The LnK Spinal Fixation System /OpenLoc-L Spinal Fixation System is a non-cervical spinal fixation device intended for use as a posterior pedicle screw fixation system (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. This device is an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, and/or lordosis); tumor; stenosis, and failed previous fusion (pseudoarthrosis). The AccelFix Spinal Fixation System is a non-cervical spinal fixation device intended for use as a posterior pedicle screw fixation system (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, stenosis, and failed previous fusion (pseudoarthrosis).
Device Story
Spinal fixation systems (LnK/OpenLoc-L and AccelFix) consist of pedicle screws, rods (including Z-rods), crosslinks, connectors, and hooks; used by surgeons in clinical settings to provide stabilization as an adjunct to spinal fusion; components are implanted to support the spine in patients with degenerative disc disease, trauma, or deformities; device provides mechanical stability to the vertebral column; benefits include correction of spinal alignment and facilitation of fusion; no software or AI components involved.
Clinical Evidence
Bench testing only; no clinical data provided. Substantial equivalence is supported by comparison of design, materials, and mechanical performance to predicate devices.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) and Cobalt-28Chromium-6Molybdenum-4Vanadium ELI (ASTM F1537). Components: screws, rods, crosslinks, connectors, hooks. Sterilization: Gamma radiation (sterile) or steam autoclave (non-sterile).
Indications for Use
Indicated for skeletally mature patients requiring non-cervical spinal fixation (posterior T1-S2/ilium or anterolateral T8-L5) as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), deformities (scoliosis, kyphosis, lordosis), tumor, stenosis, and failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Olympic Posterior Spinal Fixation System (K181139)
Submission Summary (Full Text)
{0}------------------------------------------------
February 28, 2023
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
L&K BIOMED Co., Ltd. Ms. Katherine Kim RA 101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil Giheung-gu Yongin-si, Gyeonggi-do 17015 South Korea
Re: K230245
Trade/Device Name: LnK Spinal Fixation System/OpenLoc-L Spinal Fixation System: AccelFix Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP, KWQ Dated: January 28, 2023 Received: January 30, 2023
Dear Ms. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image contains the text "Colin O'neill -S". The text is arranged vertically, with "Colin" on the top line and "O'neill -S" on the bottom line. In the background, there is a faded logo of the letters FDA.
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K230245
### Device Name
LnK Spinal Fixation System /OpenLoc-L Spinal Fixation System
Indications for Use (Describe)
The LnK Spinal Fixation System /OpenLoc-L Spinal Fixation System is a non-cervical spinal fixation device intended for use as a posterior pedicle screw fixation system (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. This device is an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, and/or lordosis); tumor; stenosis, and failed previous fusion (pseudoarthrosis).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K230245
Device Name AccelFix Spinal Fixation System
### Indications for Use (Describe)
The AccelFix Spinal Fixation System is a non-cervical spinal fixation device intended for use as a posterior pedicle screw fixation system (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, kyphosis); tumor, stenosis, and failed previous fusion (pseudoarthrosis).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="display:inline-block; margin-right: 5px;"> <div style="width:15px; height:15px; border:1px solid black; position:relative; display:inline-block;"> <div style="position:absolute; top:0; left:0; width:100%; height:100%;">✓</div> </div> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;"> <div style="width:15px; height:15px; border:1px solid black; position:relative; display:inline-block;"></div> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
### 1. SUBMITTER
| Submitter's Name: | L&K BIOMED Co., Ltd. |
|------------------------|----------------------------------------------------------------------------------------------------------|
| Submitter's Address: | #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil<br>Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea |
| Submitter's Telephone: | +82-2-6717-1983 |
| Contact Person: | Katherine Kim |
| | khkim@lnkbiomed.com / ra@lnkbiomed.com |
| Prepared Date | January 27, 2023 |
#### DEVICE IDENTIFICATION 2.
| Trade or Proprietary Name | LnK Spinal Fixation System /OpenLoc-L Spinal Fixation System |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name | Spinal interlaminal fixation orthosis<br>Spinal intervertebral body fixation orthosis<br>Thoracolumbosacral pedicle screw system |
| Regulation class / Number | Class II, 21 CFR 888.3070 |
| Regulation Name | Thoracolumbosacral pedicle screw system |
| Product Code | NKB, KWP, KWQ |
| Classification Panel | Spinal Devices (DHT6B) |
| Trade or Proprietary Name | AccelFix Spinal Fixation System |
|---------------------------|----------------------------------------------|
| Common or Usual Name | Spinal interlaminal fixation orthosis |
| | Spinal intervertebral body fixation orthosis |
| | Thoracolumbosacral pedicle screw system |
| | Regulation class / Number |
| Regulation Name | Thoracolumbosacral pedicle screw system |
| Product Code | NKB, KWP, KWQ |
| Classification Panel | Spinal Devices (DHT6B) |
### 3. PREDICATE OR LEGALLY MARKETED DEVICES WHICH ARE SUBSTANTIALLY EQUIVALENT.
The additional components of the LnK Spinal Fixation System, OpenLoc-L Spinal Fixation System and AccelFix Spinal Fixation System are considered substantially equivalent to the predicate devices. The systems have same design, materials, scientific technology, and indications for use.
## LnK Spinal Fixation System /OpenLoc-L Spinal Fixation System
Primary Predicate Device: LnK Spinal Fixation System (K120270) Additional Predicate Devices: LnK Spinal Fixation System/OpenLoc-L Spinal Fixation System (K120270, K143363, K171813, K183168, K 200790, K223565) Olympic Posterior Spinal Fixation System(K181139)
## AccelFix Spinal Fixation System
Primary Predicate Device: AccelFix Spinal Fixation System (K182544) Additional Predicate Devices: AccelFix Spinal Fixation System (K200794, K223565) Olympic Posterior Spinal Fixation System(K181139)
{5}------------------------------------------------
#### MATERIALS 4.
| LnK Spinal Fixation System/ OpenLoc-L<br>Spinal Fixation System | Ti-6Al-4V ELI titanium alloy (ASTM F136) and<br>Cobalt-28Chromium-6Molybdenum-4Vanadium<br>ELI (ASTM F1537) |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| AccelFix Spinal Fixation System | |
The additional Z-Rod is manufactured from Ti-6AI-4V ELI titanium alloy (ASTM F136) and Cobalt-28Chromium-6Molybdenum-4Vanadium ELI (ASTM F1537). This this is the same material used in the predicate devices.
#### 5. DESCRIPTION OF THE DEVICE
### LNK SPINAL FIXATION SYSTEM /OPENLOC-L SPINAL FIXATION SYSTEM
The LNK SPINAL FIXATION SYSTEM, OPENLOC-L SPINAL FIXATION SYSTEM are available in various sizes. This system is comprised of screws, rods, crosslinks, connectors and hooks. The screws are available from 4.0 mm to 10.5 mm diameters with lengths ranging from 20 mm to 150 mm. The rods are available from 5.0 mm, 6.0 mm and 6.35 mm diameter with lengths ranging from 40 mm to 600 mm. Both straight rods and curved rods have four types of design that consist of standard type, hex type, stopper type and double stopper.
### ACCELFIX SPINAL FIXATION SYSTEM
The AccelFix Spinal Fixation System consists of screws, rods, crosslinks, set screws, cross-link connectors, and hooks. The screws are available from 5.0 mm, 6.0 mm, 7.0. 7.5. 8.0. 8.5. 9.0 and 9.5mm diameters with working lengths ranging from 20 mm. The rods are available from 5.5 mm, 6.0mm and 6.35mm diameter with lengths ranging from 40 mm to 600 mm. Both straight rods and curved rods have four types of design that consist of standard type, hex type, stopper type and double stopper.
#### INDICATION FOR USE 6.
## LNK SPINAL FIXATION SYSTEM /OPENLOC-L SPINAL FIXATION SYSTEM
The LnK Spinal Fixation System /OpenLoc-L Spinal Fixation System is a non-cervical spinal fixation device intended for use as a posterior pedicle screw fixation system (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. This device is indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, and/or lordosis); tumor; stenosis, and failed previous fusion (pseudoarthrosis).
## ACCELFIX SPINAL FIXATION SYSTEM
The AccelFix Spinal Fixation System is a non-cervical spinal fixation device intended for use as a posterior pedicle screw fixation system (T1-S2/ilium), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; stenosis, and failed previous fusion (pseudoarthrosis).
{6}------------------------------------------------
#### 7. PERFORMANCE DATA
The additional components to be added through this submission do not require additional mechanical testing. None of the additional components is the worst case of the LnK Spinal Fixation System, OpenLoc-L Spinal Fixation System and AccelFix Spinal Fixation System. Therefore, we substitute mechanical test data of additional components of LnK Spinal Fixation System, OpenLoc-L Spinal Fixation System and AccelFix Spinal Fixation System with the predicate device (LnK Spinal Fixation System, OpenLoc-L Spinal Fixation System -K120270, K143363, K171813, K183168, K200790,K223560 / AccelFix Spinal Fixation System: K182544, K200794,K223565).
#### SUMMARY OF TECHNICAL CHARACTERISTICS WITH THE PREDICATE DEVICE 8.
The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:
- · Instruction for use
- · Material
- · Design with components
- · Dimension
- Sterilization Method .
The following technological similarities and differences exist between the subject and predicate devices:
### LNK SPINAL FIXATION SYSTEM, OPENLOC-L SPINAL FIXATION SYSTEM
| INDICATION FOR USE | | | |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------|--|----------------------------------------------------------|
| Devices | | | Similarities |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System | Subject | | |
| LnK Spinal Fixation System(K120270) | Primary<br>Predicate | | |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System (K120270, K143363,<br>K171813, K183168, K 200790, K223565) | Additional<br>Predicate | | Instruction for use including indication are<br>similar. |
| Olympic Posterior Spinal Fixation<br>System(K181139) | | | |
| Materials | Devices | Similarities |
|------------------------------------------------------------------------------------------------------------------------------|-------------------------|------------------------------------------------------------------------------------------------------------------------|
| | | |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System | Subject | Similar materials<br>Ti-6Al-4V ELI titanium alloy (ASTM F136) Cobalt-28Chromium-6Molybdenum-4Vanadium ELI (ASTM F1537) |
| LnK Spinal Fixation ystem(K120270) | Primary<br>Predicate | |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System (K120270,<br>K143363, K171813, K183168, K<br>200790, K223565) | Additional<br>Predicate | |
| Olympic Posterior Spinal Fixation<br>System(K181139) | | |
{7}------------------------------------------------
| Design with Components | | Similarities | Dissimilarities |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------|-------------------------------------------------------------------------------------|------------------------------|
| Devices | | | |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System | Subject | Screws, Set Screws,<br>Rods, Crosslinks,<br>Connectors, Hooks<br>Rods include Z-rod | |
| LnK Spinal Fixation System(K120270) | Primary<br>Predicate | Screws, Set Screws,<br>Rods, Crosslinks,<br>Connectors, Hooks | Rods do not include<br>Z-rod |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System (K120270, K143363,<br>K171813, K183168, K 200790, K223565) | Additional<br>Predicate | Screws, Set Screws,<br>Rods, Crosslinks,<br>Connectors, Hooks | Rods do not include<br>Z-rod |
| Olympic Posterior Spinal Fixation<br>System(K181139) | Additional<br>Predicate | Screws, Set Screws,<br>Rods, Crosslinks,<br>Connectors, Hooks<br>Rods include Z-rod | |
| Dimension | Devices | | Similarities | |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------|--------|----------------|----------------------------------|
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System | Subject | Screws | OD Ø<br>Length | 4.0~10.5 mm<br>20~150mm |
| | | Rods | OD Ø<br>Length | 5.0/5.5/6.0/6.35 mm<br>40~600 mm |
| | | Z-Rod | OD Ø<br>Length | 5.5/6.0mm<br>320mm |
| LnK Spinal Fixation System(K120270) | Primary<br>Predicate | Screws | OD Ø<br>Length | 4.0~10.5 mm<br>20~150mm |
| | | Rods | OD Ø<br>Length | 5.0/5.5/6.0/6.35 mm<br>40~600 mm |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System (K120270, K143363,<br>K171813, K183168, K 200790, K223565) | Additional<br>Predicate | Screws | OD Ø<br>Length | 4.0~10.5 mm<br>20~150mm |
| | | Rods | OD Ø<br>Length | 5.0/5.5/6.0/6.35 mm<br>40~600 mm |
| Olympic Posterior Spinal Fixation<br>System(K181139) | Additional<br>Predicate | Screws | OD Ø<br>Length | 4.0~10 mm<br>20~120mm |
| | | Rods | OD Ø<br>Length | 5.0/5.5/6.0/6.35 mm<br>- |
| | | Z-Rod | OD Ø<br>Length | 5.5/6.0mm<br>- |
{8}------------------------------------------------
| Sterilization Method | Devices | Similarities |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------|------------------------------------------------------------------------------------------------------------------|
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System | Subject | Sterile device: Gamma radiation Non-sterile device: recommended steam<br>sterilization (autoclave sterilization) |
| LnK Spinal Fixation System(K120270) | Primary<br>Predicate | Sterile device: Gamma radiation Non-sterile device: recommended steam<br>sterilization (autoclave sterilization) |
| LnK Spinal Fixation System/OpenLoc-L<br>Spinal Fixation System (K120270, K143363,<br>K171813, K183168, K 200790, K223565) | Additional<br>Predicate | Sterile device: Gamma radiation Non-sterile device: recommended steam<br>sterilization (autoclave sterilization) |
| Olympic Posterior Spinal Fixation<br>System(K181139) | Additional<br>Predicate | Non-sterile |
# ACCELFIX SPINAL FIXATION SYSTEM
| INDICATION FOR USE | | |
|-------------------------------------------------------|-------------------------|----------------------------------------------------------|
| Devices | | Similarities |
| AccelFix Spinal Fixation System | Subject | |
| AccelFix Spinal Fixation System (K182544) | Primary<br>Predicate | |
| AccelFix Spinal Fixation System<br>(K200794, K223565) | Additional<br>Predicate | Instruction for use including indication are<br>similar. |
| Olympic Posterior Spinal Fixation<br>System(K181139) | | |
| Materials | Devices | Similarities |
|-----------|----------------------------------------------------|----------------------------------------------------------|
| | AccelFix Spinal Fixation System | Similar materials |
| | AccelFix Spinal Fixation System (K182544) | Ti-6Al-4V ELI titanium alloy (ASTM F136) |
| | AccelFix Spinal Fixation System (K200794, K223565) | Cobalt-28Chromium-6Molybdenum-4Vanadium ELI (ASTM F1537) |
| | Olympic Posterior Spinal Fixation System(K181139) | |
| | Subject | |
| | Primary Predicate | |
| | Additional Predicate | |
| Design with Components | | | |
|-------------------------------------------------------|-------------------------|----------------------------------------------------------------------------------|------------------------------|
| Devices | Similarities | Dissimilarities | |
| AccelFix Spinal Fixation System | Subject | Screws, Set Screws, Rods,<br>Crosslinks, Connectors, Hooks<br>Rods include Z-rod | |
| AccelFix Spinal Fixation System (K182544) | Primary<br>Predicate | Screws, Set Screws, Rods,<br>Crosslinks, Connectors, Hooks | Rods do not<br>include Z-rod |
| AccelFix Spinal Fixation System<br>(K200794, K223565) | Additional<br>Predicate | Screws, Set Screws, Rods,<br>Crosslinks, Connectors, Hooks | Rods do not<br>include Z-rod |
| Olympic Posterior Spinal Fixation<br>System(K181139) | Additional<br>Predicate | Screws, Set Screws, Rods,<br>Crosslinks, Connectors, Hooks<br>Rods include Z-rod | |
{9}------------------------------------------------
| Dimension | Similarities | | |
|----------------------------------------------------|----------------------|-------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Devices | | | |
| AccelFix Spinal Fixation System | Subject | Screws<br>Rods<br>Z-Rod | OD Ø $4.0~10.5$ mm<br>Length $20~150$ mm<br>OD Ø $5.5/6.0/6.35$ mm<br>Length $40~600$ mm<br>OD Ø $5.5/6.0$ mm<br>Length $320$ mm |
| AccelFix Spinal Fixation System (K182544) | Primary Predicate | Screws<br>Rods | OD Ø $4.0~10.5$ mm<br>Length $20~150$ mm<br>OD Ø $5.5/6.0/6.35$ mm<br>Length $40~600$ mm |
| AccelFix Spinal Fixation System (K200794, K223565) | Additional Predicate | Screws<br>Rods | OD Ø $4.0~10.5$ mm<br>Length $20~150$ mm<br>OD Ø $5.5/6.0/6.35$ mm<br>Length $40~600$ mm |
| Olympic Posterior Spinal Fixation System(K181139) | Additional Predicate | Screws<br>Rods<br>Z-Rod | OD Ø $4.0~10$ mm<br>Length $20~120$ mm<br>OD Ø $5.5/6.0/6.35$ mm<br>Length -<br>OD Ø $5.5/6.0$ mm<br>Length - |
| Sterilization Method | | |
|-------------------------------------------------------|----------------------|---------------------------------------------------------------------------------------------------------------|
| Devices | | Similarities |
| AccelFix Spinal Fixation System | Subject | Sterile device: Gamma radiation Non-sterile device: recommended steam sterilization (autoclave sterilization) |
| AccelFix Spinal Fixation System (K182544) | Primary Predicate | Sterile device: Gamma radiation Non-sterile device: recommended steam sterilization (autoclave sterilization) |
| AccelFix Spinal Fixation System<br>(K200794, K223565) | Additional Predicate | Sterile device: Gamma radiation Non-sterile device: recommended steam sterilization (autoclave sterilization) |
| Olympic Posterior Spinal Fixation<br>System(K181139) | Additional Predicate | Non-sterile |
#### 9. SUBSTANTIAL EQUIVALENCE AND CONCLUSION
The subject additional components of LnK Spinal Fixation System/ OpenLoc-L Spinal Fixation System and AccelFix Spinal Fixation System have been found to be substantially equivalent to the predicate devices with respect to technical characteristics and intended use.
The information provided within this premarket notification supports substantial equivalence of the subject device to the predicate devices. The overall data lead to the conclusion that the additional components of LnK Spinal Fixation System/ OpenLoc-L Spinal Fixation System and AccelFix Spinal Fixation System are substantially equivalent to the predicate devices (K120270, K143363, K171813, K183168, K200790, K182544, K200794, K223565).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.