K230271 · Varian Medical Systems, Inc. · GEH · Mar 28, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230271
Device Name
ISOLIS Cryoprobe
Applicant
Varian Medical Systems, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Mar 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ISOLIS cryoprobes, used with the Cryocare Systems, are intended for use in open, minimally invasive or endoscopic surgical procedures in the areas in general surgery, urology, gynecology, oncology, neurology, dermatology, ENT, proctology, pulmonary surgery and thoracic surgery. The systems are designed to freeze/ablate tissue by the application of extreme cold temperatures including prostate and kidney tissue, tumors, skin lesions, and warts. In addition, the systems are intended for use in the following indications: General Surgery - Destruction of warts or lesions - Palliation of tumors of the oral cavity, rectum and skin - Ablation of breast fibroadenomas - Ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions Urology - Ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia Gynecology - Ablation of malignant neoplasia or benign dysplasia of the female genitalia Oncology - Ablation of cancerous or malignant tissue - Ablation of benign tumors - Palliative intervention Neurology - Freezing of nerve tissue in pain management/cryoanalgesia Dermatology - Ablation or freezing of skin cancers and other cutaneous disorders Proctology - Ablation of benign or malignant growths of the anus or rectum - Ablation of hemorrhoids Thoracic Surgery - Ablation of cancerous lesions
Device Story
ISOLIS cryoprobe is a single-use, disposable surgical instrument used with Varian CryoCare Systems; facilitates cryoablation via high-pressure argon gas (Joule-Thomson effect) and active thawing via helium gas. Distal tip (stainless steel) delivers extreme cold to target tissue; features include trocar tip for insertion, vacuum sleeve handle for insulation, and return hose clips for stability during imaging. Operated by physicians in OR/clinic settings; provides localized tissue destruction/ablation. Output is physical tissue freezing/thawing; healthcare providers monitor via system interface to guide clinical decision-making for tumor/lesion management. Benefits include precise tissue ablation and improved handling/stability compared to previous probe designs.
Clinical Evidence
No clinical data or animal studies included. Substantial equivalence supported by bench testing, including functional, electrical safety, performance, physical, usability, packaging, bioburden, sterilization, and biocompatibility evaluations.
Indicated for patients requiring tissue ablation or freezing in general surgery, urology, gynecology, oncology, neurology, dermatology, ENT, proctology, pulmonary, and thoracic surgery. Includes treatment of prostate/kidney tissue, tumors, skin lesions, warts, breast fibroadenomas, hemorrhoids, and nerve tissue for pain management.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
March 28, 2023
Varian Medical Systems, Inc. Peter Coronado Director Regulatory Affairs 3100 Hansen Way Palo Alto, California 94304-1038
Re: K230271
Trade/Device Name: ISOLIS Cryoprobe Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: January 31, 2023 Received: January 31, 2023
Dear Peter Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Carr -S
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230271
Device Name ISOLIS cryoprobe
## Indications for Use (Describe)
The ISOLIS cryoprobes, used with the Cryocare Systems, are intended for use in open, minimally invasive or endoscopic surgical procedures in the areas in general surgery, urology, gynecology, neurology, dermatology, ENT, proctology, pulmonary surgery and thoracic surgery. The systems are designed to freeze/ablate tissue by the application of extreme cold temperatures including prostate and kidney tissue, tumors, skin lesions, and warts.
In addition, the systems are intended for use in the following indications:
General Surgery
- · Destruction of warts or lesions
- · Palliation of tumors of the oral cavity, rectum and skin
- · Ablation of breast fibroadenomas
- Ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions
Urology
· Ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia
Gynecology
- · Ablation of malignant neoplasia or benign dysplasia of the female genitalia
Oncology
- · Ablation of cancerous or malignant tissue
- · Ablation of benign tumors
- · Palliative intervention
Neurology
- · Freezing of nerve tissue in pain management/cryoanalgesia
- Dermatology
- · Ablation or freezing of skin cancers and other cutaneous disorders Proctology
- Ablation of benign or malignant growths of the anus or rectum
- · Ablation of hemorrhoids
- Thoracic Surgery
- · Ablation of cancerous lesions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "varian" in a bold, sans-serif font. The letters are all lowercase and evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
Varian Medical 5 Spectrum Drive
## PREMARKET NOTIFICATION
510(k) Summary
ISOLIS™ Cryoprobe
The following information is provided as required by 21 CFR 807.92
#### l. Submitter's Information:
Name and Address: Varian Medical Systems Inc. 9825 Spectrum Drive, Building 2 Austin, TX 78717 Contact Name: Mr. Peter J. Coronado, Senior Director Regulatory Affairs Phone: 650-424-6320 | Fax: 650-646-9200 E-mail: submissions.support@varian.com Date Prepared: 31st January 2023
#### II. Device Information:
| Proprietary Name: | ISOLIS™ Cryoprobe |
|----------------------|-----------------------------------|
| Common/ Usual Name: | Cryosurgical Unit and accessories |
| Classification Name: | Cryosurgical unit and accessories |
| Regulation Number: | 21 CFR 878.4350 |
| Product Code: | GEH |
#### III. Predicate Device:
CRYOCARE TOUCH™ System and Accessories: (K201588)
#### IV. Device Description:
The ISOLIS cryoprobe is a single use, disposable device designed for use with CryoCare® Systems. ISOLIS cryoprobes are designed to deliver cold temperatures for cryoablation using high-pressure argon gas circulated through the cryoprobe, followed by active thawing using helium gas. The refrigerative and warming capability is limited to the distal end of the probe shaft that is comprised of stainless steel and is designed with a pointed tip to facilitate cryoprobe insertion and placement.
| Model No | Description |
|----------|-----------------------------------------------------------------|
| PC-2.1E | Varian 2.1mm (14 G) ISOLIS™ Cryoprobe, 15cm Length, Ellipsoidal |
| PC-2.1S | Varian 2.1mm (14 G) ISOLIS™ Cryoprobe, 15cm Length, Spherical |
| PC-2.1EL | Varian 2.1mm (14 G) ISOLIS™ Cryoprobe, 20cm Length, Ellipsoidal |
| PC-2.1SL | Varian 2.1mm (14 G) ISOLIS™ Cryoprobe, 20cm Length, Spherical |
ISOLIS Cryoprobe comprises 4 variants as listed below:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "varian" in a bold, sans-serif font. The letters are black and the background is white. The word is centered in the image and takes up most of the space.
Varian Medical
Systems, Inc.
9825 Spectrum Drive
Building 2,
Austin, TX 78717
#### Intended Use/ Indications of Use Statement: V.
The ISOLIS cryoprobes, used with the CryoCare Systems, are intended for use in open, minimally invasive or endoscopic surgical procedures in the areas in general surgery, urology, gynecology, oncology, neurology, dermatology, ENT, proctology, pulmonary surgery and thoracic surgery. The systems are designed to freeze/ablate tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. In addition, the systems are intended for use in the following indications:
## General Surgery
- · Destruction of warts or lesions
- Palliation of tumors of the oral cavity, rectum and skin
- Ablation of breast fibroadenomas
• Ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions
## Urology
- Ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia
## Gynecology
· Ablation of malignant neoplasia or benign dysplasia of the female genitalia
## Oncology
- · Ablation of cancerous or malignant tissue
- Ablation of benign tumors
- Palliative intervention
## Neurology
- Freezing of nerve tissue in pain management/cryoanalgesia
## Dermatology
- Ablation or freezing of skin cancers and other cutaneous disorders
## Proctology
- Ablation of benign or malignant growths of the anus or rectum
- Ablation of hemorrhoids
## Thoracic Surgery
- Ablation of cancerous lesions
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "varian" in a bold, sans-serif font. The letters are all lowercase and evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
#### Comparison of Technological Characteristics with the Predicate Device: VI.
Technological differences between the subject and predicate devices are summarized below:
#### . Probe Shaft Tip Angle
The trocar tip angle in the subject device has been reduced for a slightly sharper tip which will aid in piercing hard tumors.
#### ● Handle Construction
The subject device has a vacuum sleeve handle in comparison to the molded plastic handle in the predicate device to reduce overall size of the handle to provide stability and additional insulation.
#### . Handle Materials
The subject device handle is made of heat shrink while the predicate device handle is made of plastic ABS injection molded handle covering a stainless-steel tube for a smaller, slimmer probe handle and provides better improved insulation.
#### . Heat Exchanger
In comparison to the predicate device, the subject device's heat exchanger increases the surface area for thermal exchange and maximizes the freeze performance of the probes.
#### ● Accessories - Addition of Return Hose Clips
The subject devices have two Return Hose Clips added to the Return Hose to aid in the stability of the Return Hose and Probe and to maintain the position of the probe in the treatment location particularly during CT imaging that requires the patient to move in and out of a gantry.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Varian Medical Systems, Inc. The logo consists of the word "varian" in a bold, sans-serif font on the left, and the company name and address on the right. The address is 9825 Spectrum Drive, Building 2, Austin, TX 78717.
#### Summary of the technological characteristics between subject and predicate devices VII.
| Feature and/or<br>Specification | PREDICATE DEVICE(S)<br>CRYOCARE TOUCH™ System and Accessories (K201588) | SUBJECT DEVICE(S)<br>ISOLIST™ Cryoprobe | Comparison between<br>predicate and subject<br>devices and Rationale |
|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Slimline Cryoprobes | ISOLIST™ Cryoprobe | The subject devices are a<br>line extension to the<br>cryoprobes compatible with<br>Varian Cryoablation Systems |
| Prescription/ Rx | Yes | Yes | Same |
| Compatible Varian<br>Cryoablation Systems | CRYOCARE TOUCH System (K201588)<br>CRYO-CS-3 Basic (K153489)<br>Cryocare CS-3 V5 (K151968)<br>Cryocare CS-3 V4 (K141110)<br>Cryocare CS-3 (K101333)<br>Cryocare SL (K011074)<br>Cryocare-CS (K060279) | CRYOCARE TOUCH System (K201588)<br>CRYO-CS-3 Basic (K153489)<br>Cryocare CS-3 V5 (K151968)<br>Cryocare CS-3 V4 (K141110)<br>Cryocare CS-3 (K101333)<br>Cryocare SL (K011074)<br>Cryocare-CS (K060279) | Same |
| Intended Use /<br>Indications of Use of the<br>compatible Varian<br>Cryoablation Systems | Intended for use in open, minimally invasive or endoscopic<br>surgical procedures in the areas in general surgery, urology,<br>gynecology, oncology, neurology, dermatology, ENT,<br>proctology, pulmonary surgery, and thoracic surgery. The<br>system is designed to freeze/ablate tissue by the application<br>of extreme cold temperatures including prostate and kidney<br>tissue, liver metastases, tumors, skin lesions, and warts.<br><br>In addition, the CRYOCARE TOUCH System is intended for use<br>in the following indications:<br><br><b>General Surgery</b><br>· Destruction of warts or lesions<br>· Palliation of tumors of the oral cavity, rectum and skin<br>· Ablation of breast fibroadenomas<br>· Ablation of leukoplakia of the mouth, angiomas, sebaceous<br>hyperplasia, basal cell tumors of the eyelid or canthus area. | Intended for use in open, minimally invasive or endoscopic<br>surgical procedures in the areas in general surgery, urology,<br>gynecology, oncology, neurology, dermatology, ENT,<br>proctology, pulmonary surgery, and thoracic surgery. The<br>system is designed to freeze/ablate tissue by the application<br>of extreme cold temperatures<br>including prostate and kidney tissue, liver metastases, tumors,<br>skin lesions, and warts.<br><br>In addition, the systems are intended for use in the following<br>indications:<br><br><b>General Surgery</b><br>· Destruction of warts or lesions<br>· Palliation of tumors of the oral cavity, rectum and skin<br>· Ablation of breast fibroadenomas<br>· Ablation of leukoplakia of the mouth, angiomas, sebaceous<br>hyperplasia, basal cell tumors of the eyelid or canthus area. | Same |
| | | | |
| | ulcerated basal cell tumors, dermatofibromas, small<br>hemangiomas, mucocele cysts, multiple warts, plantar warts,<br>hemorrhoids, anal fissures, perianal condylomata, pilonidal<br>cysts, actinic and seborrheic keratoses, cavernous<br>hemangiomas, recurrent cancerous lesions | | |
| | Urology | Urology | |
| | · Ablation of prostate tissue in cases of prostate cancer and<br>benign prostatic hyperplasia | · Ablation of prostate tissue in cases of prostate cancer and<br>benign prostatic hyperplasia | |
| | Gynecology | Gynecology | |
| | · Ablation of malignant neoplasia or benign dysplasia of the<br>female genitalia | · Ablation of malignant neoplasia or benign dysplasia of the<br>female genitalia | |
| | Oncology | Oncology | |
| | · Ablation of cancerous or malignant tissue | · Ablation of cancerous or malignant tissue | |
| | · Ablation of benign tumors | · Ablation of benign tumors | |
| | · Palliative intervention | · Palliative intervention | |
| | Neurology | Neurology | |
| | · Freezing of nerve tissue in pain management/ cryoanalgesia | · Freezing of nerve tissue in pain management/ cryoanalgesia | |
| | Dermatology | Dermatology | |
| | · Ablation or freezing of skin cancers and other cutaneous<br>disorders | · Ablation or freezing of skin cancers and other cutaneous<br>disorders | |
| | Proctology | Proctology | |
| | · Ablation of benign or malignant growths of the anus or<br>rectum | · Ablation of benign or malignant growths of the anus or<br>rectum | |
| | · Ablation of hemorrhoids | · Ablation of hemorrhoids | |
| | Thoracic Surgery | Thoracic Surgery | |
| | · Ablation of cancerous lesions | · Ablation of cancerous lesions | |
| Software | Not Applicable; Probes do not have software. | Not Applicable; Probes do not have software. | Not Applicable |
| Fundamental Technology | Joule-Thomson Effect | Joule-Thomson Effect | Same |
| Mechanism of Action<br>(Freezing/Thawing<br>Technology) | Operate on the Joule-Thomson Effect. | Operate on the Joule-Thomson Effect. | Same |
| Stick Function for<br>ablation | Yes, via compatible Cryocare system | Yes, via compatible Cryocare system | Same |
| Activation Method | Software (Freeze, thaw, stop) via compatible Cryocare system | Software (Freeze, thaw, stop) via compatible Cryocare system | Same |
| Size of Isotherm | Size of isotherm is model specific as indicated in IFU<br>Maximum diameter at -20°C = 25mm | Size of isotherm is model specific as indicated in IFU<br>Maximum diameter at -20°C = 38mm | Improved isotherm<br>performance |
| | | | |
| Freeze Gas | Argon | Argon | Same |
| Thaw Gas | Helium | Helium | Same |
| Packaged Single-Use,<br>Sterile and Disposable | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.