K230408 · Sb-Kawasumi Laboratories, Inc. · JKA · Jun 1, 2023 · Clinical Chemistry
Device Facts
Record ID
K230408
Device Name
K-SHIELD Zen (Model Numbers: PBMH/ PBMA/ PBM)
Applicant
Sb-Kawasumi Laboratories, Inc.
Product Code
JKA · Clinical Chemistry
Decision Date
Jun 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Attributes
Pediatric
Indications for Use
K-SHIELD Zen is designed for accessing peripheral veins to collect blood. The Press Button anti-needle stick protector is an integral safety device intended to minimize accidental needlestick injuries.
Device Story
Sterile, single-use winged blood collection set; includes cannula, flexible tube, connection parts, and press-button anti-needle stick protector. Used in hospitals/clinics by healthcare professionals for peripheral venous access. User selects configuration (Luer connector, multiple sample Luer adapter, or pre-attached holder) based on procedure. After blood collection, pressing activation button retracts needle into housing to prevent needlestick injuries. Facilitates blood collection via syringes or vacuum tubes/bottles. Benefits include reduced risk of accidental needlestick injuries for clinicians.
Clinical Evidence
Bench testing only. No clinical data provided. Biocompatibility and chemical characterization testing (DEHP extraction, risk analysis, EtO residuals) performed per ISO 10993-7, 10993-17, and 10993-18 to support safety for pediatric use. All tests passed.
Technological Characteristics
Materials: Stainless steel needle, PP (wing, protector, hub), Isoprene rubber (sheath), PVC (tubing), ABS (Luer connector), heat-curing epoxy adhesive. Sterilization: Ethylene Oxide (ISO 11135:2014). Form factor: Winged set with 21G, 23G, 25G needle gauges. Connectivity: None. Software: None.
Indications for Use
Indicated for accessing peripheral veins to collect blood in adult and pediatric patient populations. Includes an integral safety device to minimize accidental needlestick injuries.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo features the Department of Health & Human Services logo on the left. To the right of that is the FDA logo in blue, with the words "U.S. Food & Drug Administration" written in blue text.
June 1, 2023
SB-Kawasumi Laboratories, Inc. % Valerie Followell Official Correspondent for SB-KAWASUMI LABORATORIES, INC. Regulatory Compliance Associates, Inc. (RCA) 10411 Corporate Drive, Suite 102 Pleasant Prairie, Wisconsin 53158
Re: K230408
Trade/Device Name: K-SHIELD Zen (Model Numbers: PBMH/ PBMA/ PBM) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: May 2, 2023 Received: May 3, 2023
Dear Valerie Followell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Page 2
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230408
Device Name
K-SHIELD Zen (Model Numbers: PBMH/ PBMA/ PBMA/ PBM)
Indications for Use (Describe)
K-SHIELD Zen is designed for accessing peripheral veins to collect blood. The Press Button anti-needle stick protector is an integral safety device intended to minimize accidental needlestick injuries.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary -K230408
#### DATE: 06/01/2023
#### I. SUBMITTER
#### Sponsor/Manufacturer
SB-KAWASUMI LABORATORIES, INC. 3-25-4 Tonomachi, Kawasaki-ku, Kawasaki-shi, Kanagawa 210-8602, Japan Tel:+81-44-589-9750 Fax:+81-44-589-9768
#### US Contact and Official Correspondent for SB-KAWASUMI LABORATORIES, INC.
Valerie Followell Official Correspondent Regulatory Compliance Associates®, Inc. 10411 Corporate Drive, Suite 102 Pleasant Prairie, Wisconsin 53158 Phone: (847) 400-6187 Email: v.followell@rcainc.com
#### II. DEVICE
| Device Trade/Proprietary Name: | K-SHIELD Zen (Model Numbers: PBMH/ PBMA/ PBM) |
|-----------------------------------|----------------------------------------------------------------------------------|
| Device Common or Usual Name: | Blood Specimen Collection Device |
| Device Classification Name: | Tubes, Vials, Systems, Serum Separators, Blood Collection |
| Device Regulatory Classification: | Class II |
| Device Classification Regulation: | 21 CFR 862.1675 |
| Product Code: | JKA - Blood Specimen Collection Device<br>FMI – Needle, Hypodermic, Single Lumen |
| Submission Type: | 510(k) |
| Classification Panel: | General Hospital |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
K-SHIELD Zen (K220799) Predicate Device: ●
#### IV. DEVICE DESCRIPTION
The K-SHIELD Zen is a sterile, single use winged blood collection set consisting with a cannula, flexible tube, connection parts and the Press Button anti-needle stick protector designed to minimize accidental needlestick injuries when accessing peripheral veins to collect patient blood. The Press Button anti-needle stick protector stores the needle within the housing when the activation button is pressed after blood collection procedure.
This product has a variety of Needle gauge sizes (21G, 23G, 25G) and a lineup of 3 types of connection parts, Luer connector (PBM), Multiple Sample Luer Adapter (PBMA) and Preattached Holder (PBMH). During blood collection, the Luer connector is connected to a syringe. Multiple Sample Luer Adapter and Pre-attached holder are connected to vacuum blood collection tubes / blood culture bottles to obtain multiple samples. The different configurations (connection options) are chosen by the user based on facility blood collection procedure.
The mechanism of action for the device, all configurations, is blood collection using syringes or vacuum blood collection tubes / blood culture bottles to assist the user with obtaining blood from a patient.
The K-SHIELD Zen is a single use, disposable Ethylene Oxide Sterilized medical device. Sterilization Validation was conducted per ISO 11135: 2014 – Sterilization of Health-Care Products – Ethylene Oxide –Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices.
The K-SHIELD Zen is considered blood path, indirect, limited contact (< 24 hours) per /SO 10993-1: 2018 - Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process. All device materials comply with ISO 10993-1.
The K-SHIELD Zen is a sterile, single-use device that consists of the following components:
●
- . Winged needle with Press Button Anti-needle Stick Protector
- Luer Connector
- . Main Tube
- Clip
- . Connector Tube
- Locking Cap (optional) Multiple Sample Luer Adapter (optional) ●
- Pre-attached Holder (optional) ●
{5}------------------------------------------------
The K-SHIELD Zen is intended for use at healthcare facilities or in hospitals.
## V. INTENDED USE
K-SHIELD Zen is designed for accessing peripheral veins to collect blood. The Press Button antineedle stick protector is an integral safety device intended to minimize accidental needlestick injuries.
## VI. INDICATIONS FOR USE
K-SHIELD Zen is designed for accessing peripheral veins to collect blood. The Press Button antineedle stick protector is an integral safety device intended to minimize accidental needlestick injuries.
## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
- The comparison of the similarities and differences between the K-SHIELD Zen (subject device) and the ● K-SHIELD Zen (K220799) (predicate device) is provided in the Substantial Equivalence (SE) Comparison Table below.
| Substantial Equivalence (SE) Comparison Table | | | |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| | K-SHIELD Zen<br>(Subject Device) | K-SHIELD Zen<br>(K220799)<br>(Predicate Device) | Same/Similarities/<br>Differences |
| Manufacturer | SB-KAWASUMI LABORATORIES, INC | SB-KAWASUMI LABORATORIES, INC | -- |
| Device Trade or<br>Proprietary<br>Name | K-SHIELD Zen | K-SHIELD Zen | -- |
| 510(k)<br>Number | K230408 | K220799 | -- |
| Device Class | Class II | Class II | Same |
| Device<br>Classification<br>Name | Tubes, Vials, Systems, Serum<br>Separators, Blood Collection | Tubes, Vials, Systems, Serum<br>Separators, Blood Collection | Same |
| Device Common<br>Name | Blood specimen collection device | Blood specimen collection device | Same |
| Product Code | JKA and FMI | JKA and FMI | Same |
| Regulation<br>Number | 21 CFR 862.1675 | 21 CFR 862.1675 | Same |
| Indications for<br>Use | K-SHIELD Zen is designed for<br>accessing peripheral veins to collect<br>blood. The Press Button anti-needle<br>stick protector is an integral safety<br>device intended to minimize<br>accidental needlestick injuries. | K-SHIELD Zen is designed for accessing<br>peripheral veins to collect<br>blood. The Press Button anti-needle<br>stick protector is an integral safety<br>device intended to minimize<br>accidental needlestick injuries. | Same |
| Intended Use | Same as Indications for Use | Same as Indications for Use | Same |
| Patient<br>Population | Adults and Pediatric | Adults Only | Different<br>(Biocompatibility<br>assessments (DEHP<br>extraction testing and<br>risk assessment<br>supports comparison) |
| Device Description | The K-SHIELD Zen is a sterile, single use<br>winged blood collection set<br>consisting with cannula, flexible<br>tube, connection parts and the Press<br>Button anti-needle stick protector<br>designed to minimize accidental<br>needlestick injuries when accessing<br>peripheral veins to collect patient<br>blood. The Press Button anti-needle<br>stick protector stores the needle<br>within the housing when the<br>activation button is pressed after<br>blood collection procedure. | The K-SHIELD Zen is a sterile, single<br>use winged blood collection set<br>consisting with cannula, flexible<br>tube, connection parts and the Press<br>Button anti-needle stick protector<br>designed to minimize accidental<br>needlestick injuries when accessing<br>peripheral veins to collect patient<br>blood. The Press Button anti-needle<br>stick protector stores the needle<br>within the housing when the<br>activation button is pressed after<br>blood collection procedure. | Same |
| Press Button<br>Feature | Yes | Yes | Same |
| Anti-Needle Stick<br>Prevention | Yes | Yes | Same |
| Color-Coded<br>Wing | Yes | Yes | Same |
| Needle | Stainless Steel | Stainless Steel | Same |
| Needle Gauge<br>Diameter OD | 21G, 23G, 25G | 21G, 23G, 25G | Same |
| Needle<br>Diameter ID | Ultra-Thin Wall | Ultra-Thin Wall | Same |
| Needle Protector | PP | PP | Same |
| Needle Sheath<br>Cover | Isoprene Rubber | Isoprene Rubber | Same |
| Wing | PP | PP | Same |
| Tubing | PVC | PVC | Same |
| Female Luer<br>Connector | ABS | ABS | Same |
| Luer Adaptor<br>Hub | PP | PP | Same |
| Adhesive | Heat Curing Epoxy | Heat Curing Epoxy | Same |
| Packaging | Envelope Pouch | Envelope Pouch | Same |
| Materials Comply<br>with<br>ISO 10993-1 | Materials comply with<br>ISO 10993-1 | Materials comply with ISO<br>10993-1 | Same |
| Non-Pyrogenic | Non-Pyrogenic | Non-Pyrogenic | Same |
| Non-Toxic | Non-Toxic | Non-Toxic | Same |
| Prescription (Rx)<br>or Over- the<br>Counter<br>(OTC) Use | Prescription Use | Prescription Use | Same |
| Use<br>Environment | Hospital / Clinic | Hospital / Clinic | Same |
| Single-Use | Single-Use Only | Single-Use Only | Same |
| Sterile | Sterile | Sterile | Same |
| Sterility<br>Assurance Level<br>(SAL 10-6) | SAL 10-6 | SAL 10-6 | Same |
| Sterilization<br>Method | Ethylene Oxide Sterilization (EtO) | Ethylene Oxide Sterilization (EtO) | Same |
| Shelf-Life | 3 Year Shelf-Life | 3 Year Shelf-Life | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
### VIII. SUMMARY OF PERFORMANCE DATA AND PERFORMANCE TEST CONCLUSIONS
The following Performance Data were provided in support of the substantial equivalence (SE) determination.
| Summary of Non-Clinical Bench Performance Testing | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| This submission is for the same device but for an expanded patient population and non-clinical performance and<br>biocompatibility data is leveraged/referenced from K220799. | |
| The table below includes the list of the testing results submitted specific to supporting this expanded patient population. | |
| Biocompatibility Testing | |
| Biocompatibility Testing including:<br>• DEHP Extraction test<br>• Risk Analysis Report of DEHP Exposure<br>FDA Recognized Testing Standards:<br>• ISO 10993-17:2002(First Edition 2002-12-01) – Biological Evaluation of Medical<br>Devices – Part 17: Establishment of allowable limits for leachable substances<br>• ISO 10993-18:2020 (Second edition 2020-01 Amendment 1 2022-05) -<br>Biological evaluation of medical devices - Part 18: Chemical characterization of<br>medical device materials within a risk management process [Including<br>Amendment 1 (2022)] | Test Results: PASSED<br>The results of these Non-<br>Clinical Bench Performance<br>Data were provided in<br>support of the substantial<br>equivalence determination. |
| Chemical Testing | |
| Chemical Testing<br>• EtO Residuals<br>FDA Recognized Testing Standards:<br>• ISO 10993-7:2008 (Second Edition 2008-10-15) –<br>Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization<br>Residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability<br>of allowable limits for neonates and infants (2019)] | Test Results: PASSED<br>The result of the Non-<br>Clinical Bench Performance<br>Data was provided in<br>support of the substantial<br>equivalence determination. |
### IX. OVERALL CONCLUSIONS
- The K-SHIELD Zen [subject device] is substantially equivalent to the K-SHIELD Zen (K220799) [predicate device] with respect to indications for use, treatment method and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.