CoverScan is a medical image management and processing software package that allows the display, analysis and post-processing of DICOM compliant medical images and MR data. CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple organs such as the heart, lungs, liver, spleen, pancreas and kidney. CoverScan provides measurements in different organs to be used for the assessment of longitudinal and transversal relaxation time (T1, srT1, T2), fat content (proton density fat fraction or PDFF) and metrics of organ function (e.g., left ventricular ejection fraction and lung fractional area change on deep inspiration). These metrics derived from the images, when interpreted by a licensed physician, yield information that may assist in diagnosis, clinical management and monitoring of patients.
Device Story
CoverScan v1.1 is a post-processing software system for DICOM-compliant MR data; processes images to quantify tissue characteristics and organ function (heart, lungs, liver, spleen, pancreas, kidney). Inputs: MR images from Siemens 1.5T/3T scanners. Operation: software modules perform segmentation, relaxometry (T1, T2, T2*), and fat fraction (PDFF) mapping using IDEAL methodology; cardiac modules calculate ejection fraction, volumes, and mass. Used in clinical settings by trained internal operators at Perspectum image centers; outputs are quantitative reports for clinician interpretation. Benefits: assists in diagnosis, clinical management, and monitoring of patients by providing standardized, quantified organ metrics.
Clinical Evidence
Bench testing only. Performance assessed using purpose-built phantoms and in-vivo acquired data from volunteers covering a range of physiological values for sT1, cT1, T1, and PDFF. Verification and validation testing confirmed product specifications and system capabilities.
Technological Characteristics
Post-processing software; runs on Mac OS workstations. Supports DICOM 3.0 MR data. Implements IDEAL (Iterative Decomposition of water and fat with Echo Asymmetric and Least-squares estimation) for PDFF. Standards: IEC 62304, IEC 62366, ISO 14971, ISO 13485.
Indications for Use
Indicated for patients requiring analysis of DICOM-compliant MR images of the heart, lungs, liver, spleen, pancreas, and kidney. Not intended for asymptomatic screening. Intended for use with Siemens 1.5T and 3T MRI scanners.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
March 3, 2023
Perspectum Ltd. % Ioan Wigley Head of Regulatory Affairs Gemini One 5220 John Smith Drive Oxford, Oxfordshire OX4 2LL United Kingdom
Re: K230294
Trade/Device Name: CoverScan (CoverScan v1.1) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: November 28, 2022 Received: February 2, 2023
Dear Ioan Wigley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Dghk
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K230294
Device Name
CoverScan (CoverScan v1.1)
Indications for Use (Describe)
CoverScan is a medical image management and processing software package that allows the display, analysis and post-processing of DICOM compliant medical images and MR data.
CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple organs such as the heart. lungs. liver, spleen, pancreas and kidney.
CoverScan provides measurements in different organs to be used for the assessment of longitudinal and transversal relaxation time (T1. srT1. T2), fat content (proton density fat fraction or PDFF) and metrics of organ function (e.g., left ventricular ejection fraction and lung fractional area change on deep inspiration).
These metrics derived from the images, when interpreted by a licensed physician, yield information that may assist in diagnosis, clinical management and monitoring of patients.
CoverScan is not intended for asymptomatic screening. This device is intended for use with Siemens 1.5T and Siemens 3T MRI scanners.
Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
RA0913
Image /page/3/Picture/2 description: The image shows the logo for Perspectum. The word "Perspectum" is written in a bold, sans-serif font in blue color. To the right of the word is a circular graphic divided into four colored sections: light blue, green, pink, and dark blue. The logo has a clean and modern design.
K230294
| Date Prepared: | 28 November 2022 |
|----------------|------------------|
|----------------|------------------|
## Submitter Details
Owner Address: Gemini One, 5520 John Smith Drive, Oxford Business Park, Oxford. OX4 2LL
Perspectum Ltd
Contact Person:
loan Wigley
# Subject and Predicate Device
| | Predicate Device | Subject Device |
|-------------------------|----------------------------------------|----------------------------------------|
| 510(k) number | K212565 | Unknown |
| Legal Manufacturer | Perspectum Ltd. | Perspectum Ltd. |
| Owner/Operator Number | 10056574 | 10056574 |
| Device Name | CoverScan v1.0 | CoverScan v1.1 |
| Proprietary/Common name | CoverScan | CoverScan |
| 510k Review Panel | Radiology | Radiology |
| Regulation Number | 892.2050 | 892.2050 |
| Risk Class | Class II | Class II |
| Product Class code | LLZ | LLZ |
| Classification | System, Image Processing, Radiological | System, Image Processing, Radiological |
## Device Name
| Device Trade Name | Common Name |
|-------------------|-------------------------|
| CoverScan v1.1 | CoverScan, CoverScan v1 |
# Device Description
CoverScan is a post-processing software system comprised of several software modules. It uses acquired MR data to produce metrics of quantified tissue characteristics of the heart, lungs, liver, kidneys, pancreas and spleen.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Perspectum" in a bold, sans-serif font. The word is colored in a dark blue. To the right of the word is a circular logo that is divided into four sections. The top three sections are colored light blue, green, and pink, while the bottom section is colored dark blue.
Metrics produced by CoverScan can be used by healthcare professionals (HCPs) in a clinical setting for the purposes of assessing multiple organs.
CoverScan v1.1 is a new version of CoverScan that supports the use of Siemens 3T scan data to enable more scan sites to utilise CoverScan.
# Intended Use
CoverScan is a medical image management and processing software package that allows the display, analysis and post-processing of DICOM compliant medical images and MR data.
CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple organs such as the heart, lungs, liver, spleen, pancreas and kidney.
CoverScan provides measurements in different organs to be used for the assessment of longitudinal and transversal relaxation time (T1, cT1, T2), fat content (proton density fat fraction or PDFF) and metrics of organ function (e.g., left ventricular ejection fraction and lung fractional area change on deep inspiration).
These metrics derived from the images, when interpreted by a licensed physician, yield information that may assist in diagnosis, clinical management and monitoring of patients.
# Indications for use
CoverScan is not intended for asymptomatic screening. This device is intended for use with Siemens 1.5T and 3T MRI scanners.
# Substantial equivalence
The modified CoverScan v1.1 device has the following similarities to the previously cleared CoverScan v1 device.
| Characteristic | Subject and Predicate Device Comparison | |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CoverScan v1.0 (Predicate device) | CoverScan v1.1 (Subject Device) |
| Product Code | LLZ | LLZ |
| Regulation<br>Number | 892.2050 | 892.2050 |
| Class | II | II |
| Intended Use &<br>Indications for<br>Use | CoverScan is a medical image<br>management and processing software<br>package that allows the display, analysis<br>and postprocessing of DICOM compliant<br>medical images and MR data.<br>CoverScan provides both viewing and<br>analysis capabilities to ascertain quantified<br>metrics of multiple organs such as the<br>heart, lungs, liver, spleen, pancreas and<br>kidney.<br>CoverScan provides measurements in<br>different organs to be used for the<br>assessment of longitudinal and transversal<br>relaxation time and rate (T1, SR-T1, cT1, T2),<br>fat content (proton density fat fraction or<br>PDFF) and metrics of organ function (e.g.,<br>left ventricular ejection fraction and lung<br>fractional area change on deep<br>inspiration).<br>These metrics derived from the images,<br>when interpreted by a licensed physician,<br>yield information that may assist in<br>diagnosis, clinical management and<br>monitoring of patients. | CoverScan is a medical image<br>management and processing<br>software package that allows the<br>display, analysis and post-processing<br>of DICOM compliant medical images<br>and MR data.<br>CoverScan provides both viewing<br>and analysis capabilities to ascertain<br>quantified metrics of multiple organs<br>such as the heart, lungs, liver, spleen,<br>pancreas and kidney.<br>CoverScan provides measurements<br>in different organs to be used for the<br>assessment of longitudinal and<br>transversal relaxation time (T1, SR-T1,<br>cT1, T2), fat content (proton density<br>fat fraction or PDFF) and metrics of<br>organ function (e.g., left ventricular<br>ejection fraction and lung fractional<br>area change on deep inspiration).<br>These metrics derived from the<br>images, when interpreted by a<br>licensed physician, yield information<br>that may assist in diagnosis, clinical<br>management and monitoring of<br>patients. |
| Indications for<br>use | CoverScan is not intended for<br>asymptomatic screening. This device is<br>intended for use with Siemens 1.5T MRI<br>scanners | CoverScan is not intended for<br>asymptomatic screening. This device<br>is intended for use with Siemens 1.5T<br>and Siemens 3T MRI scanners. |
| The Subject device underwent additional performance testing to show that the performance was the<br>same as the predicate. | | |
| Limitations of<br>Use | Indicated where MRI is not<br>contraindicated. | Same as predicate |
| Device Users | Trained Perspectum internal operators. | Same as predicate |
| Use Environment | Installation of Modules 1-5 of CoverScan<br>are installed on general purpose<br>workstations at Perspectum's image | |
| | Module 6 of CoverScan is hosted on<br>Amazon Web Services (AWS) there is no<br>user interface for these modules. | |
| Clinical Setting | CoverScan is a software device that is<br>intended to be installed on general<br>workstations at Perspectum's image<br>centre. The intended device users will log<br>on to the workstations, access the device,<br>and use the device on general-use HD<br>monitors.<br>CoverScan is a post-processing software,<br>the intended device users are trained<br>Perspectum internal operators. Operators<br>use CoverScan to conduct quantitative<br>analysis of tissue characteristics and<br>function to produce a quantitative report.<br>The end-users for the output from the<br>device, the report, are clinicians who<br>receive and interpret reports. | Same as predicate |
| Principles of<br>Operation | CoverScan offers comprehensive<br>functionality for image analysis and<br>visualisation, CoverScan contains multiple<br>modules for the quantitative analysis of<br>tissue characteristics and function.<br>Visualisation and quantification tools for<br>image analysis depend on the module. | Same as predicate |
| | Module 1 (Liver module)<br>cT1<br>Full segmentation of the outer liver<br>contour and liver vasculature of the cT1<br>parametric map.<br>ROI placed method on the cT1 map with<br>IQR and median metrics from the placed<br>ROIs potentially across multiple acquired<br>slices.<br>PDFF<br>Full liver segmentation of the PDFF<br>parametric map where IQR and median<br>metrics are reported from the<br>segmentation.<br>ROI placed method on the PDFF map with<br>IQR and median metrics from the placed<br>ROIs potentially across multiple acquired<br>slices | Same as predicate |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Perspectum" in a bold, dark blue font. To the right of the word is a circular logo that is divided into several colored sections. The top half of the circle is divided into three sections colored light blue, green, and pink. The bottom half of the circle is a solid dark blue color.
analysis centre by specialist members of staff. Workstations need to meet the minimum technical requirements.
Same as predicate.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Perspectum" in a bold, sans-serif font. The word is in a dark blue color. To the right of the word is a circular logo that is divided into four sections. The top left section is light blue, the top right section is pink, the bottom section is dark blue, and the middle section is green.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Perspectum" in a dark blue font. To the right of the word is a circular logo that is divided into four colored sections: light blue, green, pink, and dark blue. The dark blue section of the logo also contains a smaller, concentric circle within it.
| PDFF parametric maps are calculated<br>using the LMS IDEAL method (1) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Module 2 (Pancreas module)<br>sT1 | |
| ROI placed method on the T1 map with IQR<br>and median metrics from the placed ROIs<br>potentially across multiple acquired slices<br>PDFF<br>ROI placed method on the PDFF map with<br>IQR and median metrics from the placed<br>ROIs potentially across multiple acquired<br>slices<br>PDFF parametric maps are calculated<br>using the LMS IDEAL method (1) | Same as predicate |
| Module 3 (Kidney module)<br>T1<br>ROI placed method on the T1 map with IQR<br>and median metrics from the placed ROIs<br>potentially across multiple acquired slices | Same as predicate |
| Module 4 (Cardiac function module)<br>Operators may use modules within<br>CoverScan to analyse and quantify cardiac<br>images with the below capabilities:<br>Left Ventricular function<br>Ejection fraction<br>End Diastolic Volume – left ventricle<br>End Systolic Volume - left ventricle | Same as predicate |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Perspectum" in a bold, dark blue font. To the right of the word is a circular logo that is divided into four sections. The top three sections are colored light blue, green, and pink, while the bottom section is dark blue. The logo is meant to represent a global perspective.
| | Stroke Volume<br>Left Ventricle Muscle Mass<br>Left Ventricular Wall Thickness<br>Global and regional LV function and<br>volume analysis<br>Global RV function analysis<br>T1 mapping Module<br>Assessment of native and post contrast T1<br>Relaxation times T1, T1*and R2 maps with<br>customizable color LUT and polar map<br>display<br>Assessment of ECV % per slice and<br>segment including polar map display and<br>map generation with customizable color<br>LUT<br>T2 Mapping Module<br>Global and Regional Segmental T2 times | |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| | Module 5 (Lung)<br>Basic calculations to determine the<br>percentage change in area from<br>inspiration to expiration from datasets<br>exported from analysis conducted in Osirix<br>MD. | Same as predicate |
| | Module 6 (Metric consolidation)<br>A compiled clinical report containing<br>metrics from modules 1-5, rounding of<br>these numbers along with a reference<br>range. Basic logic algorithm applied to<br>determine if there is evidence of heart<br>impairment from cardiac metrics exported<br>from analysis conducted. | Same as predicate |
| Performance<br>Features | Main software features:<br>Post-processing, display and allow<br>manipulation of medical MR images<br>Image loading and saving<br>Session file loading and saving<br>Image viewing<br>Image manipulation<br>Image analysis<br>Image processing<br>Relaxometry post-processing<br>Fat fraction postprocessing<br>Segmentation of regions of interest | Same as predicate |
| Design: MR<br>Relaxometry | Relaxometry post-processing (T1, T2 and<br>T2*). And subsequently cT1 and sT1. | Same as predicate |
| Design: Liver Fat<br>Quantification | Fat fraction postprocessing (PDFF) | Same as predicate |
| Design:<br>Parametric Maps | An operator can use modules 1-3 of<br>CoverScan to generate T2*, T2 and T1<br>relaxometry maps fitted from an<br>appropriate set of MR Inversion Recovery<br>images which are Gradient Echo (GRE) and<br>MOLLI acquisition protocols respectively.<br>Modules within CoverScan may also be<br>used to generate fat signal fraction (PDFF)<br>maps calculated from an appropriate set of<br>MR GRE images using the IDEAL (iterative<br>decomposition of water and fat with echo<br>asymmetric and least-squares estimation)<br>methodology (13).<br><br>Module 5 within CoverScan can be used to<br>calculate percentage change in area from<br>inspiration to expiration change using the<br>interface that offers tools:<br><br>3D rendering tools, such as Multiplanar<br>Reconstructions, Curved Reconstructions,<br>3D Volume Rendering, 3D Surface<br>Rendering, 3D Endoscopy<br>3D sculpting tools<br>Measure distance in 3D Volume<br>Rendering, 3D Curved-MPR or 3D<br>Orthogonal MultiPlanar<br>3D rigid registration<br><br>Module 4 within CoverScan can be used to<br>analyse and calculate cardiac metrics to<br>report:<br>Left Ventricular function<br><br>Ejection fraction<br>End Diastolic Volume – left ventricle<br>End Systolic Volume – left ventricle<br>Stroke Volume<br>Left Ventricle Muscle Mass<br>Left Ventricular Wall Thickness<br>Global and regional LV function and<br>volume analysis<br>Global RV function analysis<br>Assessment of native and post contrast T1<br>Relaxation times T1, T1* and R2 | Same as predicate |
| | Assessment of ECV % per slice and<br>segment<br>Global and Regional Segmental T2 times | |
| Design:<br>Visualisation | Offers numerous views within modules 1-5<br>of the CoverScan interface can be used to<br>assist in analysis, T1, T2* and Proton<br>Density Fat Fraction (PDFF)) parametric<br>maps can be created from all supported<br>scanners. R2 maps can also be utilised to<br>assess the quality of the map fitting.<br>Colormaps in the parametric maps are<br>designed to have maximum contrast on<br>organ tissue. | Same as predicate |
| Design:<br>Outputted data | Quantified metrics and images derived<br>from the analysis of liver tissue<br>characteristic on parametric maps are<br>collated into a datafile that may<br>subsequently be assembled into a report<br>for evaluation and interpretation by a<br>clinician.<br>Based on the placed ROI's (during analysis<br>in Modules 1-3), for each metric the median<br>and IQR are given as well as a 'reference<br>range'. | Same as predicate |
| Design:<br>Supported<br>Modalities | DICOM 3.0 compliant MR data from<br>supported MRI scanners. | Same as predicate |
| Performance<br>Testing | Perspectum has conducted extensive<br>validation testing of CoverScan, a medical<br>image management and processing<br>system (MIMPS), that is capable of<br>providing reliable post-processing and<br>display of images for instantaneous multi-<br>parametric analysis.<br><br>Internal verification and validation testing<br>confirms that the product specifications<br>are met.<br><br>All of the different components of the<br>CoverScan software have been stress<br>tested to ensure that the system as a<br>whole provides all the capabilities<br>necessary to operate according to its<br>intended use.<br><br>The main groups of tests performed<br>include:<br>Product Risk Assessment<br>Software modules verification tests<br>Software validation test<br><br>Device performance was assessed with<br>purpose-built phantoms and in-vivo<br>acquired data from volunteers covering a<br>range of physiological values for sT1, cT1,<br>T1 and PDFF. | Same as predicate |
| Human Factors | Assessed in accordance with IEC 62366<br>and FDA guidance document 'Applying<br>Human<br>Factors and Usability Engineering to<br>Medical Devices' | Same as predicate |
| Standards | IEC 62304, IEC 62366, DICOM 3.0, ISO<br>14971, ISO 13485 | Same as predicate |
| System/Operatin<br>g System | Mac OS | Same as predicate |
| Materials | Not applicable, post-processing software | Not applicable, post-processing<br>software |
| Energy Source | Not applicable, post-processing software | Not applicable, post-processing<br>software |
| Biocompatibility | Not applicable, post-processing software | Not applicable, post-processing<br>software |
| Sterility | Not applicable, post-processing software | Not applicable, post-processing<br>software |
| Electrical Safety | Not applicable, post-processing software…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.