Durio Surgical Face Mask, Models 501 (3-ply, blue) and 545 (4-ply,blue)
K230395 · Durio Ppe Sdn Bhd · FXX · Jul 17, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230395
Device Name
Durio Surgical Face Mask, Models 501 (3-ply, blue) and 545 (4-ply,blue)
Applicant
Durio Ppe Sdn Bhd
Product Code
FXX · General, Plastic Surgery
Decision Date
Jul 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.
Device Story
Non-sterile, single-use, flat-pleated surgical mask; available in 3-ply (Model 501) and 4-ply (Model 545) configurations. Functions as a physical barrier to protect patients and healthcare personnel from microorganisms, body fluids, and particulates. Features ear loops (polyester/spandex) and an adjustable zinc electro-galvanized steel nose strip. Outer layer (spunbond polypropylene) repels fluids; middle layer (melt-blown polypropylene) provides filtration via mechanical blocking and electrostatic attraction. Used in clinical settings or general infection control. Healthcare providers wear the mask to mitigate cross-contamination risks. Benefits include reduced exposure to infectious agents and body fluids.
Clinical Evidence
Bench testing only. No clinical or animal data provided. Performance validated via biocompatibility (ISO 10993-5, -10), Bacterial Filtration Efficiency (ASTM F2101-19), Sub-micron Particulate Efficiency (ASTM F2100-19), synthetic blood penetration resistance (ASTM F1862/F1862M-17), differential pressure (ASTM F2100-19), and flame spread (16 CFR 1610).
Technological Characteristics
Flat-pleated mask; 3 or 4-ply construction. Materials: Spunbond polypropylene (outer), melt-blown polypropylene (middle), hydrophilic spunbond polypropylene (inner), polyester/spandex ear loops, zinc electro-galvanized steel nose wire. Dimensions: 175mm x 95mm. Non-sterile. Filtration via mechanical blocking and electrostatic attraction. Biocompatible per ISO 10993.
Indications for Use
Indicated for use by patients and healthcare personnel as a physical barrier to reduce the transfer of microorganisms, body fluids, and particulate material, and to reduce exposure to blood and body fluids in infection control practices.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
CAREWE Surgical Face Mask Models N001-AW, N002-AW, and N003-AW (K203078)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
July 17, 2023
Durio PPE SDN BHD % Ms. Sarah Fitzgerald Senior Consultant, Regulatory and Quality Affairs Emergo Global Consulting, LLC 2500 Bee Caves Rd., Bldg. 1, Ste. 300 Austin, Texas 78746
Re: K230395
Trade/Device Name: Durio Surgical Face Mask, Models 501 (3-ply, blue) and 545 (4-ply,blue) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: June 12, 2023 Received: June 12, 2023
Dear Ms. Fitzgerald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230395
Device Name Durio Surgical Face Mask
Indications for Use (Describe)
Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K230395: 510(k) Summary
## Durio Surgical Face Masks:
## Model 501 (3-ply, blue) and Model 545 (4-ply,blue)
## 1. Submission Sponsor
Durio PPE SDN BHD 16, Jalan Temenggong Johor Bahru, Johor 81100 Malaysia Contact: Xavier Kong Title: Deputy General Manager
## 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement@ul.com Contact: Sarah Marie Fitzgerald Title: Senior Consultant, Regulatory and Quality Affairs
## 3. Date Prepared
June 16, 2023
#### 4. Device Identification
| Trade/Proprietary Name: | Durio Surgical Face Mask, Models 501 (3-ply, blue) and 545 (4-ply,blue) |
|-------------------------|-------------------------------------------------------------------------|
| Common/Usual Name: | Surgical Mask |
| Classification Name: | Surgical apparel |
| Regulation Number: | 21 CFR 878.4040 |
| Product Code: | FXX |
| Class: | II |
| Classification Panel: | General Hospital |
## 5. Legally Marketed Predicate and Reference Device(s)
| Device name: | Surgical Mask |
|----------------|---------------|
| 510(k) number: | K212344 |
| Type: | Predicate |
K230395 510k Summary Page 1 of 4
{4}------------------------------------------------
| Manufacturer: | Xiantao Dingcheng Non-Woven Product Co., Ltd. |
|----------------|----------------------------------------------------------------|
| Device name: | CAREWE Surgical Face Mask Models N001-AW, N002-AW, and N003-AW |
| 510(k) number: | K203078 |
| Type: | Reference |
| Manufacturer: | Guangdong Kaidi Garments Co., Ltd. |
### 6. Indication for Use Statement
Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.
## 7. Device Description
The Durio Surgical Face Mask is a non-sterile, single use, flat-pleated style surgical mask with ear loops and nose piece.
The Durio Surgical Face Mask is available in three or four ply models. In all cases, the inner and outer layers of the Durio Surgical Face Mask are made of non-woven spunbond polypropylene (PP). The outer layer repels fluid and droplets. The middle layer is made of non-woven melt blown PP fabric for filtration by mechanical blocking and electrostatic attraction. For the four-ply model, there is an additional middle layer made of non-woven spunbound PP. The ear loops are made of polyester covered spandex yarn. The elastic ear loops pull inward from welding on the outer layer to reduce gaps as the side between the face and mask. The nose strip is made of zinc electro galvanized steel wire allowing adjustability and to let it stand fixed for better fit.
The Durio Surgical Face Mask is provided in blue. The device is not made from any natural rubber latex.
## 8. Substantial Equivalence Discussion
The following table compares the Durio Surgical Face Mask to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
| Attribute | Subject Durio Surgical Face<br>Mask | Predicate Surgical Mask<br>(K212344) | Comparison |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Product Code | FXX | FXX | Same |
| Regulation | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Indications for Use | Surgical Mask is intended to be<br>worn to protect both the patient<br>and healthcare personnel from<br>transfer of microorganisms, | Surgical Mask is intended to be<br>worn to protect both the patient<br>and healthcare personnel from<br>transfer of microorganisms, | Same |
| Attribute | Subject Durio Surgical Face<br>Mask | Predicate Surgical Mask<br>(K212344) | Comparison |
| | body fluids and particulate<br>material. It is intended for use<br>in infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This is<br>a single use, disposable<br>device(s), provided non sterile. | body fluids, and particulate<br>material. It is intended for use in<br>infection control practices to<br>reduce the potential exposure<br>to blood and body fluids. This is<br>a single use, disposable<br>device(s), provided non sterile. | |
| Mechanism of<br>Action | Physical barrier | Physical barrier | Same |
| Single-Use | Yes | Yes | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Composite | Flat Pleated, 3 or 4 Layers | Flat Pleated, 3 Layers | Equivalent*<br>Same as reference<br>K203078 |
| Materials: Outer | Spunbond Polypropylene (SBPP) | Spunbond Polypropylene (SBPP) | Same |
| Materials:<br>Middle | Melt Blown Polypropylene<br>(MBPP) | Melt Blown Polypropylene<br>(MBPP) | Same |
| Materials: 4th<br>Layer | None or MBPP | None | Equivalent*<br>Same as reference<br>K203078 |
| Materials:<br>Inner | Hydrophilic Spunbond<br>Polypropylene (HPP) | Spunbond Polypropylene (SBPP) | Equivalent* |
| Materials:<br>Nose-Piece | Galvanized steel wire, flexible | Galvanized iron wire, flexible | Equivalent* |
| Materials:<br>Earloop | Polyester and Spandex | Polyester and Spandex | Same |
| Dimensions | 175mm x 95mm | 175mm x 95mm | Same |
| Color | Blue | Blue | Same |
| OTC Use | Yes | Yes | Same |
| ASTM F2100<br>Level | Level 2 (3-Ply)<br>Level 3 (4-Ply) | Level 3 | Equivalent** |
| Biocompatibility | Biocompatible per ISO 10993<br>Series Testing | Biocompatible per ISO 10993<br>Series Testing | Same |
## Table 5A - Comparison of Characteristics
{5}------------------------------------------------
* The devices have minor differences in materials. The differences do not raise new or different questions related to the safety or effectiveness of the device, and performance and biocompatibility evaluation and testing addresses these differences; therefore the devices are substantially equivalent.
** Testing was conducted within alignment with the labeled protection level; therefore the devices are substantially equivalent.
#### 9. Non-Clinical Performance Data
To demonstrate safety and effectiveness of the Durio Surgical Face Mask and to show substantial equivalence to the predicate device, Durio completed various bench performance tests. No animal or clinical testing was required. Results confirm that the design inputs and performance specification of the device are met. Durio Surgical Face Masks passed the testing in accordance with internal requirements
{6}------------------------------------------------
and standards as shown below, supporting its safety, effectiveness, and substantial equivalence to the predicate device.
- Biocompatibility Testing (Cytotoxicity, Irritation, Sensitization) per ISO 10993-5 and ISO 10993-10
- Bacterial Filtration Efficiency (BFE) per ASTM F2101-19
- Sub-micron Particulate Efficiency (PFE) per ASTM F2100-19
- Resistance to Penetration by Synthetic Blood per ASTM F1862/F1862M-17 ●
- Differential Pressure per ASTM F2100-19
- Flame Spread per 16 CFR 1610 (Class 1)
- Shipping Performance per ASTM D4169-16 ●
### 10. Clinical Performance Data
No animal or clinical testing was required.
## 11. Statement of Substantial Equivalence
The Durio Surgical Face Mask has the same indications for use as the Surgical Mask (K212344). Any minor differences in the technological characteristics of the subject device when compared to the predicate device have been successfully evaluated through appropriate safety and performance testing which demonstrates that the subject device, when compared to the predicate device, does not raise any new questions of safety and effectiveness. Therefore, the Durio Surgical Face Mask has been determined to be substantially equivalent to predicate Surgical Mask (K212344).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.