Sonix Health is intended for quantifying and reporting echocardiography for use by or on the order of a licensed physician. Sonix Health accepts DICOM-compliant medical images acquired from ultrasound imaging devices. Sonix Health is indicated for use in adult populations. Ultrasound images are acquired via the B (2D), M, Pulsed-wave Doppler, and Continuous-wave Doppler modes.
Device Story
Sonix Health is a software-only medical device installed on local PC hardware; it processes DICOM-compliant ultrasound images (B-mode, M-mode, Pulsed-wave/Continuous-wave Doppler). The device employs a two-step AI algorithm: first, it identifies the specific echocardiographic view; second, it performs deep learning-based anatomical segmentation and automated measurements according to ASE guidelines. The system operates within a local network in hospitals, clinics, or physician offices. Clinicians review the automated output, verify results, and manually adjust measurements as needed before generating a final report. By automating routine quantification tasks, the device aims to increase clinical efficiency and support physician decision-making in adult echocardiography.
Clinical Evidence
No clinical testing conducted. Bench testing performed on 2,744 clinical images (476 B-mode, 243 M-mode, 2,025 Doppler) from diverse sources (U.S. and South Korea). Performance metrics: 98.22% average accuracy for view recognition and 93.98% average correlation coefficient for automated measurements compared to expert manual ground truth. Validation data included LVEF range 14-76%.
Technological Characteristics
Software-only device for local PC installation. DICOM-compliant. Utilizes deep learning algorithms for view classification and anatomical segmentation. Complies with IEC 62304 (software life cycle), ISO 14971 (risk management), and IEC 62366-1 (usability).
Indications for Use
Indicated for adult patients requiring echocardiography quantification and reporting. Used by or on the order of a licensed physician to support diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Ontact Health Co., Ltd. Hyuk-Jae Chang CEO 5f, 50-5, Ewhayeodae-gil, Seodaemun-gu Seoul, KS013/03764 Korea. South
Re: K230209
October 20, 2023
Trade/Device Name: Sonix Health Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: October 19, 2023 Received: September 21, 2023
Dear Hyuk-Jae Chang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Jessica Lamb
Jessica Lamb Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230209
Device Name Sonix Health
Indications for Use (Describe)
Sonix Health is intended for quantifying and reporting echocardiography for use by or on the order of a licensed physician. Sonix Health accepts DICOM-compliant medical images acquired from ultrasound imaging devices. Sonix Health is indicated for use in adult populations.
Ultrasound images are acquired via the B (2D), M, Pulsed-wave Doppler, and Continuous-wave Doppler modes.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
### 1. Date
October 19, 2023
#### Submitter's Information & Contact Person 2.
- Submitter: ONTACT HEALTH Co., Ltd.
- Address: 5F Erom Tower, 50-5, Ewhayeodae-gil, Seodaemun-gu, Seoul, 03764, Republic of Korea
- Contact Person: Hyuk-Jae Chang / CEO
- Telephone No.: +82-2- 362-9610
- Email Address: ontact(01@ontacthealth.com
#### Trade Name, Common Name, Classification 3.
- Workstation Software for cardiac ultrasound image review, analysis, and reporting - Common name:
- Trade name: Sonix Health
| Classification Description | 21 CFR Section | Product Code |
|--------------------------------------------------|-----------------|--------------|
| Automated Radiological Image Processing Software | 21 CFR 892.2050 | QIH, LLZ |
As stated in 21 CFR parts 892.2050 of the device has been classified as Class II.
#### 4. Identification of Predicate Device(s)
The identified predicate devices within this submission are shown as follow:
### Predicate Device
- 510(k) Number: K213544 .
- Trade/Device name: TOMTC-ARENA
- Manufacturer: TOMTEC Imaging Systems GmbH .
### Reference Device
- . 510(k) Number: K220975
- . Trade/Device name: V8 Diagnostic Ultrasound System,
- Manufacturer: Samsung Medison Co., Ltd. .
{4}------------------------------------------------
#### Description of the Device 5.
Sonix Health comes with the following functions:
- Checking ultrasound multiframe DICOM
- Echocardiography multiframe DICOM classification and automatic measurement.
- Verification of the results and making adjustments manually.
- Providing the report for analysis
Sonix Health will be offered as SW only, to be installed directly on customer PC hardware. Sonix Health is DICOM compliant and is used within a local network.
Image /page/4/Figure/7 description: The image shows a diagram of a local area network with a personal computer connected to a NAS. The personal computer has a Sonix Health application running on it. The Sonix Health application has a Windows application and an AI Engine. The Windows application and AI Engine are connected to each other.
Sonix Health utilizes a two-step algorithm. A single identification model identifies a view in the first step. The second step performs the deep learning according to the view. The deep learning algorithms for the second step are categorized as B-mode, and Doppler algorithms. The main algorithm of Sonix Health is to identify the view and segment the anatomy in the image.
#### 6. Indications for Use
Sonix Health is intended for quantifying and reporting echocardiography for use by or on the order of a licensed physician. Sonix Health accepts DICOM-compliant medical images acquired from ultrasound imaging devices. Sonix Health is indicated for use in adult populations.
Ultrasound images are acquired via the B (2D), M, Pulsed-wave Doppler, and Continuous-wave Doppler modes.
{5}------------------------------------------------
# 7. Determination of Substantial Equivalence
The identified predicate devices within this submission is shown in the following table:
| Description | Decision | Subject Device (K230209) | Predicate Device (K213544) | Reference Device (K220975) |
|----------------------------------------------------|----------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device name | | Sonix Health | TOMTEC-ARENA | V8 Diagnostic Ultrasound System |
| Product Code | | QIH (Subsequent Product Code: LLZ) | QIH (Subsequent Product Code: LLZ) | IYN, IYO, ITX |
| Regulatory Class | SE | 2 | 2 | 2 |
| Regulation Number | | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.1550, 892.1560, 892.1570 |
| Intended use | SE | Sonix Health is intended for quantifying<br>and reporting echocardiography for use by<br>or on the order of a licensed physician.<br>Sonix Health accepts DICOM-compliant<br>medical images acquired from ultrasound<br>imaging devices. Sonix Health is indicated<br>for use in adult populations.<br><br>Ultrasound images are acquired via the B<br>(2D), M, Pulsed-wave Doppler, and<br>Continuous-wave Doppler modes. | TOMTEC-ARENA software is a<br>clinical software package designed for<br>review, quantification and reporting of<br>structures and function based on multi-<br>dimensional digital medical data<br>acquired with different modalities.<br>TOMTEC ARENA is not intended to<br>be used for reading of mammography<br>images. | The V8 / V7 diagnostic ultrasound system and<br>probes are designed to obtain ultrasound images<br>and analyze body fluids.<br><br>The clinical applications include: Fetal/Obstetrics,<br>Abdominal, Gynecology, Intra-operative, Pediatric,<br>Small Organ, Neonatal Cephalic, Adult Cephalic,<br>Trans-rectal, Trans-vaginal, Muscular-Skeletal<br>(Conventional, Superficial), Urology, Cardiac<br>Adult, Cardiac Pediatric, Thoracic, Trans-<br>esophageal (Cardiac) and Peripheral vessel.<br><br>It is intended for use by, or by the order of, and |
| Indications for use | SE | | Indications for use of TOMTEC-<br>ARENA TTA2 software are<br>quantification and reporting of<br>cardiovascular, fetal, abdominal<br>structures and function of patients with<br>suspected disease to support the<br>physician in the diagnosis | under the supervision of, an appropriately trained<br>healthcare professional who is qualified for direct<br>use of medical devices. It can be used in hospitals,<br>private practices, clinics and similar care<br>environment for clinical diagnosis of patients.<br><br>Modes of Operation: 2D mode, Color Doppler<br>mode, Power Doppler (PD) mode, M mode, Pulsed<br>Wave (PW) Doppler mode, Continuous Wave<br>(CW) Doppler mode, Tissue Doppler Imaging<br>(TDI) mode, Tissue Doppler Wave (TDW) mode,<br>ElastoScan Mode, Combined modes, Multi-Image<br>mode (Dual, Quad), 3D/4D mode. |
| Where used<br>(hospital, home,<br>ambulance, etc.) | SE | Inside of hospitals, clinics, and physician's<br>offices. | Inside and outside of Hospitals,<br>Clinics, and Physician's offices. | Inside of hospitals, clinics, and physician's offices. |
| Application<br>description | SE | Sonix Health utilizes artificial intelligence<br>to automate previous manual quantification<br>tasks, resulting in increased efficiency for<br>users. Our system performs view | IMAGE-COM is a basic module for<br>reviewing and measuring digital<br>medical data. It supports routine<br>workflows for loading, analyzing and | The V8 / V7 are a general purpose, mobile,<br>software controlled, diagnostic ultrasound system.<br>Their function is to acquire ultrasound data and to<br>display the data as 2D mode,Color Doppler mode, |
| Description | Decision | Subject Device (K230209) | Predicate Device (K213544) | Reference Device (K220975) |
| | | classification and measurements according<br>to the US ASE guidelines, and the users<br>can review and modify the results if<br>necessary. | saving medical studies, e.g. for the<br>purpose of creating reports. IMAGE-<br>COM is where basic measurements can<br>be performed and the entry point for<br>advanced analysis modules. Study<br>related routine measurements can be<br>imported, displayed, edited and<br>exported to accompanying reporting<br>systems. | Power Doppler (PD) mode, M mode, Pulsed Wave<br>(PW) Doppler mode, Continuous Wave (CW)<br>Doppler mode, Tissue Doppler Imaging (TDI)<br>mode, Tissue Doppler Wave (TDW) mode,<br>ElastoScan Mode, Combined modes, Multi-Image<br>mode(Dual, Quad), 3D/4D mode. The V8 / V7 also<br>give the operator the ability to measure anatomical<br>structures and offer analysis packages that provide<br>information that is used to make a K223387<br>Traditional 510(k) diagnosis by competent health<br>care professionals. The V8 / V7 have a real time<br>acoustic output display with two basic indices, a<br>mechanical index and a thermal index, which are<br>both automatically displayed. |
| Semi-automated<br>view classification | SE | PLAX LV, A4C, A4C Zoomed LV, A2C,<br>A2C zoomed LV, M-mode LA/Ao, M-<br>mode LV, CW Doppler MS, CW Doppler<br>MR, PW Doppler MV, CW Doppler AV,<br>CW Doppler AR, CW Doppler TR, CW<br>Doppler PV, CW Doppler PR, PW Doppler<br>RVOT, PW Doppler LVOT, DTI MV<br>annulus | PLAX LV, A4C, A4C Zoomed LV, A2C, A2C<br>zoomed LV, M-mode LA/Ao, M-mode LV, CW<br>Doppler MS, CW Doppler MR, PW Doppler MV,<br>CW Doppler AV, CW Doppler AR, CW Doppler<br>TR, CW Doppler PV, CW Doppler PR, PW<br>Doppler RVOT, PW Doppler LVOT, DTI MV<br>annulus | PLAX LV, A4C, A4C Zoomed LV, A2C, A2C<br>zoomed LV, M-mode LA/Ao, M-mode LV, CW<br>Doppler MS, CW Doppler MR, PW Doppler MV,<br>CW Doppler AV, CW Doppler AR, CW Doppler<br>TR, CW Doppler PV, CW Doppler PR, PW<br>Doppler RVOT, PW Doppler LVOT, DTI MV<br>annulus |
| Semi-automated<br>measurements | SE | IVSd (2D), IVSs (2D), LVIDd (2D),<br>LVIDs (2D), LVPWd (2D), LVPWs (2D),<br>LA (2D), Ao (2D), LVESV / LVEDV<br>(4CH), LAV (4CH), LVESV / LVEDV<br>(4CH), LVESV / LVEDV (2CH), LAV<br>(2CH), LVESV / LVEDV (2CH), LA (m-<br>mode), Aorta (m-mode), IVSd (m-mode),<br>LVIDd (m-mode), LVPWd (m-mode),<br>IVSs (m-mode), LVIIDs (m-mode),<br>LVPWs (m-mode), MV Peak E Vel, MV<br>Peak A Vel, MV Decel Time, e' (med), a'<br>(med), s' (med), LVOT Vmax, LVOT VTI,<br>RVOT Vmax, RVOT VTI, PV AccT, TR<br>Vmax, TR VTI, AR Vmax / PR Vmax, AR<br>PHT / PR PHT, MV Vmax, MV VTI, MV,<br>PHT, MR Vmax, MR VTI, AV Vmax / PV<br>Vmax, AV VTI / PV VTI, RVOT (2D) | B-Mode<br>- Ao Asc diam, Ao Ann diam, Ao STJ<br>diam, Ao SV diam, IVSd, LA diam<br>systole, LVIDd, LVIDs, LVOT diam,<br>LVPWd, RVDd base (RVD1), RVDd<br>mid (RVD2), RVLd, RVOT diam lax,<br>RVOT diam prox, TV Ann diam ant-<br>post, IVSd-LVIDd-LVPWd (Same<br>line)<br><br>DOPPLER Mode<br>- MV A Vel, MV E Vel, MV E/A<br>Slope, (MV A Vel, MV E Vel, MV<br>Time), MV Dec. Slope, (MV Dec<br>Time, MV E Vel), LVOT VTI, AV<br>VTI, PV VTI, TR Vmax, LV E'(l), LV<br>A'(l), LV E'(s), LV A'(s), RV A'(l),<br>RV E'(l), RV S'(l) | (PLAX LV) - Interventricular septum diameter at the<br>diastole and systole phase, LV internal diameter at<br>the diastole and systole phase, LV posterior wall<br>diameter at the diastole and systole phase, Aorta<br>diameter and LA diameter, (A4C) - LA Volume, LV<br>Volume, (A4C Zoomed LV) - LV Volume, (A2C) -<br>LA Volume, LV Volume, (A2C zoomed LV) - LV<br>Volume, (M-mode LA/Ao) - LA diameter and Aorta<br>diameter, (M-mode / LV) - Interventricular septum<br>diameter at the diastole and systole phase, LV<br>internal diameter at the diastole and systole phase,<br>LV posterior wall diameter at the diastole and systole<br>phase, (CW Doppler MS) - Vmax, VTI, PHT, (CW<br>Doppler MR) - Vmax, VTI, (PW Doppler MV) - E.<br>A, DT, (CW Doppler AV) - Vmax, VTI, (CW<br>Doppler AR) - Vmax, PHT, (CW Doppler TR) -<br>Vmax, VTI, (CW Doppler PV) - Vmax, VTI, (CW<br>Doppler PR) - Vmax, EDV, (PW Doppler RVOT) –<br>Vmax, VTI, (PW Doppler LVOT) - Vmax, VTI,<br>(DTI MV annulus) - S', E', A' |
| Description | Decision | Subject Device (K230209) | Predicate Device (K213544) | Reference Device (K220975) |
| Semi-automation<br>technology | SE | After the automated analysis is complete,<br>users can review and modify the results as<br>needed. The product is intended for use by,<br>or on the order of a licensed physician. | Semi-automated adult<br>echocardiography 2D and Doppler<br>measurements are generated using and<br>artificial intelligence (AI) detection<br>algorithm without user interaction.<br>After measurement is generated, the<br>user can edit (manually adjust the<br>caliper positions), accept, or reject the<br>measurements. The automation of<br>measurements is constrained to the<br>specific imaging mode (2D, Doppler)<br>as recommended by ASE guidelines. | After the analysis is complete, users can review and<br>modify the results as needed. The product is for use<br>by, or on the order of a licensed physician. |
{6}------------------------------------------------
{7}------------------------------------------------
* It is necessary for the end user to verify, confirm, and modify the analysis results when using Subject Device.
{8}------------------------------------------------
#### 8. Non-Clinical Test Summary
### Performance Standards
Digital Imaging and Communications in Medicine (DICOM) Set (Ps3.1 - .20) IEC 62304:2006, Medical Device Software - Software Life Cycle Processes. ISO 14971 Second edition 2007-03-01, Medical devices - Application of risk management to medical devices. IEC 62366-1 Edition 1.1 2020-06, Medical devices-Part 1 Application of usability engineering to medical devices.
Sonix Health is classified as a Moderate Level of Concern. The software verification and validation were conducted in compliance with IEC 62304 "Medical device software life cycle process" and the internal design control process to ensure safety and effectiveness. Throughout the verification and validation process, traceability was maintained, encompassing risk management (including Cyber Security and Usability).
Sonix Health was designed to assist clinicians in adult echocardiography examinations. The training data was collected from six centers. Among these, the demographic data from the representative institution showed 50.2% male participants, an average body mass index of 22.2, and 67.0% left ventricle ejection fraction.
The training data and validation data are distinct and independent. The ground truth annotation for the test was performed by two experienced sonographers with a Registered Diagnostic Cardiac Sonographer (RDCS) certification. The annotation was supervised by two experienced cardiologists and the consensus annotation was used as the final ground truth.
For performance testing, we used 476 B-mode clinical images, 243 M-mode clinical images, and 2,025 Doppler images, resulting in a total of 2,744 test images to assess these functions. Of these, 2,648 (96.5%) were from American participants. The data was respectively collected with 1,264 (47.7 %) from a U.S. hospital and 1,384 (52.3%) from a South Korea hospital. The gender distribution for the American participants was 65% male and 35% female. Data was acquired using equipment from four manufacturers.
The test was conducted blinded to private data, including patients' disease status. However, we measured LVEF based on the ground-truth of the validation datasets. The results showed that the left ventricle ejection fraction (LVEF) ranged from 14% to 76%, with a mean LVEF of 58% and a standard deviation of 11%. The normal LVEF (%) range for females is 54% to 74%, while for males it is 52% to 72%.
Sonix Health passed the test in two categories: an average accuracy of 98.22% for view recognition and a 93.98% average correlation coefficient when compared to manual measurements by participating experts.
Through verification and validation activities, Sonix Health is confirmed to meet all the design and performance requirements, ensuring safety and effectiveness without raising adverse issues. In conclusion, it demonstrates substantial equivalence.
### Clinical Test Summary
No clinical testing conducted in support of substantial equivalence when compared to the predicate and reference devices.
#### 9. Conclusion
In conclusion, all verification and validation activities demonstrated that the design specifications and technological characteristics of Sonix Health meet the applicable requirements and standards for the safety and effectiveness of the device for its intended use. As a result. Sonix Health is substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.