Retrospective clinical cases were used to validate the performance of the ClearInfinity deep learning image reconstruction algorithm by comparing it against Filtered Back Projection (FBP) via radiologist reader studies.
The NeuViz 128 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
Device Story
NeuViz 128 is a whole-body CT scanner featuring a 72cm bore, rotating X-ray tube, and 64-row solid-state GOS ceramic detector array. It acquires X-ray transmission data, which is reconstructed into cross-sectional images via FBP, ClearView (iterative), or ClearInfinity (deep learning). Operated by clinicians in a clinical setting, the device supports various scan protocols and advanced post-processing (e.g., MPR, 3D, cardiac analysis). ClearInfinity uses a trained neural network to reduce noise and improve low-contrast detectability, potentially allowing for reduced radiation dose. Output is viewed by radiologists to assist in diagnostic decision-making. The system includes intelligent positioning and automated FOV features to assist in scan planning.
Clinical Evidence
No prospective clinical trials were conducted. Evidence consists of bench testing using CCT189/CCT191 MITAIQ LC phantoms and retrospective clinical image evaluations. Reader studies (n=30 and n=20 cases) assessed image quality via 5-point Likert scales, with 3 board-certified radiologists confirming ClearInfinity is equivalent or superior to FBP. Bench testing demonstrated dose reduction (up to 85% for body), noise reduction (up to 90%), and improved low-contrast detectability/spatial resolution compared to FBP.
Indicated for whole-body computed tomography X-ray imaging in patients requiring cross-sectional diagnostic imaging. No specific age or gender contraindications provided, though pediatric use requires radiologist/physicist consultation for scan parameters.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
NeuViz 128 Multi-Slice CT Scanner System (K151383)
Reference Devices
NeuViz Prime Multi-Slice CT Scanner System (K171201)
Submission Summary (Full Text)
{0}------------------------------------------------
October 11, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized representation of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
Neusoft Medical Systems Co., Ltd. % Tian Yuehui Q&R Manager No. 177-1 Chuangxin Road, Hunnan District SHENY ANG, LIAONING 110167 CHINA
Re: K230220
Trade/Device Name: NeuViz 128 Multi-Slice CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: September 8, 2023 Received: September 11, 2023
Dear Tian Yuehui:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230220
Device Name NeuViz 128 Multi-Slice CT Scanner System
### Indications for Use (Describe)
The NeuViz 128 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;"> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span style="font-size: 10pt;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92
## 1. General Information:
| Manufacturer: | Neusoft Medical Systems Co., Ltd.<br>No.177-1 Chuangxin Road, Hunnan District,<br>Shenyang, Liaoning, China, 110167. |
|-----------------|----------------------------------------------------------------------------------------------------------------------|
| Contact person: | Tian Yuehui<br>Title : Q&R Manager<br>Tel : 86-24-23358105<br>E-Mail : tianyh@neusoftmedical.com |
Date of Preparation: October 8, 2023
## 2. Device Name and Classification:
| Trade Name: | NeuViz 128 Multi-Slice CT Scanner System |
|-----------------------|------------------------------------------------------------------|
| Common Name: | CT Scanner |
| Regulation Number: | 21 CFR 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Product Code: | JAK |
| Classification: | Class II |
| Performance Standard: | 21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard |
### 3. Predicate device:
| Trade Name: | NeuViz 128 Multi-Slice CT Scanner System |
|----------------------|--------------------------------------------------------------------------------------|
| 510(k) number: | K151383 |
| Clearance Date: | 11/04/2015 |
| Regulation Number: | 21 CFR 892.1750 |
| Classification Name: | Computed tomography x-ray system |
| Product Code: | JAK |
| Classification: | Class II |
| Manufacturer: | Neusoft Medical Systems Co., Ltd. |
| Recall Information: | All predicate device recalls have been considered in the<br>subject<br>device design |
### 4. Reference Device:
| Trade Name: | NeuViz Prime Multi-Slice CT Scanner System |
|--------------------|--------------------------------------------|
| 510(k) number: | K171201 |
| Clearance Date: | 09/13/2017 |
| Regulation Number: | 21 CFR 892.1750 |
{4}------------------------------------------------
# Neusoft 东软医疗
Classification Name: Product Code: Classification: Manufacturer:
Computed tomography x-ray system JAK Class II Neusoft Medical Systems Co., Ltd.
### 5. Reason for submission:
Modification of existing medical device
### 6. Device Description:
The proposed NeuViz 128 Multi-slice CT Scanner System is composed of a gantry, a patient couch, an operator console and includes image acquisition hardware and software, and associated accessories. It is designed to be a head and whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle.
Same as the predicate, The gantry has a 72cm bore with a maximum FOV of 50 cm and available rotation speeds are still 0.374s,0.5s, 0.6s.0.8s,1.0s,1.5s,2.0s per 360° rotation.
The proposed NeuViz 128 has three types of reconstruction methods available: FBP(filter back-projection), ClearView(iterative reconstruction algorithm), and Clearlnfinity . ClearInfinity is a new added recon mode where the system uses a trained deep learning neural network to generate noise reduction images and improve low contrast detectability with reduced dose compared with standard FBP recon mode.
### 7. Indications for use
The NeuViz 128 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
### 8. Indications for use comparison
The indications for use of the proposed Neu\/iz 128 is the same as that of the predicate device.
### 9. Comparison of Technological Characteristics with the Predicate Devices
The proposed NeuViz 128 includes most of the available features on the current production predicate device and employs the same fundamental technologies as those of the predicate device.
A tabular summary of the comparable hardware ,software properties and features between the proposed device with predicate device are listed in Table 1and 2 below (modifications are in gray shaded sections).
{5}------------------------------------------------
| Hardware property | Proposed device<br>NeuViz 128 | Predicate device<br>NeuViz 128 (K151383) |
|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Gantry<br>Aperture | 720mm | 720mm |
| Gantry<br>Tilt | +/-30° | +/-30° |
| Gantry<br>Scan Speed (s / 360°) | 0.374s、0.5s、0.6s、0.8s、<br>1.0s、1.5s、2.0s | 0.374s、0.5s、0.6s、0.8s、<br>1.0s、1.5s、2.0s |
| Detector Type | Solid-state GOS ceramic | Solid-state GOS ceramic |
| Detector Number of<br>Detector Rows | 64 | 64 |
| Maximum slices generated<br>per rotation (multislice<br>capability) | 128 | 128 |
| Generator Max. Power | 80kW | 80kW |
| Generator mA Range | 10mA~667mA | 30mA~667mA |
| Generator kV Settings | 80kV, 100kV, 120kV,<br>140kV | 80kV, 100kV, 120kV,<br>140kV |
| Tube<br>Focal Spots | 0.6×1.2 (small)<br>1.1×1.2 (large) | 0.6×1.2 (small)<br>1.1×1.2 (large) |
| Tube<br>heat capacity | 8.0MHU | 8.0MHU |
| Couch<br>Type/max. moving<br>length/max. table<br>loading/max. horizontal<br>speed | standard couch(option):<br>211kg/225mm/s<br>long couch(option):<br>211kg/310mm/s<br>300kg couch(option): | standard couch:<br>205kg/160mm/s<br>300kg couch (option):<br>300kg/160mm/s |
| | 300kg/225mm/s | |
| Intelligent Positioning<br>Device<br>(optional) | option for Intelligent<br>Positioning feature | N/A |
Table 1:Technical hardware characteristics for subject device compared to the predicate devices.
{6}------------------------------------------------
Table 2: Software characteristics and features for subject device compared to the predicate devices.
| Software property<br>and Feature | Proposed device<br>NeuViz 128 | Predicate device<br>NeuViz 128 (K151383) |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Recon FOV | 50mm~500mm;<br>Extended FOV 700mm(Optional) | 50mm~500mm |
| Spiral Scan Range of<br>Pitch | 0.1-2.1 | 0.13-1.5 |
| Reconstruction<br>Algorithm | Filtered back-projection (FBP);<br>ClearView;<br>ClearInfinity (Optional); | Filtered back-projection<br>(FBP) ;<br>ClearView |
| O-Dose | Support | Support |
| Bolus tracking | Support | Support |
| SAS | Support | Support |
| Home | Support | Support |
| Film | Support | Support |
| Report | Support | Support |
| Image Review | Support | Support |
| MPR | Support | Support |
| 3D | Support | Support |
| Virtual Endoscopy | Support | Support |
| Dental Analysis | Support | Support |
| (Optional) | | |
| Vessel Analysis | Support | Support |
| DICOM Viewer | Support | Support |
| Virtual<br>Colonoscopy(Optional<br>) | Support | Support |
| Brain<br>Perfusion(Optional) | Support | Support |
| Body<br>Perfusion(Optional) | Support | Support |
| Lung Nodule<br>Analysis(Optional) | Support | Support |
| Lung<br>Density(Optional) | Support | Support |
| Coronary<br>Analysis(Optional) | Support | Support |
| Cardiac Calcium<br>Scoring(Optional) | Support | Support |
| Cardiac Function<br>Analysis(Optional) | Support | Support |
| Cardiac<br>Viewer(Optional) | Support | Support |
| Fat Analysis(Optional) | Support | Support |
| CTDSA(Optional) | Support | Support |
| Tumor<br>Assessment(Optional) | Support | Support |
| Real-time<br>MPR(Optional) | Support. Real-time MPR reconstruction is a recon mode, in the clinical scanning, automatically generate sagittal or coronal MPR images based on the CT axial images. Reduce secondary loading reconstruction time. | N/A |
| 4D scanning(Optional) | Support.<br>It is a kind of fast spiral perfusion<br>scan.The table top of couch<br>moves back and forth repeatedly<br>in the area of interest when the<br>patient is injected with contrast<br>agent. So as to track the dynamic<br>images of contrast agent<br>changing in the same part at<br>different times. | N/A |
| Dual Energy Imaging<br>( Prism Imaging)<br>(Optional) | Support.<br>Dual-energy imaging uses two<br>energies during one CT<br>examination, performs image<br>reconstruction of data originated<br>by the two energy separately and<br>then decomposes the two kinds<br>of images into a set of basis<br>images of pre-defined materials<br>(the decomposition parameters<br>are pre-set by the manufacturer) ,<br>according to which the next<br>process is conducted. | N/A |
| Prism<br>Viewer(Optional) | Support<br>Prism Viewer application is used<br>to view and analysis multi energy<br>images. Display a variety of<br>images with different parameters<br>and provide visual tools, to help<br>users to locate the lesion<br>accurately. | N/A |
| Intelligent<br>Positioning(Optional) | Support.<br>The system collects and displays<br>the natural image information of<br>the human body. The system can<br>use the human body image<br>information to automatically<br>calculates the scanning position<br>according to the scanning<br>protocol and using artificial<br>intelligence<br>technology. At the same time,<br>User can also manually draw the<br>scan frame on the human<br>body Image, And Then, the<br>system calculates the scanning<br>position by the scan frame. | N/A |
| | According to the scanning<br>protocol, the system can<br>automatically adjust the couch<br>height from intelligence setting.<br>The system support patient<br>position, posture and collision<br>Detection. | |
| Coronary Motion<br>Clear(Optional) | Support.<br>Coronary Motion Clear functi<br>on is used to generate imag<br>e dateset with less motion ar<br>tefact in coronary area. | N/A |
| Metal artifact<br>removal(Optional) | MAR,MAR+.<br>MAR/MAR+ stands for metal<br>artifact reduction. It's the post<br>processing algorithm that can<br>remove the artifacts caused by<br>metal or high CT value. MAR<br>method only needs image<br>data, while MAR+ method needs<br>both image data and raw data. | MAR |
| Arrhythmia<br>Handling(Optional) | Support.<br>Recognize and ignore the ar<br>rhythmia R-peak during card<br>iac scan, and trigger the sca<br>n while normal R-peak recog<br>nized. | N/A |
| Auto FOV+(Optional) | Support.<br>Auto FOV is automatically mark t<br>he FOV range on the surview im<br>age based on Al technology, and<br>the FOV range can be adjusted<br>manually. Supported scan parts i<br>nclude head and lungs | N/A |
{7}------------------------------------------------
# Neusoft 东软医疗
{8}------------------------------------------------
# Neusoft 东软医疗
{9}------------------------------------------------
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject device is substantially equivalent to the predicate devices.
### Note:
* Auto FOV has not been thoroughly tested on pediatric under 3 years old. A consultation with a radiologist and a physicist should be made to determine the appropriate scan FOV for perform the particular clinical task.
{10}------------------------------------------------
### 10. Performance Data
### Summary of Non-Clinical Testing:
The device has successfully completed all testing per our quality system as well as addition engineering bench testing in support of this submission. It was designed and is manufactured under the Quality System Regulations of 21 CFR 820 and ISO 13485 .
This device is in conformance with the applicable parts of the following standards: --ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
--ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)] --IEC 60601-1-3 Edition 2.2 2021-01 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
--IEC 60601-1-6 Edition 3.2 2020-07 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability --IEC 60601-2-28 Edition 3.0 2017-06 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
--IEC 60601-2-44 Edition 3.2: 2016 Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography
--IEC 60825-1 Edition 2.0 2007-03 Safety of laser products - Part 1: Equipment classification, and requirements
-- IEC 61223-3-5:2019 Evaluation and routine testing in medical imaging departments -Part 3-5: Acceptance and constancy tests - Imaging performance of computed tomography X-rav equipment
--IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes --IEC 62366-1 2020 Medical devices Part 1: Application of usability enqineering to medical devices
--ISO 14971:2019 Medical devices - Application of risk management to medical devices --NEMA XR 25: 2019, Computed tomography dose check
--NEMA XR 28: 2013, Supplemental Requirements for User Information and System Function Related to Dose in CT
Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
Risk analysis and verification/validation activities conducted through bench testing which is included in this submission demonstrate that the established specifications for the device have been met. Additional performance testing, using phantom studies, were conducted to assess the improvements to existing features. Results of all these studies demonstrate that the features included in this submission meet specifications and perform as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document,"Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also referenced as for this submission.
{11}------------------------------------------------
Neusoft Medical Systems Co., Ltd conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices issues on October 2. 2014" is included within this submission.
### Additional Non-Clinical Testing for ClearView and ClearInfinity
### Performance Verification for ClearView*
Engineering bench testing was performed to support substantial equivalence and the product performance claims. The evaluation and analysis used the raw datasets obtained on NeuViz 128 and then applies both ClearView and Filtered Back Projection reconstruction. The content and results of the test are as follows:
Quantitative dose reduction test using the CCT189 and CCT191 MITAIQ LC Phantoms and the channelized Hotelling observer. Compared with FBP, ClearView may enable lower radiation dose of head by 40% to 50% at the same image quality*; Compared with FBP, ClearView may enable lower radiation dose of body by 45% to 60% at the same image quality*.
Quantitative low-contrast resolution improvement test using the CCT189 and CCT191 MITAIQ LC Phantoms and the channelized Hotelling observer. Compared with FBP, ClearView may enable improve low contrast detectability of head by 30% to 35% at the same dose*; Compared with FBP, ClearView may enable improve low contrast detectability of body by 35% to 45% at the same dose *.
Quantitative noise reduction test using the water layer of the QA phantom. Compared with FBP, ClearView may enable reduce image noise up to 55% at the same dose*.
### Note:
* ClearView has not been thoroughly tested on advanced post-processing application. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task Performance Verification for ClearView:
ClearView has not been thoroughly tested in pediatric population. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task;
Clinicians or dosimetrists should only be allowed to reduce patient dose in clinical practice.
* In clinical practice, the use of ClearView may reduce CT patient dose depending on the clinical task, patient size, anatomical location, and clinical practice. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task. Low Contrast Detectability (LCD), Image Noise were assessed using normal dose comparing ClearView and FBP. The LCD and Image Noise measured in smooth kernel and 0.625 mm slices. The claims are based on ClearView 90% level.
### Performance Verification for ClearInfinity*
Enqineering bench testing was performed to support substantial equivalence and the product performance claims. The evaluation and analysis used the raw datasets obtained on NeuViz 128 and then applies both Clearlnfinity and Filtered Back Projection reconstruction. The content and results of the test are as follows:
{12}------------------------------------------------
Quantitative dose reduction test using the CCT189 and CCT191 MITAIQ LC Phantoms and the channelized Hotelling observer. Compared with FBP, Clearlnfinity may enable lower radiation dose of head by 50% to 60% at the same image quality*; Compared with FBP, Clearlnfinity may enable lower radiation dose of body by 75% to 85% at the same image quality*.
Quantitative low-contrast resolution improvement test using the CCT189 and CCT191 MITAIQ LC Phantoms and the channelized Hotelling observer. Compared with FBP. Clearlnfinity mav enable improve low contrast detectability of head by 35% to 65% at the same dose*; Compared with FBP, Clearlnfinity may enable improve low contrast detectability of body by 60% to 115% at the same dose *.
Quantitative noise reduction test using the water layer of the QA phantom. Compared with FBP, ClearInfinity may enable reduce image noise up to 90% at the same dose*.
Quantitative improvement of spatial resolution test using the high-contrast test device in the physical layer and water layer of the QA phantom. Compared with FBP, ClearInfinity may enable improve spatial resolution up to 1.88X (88%) at same image noise*.
#### Note:
* US patient-based testing data for Clearlnfinity collected from three sites of OK and one site of TX, a total of 445 clinical cases of 297 independent patients;
ClearInfinity has not been thoroughly tested on advanced post-processing application. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task Performance Verification for ClearInfinity;
ClearInfinity has not been thoroughly tested in pediatric population. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task;
Clinicians or dosimetrists should only be allowed to reduce patient dose in clinical practice.
* In clinical practice, the use of Clearlnfinity may reduce CT patient dose depending on the clinical task, patient size, anatomical location, and clinical practice. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task. Low Contrast Detectability (LCD),Image Noise, Spatial Resolution were assessed using normal dose comparing ClearInfinity and FBP. The LCD and Image Noise measured in smooth kernel and 0.625 mm slices; Spatial Resolution measured in sharp kernel and 0.625 mm slices. The claims are based on ClearInfinity 90% level.
### Clinical Image Evaluation for ClearInfinity
The reader study used a total of 30 retrospectively collected clinical cases (Determine the sample size according to Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices) . The raw data from each of these cases was reconstructed with both Filtered Back Projection and ClearInfinity Deep Learning Image Reconstruction. The images generated by the two reconstruction methods of the same raw data are presented to the reader in pairs. Each image was read by 3 board-certified radiologists who provided an assessment of image quality related to diagnostic use according to a 5-point Likert scale. The results of the study indicate that ClearInfinity is equivalent or better than Filtered Back Projection in diagnostic quality. An additional study used a total of 20 retrospectively collected clinical cases which raw data were processed at simulated low doses to produce low-dose images. Each of the low dose images
{13}------------------------------------------------
was reconstructed with both Filtered Back Projection and ClearInfinity Deep Learning Image Reconstruction. Each image was read by 3 board-certified radiologists who provided an assessment of image quality related to diagnostic use according to a 5-point Likert scale. The results of the study indicate that Clearlnfinity is equivalent or better than Filtered Back Projection in diagnostic quality at low doses.
Testing for verification and validation support the claims of substantial equivalence.
### Clinical Testing
No Clinical Study is included in this submission.
### 11.Substantial equivalence Conclusions
Based on the conformance to standards, development under our quality system, and the engineering testing provided, we believe that the modified NeuViz 128 is as safe and effective, and performs in a substantially equivalent manner to the predicate device (K151383)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.