← Product Code [MTD](/productcode/MTD) · K230253

# OPTRELL™ Mapping Catheter with TRUEref™ Technology (K230253)

_Biosense Webster, Inc. · MTD · Mar 2, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230253

## Device Facts

- **Applicant:** Biosense Webster, Inc.
- **Product Code:** [MTD](/productcode/MTD.md)
- **Decision Date:** Mar 2, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The OPTRELL™ Mapping Catheter with TRUEref™ Technology is indicated for multiple electrophysiological mapping of cardiac structures in the heart, i.e., recording or simulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The cather provides location information only when used with a compatible version of the CARTO™ 3 System.

## Device Story

The OPTRELL™ Mapping Catheter is a bi-directional deflectable catheter used for electrophysiological mapping of the heart. It features six parallel 2 Fr spines forming three loops on the deflectable tip, with a 6x6 grid of 36 platinum electrodes for recording and stimulation. A magnetic location sensor in the tip provides spatial data to the CARTO™ 3 EP Navigation System. The device includes a TRUEref™ electrode at the spine confluence for internal unipolar reference and an irrigation lumen for anticoagulant fluid. Operated by physicians in a clinical setting, the catheter is inserted via an 8.5 F guiding sheath. The physician manipulates the tip using a rocker lever. Electrical signals (electrograms) are transmitted to the CARTO™ 3 system for visualization, analysis, and interpretation of heart arrhythmias. The device aids in diagnosing arrhythmias by providing high-density mapping of cardiac electrical activity.

## Clinical Evidence

Bench and animal testing only. Bench testing evaluated mechanical integrity, deflection, functionality, simulated use, electrical properties, visualization, and shelf life. Animal testing assessed device effectiveness, safety, and maneuverability. All tests met pre-defined acceptance criteria.

## Technological Characteristics

Bi-directional deflectable catheter with 6 parallel 2 Fr spines; 36 platinum electrodes (6x6 grid); magnetic location sensor; TRUEref™ unipolar reference electrode; irrigation lumen for anticoagulant fluid; 8.5 F guiding sheath compatibility; connects to CARTO™ 3 EP Navigation System.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- OPTRELL™ Mapping Catheter with TRUEref™ Technology ([K211438](/device/K211438.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 2, 2023

Biosense Webster, Inc. John Jimenez Senior RA Program Lead 31 Technology Drive, Suite 200 Irvine, California 92618

Re: K230253

Trade/Device Name: OPTRELL™ Mapping Catheter with TRUEref™ Technology Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: January 27, 2023 Received: January 31, 2023

Dear John Jimenez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230253

Device Name

OPTRELL™ Mapping Catheter with TRUEref™ Technology

Indications for Use (Describe)

The OPTRELL™ Mapping Catheter with TRUEref™ Technology is indicated for multiple electrophysiological mapping of cardiac structures in the heart, i.e., recording or simulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The cather provides location information only when used with a compatible version of the CARTO™ 3 System.

Type of Use (*Select one or both, as applicable*)

| <span> <b> <span style="text-decoration: overline;">X</span> </b> Prescription Use (Part 21 CFR 801 Subpart D)       </span> | <span>         Over-The-Counter Use (21 CFR 801 Subpart C)       </span> |
|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 2. 510(k) Summary

| Applicant:                         | Biosense Webster, Inc.<br>31 Technology Drive, Suite 200<br>Irvine, CA 92618<br>Phone: (949) 923-4774<br>Fax: 949-450-6886 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                    | John Jimenez<br>Senior Program Lead, Regulatory Affairs                                                                    |
| Date:                              | January 27, 2023                                                                                                           |
| Trade or Proprietary Name:         | OPTRELL™ Mapping Catheter with TRUEref™ Technology                                                                         |
| Common or Usual<br>Name of Device: | Electrophysiological Mapping Catheter                                                                                      |
| Classification Name:               | Electrode Recording Catheter<br>(21 CFR 870.1220, Product Code MTD)                                                        |
| Predicate Device:                  | OPTRELL™ Mapping Catheter with TRUEref™ Technology<br>510(k): K211438                                                      |
| Manufacturer:                      | Biosense Webster, Inc.<br>31 Technology Drive, Suite 200<br>Irvine, CA 92618                                               |
| Manufacturing Sites:               | Biosense Webster, Inc.<br>Circuito Interior Norte #1820<br>Parque Industrial Salvarcar 32599<br>Juarez, Chihuahua, Mexico  |

{4}------------------------------------------------

#### Substantially Equivalent To A.

The OPTRELL™ Mapping Catheter with TRUEref™ Technology is substantially equivalent to the Biosense Webster OPTRELL™ Mapping Catheter with TRUEref™ Technology cleared on March 18, 2022, under 510(k) K211438.

#### Description of the Device Subject to Premarket Notification B.

The OPTRELL™ Mapping Catheter with TRUEref™ Technology has been designed to facilitate electrophysiological mapping of the heart with the CARTO™ 3 System. The OPTRELL™ Mapping Catheter with TRUEref™ Technology is deployed in the heart through an 8.5 F guiding sheath. This bi-directional deflectable catheter includes six (6) parallel 2 Fr spines that are joined to form three (3) loops on the deflectable tip. Each spine has six (6) platinum electrodes that are used for stimulating and recording. The electrodes form a 6 by 6 (6x6) grid. A magnetic location sensor embedded in the deflectable tip transmits location information to the CARTO™ 3 EP Navigation System. Below the spines on the deflectable tip are three (3) electrodes that allow visualization of the tip on the CARTO™ 3 System. The TRUEref™ Electrode, which is embedded in the distal end of the irrigation lumen, can be used as an internal close unipolar reference electrode within the heart chamber. Rotating the catheter's Rocker Lever clockwise or counterclockwise deflects the tip; rotating the Rocker Lever to the neutral position straightens the tip. The catheter includes an irrigation lumen for connection to a source of continuous drip anticoagulant fluid. Lastly, the proposed catheter is offered in the same deflectable curve offerings as the predicate.

The OPTRELL™ Mapping Catheter with TRUEref™ Technology design uses the same technology, has the same intended use, functions, materials, and method of operation as the predicate OPTRELL™ Mapping Catheter with TRUEref™ Technology. Instead of eight (8) platinum electrodes on each spine, the proposed OPTRELL™ Mapping Catheter with TRUEref™ Technology has six (6) electrodes on each spine, forming a 6 by 6 paddle-shaped tip containing thirty-six (36) electrodes arranged across six (6) spines. Like the predicate, the proposed OPTRELL™ also includes a close uni-polar reference electrode located at the confluence of the spines. Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation System and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias. The difference in paddle size and number of electrodes between the proposed OPTRELL™ Mapping Catheter with TRUEref™ Technology and the predicate OPTRELL™ Mapping Catheter with TRUEref™ Technology have been evaluated through bench and animal testing. Results from the bench and animal data demonstrated that the proposed device is substantially equivalent to the predicate device and did not result in new questions with regards to safety and effectiveness of the device.

#### Indications for Use ﻥ

The OPTRELL™ Mapping Catheter with TRUEref™ Technology is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the

{5}------------------------------------------------

# K230253

heart. The catheter provides location information only when used with a compatible version of the CARTO™ 3 System.

#### Performance Data D.

The OPTRELL™ Mapping Catheter with TRUEref™ Technology underwent Bench and Animal Testing. Testing included mechanical integrity, deflection, device functionality, simulated use, electrical properties, visualization, shelf life, device maneuverability and signal quality, and animal testing to assess device effectiveness and safety. The catheter passed all intended criteria in accordance with appropriate test criteria and standards. No new questions of safety or effectiveness were identified. Testing demonstrated that the proposed OPTRELL™ Mapping Catheter with TRUEref™ Technology is as safe, effective, and performs as well as or better, than the predicate device. This testing program supports the determination of substantial equivalence to the predicate device.

#### Overall Performance Conclusions ட்.

The nonclinical studies demonstrate that the OPTRELL™ Mapping Catheter with TRUEref™ Technology is as safe, effective, and performs as well as or better, than the predicate OPTRELL™ Mapping Catheter with TRUEref™ Technology.

---

**Source:** [https://fda-staging.innolitics.com/device/K230253](https://fda-staging.innolitics.com/device/K230253)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
