← Product Code [KPR](/productcode/KPR) · K230241

# Jumong General (K230241)

_Sg Healthcare Co, Ltd. · KPR · Feb 23, 2023 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K230241

## Device Facts

- **Applicant:** Sg Healthcare Co, Ltd.
- **Product Code:** [KPR](/productcode/KPR.md)
- **Decision Date:** Feb 23, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Jumong General is intended for use by a qualified/trained doctor or technician on both adult and pediatic subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

## Device Story

Jumong General is a stationary X-ray system for diagnostic imaging; used by physicians or technicians in clinical settings. System captures radiographic images of skull, spine, chest, abdomen, and extremities; patient positioned sitting, standing, or lying. Device integrates X-ray generator (Delta Electronics DMP 100R), collimator (Fairy Medical Electronics CRUX FR04), tubehead (Hangzhou Kailong H1074X), and previously cleared digital X-ray acquisition panels/software. System transforms X-ray exposure into digital images for clinical review; aids healthcare providers in diagnostic decision-making. Benefits include high-quality diagnostic imaging for various anatomical regions.

## Clinical Evidence

No clinical data. Bench testing only. Includes integration testing, image quality assessment, and compliance testing for radiation safety (IEC 60601-2-54), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2).

## Technological Characteristics

Stationary X-ray system. Components: Delta Electronics DMP 100R generator (40-150 kVp), Fairy Medical Electronics CRUX FR04 collimator, Hangzhou Kailong H1074X tubehead. Digital panels: Vieworks Vivix-S (wireless/wired). Connectivity: DICOM 3.0. Power: AC line. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-54, IEC 60613, IEC 60336. Software: Unchanged from predicate (K122866).

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Jumong Series Stationary Radiographic System ([K150816](/device/K150816.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.

February 23, 2023

SG Healthcare CO LTD % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES, FL 34114

Re: K230241

Trade/Device Name: Jumong General Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR, MOB Dated: January 30, 2023 Received: January 30, 2023

Dear Mr. Daniel Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230241

Device Name Jumong General

Indications for Use (Describe)

The Jumong General is intended for use by a qualified/trained doctor or technician on both adult and pediatic subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Type of Use (Select one or both, as applicable)

| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary, 510(k) K230241 Submitter: SG Healthcare Co., Ltd. 10 Yangnyeong-ro, Moga-myeon, Icheon-Si, Gyeonggi-do 17408, KOREA Tel : +82-70-7011-6161 FAX : +82-31-737-4954 Contact: YOSEP PARK, sales@sghealthcare.com Date Prepared: February 1, 2023

- 1. Identification of the Device: System.Trade/Device Name: Jumong General Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR, MQB
- 2. Equivalent legally marketed device: K150816, System.Trade/Device Name: Jumong Series Stationary Radiographic System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR, MQB.
- 3. Indications for Use The Jumong General is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest. abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
- 4. Description of the Device: This device represents a combination of an already cleared solid state digital x-ray acquisition panel with software and diagnostic x-ray compnents required to make a complete system. Film cassettes may be employed in place of the x-ray generator has been changed to a Delta Electronics Delta DMP 100R The collimator has been changed to a Fairy Medical Electronics model CRUX FR04. The tubehead has been changed to a Hangzhou Kailong Medical Instruments Co., Ltd H1074X. The system complies with the CDRH Radiological Health performance standard in the Code of Federal Regulations, as well as the voluntary IEC standards IEC 60601-1 and IEC 60601-1-2.
- 5. Safety and Effectiveness, comparison to predicate device. This combination device has the same indications for use and very similar technological characteristics as the predicate device, and employs already 510(k) cleared digital panels and software.
- 6. Substantial Equivalence Chart: Please see the next page.

{4}------------------------------------------------

| Characteristic                                            | K150816, Jumong Series Stationary<br>Radiographic System SG Healthcare, Co.<br>Ltd.,                                                                                                                                                                                                                                                                                                                         | Modified: Jumong General                                                                                                                                                                                                                                                                                                                                                                                               |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:                                             | Jumong Series is intended for use by a<br>qualified, trained doctor or technician on<br>both adult and pediatric subjects for<br>taking diagnostic radiographic exposures<br>of the skull, spinal column, chest.<br>abdomen, extremities, and other body<br>parts. Applications can be performed<br>with the patient sitting, standing, or lying<br>in the prone or supine position. Not for<br>mammography. | Jumong General is intended for use by a<br>qualified/trained doctor or technician on<br>both adult and pediatric subjects for taking<br>diagnostic radiographic exposures of the<br>skull, spinal column, chest. abdomen,<br>extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the<br>prone or supine position. Not for<br>mammography. UNCHANGED |
| Generator                                                 | CPI CMP 200 DR kVp range: 40 to 150<br>kVp, depending on model                                                                                                                                                                                                                                                                                                                                               | Delta Electronics DMP 100R kVp range: 40<br>to 150 kVp SAME                                                                                                                                                                                                                                                                                                                                                            |
| Collimator                                                | Ralco Model R225                                                                                                                                                                                                                                                                                                                                                                                             | Fairy Medical Electronics CRUX FR04                                                                                                                                                                                                                                                                                                                                                                                    |
| Tubehead                                                  | Varian RAD14                                                                                                                                                                                                                                                                                                                                                                                                 | Hangzhou Kailong Medical Instruments Co.,<br>Ltd H1074X                                                                                                                                                                                                                                                                                                                                                                |
| Digital Panel<br>Models and<br>their clearance<br>numbers | Vieworks<br>K122865 Vivix-S Wireless<br>K122866 Vivix-S With Vxvue<br>K120020 Vivix-S                                                                                                                                                                                                                                                                                                                        | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Image<br>acquisition<br>panel<br>specifications           | FXRD-1717SA, FXRD-1717SB<br>3,072 x 3,072, 140μm or<br>FXRD-1417SA, FXRD-1417SB<br>2560 x 3072, 140μm<br>Wireless: FXRD-1417WA, FXRD-1417WB,<br>2560 x 3072, 140μm                                                                                                                                                                                                                                           | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |
| DICOM                                                     | DICOM 3                                                                                                                                                                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Image<br>acquisition<br>software                          | Vieworks as cleared in K122866 Vivix-S With<br>Vxvue                                                                                                                                                                                                                                                                                                                                                         | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Power Source                                              | AC Line, various voltages available                                                                                                                                                                                                                                                                                                                                                                          | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Characteristic                                            | K150816, Jumong Series Stationary<br>Radiographic System SG Healthcare, Co.<br>Ltd.,                                                                                                                                                                                                                                                                                                                         | Modified: Jumong General                                                                                                                                                                                                                                                                                                                                                                                               |
| Photo, floor<br>mount                                     | Image: Photo, floor mount                                                                                                                                                                                                                                                                                                                                                                                    | Image: Photo, floor mount                                                                                                                                                                                                                                                                                                                                                                                              |
| Photo, ceiling<br>mount                                   | Image: Photo, ceiling mount                                                                                                                                                                                                                                                                                                                                                                                  | Image: Photo, ceiling mount                                                                                                                                                                                                                                                                                                                                                                                            |
| Performance<br>Standard                                   | US Radiation Safety Performance<br>Standard                                                                                                                                                                                                                                                                                                                                                                  | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Electrical<br>safety/EMC<br>Standards<br>compliance       | IEC 60601-1, IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                   | SAME                                                                                                                                                                                                                                                                                                                                                                                                                   |

{5}------------------------------------------------

- 7. Summary of non-clinical testing: We performed integration and image quality testing. The device conforms to the US Performance Standard. The new generator was tested by an NRTL to comply with: IEC 60601-2-54:2009+A1+A2, IEC 60601-1:2005+Corrigendum 1+Corrigendum 2+A1, IEC 60601-1-3:2008+A1, and IEC 60601-1-6:2010+A1. EMC testing complies with IEC 60601-1-2:2014. The new collimator was tested for compliance with EN/IEC 60601-2-54 for radiation leakage. The new tubehead was tested to IEC 60613:2010 Electrical and loading characteristics of X-ray tube assemblies for medical diagnosis and IEC 60336:2005 Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Focal spot dimensions and related characteristics. Test images were acquired which showed excellent diagnostic quality. Note that the digital x-ray receptor panels and software have not changed from the predicate.
Every unit is tested for electrical safety, input power, display of operation and exposure factors, collimator operation, reproducibility, and accuracy. A risk analysis was performed with regard to the 

{6}------------------------------------------------

modifications. No software modifications were made to the imaging chain.

- 8. Summary of clinical testing: Not applicable.
- 9. Conclusion: The results of a review of bench, safety test, and software validation documentation indicates that the new device is as safe and effective as the predicate device. After analyzing safety testing data, final checkout, and test images, it is the conclusion of SG Healthcare that the modified "Jumong General" is as safe and effective as the predicate device, has few technological differences, and has identical indications for use, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K230241](https://fda-staging.innolitics.com/device/K230241)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
