>Counter | Transcutaneous<br>Electrical Nerve<br>Stimulator For Pain<br>Relief, Over The<br>Counter | Same | |
| Classification<br>Product Code | NUH | NUH | NUH | Same | |
| Subsequent<br>Product Code | NYN, NGX | NYN, NGX | GZJ, NGX, NYN | Same | |
| Elements of | | Subject Device | Predicate Device | Reference device | Remark |
| Comparison | | | | | |
| Intended Use /<br>Indications for Use | | TENS & EMS Device (<br>Model: LY-ET-01, LY-E<br>T-02, LY-ET-04):<br>- To be used for tempor<br>ary relief of pain associ<br>ated with sore and achi<br>ng muscles in the upper<br>and lower back, back o<br>f the neck, upper extre<br>mities (shoulder and ar<br>m), lower extremities (le<br>g and feet) due to strain<br>from exercise or norma<br>l household work activiti<br>es by applying current t<br>o stimulate nerve. To be<br>used for symptomatic r<br>elief and management o<br>f chronic, intractable pai<br>n and relief of pain asso<br>ciated with arthritis.<br>- To stimulate healthy m<br>uscles in order to impro<br>ve and facilitate muscle<br>performance .<br>- To temporarily increas<br>e local blood circulation<br>in healthy muscles. | To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the upper and lower<br>back, back of the neck,<br>upper extremities<br>(shoulder and arm),<br>lower extremities (leg<br>and feet) due to strain<br>from exercise or normal<br>household work<br>activities by applying<br>current to stimulate<br>nerve. To be used for<br>symptomatic relief and<br>management of chronic,<br>intractable pain and<br>relief of pain associated<br>with arthritis (Choose<br>Mode B or C).<br>To stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance<br>(Choose Mode A).<br>To temporarily increase<br>local blood circulation in<br>healthy muscles<br>(Choose Mode A). | To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the upper and lower<br>back due to strain from<br>exercise or normal<br>household work<br>activities (Choose<br>Mode A, B, or C)<br>To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the upper extremities<br>(Arms) due to strain<br>from exercise or normal<br>household work<br>activities (Choose<br>Mode A, B or C)<br>To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the lower extremities<br>(Legs) due to strain<br>from exercise or normal<br>household work<br>activities (Choose<br>Mode A, B or C)<br>To be used for<br>symptomatic relief and<br>management of<br>chronic, intractable pain<br>and relief of pain<br>associated with Arthritis<br>(Choose Mode A)<br>To stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance<br>(Choose Mode B) | Same |
| Regulatory Class | | Class II | Class II | Class II | Same |
| OTC or Rx | | OTC | OTC | OTC | Same |
| Power Source(s) | | DC 3.0V, 2 x AAA | DC 3.0V, 2 x AAA | DC 3.0V, 2 x AAA | Same |
| Method of Line<br>Current Isolation | | Type BF Applied Part | Type BF Applied Part | Use resistance to<br>isolate | Same |
| Patient<br>Leakage<br>Current | NC<br>SFC | DC: <1 μA | DC: 0.5 μA | N/A | Similar<br>Note 1 |
| | | DC: <1 μA | DC: 0.6 μA | N/A | |
| Average DC | | < 0.01 μA | < 0.01 μA | 0 μA | Same |
| Elements of<br>Comparison | Subject Device | Predicate Device | Reference device | Remark | |
| current through<br>electrodes when<br>device is on but no<br>pulses are being<br>applied | | | | | |
| Number of Output<br>Channels | 1 | 2 Channels: for models<br>P.T.S-II,<br>P.T.S-IIA, P.T.S-IIB;<br>1 Channel: for model<br>CP-I | 2 Synchronous | Similar<br>Note 2 | |
| Output Intensity<br>Level | LY-ET-01: 5 intensity<br>levels<br>LY-ET-02: 5 intensity<br>levels<br>LY-ET-04: 15 intensity<br>levels | 5 steps | Not published | Different<br>Note 2 | |
| Synchronous or<br>Alternating? | Synchronous | Synchronous | Synchronous | Same | |
| Method of<br>Channel Isolation | Parallel connection | Parallel connection | Parallel connection | Same | |
| Software/Firmware<br>/Microprocessor<br>Control? | Yes | Yes | Yes | Same | |
| Automatic<br>Overload Trip | No | No…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.