The portable (Ultrasonic) nebulizer is an ultrasonic vibrating mesh nebulizer system designed to aerosolize liquid medications for inhalation by the patient. The device may be used with pediatric (5 years and older), defined by the prescribed medication, and adult patients in hospital/ institutional settings, home care use, schools, and long term care facilities. It is not intended for use with Pentamidine.
Device Story
Portable (Ultrasonic) Nebulizer NBM-1 is a handheld, battery-powered device using ultrasonic vibrating mesh technology to aerosolize liquid medications for inhalation. It consists of a main unit, medicine cup, adult/pediatric masks, and mouthpiece. Operated by patients or caregivers in homes, hospitals, schools, or long-term care facilities. The device transforms liquid medication into a fine vapor mist, allowing deeper lung penetration for faster relief. It operates continuously during inhalation and exhalation. Healthcare providers prescribe the medication; the device facilitates delivery. Benefits include portable, efficient medication administration for respiratory conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), home healthcare environment safety (IEC 60601-1-11), aerosol performance (ISO 27427), and biocompatibility (ISO 10993 series).
Technological Characteristics
Ultrasonic vibrating mesh technology; 110 KHz ± 10KHz oscillation frequency. Materials: ABS (shell, battery cover, nozzle), silica gel (button ring), PC (cup, button), PVC (mask), PP (mouthpiece), stainless steel (nebulizer piece). Powered by 2x AA batteries (3.0V DC). IP22 ingress protection. Type BF applied part. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ISO 27427.
Indications for Use
Indicated for pediatric (5 years and older) and adult patients requiring aerosolized liquid medication inhalation. Not for use with Pentamidine.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2023
Dongguan SIMZO Electronic Technology Co.Ltd. % Reanny Wang General Manager Shenzhen Reanny Medical Devices Management Consulting., Ltd Room 1407, Jingting Building, Dongzhou Community, Guangming Street, Guangming District Shenzhen, Guangdong 518000 China
Re: K230379
Trade/Device Name: Portable (Ultrasonic) Nebulizer Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: May 9, 2023 Received: October 27, 2023
Dear Reanny Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows a digital signature. The signature is from John S. Bender -S. The date of the signature is 2023.11.27, and the time is 13:38:14 -05'00'.
for Ethan Nyberg Assistant Director DHT1C: Division of Sleep Disordered
{2}------------------------------------------------
Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K230379
Device Name Portable (Ultrasonic) Nebulizer
#### Indications for Use (Describe)
The portable (Ultrasonic) nebulizer is an ultrating mesh nebulizer system designed to aerosolize liquid medications for inhalation by the patient. The device may be used with pediatric (5 years and older), defined by the prescribed medication, and adult patients in hospital/ institutional settings, home, schools, and long-term care facilities. It is not intended for use with Pentamidine.
| Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Simzo" in a bold, sans-serif font. The letters are a dark blue color. The word is displayed horizontally and appears to be a logo or brand name.
# 510(k) Summary
The assigned 510(k) number is: K230379
# 1.0 Information of Submitter and Correspondent
## Submitter's information:
Name: Dongguan SIMZO Electronic Technology Co., Ltd. Address: No.6, Zhangzhou Road, Daojiao Town, Dongguan City, Guangdong Province, 523187, R.R. China Tel: +86-769-86339010 Fax: +86-769-22981193 Contact with: Jin Gang Website: www.simzo.net E-mail: dong@simzo.net
## Submission correspondent's information:
Shenzhen Reanny Medical Devices Management Consulting Co., Ltd. Address: Room 1407, Jingting Building, Dongzhou Community, Guangming Street, Guangming District, Shenzhen 518000, China Contact Person: Reanny Wanq E-mail: reanny@reanny.com
# 2.0 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|--------------------------------------|
| Trade Name: | Portable (Ultrasonic) Nebulizer |
| Model: | NBM-1 |
| Classification name: | Nebulizer (Direct Patient Interface) |
| Review Panel: | Anesthesiology |
| Product Code: | CAF |
| Device Class: | II |
| Regulation Number: | 868.5630 |
# 3.0 Predicate Device Information
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Simzo" in a bold, sans-serif font. The letters are a dark blue color, and the background is white. The word is slightly angled upwards from left to right.
| Sponsor: | Qingdao Future Medical Technology Co., Ltd. |
|----------------|---------------------------------------------|
| Device: | Intelligent Mesh Nebulizer |
| 510(K) Number: | K171549 |
# 4.0 Device Description
The Portable (Ultrasonic) nebulizer NBM-1 is small, handheld, internally powered general purpose nebulizer which utilizes ultrasound vibrating mesh technology to generate aerosol. The Portable (Ultrasonic) nebulizer provides a vapor mist of medicine that has much smaller particles and can travel much deeper into the patient's lungs, resulting in much faster relief. The ultrasonic nebulizer is made up of main unit, medicine cup, adult mask, kid mask and mouth piece. The shell of nebulizer, battery cover and spray nozzle are made of ABS, the button decoration ring is silica gel, the medicine cup and button are made by PC, the mask is made of PVC, the mouth piece is made of PP and the nebulizer piece is stainless steel.
The Portable (Ultrasonic) nebulizer NBM-1 has the same working principle, specification, structure, intended use and software, the only differences is appearance.
# 5.0 Indications for Use
The portable (Ultrasonic) nebulizer is an ultrasonic vibrating mesh nebulizer system designed to aerosolize liquid medications for inhalation by the patient. The device may be used with pediatic (5 years and older), defined by the prescribed medication, and adult patients in hospital/ institutional settings, home, schools, and long-term care facilities. It is not intended for use with Pentamidine.
## 6.0 Comparison to predicate device and conclusion
The subject device is substantially equivalent to predicate devices, K171549, Intelligent Mesh Nebulizer. The substantial equivalence chart is provided as follows:
| Eleme<br>nts of<br>Comparison | Predicate Device<br>(K171549) | Subject Device | Judgment |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Models | NEB001 | NBM-1 | -- |
| Company | Qingdao Future Medical Technology<br>Co., Ltd. | Dongguan SIMZO Electronic<br>Technology Co. Ltd. | -- |
| Eleme<br>nts of<br>Comparison | Predicate Device<br>(K171549) | Subject Device | Judgment |
| Models | NEB001 | NBM-1 | -- |
| Device Name | Intelligent Mesh Nebulizer | Portable (Ultrasonic) Nebulizer | -- |
| Product code | CAF | CAF | SE |
| Regulation # | 21CFR868.5630 | 21CFR868.5630 | SE |
| Intended use | The Intelligent Mesh Nebulizer<br>designed to aerosolized liquid<br>medications for inhalation by patient,<br>the device may be used with<br>pediatric (>4 years of age) and adult<br>patients in the home, hospital and<br>sub-acute care settings. It is not<br>intended for use with Pentamidine. | The ultrasonic nebulizer is an<br>ultrasonic vibrating mesh nebulizer<br>system designed to aerosolize liquid<br>medications for inhalation by the<br>patient. The device may be used<br>with pediatric (5 years and older),<br>defined by the prescribed<br>medication, and adult patients in<br>hospital/institutional settings, home<br>care use, schools, and long term<br>care facilities.<br>lt is not intended for use with<br>Pentamidine. | SE, see<br>Remark 1 |
| Principle of<br>operation | Vibrating mesh | Vibrating mesh | SE |
| Ultrasonic<br>Oscillation<br>Frequency | Approx. 110KHz | 110 KHz ± 10KHz | SE |
| Aerosolizatio<br>n | Continuous during inhalation and<br>exhalation | Continuous during inhalation and<br>exhalation | SE |
| Compressed<br>gas source | None needed | None needed | SE |
| Medicine<br>Capacity | 8ml maximum,0.5ml minimum | 6 ml | Similar<br>see<br>Remark 2 |
| Nebulization<br>rate | ≥ 0.2 ml/min | >0.2mL/min. | SE |
| Aerosol<br>Performance | Table: Comparative particle test<br>comparison | Table: Comparative particle test<br>comparison | Similar, see<br>Remark 3 |
| Duration of<br>Use | Single patient, multi-use | Single patient, multi-use | SE |
| Eleme<br>nts of<br>Comparison | Predicate Device<br>(K171549) | Subject Device | Judgment |
| Models | NEB001 | NBM-1 | -- |
| Power supply | ZN-103450 Lithium battery: 3.7Vd.c.<br>AC Adapter AC 100-240Va.c. 47-<br>63Hz 0.4-0.2A | DC3.0V, 2×AA batteries | Difference,<br>see<br>Remark 4 |
| Waterproof | IP22 | IP22 | SE |
| Degree of<br>protection<br>against<br>electric shock | Type BF applied part | Type BF applied part | SE |
| Type of<br>protection<br>against<br>electric shock | Internally power equipment | Internally power equipment | SE |
| Model of<br>operation | Continuous operation | Continuous operation | SE |
| Power off | ON / OFF Button | On/Off button | SE |
| Operation<br>environment | 5°C to 40°C, 15% to 90% RH | 10 -40°C, 10%-95%RH | Similar,<br>see<br>Remark 5 |
| Storage<br>environment | -25°C to 70°C, ≤90% RH | -20°C to 70°C, 10%-95% RH | Similar,<br>see<br>Remark 5 |
| Patient<br>Connector | Mouthpiece or masks | Mouthpiece or masks | SE |
| Biocompatibil<br>ity | All the patient contacting materials<br>are compliance with ISO 10993 | All the patient contacting materials<br>are compliance with ISO 10993 | SE |
| Electrical<br>Safety | Compliance with IEC 60601-1 and<br>IEC 60601-1-11 | Compliance with IEC 60601-1 and<br>IEC 60601-1-11 | SE |
| EMC | Compliance with IEC 60601-1-2 | Compliance with IEC 60601-1-2 | SE |
| Dimensions<br>(mm) | 50mm(L)×74mm(W)×111mm(H) | 60mm×55mm×130mm | Similar,<br>see<br>Remark 6 |
| Weight (kg) | Approx. 106g | Approx. 110g | Similar,<br>see<br>Remark 6 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Simzo" in a bold, sans-serif font. The letters are all capitalized and in a dark blue color. The background is white, which makes the text stand out. The font appears to be a stylized version of a common typeface.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Simzo" in a bold, sans-serif font. The letters are all capitalized and in a dark blue color. The background is white, which makes the text stand out. The font appears to be a stylized version of a common typeface.
{8}------------------------------------------------
Simzo
| Item | Adult | | | Pediatric | | | | | | | | |
|-------------------------------------------|-----------------------------------|---------------------------------|----------------------------------|----------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------|------------------------|--------------------|-------------|-----------|--------------|
| | Subject<br>(Ultrasonic)<br>NBM-1) | device-<br>Nebulizer<br>(Model: | Portable<br>(Model:<br>NEB001) - | Predicate<br>Intelligent<br>Mesh<br>(K171549)- Nebulizer | Subject device- Portable (Ultrasonic)<br>Nebulizer (Model: NBM-1) | Predicate<br>Intelligent Mesh<br>(K171549)- Nebulizer<br>(model: NEB001) - | | | | | | |
| | albuterol<br>sulfate | ipratropium<br>bromide | cromolyn<br>sodium | albuterol<br>sulfate | ipratropium<br>bromide | cromolyn<br>sodium | albuterol<br>sulfate | ipratropium<br>bromide | cromolyn<br>sodium | | | |
| Particle Size<br>(MMAD) | 2.09±0.19 | 2.10±0.14 | 2.07±0.19 | 2.27±0.21 | 2.37±0.35 | 2.27±0.25 | 3.66±0.22 | 3.48±0.30 | 3.64±0.21 | 3.73±0.35 | 3.60±0.17 | 3.73±0.21 |
| Geometric<br>Standard<br>Deviation | 1.94±0.21 | 1.91±0.15 | 1.78±0.13 | 1.80±0.20 | 1.73±0.12 | 1.67±0.06 | 2.01±0.23 | 1.98±0.18 | 1.97±0.17 | 2.10±0.17 | 2.10±0.17 | 1.90±0.10 |
| Total Dose<br>Delivered by<br>device | 1,529±65 | 235±20 | 7,236±753 | 1,541±65 | 242±4 | 6,983±852 | 1593.1±113.3 | 250.2±10.4 | 7394.3±445.5 | 1624.2±81.2 | 247.4±5.2 | 7279.3±545.9 |
| Total<br>Respirable<br>Dose (0.5-<br>5um) | 1,048±106 | 156±13 | 5,341±589 | 1,133±109 | 164±7 | 5,157±629 | 961.6±129.3 | 150.8±7.1 | 4581.7±571.2 | 932.3±115.2 | 145.4±6.2 | 4369.4±201.8 |
| Coarse<br>Particle Dose<br>(>4.7um) | 397±57 | 66±9 | 1,626±224 | 333±52 | 66±11 | 1,507±198 | 618.3±62.3 | 95.5±7.8 | 2715.4±198.6 | 665.9±48.7 | 98.1±4.7 | 2790.8±338.0 |
| Fine Particle<br>Dose<br>(<4.7um) | 1,132±106 | 169±14 | 5,610±647 | 1,208±115 | 175±11 | 5,477±745 | 974.7±126.2 | 154.8±6.1 | 4679.0±597.7 | 958.3±114.8 | 149.3±7.4 | 4488.5±226.0 |
| Ultra-fine<br>Particle Dose<br>(<1.0um) | 258±47 | 41±4 | 1,222±128 | 243±28 | 36±8 | 1,091±139 | 81.0±21.0 | 12.0±2.8 | 384.4±88.8 | 98.0±6.9 | 13.4±1.5 | 353.0±26.9 |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the word "Simzo" in a bold, blue font. The letters are all capitalized and evenly spaced. The background is white, which makes the blue text stand out.
### Remark 1:
The indications for use of subject device and predicate device are similar, the intended patient population of subject device is same as predicate device, they are pediatric (>4 years of age=5 years and older) and adult patients.
The handheld nebulizers are common and usual used for Hospital/institutional settings, home care use, schools and long-term care facilities. They are both compliance with IEC 60601-1, IEC 60601-1-2 and IEC60601-1-11 standards. The propose environments of subject device is identical for the predicate device, it will not raise any problems of safety or effectiveness.
### Remark 2:
The "Medicine Capacity" of subject is within the predicate device. The subject device has tested on safety and performance tests and the test results were complied with the test requests. Therefore, the subject device didn't raise any problems of safety or effectiveness.
#### Remark 3:
From the Table: Comparative particle test comparison, all of the parameters were similar with predicate device, there were no statistically measurable differences between the subject device and predicate device, they are substantially equivalent, and the subject device didn't raise any problems of safety or effectiveness.
#### Remark 4:
The "power supply" of subject device is difference with predicate device, the subject device use AA battery to supply power, the risk of AA battery is lower than Lithium battery. Otherwise, the subject device has been passed the test according to IEC 60601-1 standard, it will not raise any safety or effectiveness issue.
### Remark 5:
The operation and storage environment of subject devices is difference with predicate device, and they are both compliance with IEC60601-1-11 standard, it will not raise any safety or effectiveness issue.
#### Remark 6:
The subject device has the similar specifications with the predicate device, such as dimensions, Weight, based on the nonclinical tests performed, those minor differences for specifications do not affects the safety and effective of the device.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "Simzo" in a bold, sans-serif font. The color of the text is a dark blue. The letters are closely spaced together, giving the word a compact appearance.
## Conclusion:
The subject devices have all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject devices.
Thus, the subject devices are substantially equivalent to the predicate device.
# 7.0 Summary of Non-clinical performance testing
## 7.1 Non-clinical testing
A series of safety and performance tests were conducted on the subject device, Portable (Ultrasonic) Nebulizer NBM-1.
## Electrical Safety and Electromagnetic Compatibility (EMC):
Electrical safety and EMC testing were conducted on the device. Testing established that, with respect to electrical safety, the device meets the applicable requirements of:
- A AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005+A1:2012, MOD)
- A IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- A IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
## Safety and performance:
Safety and performance testing of general purpose nebulizing systems intended for continuous or breath-actuated delivery of liquids, in aerosol form, to humans through the respiratory system were conducted on the device, and three kind of drugs used for testing, albuterol sulfate, ipratropium bromide, cromolyn sodium. Testing established that, with respect to safety and performance for nebulizing systems, the device meets the applicable requirements of:
- A ISO 27427:2013 Anaesthetic and respiratory equipment-Nebulizing systems and components
- A FDA Guidance: Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators
{11}------------------------------------------------
### Biocompatibility testing
The nebulizer's mouthpiece, adult mask, kid mask, medicine cup and cover, button decoration ring, nebulizer piece are patient-contacting device components.
According to provision of ISO 10993-1:2018 annex A table A.1 and FDA's 2020 Guidance entitled,
" Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" evaluation testing, and consider the product character of Portable (Ultrasonic) Nebulizer, the following biological compatibility test need to be executed as follow:
- � In Vitro Cytotoxicity Test -ISO10993-5
- � Skin Sensitization Test(polar and non polar ) -ISO10993-10
- Intracutaneous Reactivity Test-ISO10993-10 �
- � Acute Systemic Toxicity Test (polar and non polar) -ISO10993-11
- � Material-mediated pyrogens Test -ISO10993-11
- � Ames Test (polar and DMSO Extract)--ISO10993-3
- � In Vitro Mammalian Chromosomal Aberration test -- ISO10993-3
- � Subchronic systemic toxicity Test (polar and non polar) -ISO10993-11
- � Subacute systemic Toxicity Test (polar and non polar) -ISO10993-11
- � Muscle implant Test -ISO10993-6
- � Chemical characterization study- ISO10993-18
- � Toxicological risk assessment study-ISO10993-17
- Biocompatibility evaluation- ISO 18562-1
- Test for emissions of particulate matter ISO 18562-2
- Test for emissions of VOCs -ISO 18562-3
Under the parameters of the tests, it is concluded that they are biocompatible, and that there are no new issues of safety regarding their use as intended.
## Software Verification and Validation Testing
The software verification and validation testing was conducted in accordance with, and documentation was provided as recommended by FDA ' s Guidance , " Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 1, 2005" and IEC 62304:2015 Medical device software - Software life-cycle processes.
## Device life testing, shelf-life testing, Battery life testing
Device life, shelf life and battery life were tested with simulation method, the device meets the applicable requirements.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "Simzo" in a bold, sans-serif font. The color of the text is a dark blue. The letters are closely spaced together, creating a solid block of text. The background is plain white.
### Cleaning Validation
The cleaning was conducted, the test result meets the applicable requirements.
All the test results demonstrate Portable (Ultrasonic) Nebulizer NBM-1 meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate devices.
### 7.2 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
# 8.0 Conclusions
Portable (Ultrasonic) Nebulizer NBM-1 has the same intended use and similar characteristics as the predicate device. Moreover, the subject device demonstrates product safety by successful completion of testing to the IEC 60601-1, IEC60601-1-11 standards and electromagnetic standard IEC 60601-1-2. The performance test demonstrates the nebulizer meet the ISO 27427 standard and confirm the aerosol performance. These conclude that any differences in their characteristics do not raise any safety and effectiveness issues.
From the above information we conclude the subject device, NBM-1 is substantially equivalent to the predicate devices Intelligent Mesh Nebulizer.
# 9.0 Summary prepared date
October 27, 2023
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.