NIM Standard Reinforced EMG Endotracheal Tube; CONTACT Reinforced EMG Endotracheal Tube
K230320 · Medtronic Xomed, Inc. · ETN · Oct 26, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K230320
Device Name
NIM Standard Reinforced EMG Endotracheal Tube; CONTACT Reinforced EMG Endotracheal Tube
Applicant
Medtronic Xomed, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Oct 26, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.1820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EMG Endotracheal tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the intrinsic laryngeal musculature when connected to an appropriate EMG monitor.
Device Story
Flexible, reinforced silicone endotracheal tube with inflatable cuff; features integrated stainless steel wire electrodes on main shaft. Used during surgery to provide airway ventilation and monitor laryngeal musculature EMG activity. Connected to external EMG monitoring device; electrodes contact laryngeal muscles near vocal cords. Operated by anesthesiologists or nurse anesthetists in surgical settings. Output displayed on external monitor to assist clinicians in identifying/mapping recurrent laryngeal nerve integrity. Helps prevent nerve damage during procedures. Modifications involve updated labeling/IFU regarding airway obstruction management and cuff inflation tools.
Clinical Evidence
No clinical data. Substantial equivalence supported by nonclinical summative usability validation and labeling design validation. Testing confirmed anesthesiologists/nurse anesthetists could perform intubation and maintain airway without unmitigated use errors, and that labeling effectively communicates risk controls.
Technological Characteristics
Silicone elastomer tube and cuff; stainless steel wire electrodes; stainless steel reinforcement. EtO sterilization. Dimensions: ~12.8 inches length, 6-8mm ID. Standalone device requiring connection to external EMG monitor.
Indications for Use
Indicated for continuous monitoring of nerves supplying laryngeal musculature during surgical procedures. Not intended for postoperative use.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 26, 2023
Medtronic Xomed, Inc. Emily Davis Principal Regulatory Affairs Specialist 6743 Southpoint Drive North Jacksonville, Florida 32216
Re: K230320
Trade/Device Name: NIM Standard Reinforced EMG Endotracheal Tube; CONTACT Reinforced EMG Endotracheal Tube Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: ETN Dated: August 29, 2023 Received: August 30, 2023
Dear Emily Davis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230320
Device Name
NIM™ Standard Reinforced EMG Endotracheal Tube and NIM CONTACT™ Reinforced EMG Endotracheal Tube
Indications for Use (Describe)
The EMG tube is indicated for use where continuous monitoring of the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Medtronic
Medtronic Xomed, Inc. 6743 Southpoint Drive N. Jacksonville, FL 32216 United States of America www.medtronic.com Tel. 904-296-9600
### 510(k) Summary
This 510(k) Summary was prepared in accordance with 21 CFR 807.92.
| Submitted By: | Medtronic Xomed, Inc.<br>6743 Southpoint Drive North<br>Jacksonville, Florida 32216 USA<br>Telephone Number: (904) 296-9600 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K230320 |
| Date Prepared: | 10/26/2023 |
| Official Correspondent: | Alexandra Oliver<br>Regulatory Affairs Manager<br>Telephone: (904) 332-8936<br>Email Address:<br>Alexandra.j.oliver@medtronic.com |
| Alternate Contact: | Emily Davis<br>Principal Regulatory Affairs Specialist<br>Telephone: (949) 636-4621<br>Email Address:<br>emily.r.davis@medtronic.com |
| I. DEVICE | |
| Proprietary (Trade) Names: | NIMTM Standard Reinforced EMG Endotracheal Tube,<br>NIM CONTACTTM Reinforced EMG Endotracheal Tube<br>Stimulator |
| Common Name: | Surgical nerve stimulator/locator |
| Regulation Name: | 21 CFR 874.1820 |
| Regulation Number: | ETN |
| Product Code: | II |
| Device Classification: | Ear, Nose, & Throat (ENT) |
| Review Panel: | |
### II. PREDICATE DEVICE(S)
The NIM™ Standard Reinforced EMG Endotracheal Tube and NIM CONTACT™ EMG Reinforced Endotracheal Tube are substantially equivalent in intended use and technological characteristics to the following predicate device(s):
| 510(k) Number | 510(k) Device Name | Product<br>Code | Regulation<br>Number | 510(k)<br>Clearance Date |
|---------------|---------------------------------------------|-----------------|----------------------|--------------------------|
| K925640 | Xomed-Treace EMG<br>Endotracheal Tube | ETN | 21 CFR 874.1820 | August 17, 1993 |
| K050162 | CONTACT™ EMG<br>Rotatable Endotracheal Tube | ETN | 21 CFR 874.1820 | March 3, 2005 |
{4}------------------------------------------------
# Medtronic
## III. DEVICE DESCRIPTION
### Device Description
Medtronic Xomed, Inc.'s NIM™ Standard Reinforced and NIM CONTACT™ Reinforced Endotracheal Tubes are flexible, reinforced endotracheal tubes with inflatable cuffs. The NIM EMG ET Tubes are made from silicone elastomer. Each tube is fitted with electrodes on the main shaft, which are exposed only for a short distance, slightly superior to the cuff. The electrodes are designed to make contact with the laryngeal muscles around the patient's vocal cords to facilitate electromyographic (EMG) monitoring of the laryngeal musculature during surgery when connected to an EMG monitoring device. Both the tube and cuff are manufactured from material that allows the tube to conform readily to the shape of the patient's trachea with minimal trauma to tissues.
### Intended Use
The EMG Endotracheal tube is intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the intrinsic laryngeal musculature when connected to an appropriate EMG monitor.
#### IV. INDICATIONS FOR USE
### Indications for Use
The EMG tube is indicated for use where continuous monitoring of the nerves supplying the laryngeal musculature is required during surgical procedures. The EMG tube is not intended for postoperative use.
#### V. TECHNOLOGICAL COMPARISON TO THE PREDICATE DEVICE(S)
The EMG Endotracheal tube is substantially equivalent to the predicate devices based on comparison of indications for use and technological characteristics where additions to the subject devices instructions for use were made being the only difference. Technological Characteristics of our device compared to each predicate is shown in the following table and in the Substantial Equivalence section of this submission.
{5}------------------------------------------------
# Medtronic
| Feature/Attribute | NIM™ Standard<br>Reinforced EMG<br>Endotracheal Tube<br>& NIM<br>CONTACT™<br>Reinforced EMG<br>Endotracheal Tube<br>(Subject Device(s)) | Xomed-Treace EMG<br>Endotracheal Tube /<br>K925640<br>(Predicate Device) | CONTACT TM EMG<br>Rotatable ET Tube / K050162<br>(Predicate Device) |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | ETN | ETN | ETN |
| Regulation Number | 21 CFR 874.1820 | 21 CFR 874.1820 | 21 CFR 874.1820 |
| Regulation<br>Description | Surgical nerve<br>stimulator/locator | Surgical nerve<br>stimulator/locator | Surgical nerve stimulator/locator |
| Classification | Class II | Class II | Class II |
| Common Name | Nerve Stimulator | Nerve Stimulator | Nerve Stimulator |
| Indications for Use | The EMG tube is<br>indicated for use where<br>continuous monitoring<br>of the nerves supplying<br>the laryngeal<br>musculature is required<br>during surgical<br>procedures. The EMG<br>tube is not intended or<br>postoperative use. | Intraoperative<br>electromyographic<br>monitoring of the vocal<br>cords to facilitate locating<br>and mapping of the<br>recurrent laryngeal nerve<br>in surgical procedures<br>involving the laryngeal<br>region of the throat. | The CONTACTTM EMG<br>Rotatable Endotracheal Tube is<br>intended for use as a means of<br>providing both an open airway for<br>patient ventilation and for<br>intraoperative monitoring of EMG<br>activity of the intrinsic laryngeal<br>musculature when connected to an<br>appropriate EMG monitor.<br>The EMG Tube is indicated<br>for use where continuous<br>monitoring of the integrity<br>of the nerves supplying the<br>laryngeal musculature is<br>required during surgical<br>procedures. |
| Tube Lengths | Approximately 12.8<br>inches | Approximately 12.8<br>inches | Approximately 12.8 inches |
| Tube Inner<br>Diameter (ID) | 6mm-8mm | 4mm-8mm | 6mm-8mm |
| Number of<br>Laryngeal Surface<br>Electrodes | 4 | 4 | 4 |
| Electrode Surface | Stainless steel wire | Stainless steel wire | Stainless steel wire |
| Tube Material | Silicone | Silicone | Silicone |
| Cuff Material | Silicone | Silicone | Silicone |
| Reinforcing<br>Material | Stainless Steel | Stainless Steel | Stainless Steel |
| Sterilization Method | EtO | EtO | EtO |
{6}------------------------------------------------
### Substantial Equivalence Discussion
The purpose of this submission is to gain clearance for proposed modifications to the labeling (Instructions for Use (IFU) and product labels) for the NIM™ Standard Reinforced and NIM CONTACT™ Reinforced Endotracheal Tubes. The NIM™ EMG ET Tubes are intended for use as a means of providing both an open airway for patient ventilation and for intraoperative monitoring of EMG activity of the intrinsic laryngeal musculature when connected to an appropriate EMG monitor. As the intended use remains the differences in labeling between the subject and predicate devices are additional warnings and precautions in the IFU for actions to be taken if airway obstruction is encountered, with additional symbols on product labeling for clarification on EMG ET Tube material and recommended tool for cuff inflation. These modified labeling differences are critical to the intended surgical use of the device when used as labeled and have been evaluated by Human Factors Testing and a Labeling Design Validation study which were found to improve the safety and efficacy. Both the subject device(s) and the predicate device(s) share similar characteristics (design, materials, operating principle, energy source, and performance).
## VI. PERFORMANCE DATA
### Performance Testing Discussion
The following nonclinical testing was performed to establish substantial equivalence to the predicates: Summative Usability Validation and Labeling Design Validation. The Objective of this Summative Usability Validation testing was to accomplish:
- 1. Achieve the Usability Goal: The Anesthesiologist/Nurse Anesthetist shall be able to intubate the patient and maintain the airway without introducing any unrealized use errors or critical tasks;
- 2. Confirm the Critical tasks were completed without any unacceptable Use Error that may have resulted in unmitigated potential hazards; and
- 3. To show the risk mitigations were effective in regard to labeling and training.
Additionally, the objective of the Labeling Design Validation was to validate that the NIM EMG Tube IFU and product labels met the User Need which was that the product labeling was to be understandable and provide needed information for proper safe and effective use of the device.
The results of these validations with the modified proposed labeling demonstrated that the usability goal was achieved, all critical tasks were completed without introducing any additional unmitigated hazards, the user need and risk control measures were met and the training mitigations proposed were effective. This nonclinical testing demonstrated that the subject device is substantially equivalent to the predicate device.
### Clinical Testing
No clinical test data is required of the proposed device.
{7}------------------------------------------------
### VII. CONCLUSION
The NIM™ Standard Reinforced and NIM CONTACT™ Reinforced Endotracheal Tubes is concluded to be substantially equivalent to the predicates as the identified differences do not raise new or different concerns or questions of safety and effectiveness relative to the predicate device(s). Therefore, it is the opinion of the submitter that the NIM™ Standard Reinforced and NIM CONTACT™ Reinforced Endotracheal Tubes are as safe, as effective and performs as well as the legally marketed predicated devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.