K230418 · Merit Medical Systems, Inc. · DQX · Oct 28, 2023 · Cardiovascular
Device Facts
Record ID
K230418
Device Name
Mighty Wire Guide Wire
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Oct 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Merit Mighty Wire is intended to be used to facilitate the placement of devices during diagnostic and interventional procedures in the central circulatory system, excluding the coronary arteries and neurovasculature.
Device Story
The Merit Mighty Wire is a guide wire used to facilitate the placement of diagnostic and interventional devices within the central circulatory system. It consists of a PTFE-coated stainless-steel core wire with a grind profile and a flexible distal tip. A PTFE-coated stainless-steel distal coil is laser-welded to the proximal end and plasma arc-welded to the distal end of the core wire. For longer lengths (230, 260, 300cm), a radiopaque Platinum-Tungsten marker coil is added to the distal tip. The device is used by clinicians in a clinical setting to navigate vessels and guide catheters or other interventional tools. The radiopaque tip allows for visualization under fluoroscopy, assisting the physician in precise positioning. By providing a stable track, the wire enables the safe delivery of therapeutic or diagnostic devices, potentially improving procedural outcomes in peripheral vascular interventions.
Clinical Evidence
Bench testing only. No clinical data was provided. Performance testing included dimensional verification, surface finish, particulate evaluation, coating lubricity, fracture/flexing tests, tensile/torque strength, corrosion resistance, kink resistance, and tip flexibility. Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, genotoxicity, and hemocompatibility (hemolysis, thrombogenicity, complement activation). All results met predetermined acceptance criteria.
Indicated for patients requiring device placement during diagnostic and interventional procedures in the central circulatory system, excluding coronary arteries and neurovasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 28, 2023
Merit Medical Systems, Inc. % James Kenny Senior Regulatory Affairs Specialist Merit Medical Ireland, Ltd. Parkmore Business Park West Galway, Ireland
Re: K230418
Trade/Device Name: Mighty Wire Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: September 29, 2023 Received: September 29, 2023
Dear James Kenny:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brian D. Pullin -S
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230418
Device Name Mighty Wire Guide Wire
Indications for Use (Describe)
The Merit Mighty Wire is intended to be used to facilitate the placement of devices during diagnostic and interventional procedures in the central circulatory system, excluding the coronary arteries and neurovasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K230418
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(+353) 91 703798<br>(+353) 91 680104<br>James Kenny<br>1721504 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| | Correspondent Name:<br>Address: | Merit Medical Ireland Ltd.<br>Parkmore Business Park<br>Parkmore, Galway, Ireland |
| | Telephone Number: | (+353) 91 703798 |
| | Fax Number:<br>Contact Person: | (+353) 91 680104<br>James Kenny |
| | Date of Preparation: | 14 February 2023 |
| | Registration Number: | 9616662 |
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Class:<br>Product code:<br>Classification Name:<br>Regulation Number: | Mighty Wire Guide Wire<br>Guide Wire<br>II<br>DQX<br>Wire, Guide, Catheter<br>21 CFR 870.1330 |
| | Regulation Medical Specialty Cardiovascular | |
| Predicate<br>Device | Trade Name:<br>Class:<br>Product code:<br>Classification Name:<br>Regulation Number:<br>Regulation Medical Specialty Cardiovascular<br>Premarket Notification:<br>Manufacturer: | Lunderquist Extra-Stiff Wire Guide<br>II<br>DQX<br>Wire, Guide, Catheter<br>21 CFR 870.1330<br>Cardiovascular<br>K220137<br>Cook Medical |
| | | |
# 510(k) Summary
{4}------------------------------------------------
| | 510(k) Summary<br>1 |
|--|---------------------|
|--|---------------------|
| | The Merit Mighty Wire Guide Wire family all have 0.035" outer<br>diameter with length configurations available in 90, 145, 180, 230,<br>260, and 300cm. The distal tip configurations are Straight, Curved<br>(7.5mm J Radius), Double Curved (75mm/15mm Radius), and<br>Extended Double Curved (55mm/15mm Radius). |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description | The wire is composed of a PTFE (polytetrafluoroethylene) coated<br>stainless-steel core wire with a grind profile ending in a flexible<br>distal tip. A PTFE coated stainless steel distal coil is laser-welded<br>to the proximal end of the core wire grind profile and plasma arc-<br>welded to the distal end of the core wire profile. The PTFE coating<br>covers 100% of the guide wire surface. For wire lengths 230, 260,<br>and 300cm, an additional radiopaque Platinum-Tungsten marker<br>coil is welded to the core wire flexible distal tip. The distal tip is<br>radiopaque. |
| Intended Use | The Merit Mighty Wire is intended to be used to facilitate the<br>placement of devices during diagnostic and interventional<br>procedures in the central circulatory system, excluding the<br>coronary arteries and neurovasculature. |
{5}------------------------------------------------
The technological characteristics of the subject Merit Mighty Wire Guide Wire are substantially equivalent to those of Predicate Device, Lunderquist Extra-Stiff Wire Guide [K220137]. The Merit Mightv Wire Guide Wire was developed based on a full characterization of the Lunderquist Extra-Stiff Wire Guide: The wire is composed of a PTFE (polytetrafluoroethylene) coated stainless-steel core wire with a grind profile ending in a flexible distal tip. A PTFE coated stainless steel distal coil is laser-welded to the proximal end of the core wire grind profile and plasma arcwelded to the distal end of the core wire profile. The PTFE coating covers 100% of the quide wire surface. For wire lengths 230. 260. and 300cm, an additional radiopaque Platinum-Tungsten marker coil is welded to the core wire flexible distal tip. The distal tip is radiopaque. The subject device has the same basic design as the predicate. The Comparison Lunderquist Extra-Stiff Wire Guide is comprised of a PTFE coated to Predicate stainless-steel core wire and a PTFE coated stainless-steel coil. The PTFE coated coil is welded to the core on the very distal tip. Wires with lengths 260 and 300cm include an inner gold radiopaque coil. The fundamental technology and operating principles of the subject and the predicate are the same. While the indications for use wordings are not identical, both are intended to facilitate the placement of devices during diagnostic and interventional procedures. The Mighty Wire is for use in the "central circulatory system, excluding the coronary arteries and neurovasculature". The Lunderquist Extra-Stiff Wire Guide is for use in the "major vessels, the aorta and vena cava, including their access vessels and adjacent vessels". The intended use is the same for both the subject Mighty Wire Guide Wire and predicate Lunderquist Extra-Stiff Wire Guide, with both used for the placement and exchange of devices during diagnostic and interventional procedures. No performance standards have been established under section 514 of the Food, Drug and Cosmetic Act for these devices. A battery of testing was conducted, on the Merit Mighty Wire Guide Wire, in accordance with protocols based on requirements outlined in Safety & quidance's and industry standards and these were shown to meet Performance the acceptance criteria that were determined to demonstrate Tests substantial equivalence.
Where appropriate, the tests were based on the requirements of the
{6}------------------------------------------------
following documents:
- FDA Guidance "Coronary, Peripheral, and Neurovascular ● Performance Tests Guidewires and Recommended Labelling" - October 2019
- FDA Guidance "Intravascular Catheters, Wires, and ● Delivery Systems with Lubricious Coatings - Labeling Considerations" - October 2019
- FDA Guidance "Submission and Review of Sterility ● Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile" - January 2016
- FDA Guidance - "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" -September 2020
- ISO 11070:2014/Amd.1:2018, Sterile single-use intravascular introducers, dilators and guidewires
- ISO 11135:2014/Amd 1:2018, Sterilization of health-care ● products - Ethylene oxide: Requirements for the development, validation and routine control of a sterilization process for medical devices
- ISO 10993-1:2018, Biological evaluation of medical devices ● - Part 1: Evaluation and testing within a risk management process
- ISO 11607-1:2019, Packaging for terminally sterilized ● medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ISO 11737-1:2018/Amd 1:2021, Sterilization of health care ● products - Microbiological methods - - Part 1: Determination of a population of microorganisms on products
- ISO 10993-4:2017. Biological evaluation of medical ● devices-Part 4: Selection of tests for interactions with blood
- ISO 10993-5:2009/(R)2014, Biological evaluation of medical ● devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021. Biological evaluation of medical ● devices - Part 10: Tests for skin sensitization
- ISO 10993-11:2017, Biological evaluation of ● medical devices - Part 11: Tests for systemic toxicity
{7}------------------------------------------------
- ISO 10993-12:2021, Biological evaluation of medical ● devices - Part 12: Sample preparation and reference materials
- ISO 10993-18:2020, Biological evaluation of medical ● devices – Part 18: Chemical characterization of materials
- . ISO 10993-19:2020, Biological evaluation of medical devices - Part 19: Physico-chemical, morphological, and topographical characterization of materials
- ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation
- . ASTM F2475-20, Standard Guide for Biocompatibility of Medical Device Packaging Materials
- ISO 10993-7:2008/Amd.1:2019(E), Biological evaluation of ● medical devices - Part 7: Ethylene oxide sterilization residuals
- ASTM F1980-21, Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices
- ASTM D4169-22, Standard Practice for Performance ● Testing of Shipping Containers and Systems
- ANSI ST72:2019, Bacterial endotoxins Test methods, ● routine monitoring, and alternatives to batch testing
- . ISO 2233:2000, Packaging - Complete, filled transport packages and unit labels - Conditioning for testing
- TIR42:2021, Evaluation of Particulates Associated with ● Vascular Medical Devices
- TIR28:2016, Product adoption and process equivalence for ethylene oxide sterilization
The following is a list of all testing that was successfully completed:
{8}------------------------------------------------
## Performance Non-Clinical Testing-Bench
- Dimensional Verification
- . Surface Finish
- Particulate Evaluation ●
- Coating Lubricity ●
- Fracture Test ●
- Flexing Test ●
- . Guidewire Tip Radiopacity
- Guidewire Joints Tensile Strength ●
- Guidewire Torque Strength ●
- Corrosion Resistance ●
- Tip Flexibility ●
- Kink Resistance ●
- Coating Integrity ●
- Ancillary Device Compatibility ●
- Guidewire Preparation ●
- . Guidewire Maneuverability in Catheter
- Tracking and Positioning Ancillary devices ●
- Guidewire Integrity
- J-Straightener Function ●
- Guidewire Body Stiffness ●
- Guidewire Distal Tip Flexibility ●
- Guidewire Trackability ●
- Guidewire Tip Shape Retention ●
### Biocompatibility
- Cytotoxicity ●
- Sensitization
- Irritation ●
- Acute Systemic Toxicity ●
- Pyrogenicity ●
- Genotoxicity ●
- Hemocompatibility ●
- o Hemolysis
- Thrombogenicity o
{9}------------------------------------------------
- o Complement Activation All test results were comparable to the predicate device and the subject Merit Mighty Wire Guide Wire met the predeterminded acceptance criteria.This demonstrated that the subject device is substantially equivalent to the predicate device. Based on the comparisons noted, the subject Merit Mighty Wire Summary of Guide Wire meets the requirements that are considered essential Substantial for its intended use and is substantively equivalent to the Predicate Device, the Lunderquist Extra-Stiff Wire Guide [K220137] Equivalence manufactured by Cook Medical.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.