K230258 · Excelent, Inc. · LRC · May 25, 2023 · Ear, Nose, Throat
Device Facts
Record ID
K230258
Device Name
BB 8 Sinus Dilation Kit
Applicant
Excelent, Inc.
Product Code
LRC · Ear, Nose, Throat
Decision Date
May 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
To access and treat the maxillary ostia/ethmoid infundibula in patients 2 years and older, and frontal ostia/recesses and sphenoid sinus ostia in patients 12 years and older using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.
Device Story
BB 8 Sinus Dilation Kit is a manual ENT surgical instrument for sinus ostia dilation. Device inputs include manual positioning by a physician using an integrated LED light fiber for transillumination-guided placement. The device features a shapeable catheter with an inflatable balloon, suction, and irrigation lines. The physician uses a bend template to shape the catheter tip for specific sinus anatomy. The balloon is manually inflated via a provided syringe to a maximum of 12 ATM to remodel bony sinus outflow tracts. Suction and irrigation capabilities allow for field clearing and visualization. Used in a clinical setting by ENT specialists, the device enables minimally invasive sinus access, potentially reducing recovery time and improving sinus drainage for patients with obstructive sinus conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Verification testing included visual, dimensional, and functional assessments (e.g., balloon burst pressure, fatigue, LED visibility, suction/irrigation rates), electrical safety, EMC, and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Manual ENT surgical instrument; shapeable catheter with inflatable balloon; integrated LED light fiber (battery-powered); suction and irrigation channels. Materials are biocompatible per ISO 10993-1. Sterilization via Ethylene Oxide (EtO). Maximum inflation pressure 12 ATM. Includes bend template and inflation syringe with pressure gauge.
Indications for Use
Indicated for patients 2+ years for maxillary ostia/ethmoid infundibula treatment and patients 12+ years for frontal ostia/recesses and sphenoid sinus ostia treatment via trans-nasal approach.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
Predicate Devices
Entellus Xpress Multi-Sinus Dilation System (K152434)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 25, 2023
Excelent Inc. % Bryan Brosseau President Brosseau Consulting LLC 2352 Kennesaw Oaks Trl NW Kennesaw, Georgia 30152
Re: K230258
Trade/Device Name: BB 8 Sinus Dilation Kit Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, and Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: April 25, 2023 Received: April 26, 2023
Dear Bryan Brosseau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230258
Device Name BB 8 Sinus Dilation Kit
Indications for Use (Describe)
To access and treat the maxillary ostia/ethmoid infundibula in patients 2 years and older, and frontal ostia/recesses and sphenoid sinus ostia in patients 12 years and older using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k): EXCELENT INC. 68 TW Alexander Drive PO Box 13628 Research Triangle Park Durham, NC 27709 510(k) K230258
# 510(k) Summary – K230258
Date Prepared:
May 23, 2023
#### Submitted By:
EXCELENT INC. 68 TW Alexander Drive PO Box 13628 Research Triangle Park Durham, NC 27709
## Contact:
| Name: | Bryan Brosseau |
|------------|-----------------------------|
| Title: | Consultant to EXCELENT INC. |
| Telephone: | (404) 610-7215 |
| Email: | bryan@brosseauconsult.com |
#### Device:
| Trade Name: | BB 8 Sinus Dilation Kit |
|------------------------|---------------------------------|
| Common Name: | Sinuplasty Balloon Catheter |
| FDA Product Code: | LRC |
| Device Classification: | Class I |
| Classification Name: | Instrument, ENT Manual Surgical |
| Device Regulation: | 21 CFR 874.4420 |
#### Predicate Devices:
| The device is substantially equivalent to the following predicate devices: | |
|----------------------------------------------------------------------------|-------------------------------------------------------|
| Primary Predicate: | Entellus Xpress Multi-Sinus Dilation System, (K152434 |
| Secondary Predicate: | Entellus PathAssist LED Light Fiber, (K152435) |
{4}------------------------------------------------
#### Device Description:
The BB 8 Sinus Dilation Kit is intended to remodel or recreate the sinus outflow tract via transnasal balloon dilation. The device combines features of a curved suction tip and a sinus ostium seeker with the tissue expansion effect of balloon dilation. Since the distal end of the device is re-shapeable, one balloon can be modified to work on multiple sinuses within the same patient. The device also allows the capability to suction and irrigate the surgical field to allow for removal of bodily secretions and to keep the field of view clean for improved visualization. The device includes an LED light with light fiber to locate, illuminate within, and transilluminate across nasal and sinus structures.
In summary, the device includes the following components and accessories:
- 1. BB 8 Sinus Dilation Kit which incorporates:
- a. shapeable catheter with inflatable balloon
- b. LED light fiber
- c. irrigation line, and
- d. suction line,
- 2. Inflation syringe,
- 3. Extension line used for inflation of the balloon (and to provide irrigation, if desired), and
- 4. A bend template for shaping the catheter tip for treating various sinus anatomy
#### Indications for Use:
To access and treat the maxillary ostia/ethmoid infundibula in patients 2 years and older, and frontal ostia/recesses and sphenoid sinus ostia in patients 12 years and older using a transnasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.
#### Technological Characteristics and Performance Data (Predicate Comparison):
The device has similar intended use, indications for use, mechanism of action, and performance compared to the predicate devices.
The balloons of the EXCELENT and Entellus devices are both positioned over a shapeable catheter and are manually inflated with saline and a syringe until the desired inflation pressure (up to 12 ATM) is achieved. The subject and predicate device principles of operation are identical. Both devices are advanced into the nasal and sinus anatomy under direct endoscopic visualization and treat the anatomy by dilating and displacing anatomic structures along the sinus drainage pathways. Both devices are used through a trans-nasal approach.
The intended use, indications for use, balloon dimensions, sterilization method, maximum inflation pressure, catheter type, light feature, inflation mechanism, and biocompatibility have been demonstrated as identical or similar between the subject and predicate devices.
{5}------------------------------------------------
Additionally, verification testing for the subject device is similar to the testing performed for the predicate devices and demonstrates that minor differences in the device characteristics between the subject and predicate devices do not raise any new questions of safety and efficacy. Verification testing performed for the subject device includes:
- l Visual Testing
- Visual Connection Indicators on Device Handle o
- Marker Band Location o
- I Dimensional Testing
- Catheter Shaft Working Length o
- Device Weight O
- Luer Connections O
- Distal Tip Length and Outer Diameter o
- Extension Line Length O
- O Balloon Working Length
- Balloon Outer Diameter o
- I Functional Testing
- Bend Radius O
- Pliable Shaft Fatique o
- Suction Force o
- Irrigation Rate O
- Balloon Inflation Time o
- Balloon Deflation Time O
- Balloon Fatigue Test O
- Balloon Burst Pressure O
- o LED Visibility
- LED Battery Life O
- Atraumatic Tip Joint Strength O
- Manifold Joint Strength O
- Manifold Torque Value o
- I Electrical Safety and EMC Testing
- l Biocompatibility Testing
- I Sterilization, Packaging, and Shelf-Life Testing
A comparison of technological characteristics and performance data to the predicate devices is provided in Table 1, below.
{6}------------------------------------------------
Traditional 510(k): EXCELENT INC. 68 TW Alexander Drive PO Box 13628 Research Triangle Park Durham, NC 27709 510(k) K230258
## Table 1: Comparison to Predicate Devices
| Product<br>Characteristic | Subject Device | Primary Predicate | Secondary<br>Predicate | Comparison of Devices |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | EXCELENT INC. | Entellus Medical Inc. | Entellus Medical Inc. | N/A |
| Product Name | BB 8 Sinus Dilation Kit | XprESS Multi-Sinus<br>Dilation System | Entellus PathAssist<br>LED Light Fiber | N/A |
| Common Name | Balloon Sinus Dilation<br>System | Balloon Sinus<br>Dilation System | Sinus Guidewire | Similar, the EXCELENT device is a balloon<br>dilation system similar to the XprESS system<br>with an integrated LED light similar to the<br>PathAssist LED light fiber. |
| Device<br>Classification<br>Name | Instrument, ENT<br>Manual Surgical | Instrument, ENT<br>Manual Surgical | Instrument, ENT<br>Manual Surgical | Identical, all devices manual ENT surgical<br>instruments. |
| 510(k) Number | K230258 | K152434 | K152435 | N/A |
| Classification<br>Regulation | Class 1, 874.4420 | Class 1, 874.4420 | Class 1, 874.4420 | N/A |
| Product Code | LRC | LRC | LRC | Identical, all products are classified under the<br>LRC product code. |
| Intended Use | To access and treat<br>the maxillary<br>ostia/ethmoid<br>infundibula in patients<br>2 years and older, and<br>frontal ostia/recesses<br>and sphenoid sinus<br>ostia in patients 12<br>years and older using<br>a trans-nasal<br>approach. | To access and treat<br>the maxillary<br>ostia/ethmoid<br>infundibula in patients<br>2 years and older,<br>and frontal<br>ostia/recesses and<br>sphenoid sinus ostia<br>in patients 12 years<br>and older using a<br>trans-nasal approach. | To locate, illuminate<br>within, and<br>transilluminate<br>across nasal and<br>sinus structures. | Similar, the EXCELENT device shares the<br>same intended use as the primary predicate<br>and includes an LED light to illuminate the<br>treatment area in a manner that is similar to the<br>secondary predicate. |
| Product<br>Characteristic | Subject Device | Primary Predicate | Secondary<br>Predicate | Comparison of Devices |
| Indications for<br>Use | To access and treat<br>the maxillary<br>ostia/ethmoid<br>infundibula in patients<br>2 years and older, and<br>frontal ostia/recesses<br>and sphenoid sinus<br>ostia in patients 12<br>years and older using<br>a trans-nasal<br>approach. The bony<br>sinus outflow tracts are<br>remodeled by balloon<br>displacement of<br>adjacent bone and<br>paranasal sinus<br>structures. | To access and treat<br>the maxillary<br>ostia/ethmoid<br>infundibula in patients<br>2 years and older,<br>and frontal<br>ostia/recesses and<br>sphenoid sinus ostia<br>in patients 12 years<br>and older using a<br>transnasal approach.<br>The bony sinus<br>outflow tracts are<br>remodeled by balloon<br>displacement of<br>adjacent bone and<br>paranasal sinus<br>structures. | To locate, illuminate<br>within, and<br>transilluminate<br>across nasal and<br>sinus structures. | Similar, the Indications for Use of the subject<br>and predicate device are identical. The use of<br>the LED feature of the subject device is similar<br>to the indications for use of the secondary<br>predicate device.<br>The integrated LED light of the subject device<br>allows the user to confirm the positioning of the<br>balloon using the transillumination technique<br>prior to inflating the balloon. |
| Sterility | Sterile (EtO) | Sterile (EtO) | Sterile (EtO) | Identical, all products are provided sterile and<br>are sterilized via EtO. |
| Type of Use | Single-Patient Use,<br>Prescription (Rx only) | Single-Patient Use,<br>Prescription (Rx only) | Single-Patient Use,<br>Prescription (Rx<br>only) | Identical, all devices are single-patient use and<br>prescription only. |
| Balloon<br>Diameter and<br>Length | 6mm (diameter) x<br>20mm (length) | 6mm (diameter) x<br>20mm (length)<br>(XpreSS Ultra<br>Configuration) | N/A | Identical, the subject device balloon<br>dimensions are identical to one configuration of<br>the primary predicate device. |
| Distal Tip<br>Diameter | 1.95 mm | 1.5 mm (XprESS<br>Ultra Configuration) | | Similar, the distal tip diameter of the subject<br>device is within the range of distal tip diameters<br>for the predicate device. |
| Product<br>Characteristic | Subject Device | Primary Predicate | Secondary<br>Predicate | Comparison of Devices |
| | | 1.75 mm (XprESS<br>LoProfile<br>Configuration) | | |
| | | 2.0 mm (XprESS Pro<br>Configuration) | | |
| Maximum<br>Balloon Inflation<br>Pressure | 12 ATM | 12 ATM | N/A | Identical, the maximum balloon inflation<br>pressure is identical to the primary predicate<br>device. |
| Catheter Type | Flexible, shapeable,<br>with atraumatic tip,<br>irrigation and suction<br>features | Flexible, shapeable,<br>with atraumatic tip,<br>irrigation and suction<br>features | N/A | Similar, the catheter portion of the subject and<br>primary predicate devices is shapeable for use<br>in multiple sinuses. Both devices incorporate<br>irrigation and suction. Both devices are<br>provided with a bending tool to provide the<br>proper bend for each sinus. |
| Light Source | Light delivered via fiber<br>optic from an<br>integrated, battery<br>powered, LED light<br>source with > 60<br>minutes battery life | N/A | Light delivered via<br>fiber optic from an<br>integrated, battery<br>powered, LED light<br>source with > 60<br>minutes battery life | Similar, both the subject device and the<br>secondary predicate provide illumination from a<br>similar electronic component. The integration<br>of the LED light system in the subject device<br>does not raise any new questions of safety or<br>efficacy. |
| Inflation<br>Mechanism | The inflation syringe is<br>provided with the<br>system and the<br>incorporated pressure<br>gauge provides a<br>visual scale of<br>pressure in 1 ATM<br>increments with even<br>increments labeled on<br>the pressure gauge (2 | The inflation syringe<br>is provided with the<br>system and provides<br>a visual indicator<br>when 12 ATM balloon<br>pressure is achieved<br>(via alignment of<br>syringe marking and<br>distal syringe seal). | N/A | Similar, both inflation mechanisms provide a<br>visual indication of balloon pressure to allow<br>the user to pressurize to the maximum<br>pressure of 12 ATM. |
| Product<br>Characteristic | Subject Device | Primary Predicate | Secondary<br>Predicate | Comparison of Devices |
| | 4, 6, etc. and including<br>12 ATM). | | | |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Similar, the subject device has undergone<br>biocompatibility testing in accordance with ISO<br>10993-1 based on the intended nature and<br>duration of contact. |
{7}------------------------------------------------
Traditional 510(k): EXCELENT INC. 68 TW Alexander Drive Durham, NČ 27709
{8}------------------------------------------------
Traditional 510(k): EXCELENT INC.
Durham, NČ 27709
{9}------------------------------------------------
Traditional 510(k): EXCELENT INC.
Durham, NČ 27709
{10}------------------------------------------------
## Conclusions:
The subject device and the predicate devices underwent evaluation that confirms equivalence in the intended use of each device, biocompatibility, performance, environment of use, and the principles of operation. Therefore, the subject device is deemed substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.