# FloNavi Open Field Fluorescence Imaging System (K230407)

_Guangdong Optomedic Technologies, Inc. · OWN · Jul 17, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230407

## Device Facts

- **Applicant:** Guangdong Optomedic Technologies, Inc.
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Jul 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

Upon intravenous administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence angiography. The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after: vascular, gastrointestinal, organ transplant, and plastic, micro- and reconstructive surgical procedures. Upon interstitial administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

## Device Story

FloNavi Open Field Fluorescence Imaging System captures and displays fluorescent and color images for intraoperative assessment of blood flow and tissue perfusion. System comprises Imaging Head (handheld or mechanical arm-mounted) and Image Processing Unit. Imaging Head uses LEDs for white light (color imaging) and 785nm NIR light (ICG fluorescence excitation). CMOS sensors capture video; Image Processing Unit processes signals for display on medical-grade monitors. Modes include white light, green fluorescence (GF) overlay, segmented fluorescence (SF) with color-coded brightness, and parallel display. Used in hospitals by surgeons to visualize perfusion and lymphatic structures; output aids real-time surgical decision-making and tissue viability assessment.

## Clinical Evidence

Bench testing and animal testing conducted. Device complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, and IEC 62471. Software validated as Moderate Level of Concern. No human clinical trial data reported.

## Technological Characteristics

System includes HD (1920x1080) and 4K (up to 4096x2160) models. CMOS sensor technology. Light sources: LEDs (Visible and 785nm NIR). Connectivity: SDI, DVI, HDMI, CVBS, S-VIDEO. Reusable, non-sterile device requiring low-level disinfection. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 62471.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- SPY Portable Handheld Imaging (SPY-PHI) System ([K200737](/device/K200737.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 17, 2023

Guangdong OptoMedic Technologies, Inc. Alice Lau Regulatory Affairs Engineer Suite 503, Building A, Golden Valley Intellicreation Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai Foshan, Guangdong 528200 China

Re: K230407

Trade/Device Name: FloNavi Open Field Fluorescence Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OWN, IZI, GCJ Dated: July 15, 2023 Received: July 17, 2023

Dear Alice Lau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

Jessica Carr, PhD Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230407

Device Name

FloNavi Open Field Fluorescence Imaging System

#### Indications for Use (Describe)

Upon intravenous administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence angiography.

The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after: vascular, gastrointestinal, organ transplant, and plastic, micro- and reconstructive surgical procedures.

Upon interstitial administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular graphic on the left, followed by the text "OptoMedic" in a bold, black font. The text is slightly glossy, giving it a three-dimensional appearance.

# K230407 510(k) Summary

## Prepared in accordance with the requirements of 21 CFR Part 807.92

Date Prepared: June 9, 2023

### I. General Information

| 510(k) Submitter/Owner: | Guangdong OptoMedic Technologies, Inc.<br>Suite 503, Building A, Golden Valley Intellicreation Community<br>No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan<br>Guangdong, 528200, P.R. China<br>Establishment Registration Number: Not yet registered |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Alice Lau<br>Regulatory Affairs Engineer                                                                                                                                                                                                                     |

| Regulatory Affairs Engineer |
|-----------------------------|
| Tel: +86 (757) 8670 2920    |
| Email: liuli@optomedic.com  |

### II. Device Identification

| Device Trade Name:    | FloNavi Open Field Fluorescence Imaging System                                                                                                                                          |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model:                | HD system: Image Processing Unit (OPTO-CAM2100),<br>Imaging Head (OPTO-CHD3100H; OPTO-CHD3100E)<br><br>4K system: Image Processing Unit (OPTO-CAM214K),<br>Imaging Head (OPTO-CHD314KE) |
| Common or Usual Name: | Fluorescence Angiographic System                                                                                                                                                        |
| Regulation Name:      | Endoscopes and accessories                                                                                                                                                              |
| Regulation Number:    | 21 CFR § 876.1500                                                                                                                                                                       |
| Regulatory Class:     | Class II                                                                                                                                                                                |
| Product Code:         | OWN, IZI, GCJ                                                                                                                                                                           |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circle on the left, followed by the word "OptoMedic" in black. The word "Opto" is in a smaller font than the word "Medic". The logo is simple and modern.

### III. Predicate Device

| 510(k) Number: | K200737                                        |
|----------------|------------------------------------------------|
| Product Name:  | SPY Portable Handheld Imaging (SPY-PHI) System |

The predicate device has not been subject to a design-related recall.

### IV. Device Description

The FloNavi Open Field Fluorescence Imaging System is an imaging system used in hospitals for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue transfer circulation for use in imaging during various surgical procedures.

The FloNavi Open Field Fluorescence Imaging System has two system models: HD system and 4K system. Each system consists of the following main components: Imaging Head, Image Processing Unit, power supply cord and video cables.

The Imaging Head may be either handheld or attached to a mechanical arm and provides illumination of the regions of a patient's body to be observed with near infrared light to excite ICG fluorescence. Alternatively, the Imaging Head provides white light illumination of the regions of a patient's body to be observed for color imaging. The cameras in the Imaging Head capture the fluorescent image under near infrared illumination or a color image under white light illumination. The Image Processing Unit receives the video signal from the Imaging Head and processes and outputs the video image to a medical grade video monitor and/or video recorder. Adjustments to the operation of the FloNavi Open Field Fluorescence Imaging System are possible through switches at either the Imaging Head or the Image Processing Unit.

The FloNavi Open Field Fluorescence Imaging System is a reusable device and provided nonsterile. Its components should be cleaned and low-level disinfected prior to the first use and after every subsequent use.

### V. Indications for Use

Upon intravenous administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence angiography.

The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after: vascular, gastrointestinal, organ transplant, and plastic, micro- and reconstructive surgical procedures.

Upon interstitial administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Image /page/4/Figure/15 description: The image contains a single digit, specifically the number 2. The number is written in a simple, sans-serif font. The digit is black and stands out against a white background.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "OptoMedic" in a stylized font. The "Opto" portion of the word is green, while the "Medic" portion is black. To the left of the word is a green circle that is not fully connected.

.

### VI. Comparison of Technological Characteristics with The Predicate Device

Table 1 General Comparison

| Table 1 General Comparison               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description                              | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                          | FloNavi Open Field Fluorescence Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | SPY Portable Handheld Imaging (SPY-PHI) System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 510(k) Holder /<br>Manufacturer          | Guangdong OptoMedic Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Novadaq Technologies ULC. (now a part<br>of Stryker)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Submission Reference                     | K230407                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K200737                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Combination Product                      | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code                             | OWN, IZI, GCJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | OWN, IZI, GCJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation Number                        | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Classification<br>Name            | Confocal Optical Imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Confocal Optical Imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Indications for Use                      | Upon intravenous administration and use of<br>an ICG consistent with its approved labeling,<br>the FloNavi Open Field Fluorescence<br>Imaging System is used to perform<br>intraoperative fluorescence angiography.<br><br>The FloNavi Open Field Fluorescence<br>Imaging System is indicated for fluorescence<br>imaging of blood flow and tissue perfusion<br>before, during, and after: vascular,<br>gastrointestinal, organ transplant, and plastic,<br>micro- and reconstructive surgical<br>procedures.<br><br>Upon interstitial administration and use of<br>an ICG consistent with its approved labeling,<br>the FloNavi Open Field Fluorescence<br>Imaging System is used to perform<br>intraoperative fluorescence imaging and<br>visualization of the lymphatic system,<br>including lymphatic vessels and lymph<br>nodes. | Upon intravenous administration of SPY<br>AGENTTM GREEN (indocyanine green for<br>injection, USP) the SPY-PHI System is<br>used with SPY AGENTTM GREEN to<br>perform intraoperative fluorescence<br>angiography. The SPY-PHI System is<br>indicated for use in adult and pediatric<br>patients one month of age and older.<br><br>The SPY-PHI System is indicated for<br>fluorescence imaging of blood flow and<br>tissue perfusion before, during, and after:<br>vascular, gastrointestinal, organ transplant,<br>and plastic, micro- and reconstructive<br>surgical procedures.<br><br>Upon interstitial administration of SPY<br>AGENTTM GREEN, the SPY-PHI System<br>is used to perform intraoperative<br>fluorescence imaging and visualization<br>of the lymphatic system, including<br>lymphatic vessels and lymph nodes. |
| Operating Principle                      | Full color visible light and NIR fluorescence<br>video imaging. The cameras (VIS+NIR) in<br>the Imaging Head capture the fluorescent<br>image under near infrared light illumination<br>or a color image under white light<br>illumination. The Image Processing Unit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Full color visible light and NIR<br>fluorescence video imaging. The CMOS<br>camera in the Imaging Head captures the<br>fluorescent image under near infrared light<br>illumination or a color image under white<br>light illumination. The VPI receives the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                          | Head and processes and outputs the video<br>image to a medical grade video monitor<br>and/or video recorder.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | processes and outputs the video image to a<br>medical grade video monitor and/or video<br>recorder.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Major Components                         | HD system:<br>Image Processing Unit (OPTO-CAM2100),<br>Imaging Head (OPTO-CHD3100E/OPTO-<br>CHD3100H)<br><br>4K system:<br>Image Processing Unit (OPTO-CAM214K),<br>Imaging Head (OPTO-CHD314KE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Imaging Head/Imager (HH9030), Video<br>Processor/Illuminator (PC9001)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Image modes                              | White light image - White light image is<br>displayed in full color.<br><br>GF image - NIR fluorescence is<br>superimposed in pseudo-color (green) on a<br>white light image<br><br>SF image – In Segmented Fluorescence (SF)<br>image, a high-definition white light image is<br>displayed in full color with NIR fluorescence<br>overlaid in on a color scale. The pseudo-color<br>of fluorescence will change according to the<br>fluorescence brightness.<br><br>Parallel display image - The white light<br>image, GF image, SF image, grayscale<br>fluorescent image are displayed on the same<br>screen, with the small image on the left side<br>and the large image as the main image. The<br>default main image is the white light image<br>and can be changed through the menu to the<br>image of other display modes.     | White Light Mode - White light image is<br>displayed in full color.<br><br>Overlay Mode - an NIR fluorescence<br>image is superimposed in pseudo-color<br>(green) on a white light image.<br><br>Color Segmented Fluorescence (CSF) -<br>white light image is displayed in grayscale<br>with NIR fluorescence overlaid in a color<br>scale. Increasing fluorescence levels<br>transition from blue to yellow to red.<br><br>SPY mode - an NIR fluorescence image is<br>displayed in grayscale.                                                                                                                                                                                                                                                                                                                                   |
| Prescription/<br>Over-the-counter<br>use | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Environment of Use                       | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Video Output Signals                     | HD system: SDI, DVI, CVBS, S-VIDEO<br>4K system: DVI, HDMI, 4×3G-SDI, 12G-<br>SDI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | HD-SDI, 3G-SDI, DVI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Image Resolution                         | HD system: 1920×1080p<br>4K system: 1920×1080p / 3840×2160p /<br>4096×2160p                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 1920×1080p                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Sensor Type                              | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Light source                             | Visible (VIS): Light-emitting diodes<br>Near infrared (NIR): Light-emitting diodes,<br>wavelength 785nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Visible (VIS): Light-emitting diode array<br>Near infrared (NIR): NIR laser diode,<br>wavelength 805 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Power supply                             | Mains powered                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Mains powered                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Safety Standards                         | IEC 60601-1<br>IEC 60601-1-2<br>IEC 62471<br>IEC 60601-2-57                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60825-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Design                                   | The light source of the subject device is<br>located in the Imaging Head itself.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The light source is located in the VPI and<br>transmits light through a light guide cable.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green circular design on the left, followed by the word "OptoMedic" in bold, black letters. The circular design is made up of three curved lines that start with a yellow color and transition to green.

0

.

1

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green circular graphic on the left, followed by the word "OptoMedic" in black font. The "Opto" part of the word is in a lighter shade of black, while the "Medic" part is in a darker, more glossy shade of black.

The differences technological characteristics do not raise different questions of safety and effectiveness.

## VII. Performance data

Non clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- . IEC 60601-1:2005+A1:2012; ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance (HD system)
- . IEC 60601-1:2005+A1:2012+A2:2020; ANSI/AAMI ES60601-1:2005 + A2:2010 (R2012) + A 1:2012+A2:2021 Medical electrical equipment Part 1: General requirements for basic safety and essential performance (4K system)
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- . IEC 60601-2-57:2011 Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- . IEC 62471:2006 Photobiological safety of lamps and lamp systems

The software of the proposed device was validated as Moderate level of concern (LoC) in accordance with the following guidance documents: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff

Performance testing and animal testing were also conducted on the subject device and demonstrate that the proposed system performs according to specifications and functions as intended.

## VIII. Conclusions

The performance testing and animal testing summarized above supports a substantial equivalence determination. The performance testing and animal testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K230407](https://fda-staging.innolitics.com/device/K230407)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).