K230342 · Shanghai Omni Laser Skinology Co., Ltd. · GEX · Aug 16, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230342
Device Name
Phototherapy System (OL-PDT950)
Applicant
Shanghai Omni Laser Skinology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Phototherapy System use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions. The blue light (469mm wavelength) indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. The red light (633nm wavelength) indicated to treatment of superficial, benign vascular, and pigmented lesions. The infrared light (835nm wavelength) is generally use for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
Device Story
Phototherapy System (OL-PDT950) is a vertical LED-based device for aesthetic and dermatological management. It utilizes five LED panels to emit red (633nm), blue (469nm), and infrared (835nm) light. The device is operated by healthcare professionals in a clinical setting. The operator uses a display screen interface to select treatment parameters. The device emits light energy onto the patient's skin at a working distance of 10-15cm for a recommended 20-minute session. The output energy is intended to treat acne, vascular/pigmented lesions, or provide temporary relief for musculoskeletal pain and stiffness. The device does not involve complex AI/ML processing; it functions as a therapeutic light source.
Clinical Evidence
No clinical data was provided. Substantial equivalence is supported by non-clinical bench testing, including electrical safety, electromagnetic compatibility, photobiological safety, and performance testing per IEC standards.
Technological Characteristics
Vertical LED-based phototherapy system. Emits red (633±15nm), blue (469±15nm), and infrared (835±15nm) light via five LED panels. Power supply: 100-240V AC, 50/60Hz, 440VA. Working distance: 10-15cm. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-57 (non-laser light source performance), IEC 62471 (photobiological safety).
Indications for Use
Indicated for patients with moderate inflammatory acne vulgaris (blue light), superficial benign vascular and pigmented lesions (red light), and minor muscle/joint pain, arthritis, muscle spasms, stiffness, and need for increased local blood circulation (infrared light).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
August 16, 2023
Shanghai Omni Laser Skinology Co., Ltd. % Helen Nan General Manager New Risen Enterprise Management Consulting Co., Ltd. Room 302, Building 3, Hangqian Mansion, Hangqian Street Lucheng District Wenzhou, Zhejiang 325000 China
Re: K230342
Trade/Device Name: Phototherapy System (OL-PDT950) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 17, 2023 Received: July 17, 2023
Dear Helen Nan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jianting Wang -S
For Tanisha L. Hithe, MS, MHS Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230342
#### Device Name Phototherapy System (OL-PDT950)
Indications for Use (Describe)
he Phototherapy System use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.
The blue light (469mm wavelength) indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
The red light (633nm wavelength) indicated to treatment of superficial, benign vascular, and pigmented lesions.
The infrared light (835nm wavelength) is generally use for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a green shamrock with three leaves. The shamrock is surrounded by a light purple, curved shape that partially encircles it. The background is white.
K230342 510(k) Summary (As required by 21 CFR 807.92)
## 1.0 Submitter Information
| Company: | Shanghai Omni Laser Skinology Co., Ltd. |
|----------------|----------------------------------------------------------------------------------------------|
| Address | Floor 3, Building 3, NO. 227, Mingqiang Road, Songjiang<br>District, Shanghai, 201612, CHINA |
| Phone: | +86-021-54847192 |
| E-mail: | avril@omni-laser.com |
| Contact | Avril Ouyang |
| Title: | General Manager |
| Prepared date: | 07/17/2023 |
### 2.0 Device Information
| Trade/Device Name: | Phototherapy System |
|--------------------|-----------------------------------|
| Model: | OL-PDT950 |
| Device: | Powered Laser Surgical Instrument |
| Review Panel: | General & Plastic Surgery |
| Product Code: | GEX |
| Submission Type: | Traditional 510(k) |
| Regulation Number: | CFR 878.4810 |
| Device Class: | Class II |
### 3.0 Predicate Device Information
| Predicate Device 1: | |
|---------------------|-------------------------------------------|
| K Number: | K200751 |
| Trade/Device Name: | Photodynamic Therapy System |
| Submitter: | Shangdong Huamei Technology Co., Ltd. |
| Predicate Device 2 | |
| K Number: | K222751 |
| Trade/Device Name: | LED Light Therapy Device |
| Submitter: | Xuzhou Kernel Medical Equipment Co., Ltd. |
| Predicate Device 3 | |
| K Number: | K200104 |
| Trade/Device Name: | Oxylight |
| Submitter: | RAJA Trading Company, Inc. |
## 4.0 Indications for Use
The Phototherapy System use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.
The blue light (469nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
The red light (633nm wavelength) is generally indicated to treatment of superficial,
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a green shamrock with three leaves, which is a symbol often associated with Ireland and Saint Patrick's Day. The shamrock is surrounded by a light purple, incomplete circle, giving it a stylized and somewhat abstract appearance. The overall design is simple and clean, with a focus on the iconic shamrock shape.
benign vascular, and pigmented lesions.
The infrared light (835nm wavelength) is generally use for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
## 5.0 Device Description
The Phototherapy System OL-PDT950 is a vertical device which uses specific wavelengths of light, produced by LEDs (Light emitting diodes), to manage aesthetic conditions.
The device produces light in the red light region of the spectrum (633±15nm), in the blue light regions of the light spectrum (469±15nm) and infrared light region of light spectrum (835±15nm).
Five sets of LEDs panels are available for the device.
### 6.0 Discussion of Tests Performed
### 6.1 Clinical Tests
Clinical testing was not performed for the subject device as part of the submission.
### 6.2 Non-Clinical Tests
The following performance data are provided in support of the substantial equivalence determination:
IEC 60601-1: 2005 + CORR.1:2006 + CORR.2:2007 + A1:2012: Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance;
IEC 60601-1-2: 2014: Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility
IEC60601-2-57:2021: Medical electrical equipment -- Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use:
IEC 62471: 2006: Photobiological safety of lamps and lamp systems.
## 7.0 Software
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a green shamrock inside of a white circle. The white circle is surrounded by a light purple ring. The shamrock is a traditional Irish symbol, often associated with Saint Patrick's Day.
# 8.0 Comparison of Technological Characteristics with the Predicate Device
| Device Feature | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K230342 | K200751 | K222751 | K200104 |
| Name | Phototherapy System | Photodynamic Therapy<br>System | LED Light Therapy Device | OxyLight |
| Product Code | GEX | GEX | GEX | GEX |
| Indications for use | The Phototherapy System<br>use of the red, blue and<br>infrared regions of the<br>spectrum is intended to<br>emit energy to treat<br>dermatological conditions.<br>The blue light (469nm<br>wavelength) is generally<br>indicated to treat<br>dermatological conditions<br>and specifically indicated to<br>treat moderate<br>inflammatory acne vulgaris.<br>The red light (633nm<br>wavelength) is generally<br>indicated to treatment of<br>superficial, benign vascular,<br>and pigmented lesions.<br>The infrared light (835nm<br>wavelength) is generally<br>use for the temporary relief<br>of minor muscle and joint<br>pain, arthritis and muscle<br>spasm; relieving stiffness;<br>promoting the relaxation of | The Photodynamic Therapy<br>(PDT) Equipment use of<br>the red, blue and infrared<br>regions of the spectrum is<br>intended to emit energy to<br>treat dermatological<br>conditions.<br>The blue light (415nm<br>wavelength) is generally<br>indicated to treat<br>dermatological conditions<br>and specifically indicated to<br>treat moderate<br>inflammatory acne vulgaris.<br>The red light (630nm<br>wavelength) is generally<br>indicated to treatment of<br>superficial, benign vascular,<br>and pigmented lesions.<br>The infrared light (835nm<br>wavelength) is generally<br>use for the temporary relief<br>of minor muscle and joint<br>pain, arthritis and muscle<br>spasm; relieving stiffness; | LED Light Therapy Device<br>use of the red, blue, Yellow<br>and infrared regions of the<br>spectrum is intended to<br>emit energy to treat<br>dermatological conditions.<br>The red light (633±10nm<br>wavelength) is generally<br>indicated to treatment of<br>superficial, benign vascular,<br>and pigmented lesions.<br>The blue light (417±10 nm<br>wavelength) is generally<br>indicated to treat<br>dermatological conditions<br>and specifically indicated to<br>treat moderate<br>inflammatory acne vulgaris.<br>The Yellow light<br>(599±10nm wavelength) is<br>generally indicated to treat<br>dermatological conditions<br>and specifically indicated<br>for treatment of periorbital<br>wrinkles and rhytides. | The Oxylight is intended<br>for dermatological use by<br>physicians and healthcare<br>professionals for the<br>following:<br>LED Technology is<br>intended for:<br>Blue LED - 465nm - to<br>treat dermatological<br>conditions and specifically<br>indicated to treat moderate<br>inflammatory acne vulgaris.<br>Red LED 625nm - for<br>treatment of superficial,<br>benign vascular and<br>pigmented lesions.<br>Yellow LED 590nm -<br>treatment of periorbital<br>wrinkles and rhytides. |
Table 1 - General Comparison
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a green shamrock inside of a purple circle. The shamrock is a three-leaf clover, and it is the national symbol of Ireland. The purple circle is slightly blurred, and it appears to be a halo effect around the shamrock. The image is simple and clean, and it is likely used as a logo or icon.
# Shanghai Omni Laser Skinology Co., Ltd.
| muscle tissue; and to<br>temporarily increase local<br>blood circulation where<br>applied. | promoting the relaxation of<br>muscle tissue; and to<br>temporarily increase local<br>blood circulation where<br>applied. | The infrared light<br>(835±15nm wavelength) is<br>generally use for the<br>temporary relief of minor<br>muscle and joint pain,<br>arthritis and muscle spasm;<br>relieving stiffness;<br>promoting the relaxation of<br>muscle tissue; and to<br>temporarily increase local<br>blood circulation where<br>applied. |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
# Table 2 - Performance Comparison
| Device Feature | Subject Device (K230342) | Predicate Device 1<br>(K200751) | Predicate Device 2<br>(K222751) | Predicate Device 3<br>(K200104) |
|-----------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wavelength<br>(nm) | Red light 633nm±15nm<br>Blue light 469nm±15nm<br>Infrared light 835nm±15nm | Red light 630±15<br>Blue light 415±15<br>Infrared light 835±15 | Red light 633nm±15nm<br>Blue light 417nm±15nm<br>Infrared light 835nm±15nm | Red light 625nm±5nm<br>Blue light 465nm±5nm<br>Yellow light 590nm±5nm |
| | Panels Type | 5 panels: 2400 EA LEDs.<br>The panels may emit the<br>three light (red, blue,<br>infrared)<br>Blue: 800EA,<br>Red: 800EA,<br>Infrared: 800EA | 5panel: 300 EA LEDs.<br>The panels may emit the<br>three light (red, blue,<br>infrared) | RBY Irradiator has 5 panels:<br>Red light:465EA LEDs;<br>blue light: 470EA LEDs;<br>yellow light: 465EA LEDs;<br>RBI irradiator: has 5<br>panels:<br>Red light:465EA LEDs;<br>blue light: 470EA LEDs;<br>infrared: 465EA LEDs; The<br>panels may emit the three<br>light (red, blue infrared)<br>individual or in |
| | | | combination | Shanghai Omni Laser Skinology Co., Ltd. |
| LED power | Each LED lamp bead has 1 | Each LED lamp bead has 3 | Each panel has three<br>different kinds of light-<br>emitting diodes, and the | Unknown |
| | diodes that emit single | diodes that emit different | energy power of the diode | |
| | color, the energy power of a | colors, the Energy power of | is 0.5W | |
| | diode is: | a diode is 3W. | | |
| | Red light: 120mw | | | |
| | Blue light: 160mw | | | |
| | Infrared light: 100mw | | | |
| | Red light: 75mW/cm² | Red light: 115 mW/cm² | Red light: 20~96 mW/cm² | Red light: 100mW/cm² |
| | Blue light: 35 mW/cm² | Blue light: 120 mW/cm² | Blue light: 10~120<br>mW/cm² | Blue light: 45 mW/cm² |
| Maximum<br>power<br>density | IR: 120 mW/cm² | IR: 70 mW/cm² | Infrared: ≤ 70 mw/cm² | Yellow light: 35 mW/cm² |
| | | | Red/IR: 166mW/cm² | |
| | | | Blue/IR: 190mW/cm² | |
| Treatment area | 1228 cm² | 1410 cm² | 900 cm²±10% cm² | 500cm² and 860cm² |
| Treatment time | 20 minutes (recommended | 20 minutes (recommended | 20 minutes (recommended | 20 minutes (recommended |
| | Treatment Time) | Treatment Time) | Treatment Time) | Treatment Time) |
| Working<br>distance | 10~15cm | 10~15cm | Unknown | Unknown |
| Power supply | AC<br>100-240V<br>50/60Hz<br>440VA | AC<br>100-240V<br>50/60Hz<br>440VA | AC 100-240V 50/60Hz | Unknown |
| Operation<br>interface | Display Screen | Display Screen | Display Screen | Display Screen |
| Electrical<br>Safety | Comply with IEC 60601-1<br>and IEC 60601-1-2 | Comply with IEC 60601-1<br>and IEC 60601-1-2 | Comply with IEC 60601-1<br>and IEC 60601-1-2 | Comply with IEC 60601-1<br>and IEC 60601-1-2 |
| Radiation Safety | Comply with IEC 60601-2-<br>57 | Comply with IEC 60601-2-<br>57 | Comply with IEC 60601-2-<br>57 | Comply with IEC 60601-2-<br>57 |
| Photobiological<br>safety | Comply with IEC 62471 | Comply with IEC 62471 | Comply with IEC 62471 | Comply with IEC 62471 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a green shamrock inside of a light purple circle. The shamrock is a three-leaf clover, and it is positioned in the center of the circle. The circle is not fully closed, and it appears to be a crescent shape. The image is simple and clean, with a focus on the shamrock as the main subject.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a green shamrock with a light purple arc behind it. The shamrock is a three-leaf clover, a symbol often associated with Ireland and Saint Patrick's Day. The arc is positioned behind the shamrock, creating a sense of depth. The image is simple and clean, with a focus on the iconic shamrock symbol.
# 9.0 Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.