← Product Code [FDF](/productcode/FDF) · K230332

# Ambu® aScope™ Colon; Ambu® aBox™ 2 (K230332)

_Ambu A/S · FDF · Sep 15, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230332

## Device Facts

- **Applicant:** Ambu A/S
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** Sep 15, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Ambu® aScope™ Colon: The aScope™ Colon is a sterile, single-use, flexible colonoscope intended to be used for endoscopic access to and examination of the lower gastrointestinal anatomy. The aScope™ Colon is intended to provide visualization via a compatible Ambu displaying unit and to be used with endotherapy accessories and other ancillary equipment. Ambu® aBox™ 2: The aBox™ 2 is intended to display live imaging data from compatible Ambu visualization devices.

## Device Story

System comprises sterile, single-use flexible colonoscope (aScope Colon) and reusable displaying unit (aBox 2). Colonoscope features maneuverable tip, control handle, working channel, and integrated camera with LED illumination; provides insufflation, suction, rinsing, and water jet functions. Displaying unit processes and displays live video from colonoscope; supports snapshot/video recording, brightness/contrast/sharpness/zoom adjustments, and external monitor connectivity. Used in clinical settings for lower GI examination; operated by trained clinicians. Single-use design eliminates reprocessing requirements; portable aBox 2 includes integrated monitor. Output allows real-time visualization of GI anatomy, facilitating diagnostic or therapeutic procedures via endotherapy accessories.

## Clinical Evidence

Bench testing only. Evaluated geometrical characteristics, functional performance (insufflation, suction, rinsing, water jet), optical performance (FOV, sharpness, distortion, color, noise), photobiological safety (IEC 62471), biocompatibility (ISO 10993-1), sterilization (ISO 11135), electrical safety (IEC 60601-1, IEC 60601-2-18), and EMC (IEC 60601-1-2). All tests met specifications.

## Technological Characteristics

Flexible endoscope with bowden-wire maneuverable tip. Integrated camera module with LED illumination. Reusable displaying unit with integrated monitor. Standards: ISO 10993-1 (biocompatibility), ISO 11135 (sterilization), IEC 62471 (photobiological safety), IEC 60601-1/60601-2-18 (electrical safety), IEC 60601-1-2 (EMC).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- OLYMPUS EVIS EXERA II Colonovideoscope CF-H180AL ([K100584](/device/K100584.md))
- OLYMPUS EVIS EXERA II Light Source (CLV-180) and Video System Center (CV-180) ([K100584](/device/K100584.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 15, 2023

Ambu A/S % Sanjay Parikh Director, QA/RA Ambu Inc. 6721 Columbia Gateway Drive, Suite 200 Columbia, Maryland 21046

Re: K230332

Trade/Device Name: Ambu® aScope™ Colon; Ambu® aBoxTM 2 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF, FET Dated: August 11, 2023 Received: August 15, 2023

Dear Sanjay Parikh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Sivakami Venkatachalam -S

for

Shani P. Haugen, PhD Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K230332

Device Name Ambu® aScope™ Colon Ambu® aBox™ 2

#### Indications for Use (Describe)

Ambu® aScope™ Colon:

The aScope™ Colon is a sterile, single-use, flexible colonoscope intended to be used for endoscopic access to and examination of the lower gastrointestinal anatomy.

The aScope™ Colon is intended to provide visualization via a compatible Ambu displaying unit and to be used with endotherapy accessories and other ancillary equipment.

Ambu® aBox™ 2:

The aBox™ 2 is intended to display live imaging data from compatible Ambu visualization devices.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Submitter                                                                                          | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                            |
|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person                                                                                     | Name: Mette Andersen<br>Job Title: Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, 2750 Ballerup, Denmark<br>Business Phone: +45 5381 3820<br>Email: meta@ambu.com                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                            |
| Date Summary<br>Prepared                                                                           | February 7th, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                            |
| Device Trade<br>Name                                                                               | Ambu® aScope™ Colon,<br>Ambu® aBox™ 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                            |
| Device Common<br>Name                                                                              | Endoscopy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                            |
| Device<br>Classification                                                                           | Ambu® aScope™ Colon:<br>Colonoscope And<br>Accessories, Flexible/rigid<br>Product Codes: FDF<br>21 CFR 876.1500<br>Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Ambu® aBox™ 2:<br>Colonoscope And<br>Accessories, Flexible/rigid<br>Product Codes: FDF, FET<br>21 CFR 876.1500<br>Class II |
| Legally<br>Marketed<br>devices to which<br>the device is<br>substantially<br>equivalent            | Predicate Device:<br>OLYMPUS EVIS EXERA II<br>Colonovideoscope CF-<br>H180AL<br>K100584                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device:<br>OLYMPUS EVIS EXERA II Light<br>Source (CLV-180) and Video<br>System Center (CV-180)<br>K100584        |
| Description of<br>the Device                                                                       | The Ambu® aScope™ Colon Endoscopy System is a system used<br>for endoscopic procedures in the gastrointestinal anatomy. It<br>consists of a sterile, single-use, flexible endoscope, the Ambu®<br>aScope™ Colon, and a displaying unit, the Ambu® aBox™ 2.<br><br>The Ambu® aScope™ Colon is a sterile, single-use flexible<br>colonoscope for accessing and examining the lower<br>gastrointestinal anatomy. The endoscope provides a working<br>channel for use of endotherapy accessories, as well as insufflation<br>suction, rinsing and a water jet function. Visualization is realised |                                                                                                                            |
|                                                                                                    | via an integrated camera module with bulit-in LEDs for<br>illumination.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                            |
|                                                                                                    | The Ambu® aBox™ 2 displaying unit has the following physical and<br>performance characteristics:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                            |
|                                                                                                    | Displays the image from Ambu® aScope™ Colon endoscope on the screen Can record snapshots or video of image from Ambu® aScope™ Colon endoscope Can connect to an external monitor Is a reusable device                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                            |
| Indications for<br>Use                                                                             | The Ambu® aScope™ Colon is a sterile, single-use flexible<br>colonoscope intended to be used for endoscopic access to and<br>examination of the lower gastrointestinal anatomy.<br>The Ambu® aScope™ Colon is intended to provide visualization via<br>a compatible Ambu displaying unit and to be used with<br>endotherapy accessories and other ancillary equipment.<br>The aBox™ 2 is intended to display live imaging data from<br>compatible Ambu visualization devices.                                                                                                                 |                                                                                                                            |
| Summary of the<br>technological<br>characteristics<br>in comparison to<br>the predicate<br>devices | The Ambu® aScope™ Colon and its predicate device have the<br>following technological characteristics:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                            |
|                                                                                                    | Both are flexible endoscopes with maneuverable tip, a control section (handle) and an umbilical cord Both control the tip bending via two wheels at the handle and bowden wires. Both provide a working channel Both have same technological characteristics as insertion portion length, working channel diameter, direction of view and bending angels Unlike the predicate device, the Ambu® aScope™ Colon is a sterile, single-use device and not intended to be reprocessed.                                                                                                             |                                                                                                                            |
|                                                                                                    | The Ambu® aBox™ 2 and its predicate device have the following<br>technological characteristics:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                            |
|                                                                                                    | Both are video processors displaying live video-imaging data of the connected visualization device to a monitor. Both provide video output formats, recording and data storage and data transport functions. Both share certain technical functionalities as brightness control, image contrast and sharpness adjustment as well as zoom function. Contrary to the predicate device, the Ambu® aBox™ 2 is portable and has an integrated monitor, therefore, an external monitor is not necessary                                                                                             |                                                                                                                            |
| Performance<br>Data -Bench                                                                         | The following tests to verify/validate the design and evaluate the<br>performance of the Ambu® aScope™ Colon Endoscopy System<br>were done.                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                            |
| •                                                                                                  | Geometrical characteristics including                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                            |

{4}------------------------------------------------

Page 2 of 3

{5}------------------------------------------------

- Length of insertion tube, umbilical cord, tip O
- Outer diameter of bending section, insertion tube O and overlap of both
- Tip reach O
- Bending angles o
- Working channel width о
- Functional performance including .
	- Insufflation o
	- Suction o
	- Rinsing о
	- Water Jet O
- Optical performance including
	- Field of view o
	- Direction of view O
	- Sharpness and Depth of field O
	- Geometric distortion o
	- Image intensity uniformity O
	- Color performance O
	- Noise characterization O
	- Dynamic range O
	- Camera view orientation O
- Photobiological safety according to IEC 62471 ●
- Biocompatibility according to ISO 10993-1 including ● cytotoxicity, irritation, and sensitization
- Sterilization validation according to ISO 11135
- Transport validation including packaging integrity ●
- Stability study to document shelf life
- Electrical Safety and performance according to IEC 60601-● 1 and IEC 60601-2-18
- . Electromagnetic Compatibility according to IEC 60601-1-2
- Tests to confirm procedural performance ●

In all instances, the Ambu® aScope™ Colon Endoscopy System performed as expected and met the set test specifications.

Conclusion

The Ambu® aScope™ Colon Endoscopy System, consisting of Ambu® aScope™ Colon and Ambu® aBox™2, has the same intended use and indications for use, and similar technological characteristics and principles of operation as the predicate devices.

The minor technological differences between the Ambu® aScope™ Colon Endoscopy System and its predicate devices raise no new concerns regarding safety or effectiveness.

Thus, the Ambu® aScope™ Colon Endoscopy System, consisting of Ambu® aScope™ Colon and Ambu® aBox™ 2, is substantially equivalent to its predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K230332](https://fda-staging.innolitics.com/device/K230332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).