Tigertriever 21 Revascularization Device, Tigertriever 17 Revascularization Device, Tigertriever 13 Revascularization Device

K230429 · Rapid Medical Ltd. · NRY · Apr 17, 2023 · Cardiovascular

Device Facts

Record IDK230429
Device NameTigertriever 21 Revascularization Device, Tigertriever 17 Revascularization Device, Tigertriever 13 Revascularization Device
ApplicantRapid Medical Ltd.
Product CodeNRY · Cardiovascular
Decision DateApr 17, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.

Device Story

Stentriever device for mechanical thrombectomy in neurovasculature; comprises adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire, and handle. Physician operates device under fluoroscopic guidance; pulls handle slider to expand mesh, allowing it to conform to vessel diameter and encapsulate thrombus. Device delivered via microcatheter to target site; used by trained neurointerventionalists in clinical settings. Output is physical clot removal and restoration of blood flow. Benefits include improved patient outcomes in acute ischemic stroke by facilitating rapid revascularization.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing only, including simulated use in tortuous anatomical models, durability testing (3 passes), tensile strength, and particulate generation analysis.

Technological Characteristics

Stentriever with adjustable nitinol braided mesh; stainless steel shaft; nitinol core wire. Mesh wires: Niti DFT with tantalum or platinum core. Dimensions: 20.5mm-32mm length; 2.5mm-6mm expanded OD. Manual expansion via handle slider. Sterilized via ethylene oxide. Single-use. Compliant with ISO 10555-1 for tensile strength.

Indications for Use

Indicated for patients experiencing ischemic stroke within 8 hours of symptom onset, including those ineligible for or failing IV t-PA therapy, requiring thrombus removal from large intracranial vessels.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ April 17, 2023 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. Rapid Medical Ltd. Orit Yaniv, Ph.D. VP RA/QA Carmel Building, P.O. Box 337 Yokneam, 2069205 Israel Re: K230429 Trade/Device Name: Tigertriever 21 Revascularization Device, Tigertriever 17 Revascularization Device, Tigertriever 13 Revascularization Device Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: March 20, 2023 Received: March 20, 2023 Dear Orit Yaniv: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Naira Muradyan -S Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K230429 #### Device Name Tigertriever 21 Revascularization Device, Tigertriever 17 Revascularization Device, Tigettriever 13 Revascularization Device #### Indications for Use (Describe) The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment. Type of Use (Select one or both, as applicable) | <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |-------------------------------------------------------------|------------------------------------------------------------| |-------------------------------------------------------------|------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Rapid Medical. The logo consists of a green abstract symbol on the left and the words "Rapid Medical" in bold, dark gray font on the right. The abstract symbol appears to be three curved lines of varying shades of green, arranged in a way that suggests movement or flow. ## 510(k) Summary #### K230429 | Manufacturer/Sponsor: | Rapid Medical Ltd.<br>Carmel Building, P.O. Box 337<br>Yokneam, 2069205 Israel | |-----------------------|---------------------------------------------------------------------------------------------------------------------------------| | | Phone: +972-72-250-3331<br>Facsimile: +972-72-250-3332 | | Contact: | Orit Yaniv, Ph.D.<br>VP RA/QA<br>+972-72-250-3331<br>Orit@rapid-medical.com | | Date Prepared: | April 12, 2023 | | Device Trade Name: | Tigertriever 21 Revascularization Device, Tigertriever 17<br>Revascularization Device, Tigertriever 13 Revascularization Device | | Common/Usual Name: | Catheter, Thrombus Retriever | | Classification: | 21 CFR 870.1250, Percutaneous Catheter | | Class: | II | | Product Code: | NRY | | Predicate Devices: | Tigertriever and Tigertriever 17 Revascularization Device (K203592)<br>Tigertriever 13 Revascularization Device (K220808) | #### Indications for Use: The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment. #### Device Description: The Tigertriever device is a stentriever that is comprised of an adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire and a handle. The shaft connects the mesh and the handle by the core wire that runs inside the shaft from the distal end of the mesh to the slider activation element in the handle. The mesh is expanded when the physician pulls the slider, since the wires of the mesh are completely radiopaque, the physician sees the mesh under fluoroscopy and controls it until it conforms to the vessel diameter. The design of the wire mesh is optimized to penetrate the clot and encapsulate it during retrieval. The Tigertriever Revascularization Device is supplied sterile and is intended for single-use only by physicians trained in neurointerventional procedures and the treatment {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Rapid Medical. The logo consists of a green abstract symbol on the left, followed by the words "Rapid Medical" in bold, dark gray font. The abstract symbol appears to be three curved lines that converge to a point on the right. of ischemic stroke. ## Comparison of Technological Characteristics with the Predicate Devices: The Tigertriever 21, Tigertriever 17, Tigertriever 13 Revascularization Devices subject to this submission are substantially equivalent to the predicate devices, Tigertriever and Tigertriever 17 Revascularization Device (K203592) and Tigertriever 13 Revascularization Device (K220808), based on the same indications for use, device design, materials, manufacturing, packaging and sterilization methods, and similar technological characteristics. A comparison of the subject device with the predicate devices is summarized in Table 1 below. | Predicate devices | | | Subject devices | | | | |--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | Device Name | Tigertriever TRPP7155 | Tigertriever 17 TRPP7166 | Tigertriever 13 TRPP7144 | Tigertriever 21 TRPP7155 | Tigertriever 17 TRPP7166 | Tigertriever 13 TRPP7144 | | 510(k) Number | K203592 | | K220808 | K230429 | | | | Regulation No. | 21 CFR 870.1250 | | | Same | | | | Regulation Name | Percutaneous Catheter | | | Same | | | | Classification | Class II | | | Same | | | | Product Code | NRY | | | Same | | | | Indications for Use | The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment. | | | Same | | | | Anatomical Location | Neurovasculature | | | Same | | | | <b>Technological Characteristics</b> | | | | | | | | Mode of Operation | Manual expansion of the braided distal portion into the clot using the handle component | | | Same | | | | Design of Distal Portion | Close end braided nitinol mesh, manually expandable | | | Same | | | | Stent Length (un-expanded configuration) | 32 mm | 23 mm | 20.5 mm | 32 mm | 23 mm | 20.5 mm | | Stent Size Distal OD (Unexpanded and Expanded Configuration) | Unexpanded configuration 1.5 mm<br><br>Expanded configuration 6 mm | Unexpanded configuration 0.5 mm<br><br>Expanded configuration 3 mm | Unexpanded configuration 0.5 mm<br><br>Expanded configuration 2.5 mm | Unexpanded configuration 1.5 mm<br><br>Expanded configuration 6 mm | Unexpanded configuration 0.5 mm<br><br>Expanded configuration 3 mm | Unexpanded configuration 0.5 mm<br><br>Expanded configuration 2.5 mm | | | | | | | | | | Stent Structure | Braided from<br>twelve (12)<br>Niti DFT<br>0.075 mm<br>wires with a<br>tantalum core | Braided from<br>eight (8) Niti<br>DFT 0.075 mm<br>wires with a<br>tantalum core | Braided from<br>eight (8) Niti<br>DFT 0.05 mm<br>wires with a<br>platinum core | Braided from<br>twelve (12) Niti<br>DFT 0.075 mm<br>wires with a<br>tantalum core | Braided from<br>eight (8) Niti<br>DFT 0.075 mm<br>wires with a<br>tantalum core | Braided from<br>eight (8) Niti<br>DFT 0.05 mm<br>wires with a<br>platinum<br>core | | Overall Length<br>(shaft+cable+<br>mesh+tip) | 212 cm | 208 cm | 228 cm | 218 cm | 218 cm | 228 cm | | Fluorosafe<br>Markers on Shaft | No | | Yes | | | | | Compatibility | Microcatheter<br>with an internal<br>diameter of<br>0.021 inches | Microcatheter<br>with an internal<br>diameter of<br>0.017 inches | Microcatheter<br>with an internal<br>diameter of<br>0.0165 inches | Microcatheter<br>with an internal<br>diameter of<br>0.021 inches | Microcatheter<br>with an internal<br>diameter of<br>0.0165 inches | Microcatheter<br>with an internal<br>diameter of<br>0.0165 inches | | Materials | | | | | | | | Stent (mesh) | Nitinol | | Same | | | | | Markers | 90% Platinum/ 10% Iridium | | Same | | | | | Core wire (shaft) | Nitinol core wire and 304 Stainless Steel shaft | | Same | | | | | Introducer sheath | PTFE | | Same | | | | | Packaging | | | | | | | | Sterilization<br>Method | Ethylene oxide | | Same | | | | | Single Use | Yes | | Same | | | | | Packaging | Placed into a Dispenser hoop, blister, Tyvek pouch,<br>and Carton box | | Same | | | | #### Table 1: Substantial Equivalence - Predicate Comparison {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Rapid Medical. The logo consists of a green abstract symbol on the left, followed by the words "Rapid Medical" in a bold, dark gray font. The symbol appears to be three curved lines that converge to a point on the right. The differences between the subject and predicate devices do not raise new questions of safety and effectiveness and are evaluated through the nonclinical testing referenced below. ## Nonclinical Performance Testing: # Bench Tests: ## Table2: Bench testing | Performance Bench Testing | | | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Test | Test Method Summary | Conclusions | | Simulated use test | Simulated use testing of the Tigertriever<br>Revascularization Device was performed<br>in an anatomical model which simulated<br>the tortuosity of the neurovasculature.<br>Devices were delivered through the<br>tortuous anatomical model to evaluate<br>the effectiveness of the device at<br>retrieval of firm and soft clots. | The device was tested for<br>handling and clot retrieval in an <i>in vitro</i> tortuous path anatomical<br>model, which has been used in the<br>evaluation of the predicate device.<br>The subject device effectively<br>retrieved clot and restored flow in<br>the test model. | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Rapid Medical. The logo consists of a green graphic on the left and the words "Rapid Medical" in bold, dark gray font on the right. The graphic is made up of three curved lines that converge to a point on the right. | Durability | Damage was evaluated after delivery and<br>withdrawal of the device beyond the<br>recommended number of passes and<br>resheathings recommended in the<br>instructions for use. | Devices tested demonstrated no<br>damage after delivery and<br>withdrawal testing. Durability<br>established acceptable<br>performance for 3 passes, which<br>is at least equivalent to the<br>number of passes specified in the<br>predicate labeling (2 passes per<br>device). | |---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Delivery, deployment<br>and retrieval | The delivery, deployment and retrieval<br>forces were measured during simulated<br>use of the subject device. | The device was tested for<br>delivery, deployment, and<br>retrieval in an in vitro tortuous<br>path anatomical model, which has<br>been used in the evaluation of the<br>predicate device. The subject<br>device demonstrated acceptable<br>performance with respect to<br>delivery, deployment and<br>retrieval. | | Dimensions test | Dimensional conformance to<br>specifications was confirmed. | The subject device dimensions are<br>within the range of existing<br>predicate device dimensions. The<br>minor differences in dimensions<br>do not affect performance, safety<br>or effectiveness. | | Tensile test | The minimum force to break the subject<br>device was tested for the handle joint. | The tensile strength of the device<br>met acceptance criteria based on<br>recognized standards (ISO 10555-<br>1). | | Particulate test | Particulate test was performed according<br>to the light obscuration test method. | The particulate generated by the<br>subject device was similar to the<br>particulate generated by the<br>predicate device. | ## Biocompatibility Testing: Biocompatibility tests were repeated with a representative Tigertriever device that contains the new fluorosafe markers and shares the same overall design and materials as the three subject device models and represents the worst case in terms of biocompatibility. All biocompatibility tests met the acceptance criteria. #### Table3: Biocompatibility testing results and methodology | Biological Endpoint | Test Results | Conclusion | |----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------| | Cytotoxicity | Pass. Grade 0 reactivity observed<br>48 hours post exposure to test<br>article extract. | Non-cytotoxic | | Irritation (Intracutaneous<br>Reactivity) | Pass. Difference of overall mean<br>score between test article and<br>control was 0. | Non-irritant | | Sensitization<br>(Guinea Pig Maximization<br>Test) | Pass. Grade 0, no evidence of<br>causing delayed dermal contact<br>sensitization. | Does not elicit sensitization response | | Hemocompatibility -<br>Complement activation Assay | SC5b-9 concentration of the test<br>article sample was statistically<br>less than the positive control and<br>was not statistically higher than the | Pass | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Rapid Medical. The logo consists of a green abstract symbol on the left and the words "Rapid Medical" in black on the right. The abstract symbol appears to be three curved lines that converge. | | negative control. | | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------| | Hemocompatibility -<br>In Vitro Hemolysis | Both the test article in direct<br>contact with blood and the test<br>article extract were non-<br>hemolytic. | Non-hemolytic | | Pyrogenicity (Material<br>Mediated Pyrogenicity) | The total rise of rabbit<br>temperatures during the 3-hour<br>observation period was within<br>acceptable USP limits. | Non-pyrogenic | | ISO Systemic Toxicity<br>Testing Study in Mice | No mortality or evidence of<br>systemic toxicity from the<br>extracts injected into mice. | Non-toxic | #### Sterilization and Shelf Life: The modifications to the Tigertriever Revascularization Device do not impact sterilization and established shelf-life of the product. ## Clinical Testing: A clinical study was not deemed necessary to evaluate the modifications to the Tigertriever Revascularization Device. Substantial equivalence of the subject devices has been established to the predicate devices through the results of nonclinical testing. ## Conclusion: The subject Tigertriever Revascularization Device has the same intended use, indications for use, principles of operation, and similar technological characteristics as the predicate devices. The technological differences of the subject Tigertriever Revascularization Device do not raise any new questions of safety or effectiveness. Performance data demonstrate that the subject Tigertriever Revascularization Device is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...